    Isogeometric Analysis of Hyperelastic Material Characteristics for Calcified Aortic Valve

    2024-03-23 08:16:10LongChenTingLiLiangLiuWenshuoWangXiaoxiaoDuandWeiWang

    Long Chen ,Ting Li ,Liang Liu ,Wenshuo Wang ,Xiaoxiao Du and Wei Wang

    1School of Mechanical Engineering,University of Shanghai for Science and Technology,Shanghai,200093,China

    2Department of Cardiac Surgery,Zhongshan Hospital,Fudan University,Shanghai,200093,China

    3School of Mechanical Engineering and Automation,Beihang University,Beijing,100191,China

    ABSTRACT This study explores the implementation of computed tomography(CT)reconstruction and simulation techniques for patient-specific valves,aiming to dissect the mechanical attributes of calcified valves within transcatheter heart valve replacement (TAVR) procedures.In order to facilitate this exploration,it derives pertinent formulas for 3D multi-material isogeometric hyperelastic analysis based on Hounsfield unit (HU) values,thereby unlocking foundational capabilities for isogeometric analysis in calcified aortic valves.A series of uniaxial and biaxial tensile tests is executed to obtain an accurate constitutive model for calcified active valves.To mitigate discretization errors,methodologies for reconstructing volumetric parametric models,integrating both geometric and material attributes,are introduced.Applying these analytical formulas,constitutive models,and precise analytical models to isogeometric analyses of calcified valves,the research ascertains their close alignment with experimental results through the close fit in displacement-stress curves,compellingly validating the accuracy and reliability of the method.This study presents a step-by-step approach to analyzing the mechanical characteristics of patient-specific valves obtained from CT images,holding significant clinical implications and assisting in the selection of treatment strategies and surgical intervention approaches in TAVR procedures.

    KEYWORDS Calcified aortic valves;tensile test;constitutive relations;CT reconstruction;isogeometric hyperelastic analysis

    1 Introduction

    Calcific aortic valve disease(CAVD)is a significant cardiac valve pathology,particularly afflicting the elderly population,and ranks as the third most prevalent cause of heart disease,as shown in Fig.1[1].Tongji Medical College in Hubei Province,China,conducted an extensive investigation focusing on individuals aged 60 years and older.Among the analyzed pool of 3,948 samples,a notable 1.93%exhibited severe heart disease,primarily manifesting as aortic insufficiency [2].The urgency of this issue is underscored by the aging trend in China,with data from the National Health Commission predicting that the elderly population (aged 60 years and above) will surpass 400 million by 2035.Extrapolating from these figures,it is projected that approximately 7.72 million individuals might grapple with aortic insufficiency,consequently imposing a considerable medical burden.This scenario highlights the critical importance of comprehensively and promptly addressing CAVD.

    Figure 1:Heart valves

    Transcatheter heart valve replacement(TAVR)offers a less invasive alternative for treating valve disorders via peripheral vascular pathways,thereby avoiding sternotomy and the risks associated with cardio-pulmonary circulation.However,selecting the right replacement valve without damaging the native aortic valve and successfully replacing it remains challenging for surgeons.Mu et al.demonstrated that rupture dramatically impacts hemodynamics[3].Previous investigations involving TAVR patients afflicted with aortic root rupture have conclusively identified the rupture site as a high-stress region [4].As TAVR demand grows,the computer simulation of patient-specific valve mechanical properties significantly enhances the feasibility assessment of surgical procedures and augments the success rate,also aiding in choosing the size of the replacement valves[5].

    In order to achieve a more precise simulation of patient-specific valve mechanical properties,two aspects should be considered: the selection of a constitutive model accurately reflecting the mechanical behavior of the calcified valve and employing a calculation method with high precision and computational efficiency.

    Global researchers have extensively explored the constitutive model of heart valves.Noteworthy work in 1995 by May-Newman and Yin exposed mitral valve traits,emphasizing nonlinear hyperelasticity and anisotropy [6].Li et al.found differing results in leaflet deformations,stress magnitudes,and distributions between isotropic and anisotropic models,with the maximum longitudinal normal stress increased,and the maximum transversal normal stress and in-plane shear stress decreased with the inclusion of anisotropy [7].In 2005,Sun et al.used a Fung-elastic material model that incorporated material parameters and axes derived from actual leaflet biaxial tests and measured leaflet collagen fiber structure to perform stress-strain analysis of the lobules[8].Progressing to 2009,May-Newman et al.introduced hyperelastic constitutive equations for aortic valves[9].Advancements by Sirois et al.in 2011 generalized the Fung-elastic model for fluid simulation of prosthesis valves[10].In 2012,a Mooney–Rivlin hyperelastic constitutive law was adopted at stent implantation [11,12].Almost concurrently,based on the fiber-reinforced material model,an anisotropic hyperelastic model was adopted to characterize the mechanical behaviors of aortic tissues[13,14].In 2014,Wang et al.used the anisotropic hyperelastic Holzapfel–Gasser–Ogden material model to analyze human heart tissues[15].However,there is no constitutive model that accurately captures the mechanical properties of calcified valves.

    Isogeometric analysis,presented by Hughes et al.[16],incorporates the exact geometrical description inherent in a computer-aided design(CAD)model into the analysis by employing the same spline basis functions utilized in the CAD system.Unlike the traditional finite element method,isogeometric analysis(IGA)demonstrates enhanced performance in dealing with nonlinear problems.The use of spline technology ensures a more precise construction of patient-specific valve geometry.High-order precision non-uniform rational B-splines (NURBS) surfaces facilitate effective mesh computation,even under significant deformation.Additionally,IGA obviates the need for repeated mesh refinement during nonlinear analysis,resulting in a substantial reduction in computational time.Furthermore,IGA is proficient in constructing and analyzing models involving dispersive and diffusive multimaterials by assigning material properties to control points[17–19].Recently,it has gained substantial significance in analyzing the mechanical properties of heart valves[20–24].

    Morganti et al.[25] and Xu et al.[20] employed NURBS to extract heart valve geometry from computed tomography (CT) images for model reconstruction.Additionally,in 2015,Kamenskya et al.introduced a geometric flexibility technique for computational fluid-structure coupling (FSI),successfully simulating the dynamic function of a three-valve biological valve throughout the cardiac cycle[23].In 2017,Takizawa et al.utilized the Space-Time Topology Change(ST-TC) method for moving grid calculations in flow problems,focusing particularly on the interleaflet contact of heart valves and preserving high-resolution representation near the blade surface while effectively addressing contact regions [26].In 2019,Bianchi conducted patient-specific aortic reconstruction using NURBS splines [27].Building on these advancements,Zhang et al.in 2021,presented an effective computational analysis framework for simulating primary and replacement three-leaf heart valves based on isogeometric analysis[28].Concurrently,Johnson et al.established a parameter-based approach for modeling tricuspid valves in a utility program,providing a robust and versatile method for assessing the performance and efficiency of various leaflets through isogeometric analysis[29].

    Although extensive research has been conducted on the mechanical attributes of heart valves,certain issues persist.Initially,valve materials are often oversimplified,adopting isotropic linear characteristics or described using classical constitutive models [13,30].Subsequently,the lack of consideration for valve thickening at calcification sites results in an ideal model with uniform thickness,significantly diverging from authentic valve morphology.Moreover,the precise extent and distribution of valve calcification remain inadequately captured.To address these issues,this paper proposes a method for studying the mechanical characteristics of patient-specific heart valves using CT images,aiming to maintain model authenticity and enhance the accuracy of valve mechanical analysis.The key contributions of this paper are outlined as follows:

    1.The pertinent formulas for 3D multi-material isogeometric hyperelastic analysis based on Hounsfield unit(HU)values are derived,enabling the foundational capabilities of isogeometric analysis for calcified aortic valves.The feasibility of these formulas is verified by the analysis results of ideal heart valves.

    2.Various constitutive models are fitted to accurately capture the stress-strain behavior of calcified heart valves via a range of uniaxial,equiaxial,and non-equiaxial experiments on human calcified valves.The best-performing model is chosen based on its fitting results,and its accuracy is confirmed through experimental comparisons.

    3.Aortic valve models with patient-specific material distribution are created from CT images using NURBS splines to build accurate volume parametric models for isogeometric analysis.This approach addresses challenges such as high computational costs,convergence difficulties,and mesh distortions commonly encountered in traditional dense mesh methods.

    4.The fitting quality of the displacement-load diagrams obtained from biaxial tensile tests on heart valves is compared with the displacement-load diagrams from the experimental simulation results of the reconstructed valves,validating the feasibility and accuracy of the methods and procedures employed in this study.

    The structure of this paper is as follows:Section 2 presents the derivation of basic formulas for 3D isogeometric hyperelastic analysis.Section 3 details the fitting process of constitutive equations.Section 4 presents the valve CT reconstruction procedure,accompanied by numerical examples and validation of the proposed method through real valve simulations.The outcomes and significance of the findings are elucidated in Section 5.Section 6 summarizes the conclusions and future works.

    2 Basic Formulas of Isogeometric Hyperelastic

    2.1 Basic Theories of Isogeometric Analysis

    Current CAD systems usually use NURBS for geometric representation and use basis functions instead of shape functions of finite elements.Theithpdegree(p+1 order)B-spline basis function is defined as:

    whereU=is the knot vector;[ui,ui+k+1]is the support interval of the basis functionNi,p(u).

    With the definition of the B-spline basis function,a 3D NURBS patch can be defined by threeknot vectors with the following expressions:

    whereξ,η,andζare parameter coordinates;p,q,andrare the degrees of the basic function in different directions;Pi,j,kis the control point.The expression of the basis functionRis as follows:

    whereNi,p(ξ),Mj,q(η),andQk,r(ζ)are the B-spline basis functions in each direction;ωi,j,kis the weight corresponding to the control pointPi,j,k[31].

    2.2 Isogeometric Hyperelastic Numerical Formula

    In a nonlinear system,the deformation gradientFis typically defined as the geometric relationship between the coordinates of the current geometry changexand the coordinates of the initial undeformed shapeX[32–34]:

    whereF=1 means there is no deformation.A right Cauchy–Green deformation tensorCis defined as

    Meanwhile,the Lagrangian strainEcan be defined as follows:

    Thereafter,the three-dimensional Lagrange strain variation can be derived as follows:

    whereRijdenotes the derivative ofRiwith respect to the reference configurationXin thejth direction.

    Then,the increment of the three-dimensional Lagrange strain can be defined as follows:

    The relationship between the variation and increment of the second Piola–Kirchhoff stressSand Lagrange strainEcan be derived as follows:

    whereDdenotes tangent constitutive tensor,which can be derived from the total strain energy density of different material properties.

    The strain energy can be defined as follows:

    Thereafter,strain energy variation can be derived as follows:

    Since nonlinear equations cannot be solved directly,they can be iteratively solved using the Newton-Raphson method.To find the displacement increment,one must linearize the weak form of the governing equation.

    2.3 Multi-Materials Realization Based on Hu Value

    2.3.1 Strain Energy Density Function

    When a material state can be entirely described by a given total strain,it is termed hyperelasticity.For hyperelastic materials,the stored strain energy remains constant as long as the strain within the material is constant,as demonstrated in Fig.2.Consequently,the strain energy density function emerges as the optimal choice for analyzing hyperelastic material models.In such materials,the strain energy density exists as a function of strain,allowing stress to be derived through differentiation[32,35].

    Figure 2:Stress-strain relationship of hyperelastic material

    Since the invariant does not vary with the reference coordinate system,regardless of the loading conditions an object undergoes,the strain energy density can be defined using the invariants of strain or alternatively,those of the deformation tensor.The three invariants of the right Cauchy–Green deformation tensorCare presented as follows:

    where the square root ofI3measures the volume change of the material.If the material is incompressible,it is clear thatI3=1.In Eqs.(22)–(24),λ1,λ2,λ3are the main stretch in three directionsx,y,z,which can be defined as:

    Incompressibility in materials can lead to numerous challenges in constitutive relations,particularly when combined with nonlinearities such as large displacements,large strains,and contact.In order to separate the distortion part from dilatation,it is necessary to introduce the so-called reduced invariantsJ1,J2andJ3,defined by

    Several typical hyperelastic models are listed in Table A1 of the Appendix.

    The nominal stress (first Piola-Kirchhoff stress) in the direction of stretch is obtained by differentiating the strain energy density with respect to the principal stretch as follows:

    Moreover,the second Piola-Kirchhoff stress is calculated by differentiating the strain energy density with respect to the Lagrangian strain:

    The Mooney-Rivlin model,frequently utilized as a strain energy density function,facilitates the specific solving process of the second Piola-Kirchhoff stressSand the tangent constitutive tensorD,as will be presented subsequently.

    Furthermore,the two-parameter strain energy density function of the Mooney-Rivlin model,including the incompressibility coefficient,is defined as

    The second Piola-Kirchhoff stressScan be derived as

    which can be rewritten as

    where the subscribed comma denotes derivative,i.e.,J1,C=2?J1/?C.AndI1,C,I2,C,I3,Ccan be obtained from

    Based on Eq.(14),the tangent constitutive tensorDcan be derived as follows:

    The formulas for the second Piola-Kirchhoff stressSand the tangent constitutive tensorDof several typical hyperelastic models are listed in Table A2 of the Appendix.

    2.3.2 Multi-Material Formula

    Mimics integrates both intuitive manual tools and fully automated algorithms for knee joint and cardiac segmentation,aiding medical practitioners in accurately processing patients’medical images.During calcified region annotation via thresholding in Mimics,theHuhighest was observed at the center of each calcified area,diminishing in a circular pattern outwardly,as depicted in Fig.3[36].

    Figure 3:The distribution of Hu valve

    Therefore,treating individual calcified regions of the valve as the same material results in relatively weak accuracy.Isogeometric analysis methods inherently offer advantages in multi-material model analysis.Material properties are assigned to various control points,and continuous refinement of the calcified regions allows the simulation to progressively approximate the real calcification scenario.

    This study only considers hyperelastic materials,and the material properties at control points are defined based on the gray valueHufrom the CT images.The material properties of each elementHun(ic)are represented by interpolating the values at the element’s center point,with the calculation formula as follows:

    whereHunis the gray valueHuof the control point;icis the center ofith element;ciis the control point set influencing the elementi;cijis thejth control point of setci,as shown in Fig.4;Nij(ic)is the NURBS basis function for the control point geometrycijof theith element[37].

    Figure 4:The relationship of element and control points

    Thus,the second Piola-Kirchhoff stressSand the tangent constitutive tensorDcan be rewritten as

    3 Constitutive Model Experiment

    Existing research on heart valve mechanical properties often utilizes fresh pig,cow,donkey,and other mammalian valve tissues for experimental testing[38,39].However,due to these animals’shorter lifespans,their aortic valve tissue rarely undergoes calcification,making it less suitable for this study.The heart valve tissue utilized in our experiment was provided by the Department of Cardiac Surgery at Zhongshan Hospital,affiliated with Fudan University.Previous studies have shown that valve tissue’s mechanical properties can be preserved at low temperatures [40].Obtaining human valve samples is challenging,so the valves used in this study were refrigerated prior to batch testing.The degree of calcification in heart valves can be categorized into three types[41,42]:Fig.5a shows mild calcification;Fig.5b moderate calcification;Fig.5c severe calcification.Severe calcification,which makes the valve material akin to linear material and significantly stiffer than native valves,is unsuitable for TAVR surgery and,thus,of limited significance in this study.Instead,the primary focus is on cases of mild and moderate calcification.

    Figure 5:Calcified heart valves

    3.1 Biaxial Tensile Test

    Biaxial stretching experiments provide a more precise characterization of mechanical properties than uniaxial stretching[43].May-Newman et al.have extensively described biaxial mechanical testing methods for valve leaflet tissues[44,45],forming the basis of this study.Additionally,early research discovered the anisotropic behavior and collagen fiber structure of aortic valve tissue.Stiffness is higher parallel to collagen fibers and lower in the transverse direction.Therefore,distinguishing between the circumferential and radial directions of the valve is essential during experiments.

    The experimental instrument employed in this study is the CellScale Biotester biaxial testing system,as illustrated in Fig.6.This system,specifically designed for biological materials,features a comprehensive biaxial testing setup.Its sample clamps,made of tungsten alloy and electrochemically sharpened,facilitate easy penetration of sample structures.Moreover,the system maintains the biological activity of samples by providing temperature-controlled water bath conditions during experiments.

    Figure 6:CellScale Biotester

    Given the smaller size of human valves compared to those of other animals such as pigs and cows,and considering the susceptibility of calcified valve leaflets to tearing and delamination during stretching,the researchers decided not to trim the valves in the biaxial experiments.The overall experimental procedure is outlined below:

    1.Before clamping the samples,a vernier caliper was used to measure the thickness of the leaflet in the central plane area ten times.The average of these measurements was then taken as the overall sample thickness for subsequent stress calculations.

    2.Using the measured thickness plane area from Step 1 as the clamping center,marking points were positioned at the sample’s center using carbon powder or ink for real-time displacement tracking during software processing.

    3.The sample was placed on the clamping fixture board,aligning its circumferential(Circ)and radial(Rad)directions with the main axes of the biaxial testing system.

    4.The distance between the dual suture hooks was adjusted to 7.5 mm×7.5 mm,and physiological saline was injected into the water bath tank.

    5.The sample was clamped onto the biaxial testing system,and the lifting mechanism was lowered to submerge the sample in the water environment.Fig.7 illustrates the sample clamping setup.

    6.Loading distances were set in both the axial and radial directions for equiaxial and nonequiaxial stretching experiments.In equiaxial stretching experiments,the axial and radial step distances were identical,i.e.,1:1.In non-equiaxial stretching experiments,the axial and radial step distances differed,i.e.,1:2 and 2:1.

    Figure 7:Biaxial tensile sample preparation and clamping

    Due to the reliance on the quality of carbon powder and ink marks for marker point displacement tracking,this study employs a quality filtering process to minimize the influence of this error.A square region at the center of the clamping area is divided,containing 30 mm×30 mm marker points within it.The coordinates at time 0 s are considered the initial coordinates,and the coordinates at each moment during stretching are used as the final coordinates.Fig.8 shows the trajectories of the marker points,with red points denoting the marker points and white curves representing their trajectories.It can be observed that marker points with clear ink marks exhibit smaller trajectory fluctuations.

    Figure 8:Movement trajectory of the mark point

    In order to select marker points with better trajectory detection quality,this study calculates the slope of the line connecting the end and start coordinates of each marker point at each moment.If the sum of the slope deviations does not exceed a predetermined error value,the point is deemed to have good trajectory detection quality.The distance between the marked points in the circumferential and radial directions is calculated using the Euclidean distance formula,as depicted in Fig.9.

    Assuming the sample holding area resembles a rectangular shape,andFC(N)andFR(N)are two pairs of forces acting on its edges.The initial dimensions of the rectangle areLC0(mm)andLR0(mm),which then reduce toLC(mm)andLR(mm)under stress.Additionally,the initial thickness and the thickness after deformation areh0(mm)andh(mm),respectively.With these parameters in place,the deformation of the sample during biaxial stretching can be accurately computed.Fig.10 visually represents the deformation process.

    Figure 9:Sample marking points

    Figure 10:Schematic diagram before and after sample deformation

    Cauchy stress on circumferential:

    Cauchy stress on radial:

    Stretch(Tensile ratio):

    Green-Lagrangian strain:

    Fig.11a illustrates the tensile ratio-stress curves of various samples during equiaxial stretching.The curves of the same color depict the response under circumferential and radial stretching conditions for each stretching degree.

    Figure 11:Cauchy stress-stretch curves of biaxial stretch

    Figs.11b and 11c show the tensile ratio-stress curves of various samples during non-equiaxial stretching.The curves of the same color represent the response under circumferential and radial stretching conditions for each stretching degree.

    The implementation of strain energy functions in modeling the material properties of soft tissues has demonstrated significant success in biomechanics.The coefficients of the constitutive model,shown in Table A2 of the Appendix,are intricately linked to the strain energy density function,which is unique to the material under consideration.Among the most widely adopted energy density functions used to capture the nonlinear behavior of soft tissue is the Fung strain energy density function,which has proven effective in characterizing the behavior of heart,artery,and mitral valve tissues [46,47].Despite the progress made,it is worth noting that,to date,no constitutive model has precisely captured the complex stress-strain relationship exhibited by heart valves.

    This study employs five representative constitutive equations: the Mooney-Rivlin 9-parameter,Yeoh,Ogden,Fung,and May-Newman models.Nonlinear least squares fitting codes were developed to fit these constitutive models.Experimental data from equiaxial and non-equiaxial tests were fitted with these equations,and the best-fitting model was selected for simulating valve mechanical characteristics.Fig.12 displays a portion of the experimental data and the effect of curve fitting.

    Figure 12:(Continued)

    Curve fitting errors of tensile experimental data under different models are presented in Table 1.

    Table 1:Fitting error of equiaxial and non-equiaxial tensile constitutive equation

    Curve fitting errors,compared in Table 1,reveal that the Mooney-Rivlin 9-parameter model and the Yeoh model exhibit the best fitting performance,accurately representing the stress-strain relationship of calcified heart valve leaflets.Variations may exist among different heart valve tissues and even within the same type of valve tissue among various individuals.The complexity of the model should be carefully balanced,considering the available data and computational resources,to avoid overfitting and excessive parameterization.Due to the higher number of coefficients required for the Mooney-Rivlin 9-parameter model and the similar fitting errors of both models under different stretching conditions,the Yeoh model is chosen for simulating the mechanical characteristics of calcified heart valve leaflets in this study.

    Yeoh[48]initially proposed the Yeoh model in 1993.It is a polynomial hyperelastic constitutive model that utilizes polynomial expansion to fit the stress-strain relationship of materials,expressed in the following general form:

    whereWis the strain energy density function;Ci0is the model parameter representing the material characteristics;Nis a non-negative integer;I1is the strain invariant;Jis the volume proportion of deformation.

    Fig.13 depicts the fitting results of various experimental data using the fixed Yeoh model.As the biaxial stretching experiments did not precisely align with the fiber orientation,some associated errors might be present in the experimental outcomes.Nevertheless,the fitting performance indicates satisfactory alignment with circumferential stretching.

    Figure 13:Fitting effect of the constitutive model

    3.2 Uniaxial Tensile Test

    The uniaxial tensile test,a common method for measuring the mechanical properties of heart valves [49],was used in this study.Aortic valves with pronounced calcification areas and clear distribution directions were selected for uniaxial stretching experiments.In order to ensure accurate representation,each semilunar valve was meticulously sampled,with one sample taken from each.

    The uniaxial tensile experiments were conducted using a YG026G multifunctional electronic fabric strength tester.During sample clamping,the edges were wrapped with gauze to prevent sample slippage.The clamping direction followed the direction of valve calcification to avoid tearing at the junction between the native and calcified valve during stretching.The clamping length was set to 10 mm,based on the available sample area,as shown in Fig.14.

    Figure 14:Uniaxial tensile sample preparation and clamping

    A force termination target of 5Nwas set during the experiments,with force and displacement recorded at each time point.Subsequently,the recorded data were used to calculate stress and strain values at each point using the following formulas:

    Cauchy stress:

    Stretch(Tensile ratio):

    Green-Lagrangian strain:

    The study reveals that severe calcification of the valve leads to higher stiffness and a tendency toward a linear stress-strain relationship.This trend is further illustrated in Fig.15a,where the stressstrain curve tends to become linear beyond a certain tensile extent.To more accurately determine the constitutive model coefficients of the calcified valve through uniaxial tensile testing,both the original and fully calcified valve segments were subjected to separate tensile tests.The Yeoh model was employed for curve fitting.The outcomes for these two extreme cases are depicted in Figs.15b and 15c,and the corresponding constitutive model coefficients are presented in Table 2.

    Table 2:Fitting error of uniaxial and biaxial tensile constitutive equation

    Figure 15:Uniaxial stretch

    4 Numerical Implementation

    4.1 CTA Image Model Reconstruction

    Due to the significant variations in calcified leaflet properties,especially regarding thickness,it is challenging to describe the leaflet model using a limited set of parameters.Utilizing patientspecific clinical imaging data for reconstruction,the models offer robust support for the accuracy of mechanical calculations and analysis.

    Image Data Import and Orientation Adjustment:

    The initial step involved importing the patient’s CT scan outcomes in DICOM format into Mimics.Subsequently,precise adjustments were made to ensure that the orientation of the view aligned perpendicularly with the aortic cross-section.

    Segmentation of Aortic Valve Root Area:

    Segmentation of the aortic valve root area,including both the native valve and the calcified region,was performed based on the variation inHuvalues,as illustrated in Fig.16.

    Figure 16:Threshold segmentation

    Triangular Mesh Generation and Material Data Derivation:

    A triangular mesh model of the aortic valve root area was generated through 3D reconstruction.Additionally,Huvalues corresponding to the mesh vertices and interior points were acquired.

    Leaflets Cutting and Smoothing:

    The reverse engineering software Geomagic Wrap was employed to trim and smooth the triangular mesh model of the aortic valve leaflets,as depicted in Fig.17.

    Figure 17:3D reconstruction and processing

    Upon completing the triangular mesh processing,the volume parametric geometry model of the aortic valve leaflet was created utilizing NURBS splines.Initially,the vertices of leaflets triangular mesh,also known as data points {Pl}l=0,1,···,m,were grouped into six boundary regions.Subsequently,these data points were approximated for each region using NURBS surfaces {Si}i=0,1,···,5.Lastly,the six surfaces were seamlessly joined,and a volumetric boolean sum [50] was applied to generate a volume parametric geometry model in Eq.(2).Fig.18 displays the construction process of the volume parametric geometry model.

    Figure 18:Construction of volume parametric model

    When constructing the patient-specific model,both the material model and the geometry model are equally important.The material fieldM(ξ,η,ζ)shares base functionswith the geometry model above as shown in follows:

    Atmiteration,the difference vector is calculated by subtractingM(m)(ul,vl,wl),Huvalue acquired by projection ofPl,fromMlwhich is the exactHuvalue of the pointPl.

    whereIijkis the index set that includes all cases of0.It can be demonstrated that the limiting form of the iterative results is the least square fitting of theHuvalue of data points,and the spectral radius of the iterative matrix is less than 1,indicating the convergence of the iterative results.The procedure is detailed in Algorithm 1.

    4.2 Algorithm for Nonlinear Solving

    Fig.19 displays the flowcharts of the source code for linear and nonlinear analyses,differing notably in the resolution of unknown variables.

    Figure 19:Source code flowchart

    Nonlinear equations cannot be effectively solved using direct methods to achieve a relatively precise solution.This study adopts an automatic load increment approach similar to that in ABAQUS[52],employing the Newton-Raphson method to resolve nonlinear equations.The procedure is detailed in Algorithm 2.

    The automatic load incrementation scheme is given in Algorithm 3.

    4.3 Verification of Algorithm

    In order to prove the correctness of the method in this study,two hyperelastic models are simulated by an isogeometric analysis program.

    1.The model is a square plate with a thickness much smaller than its side length.As shown in Fig.20a,the length and width of the model are 10 mm,and the thickness is 0.2 mm.Constraints are applied to the lower surface,front surface,and left surface of the plate.The right and back surfaces of the plate are given a prescribed horizontal displacementux=10 mm,uy=10 mm.Fig.20b shows the calcification of the model,and the material parameters are set as follows:

    Material 1:C10=0.01863 MPa,C01=0.00979 MPa,k=1e5

    Material 2:C10=0.5516 MPa,C01=0.1379 MPa,k=1e5

    Figure 20:Model constraint

    In order to verify the accuracy of isogeometric analysis in hyperelastic materials more accurately,two cases are considered: for a single material,results from both FEA and IGA 3D analyses are displayed in Figs.21a,21b,21d,and 21e.For multiple materials,IGA 3D analysis results are depicted in Figs.21c and 21f.

    Figure 21:Model analysis results

    2.The model represents a leaflet of an ideal aortic valve.A fixed constraint is applied at the junction of the coronary valve and the aorta (valve ring),marked in red in Fig.22b,and a displacement constraint is applied at the center of the free edge of the coronary valve,marked in blue in Fig.22a.Fig.22c illustrates the model’s calcification.The material parameters are set as follows:

    Material 1:C10=0.01863 MPa,C01=0.00979 MPa,k=1e5

    Material 2:C10=0.5516 MPa,C01=0.1379 MPa,k=1e5

    Figure 22:Model constraint and calcification region map

    The Von Mises diagram for IGA analysis without calcification is shown in Fig.23a and with calcification in Fig.23b.The stress distribution in the x-direction of IGA analysis without calcification is depicted in Fig.24a and with calcification in Fig.24b.Fig.25a displays the displacement distribution in the z-direction of IGA analysis without calcification,while Fig.25b shows this distribution in the case of calcification.The stress and displacement distribution in both calcified and non-calcified models demonstrate a consistent overall trend.However,in calcified valve regions,a distinct stress concentration pattern is observed,corresponding to actual mechanical analysis.

    When comparing the analysis results of a non-calcified square plate in FEA and IGA,similarities in stress-displacement distribution are apparent.The maximum and minimum values fall within a specific error range,confirming the effectiveness of stress analysis for a basic hyperelastic model in this study.Moreover,these two simulation examples reveal thatHueffectively distinguishes calcified regions from non-calcified ones,with notable differences inHuvalue between the two models.Stress concentration in calcified areas is visibly distinct,with stress distribution patterns reflecting the form of calcification.

    Figure 23:Von Mises

    Figure 24:x-direction stress

    Figure 25:z-direction displacement

    4.4 Analysis and Comparison of Mechanical Properties of Valve

    Considering that valve CT data is scanned while the valve is operational,the reconstructed valve shape may deviate from its form under experimental conditions.In order to facilitate a comprehensive comparison between the experiment and simulation during biaxial valve stretching,the simulation valve model is meticulously adjusted to closely emulate the experimental valve’s configuration.This involves a procedure to flatten the valve before conducting model analysis and fine-tuning the valve’s orientation to align the circumferential and radial directions parallel to the corresponding coordinate axes,as illustrated in Fig.26.

    Figure 26:Valve unfold diagram

    A square with a side length of 6.7 mm is formed around the restraint and loading ends in the circumferential and radial directions of the valve.The specific constraints and loading modes are shown in Fig.27.The two adjacent sides of the valve are fixed,and the other two sides are subjected to displacement constraints.

    Figure 27:Model constraint and load

    The following figures display the biaxial tensile simulation results of the valve under varying calcification conditions.Figs.28a–28c exhibit the calcification of one leaflet in each of the three patients.The color bar on the right represents theHuvalue in the CT image of the valve.Figs.28d–28f illustrate the displacement change in the direction of valve thickness under simulation conditions,showing a smaller displacement change in more severely calcified cases,consistent with an increase in valve stiffness after calcification.Figs.28g–28i present the valve strain diagram under simulated conditions,with the strain region indicating the biaxial tensile region and more severe calcification resulting in smaller strain.The black box in Figs.28d–28i highlight the biaxial stretching area of the valve.

    Figure 28:Biaxial tensile simulation results of the valves

    In Section 2.3,the strain energy density function is fitted to the stress-strain curve obtained from processing experimental data,resulting in the best-fit constitutive model and model coefficients.This model is then applied to simulate the valve reconstruction model in Section 3.1,replicating the valve’s biaxial tensile test.The simulation produces a displacement-load curve,which is compared with the actual experimental results.The curve’s conformity analysis verifies the simulation approach used in this study and indirectly corroborates the appropriateness of the selected constitutive model,as outlined in Section 2.3.The specific comparison process is shown in Fig.29.

    Figure 29:Comparison flow chart

    Considering the significantly greater stiffness of the calcified region of the valve compared to the non-calcified region,it is imperative in the simulation to position the maximum load point at the calcification center.Concurrently,displacement measurements should be recorded at the clamping edge proximal to this center,as delineated in Fig.30.

    Figure 30:Contrast data selection

    The displacement-load curves derived from various experimental samples at differing stretching ratios were juxtaposed with those procured from the simulation,as shown in Fig.31.

    Fig.28 reveals that within the stretching region,the valve’s strain and displacement distribution closely mirrors the actual calcified area.Furthermore,as depicted in Fig.31,a noteworthy congruence exists between displacement-load curves from both biaxial stretching simulations and experimental data.This congruence not only affirms the simulation method’s relevance but also indirectly validates the accuracy of the selected constitutive model in replicating the valve’s mechanical characteristics.

    Figure 31:Comparison of experimental and simulation displacement-load curves

    5 Discussion

    In this investigation,a thorough analysis was conducted on the mechanical properties of patientspecific calcified valves using CT imaging,thereby introducing an innovative perspective into the feasibility of TAVR surgery.

    Constitutive Model Fitting: The comparison between equiaxial and non-equiaxial stretching scenarios reveals the fitting errors of various classical constitutive models to experimental data.Notably,the Yeoh model exhibited the smallest fitting error,indicating its superior ability to express the strain energy density function of calcified valves.The favorable fitting results of the Yeoh model,as shown in Fig.13,provide empirical evidence supporting the model’s validity.

    Multi-material based on Hu value: The square plate model’s analytical outcomes indicate congruent stress-strain distributions under biaxial stretching conditions,validating the correctness of the derived three-dimensional isogeometric hyperelastic multi-material correlation formula based on the Hu value.Furthermore,the analysis of an ideal valve model underscores the significant impact of calcification patterns on valve stress-strain results,thereby emphasizing the importance of simulating valve surgeries.

    Comparison of experimental and simulation results: Fig.31 displays a satisfactory alignment between experimental and simulated displacement-load curves for three reconstructed valve models under equiaxial and non-equiaxial stretching conditions.This alignment corroborates the validity of the approach employed in this study for investigating valve mechanical properties derived from CT images.

    6 Conclusion

    In summary,this study has advanced the analysis of the mechanical characteristics of patientspecific aortic valves obtained from CT images.Compared to prior studies,this study presents three principal contributions.Firstly,it utilizes the Hu value to achieve a multi-material representation of biological tissue,subsequently deriving an isogeometric hyperelastic three-dimensional multi-material correlation formula.Secondly,investigations confirm the Yeoh model as the most suitable constitutive model for expressing the stress-strain behavior of calcified valves,validated through multiple single and double experiments.Lastly,building upon existing CT reconstruction methodologies,the paper details the comprehensive process of reconstructing calcified valves.The resulting patient-specific valve parametric model not only embodies precise geometric properties but also retains material characteristics accurately.The congruence between the results from calcified valve analysis and experimental findings substantiates and lends theoretical support to the feasibility of the proposed research framework.However,due to the intricate interplay of factors influencing TAVR surgery under real-world conditions,this study does not conduct a comprehensive analysis encompassing surgical simulation of actual cases.Despite these limitations,the study introduces an effective simulation approach for TAVR surgery.Future efforts will focus on modeling and simulating real-world case studies,which will significantly enhance the practical implications and applicability of these research findings.

    Acknowledgement:The authors would like to thank the School of Mechanical Engineering and Automation at Beihang University for providing the NLIGA source code.We also thank the Department of Cardiology,Zhongshan Hospital,Affiliated with Fudan University,for providing the valve samples and financial support.

    Funding Statement:This work was supported by the Natural Science Foundation of China (Project Nos.52075340 and 61972011) and the Shanghai Special Research Project on Aging Population and Maternal and Child Health(Project No.2020YJZX0106).

    Author Contributions:The authors confirm their contribution to the paper as follows:study conception and design: Long Chen,Ting Li,and Liang Liu;data collection: Ting Li,Liang Liu,and Wenshuo Wang;analysis and interpretation of results:Wei Wang,Xiaoxiao Du;draft manuscript preparation:Ting Li,Liang Liu.All authors reviewed the results and approved the final version of the manuscript.

    Availability of Data and Materials:The data belongs to Zhongshan Hospital,Fudan University,and is confidential.The authors do not have permission to share data.

    Conflicts of Interest:The authors declare that they have no conflicts of interest to report regarding the present study.

    Appendix A.Several typical hyperelastic models

    Table A1 lists several typical hyperelastic models.

    Table A1:Several typical hyperelastic models

    Table A2 lists the formulas for the second Piola-Kirchhoff stressSand the tangent constitutive tensorDof several typical hyperelastic models.

    Table A2:The S and D of several typical hyperelastic models

    18禁国产床啪视频网站| 久久香蕉激情| 亚洲精品中文字幕一二三四区 | 国产成人一区二区三区免费视频网站| 国产精品 欧美亚洲| 日本撒尿小便嘘嘘汇集6| 国产精品久久久久久精品电影小说| 国产人伦9x9x在线观看| 欧美 亚洲 国产 日韩一| 国产精品自产拍在线观看55亚洲 | 中国美女看黄片| 变态另类成人亚洲欧美熟女 | 亚洲精品成人av观看孕妇| 欧美精品一区二区免费开放| 午夜福利一区二区在线看| 亚洲午夜理论影院| 国产免费福利视频在线观看| 精品人妻1区二区| 国产精品 欧美亚洲| 国产一区二区三区综合在线观看| 久久青草综合色| 亚洲av国产av综合av卡| 丁香六月天网| 高潮久久久久久久久久久不卡| 亚洲中文av在线| 亚洲一区中文字幕在线| 99香蕉大伊视频| 80岁老熟妇乱子伦牲交| 国产精品久久久久成人av| 99精品久久久久人妻精品| 91麻豆精品激情在线观看国产 | 亚洲精品中文字幕在线视频| 久久精品国产99精品国产亚洲性色 | 一区二区av电影网| 久热这里只有精品99| 一进一出好大好爽视频| 色尼玛亚洲综合影院| 免费久久久久久久精品成人欧美视频| 亚洲色图 男人天堂 中文字幕| 国产精品av久久久久免费| 亚洲色图 男人天堂 中文字幕| 国产xxxxx性猛交| 国产精品一区二区精品视频观看| 国产高清videossex| av超薄肉色丝袜交足视频| 欧美激情 高清一区二区三区| 日韩免费av在线播放| 99国产精品一区二区三区| 多毛熟女@视频| 久久久久国内视频| 性高湖久久久久久久久免费观看| 丝瓜视频免费看黄片| 午夜日韩欧美国产| 一本—道久久a久久精品蜜桃钙片| 亚洲人成77777在线视频| 欧美精品高潮呻吟av久久| 法律面前人人平等表现在哪些方面| 国产精品九九99| 欧美 日韩 精品 国产| 18禁裸乳无遮挡动漫免费视频| 国产精品久久久人人做人人爽| 精品少妇一区二区三区视频日本电影| 国产精品免费大片| 国产黄色免费在线视频| 欧美黄色片欧美黄色片| 精品少妇一区二区三区视频日本电影| 色尼玛亚洲综合影院| 欧美精品一区二区免费开放| 国产在线免费精品| 久久久久久久大尺度免费视频| 色视频在线一区二区三区| 大香蕉久久成人网| 十八禁网站网址无遮挡| 飞空精品影院首页| 精品卡一卡二卡四卡免费| 两人在一起打扑克的视频| 欧美精品人与动牲交sv欧美| 国产精品香港三级国产av潘金莲| 一本综合久久免费| 欧美一级毛片孕妇| 久久久久久久久久久久大奶| 精品少妇黑人巨大在线播放| 在线观看免费视频日本深夜| 我的亚洲天堂| 国产97色在线日韩免费| 日韩欧美免费精品| 精品人妻1区二区| 午夜日韩欧美国产| 亚洲国产中文字幕在线视频| 十八禁人妻一区二区| 国产欧美日韩一区二区三| 天堂俺去俺来也www色官网| 亚洲成人手机| 国产单亲对白刺激| 免费观看人在逋| 在线亚洲精品国产二区图片欧美| 精品国内亚洲2022精品成人 | 国精品久久久久久国模美| 日韩有码中文字幕| 免费观看人在逋| 亚洲欧美一区二区三区黑人| 男女免费视频国产| 99国产精品一区二区蜜桃av | 五月开心婷婷网| 亚洲久久久国产精品| 亚洲精品久久成人aⅴ小说| 亚洲欧美一区二区三区久久| 国产成人啪精品午夜网站| 久久热在线av| 国产高清videossex| 欧美久久黑人一区二区| 国产激情久久老熟女| 搡老乐熟女国产| 99精品在免费线老司机午夜| 精品国产国语对白av| 国产亚洲精品一区二区www | 一本综合久久免费| 欧美人与性动交α欧美软件| 免费在线观看视频国产中文字幕亚洲| 老司机亚洲免费影院| 国产亚洲精品一区二区www | 91老司机精品| 国产高清videossex| 国产人伦9x9x在线观看| 一级毛片女人18水好多| 欧美中文综合在线视频| 欧美中文综合在线视频| 正在播放国产对白刺激| 黑人欧美特级aaaaaa片| 欧美大码av| 午夜成年电影在线免费观看| 搡老乐熟女国产| 久久 成人 亚洲| 亚洲成人手机| 免费在线观看日本一区| 91九色精品人成在线观看| 免费一级毛片在线播放高清视频 | 日韩大片免费观看网站| 中文欧美无线码| 超色免费av| 青草久久国产| 久久久久久亚洲精品国产蜜桃av| 在线观看免费日韩欧美大片| 中国美女看黄片| 国产成人av教育| 亚洲免费av在线视频| 法律面前人人平等表现在哪些方面| 黄色怎么调成土黄色| 在线观看免费高清a一片| 99国产精品99久久久久| 亚洲av片天天在线观看| 亚洲中文日韩欧美视频| 日韩免费高清中文字幕av| 欧美 亚洲 国产 日韩一| 97人妻天天添夜夜摸| av免费在线观看网站| 亚洲熟女毛片儿| 三上悠亚av全集在线观看| 日韩欧美一区视频在线观看| 他把我摸到了高潮在线观看 | 欧美国产精品一级二级三级| 黄片播放在线免费| 王馨瑶露胸无遮挡在线观看| 亚洲午夜精品一区,二区,三区| 中文亚洲av片在线观看爽 | 老熟妇乱子伦视频在线观看| 日本黄色视频三级网站网址 | 日韩欧美国产一区二区入口| 亚洲全国av大片| av视频免费观看在线观看| 日本精品一区二区三区蜜桃| 美女高潮到喷水免费观看| 欧美激情高清一区二区三区| 免费看十八禁软件| 操美女的视频在线观看| 狂野欧美激情性xxxx| 色视频在线一区二区三区| 亚洲va日本ⅴa欧美va伊人久久| 国产视频一区二区在线看| 悠悠久久av| 欧美精品高潮呻吟av久久| 国产91精品成人一区二区三区 | 国产精品一区二区在线不卡| 成人18禁在线播放| 免费一级毛片在线播放高清视频 | 久久久精品94久久精品| 777米奇影视久久| av一本久久久久| 中文字幕高清在线视频| 老司机在亚洲福利影院| 亚洲一区中文字幕在线| 丰满饥渴人妻一区二区三| 国产深夜福利视频在线观看| 欧美在线黄色| 91成年电影在线观看| 亚洲精品中文字幕一二三四区 | 97人妻天天添夜夜摸| 自线自在国产av| 久久精品国产亚洲av香蕉五月 | www日本在线高清视频| 又黄又粗又硬又大视频| 亚洲色图av天堂| 飞空精品影院首页| 久久国产精品男人的天堂亚洲| 十八禁网站免费在线| 免费观看av网站的网址| 纵有疾风起免费观看全集完整版| 最黄视频免费看| 高清av免费在线| 免费在线观看日本一区| 老司机在亚洲福利影院| 12—13女人毛片做爰片一| 国产高清激情床上av| 国产有黄有色有爽视频| 国产成人影院久久av| 欧美另类亚洲清纯唯美| 一区二区三区国产精品乱码| 侵犯人妻中文字幕一二三四区| 999久久久精品免费观看国产| 国产真人三级小视频在线观看| 亚洲av美国av| 日本a在线网址| 久久九九热精品免费| 免费久久久久久久精品成人欧美视频| 一级毛片精品| 下体分泌物呈黄色| 99久久人妻综合| 国产精品国产av在线观看| 精品久久久久久久毛片微露脸| a级毛片在线看网站| 亚洲成a人片在线一区二区| 久久ye,这里只有精品| 日韩免费高清中文字幕av| 悠悠久久av| 午夜福利欧美成人| 在线天堂中文资源库| 在线看a的网站| 精品熟女少妇八av免费久了| 国产精品欧美亚洲77777| 国产成+人综合+亚洲专区| 一二三四在线观看免费中文在| 欧美av亚洲av综合av国产av| 成人免费观看视频高清| 香蕉国产在线看| 成人三级做爰电影| xxxhd国产人妻xxx| 久久久久网色| 伊人久久大香线蕉亚洲五| 国产欧美日韩一区二区三| 大陆偷拍与自拍| 国产成人影院久久av| 最新在线观看一区二区三区| 久久久久久久久免费视频了| 真人做人爱边吃奶动态| 成年版毛片免费区| 自线自在国产av| 日本欧美视频一区| 视频区欧美日本亚洲| 纵有疾风起免费观看全集完整版| 久久性视频一级片| 国产亚洲欧美精品永久| 亚洲成人手机| 黑人巨大精品欧美一区二区mp4| 男女无遮挡免费网站观看| 日韩欧美免费精品| 亚洲五月婷婷丁香| 国产精品免费大片| 在线观看免费高清a一片| 亚洲欧美日韩另类电影网站| 国产又爽黄色视频| 大陆偷拍与自拍| 成人手机av| 国产成人一区二区三区免费视频网站| 制服人妻中文乱码| 欧美人与性动交α欧美软件| 飞空精品影院首页| 国产真人三级小视频在线观看| 国产欧美日韩精品亚洲av| 亚洲精品自拍成人| 亚洲avbb在线观看| 亚洲中文日韩欧美视频| 国产一区二区三区视频了| 久久狼人影院| 精品人妻熟女毛片av久久网站| 嫁个100分男人电影在线观看| 大片电影免费在线观看免费| 国产精品久久久久成人av| 黄网站色视频无遮挡免费观看| 日本欧美视频一区| 国产一区二区三区在线臀色熟女 | 黄色毛片三级朝国网站| 手机成人av网站| 国产日韩欧美亚洲二区| 国产在线视频一区二区| 欧美乱妇无乱码| av一本久久久久| 国产老妇伦熟女老妇高清| 老司机靠b影院| kizo精华| 一级片免费观看大全| 成人国产av品久久久| 777久久人妻少妇嫩草av网站| av线在线观看网站| 亚洲精品久久成人aⅴ小说| 18禁美女被吸乳视频| 一区在线观看完整版| 精品一区二区三区av网在线观看 | 免费黄频网站在线观看国产| 国产免费av片在线观看野外av| 精品国产乱码久久久久久男人| 精品一区二区三区四区五区乱码| 亚洲av成人一区二区三| 不卡一级毛片| 亚洲成人免费av在线播放| 18禁国产床啪视频网站| 国产免费福利视频在线观看| 可以免费在线观看a视频的电影网站| 99国产精品99久久久久| 嫩草影视91久久| 五月开心婷婷网| 欧美日韩一级在线毛片| 黄色怎么调成土黄色| 欧美 日韩 精品 国产| 黄片小视频在线播放| 亚洲av国产av综合av卡| 一本—道久久a久久精品蜜桃钙片| 久久精品国产综合久久久| 人人妻,人人澡人人爽秒播| 亚洲av电影在线进入| 一区二区av电影网| 无限看片的www在线观看| 丁香欧美五月| 久久中文字幕一级| 亚洲国产av新网站| 亚洲欧美一区二区三区久久| 大型黄色视频在线免费观看| 捣出白浆h1v1| 久久久国产成人免费| 两个人看的免费小视频| 午夜两性在线视频| 人妻 亚洲 视频| 欧美日韩成人在线一区二区| 最新美女视频免费是黄的| 飞空精品影院首页| 一本一本久久a久久精品综合妖精| 十八禁人妻一区二区| 午夜久久久在线观看| 在线观看66精品国产| 日韩欧美三级三区| 国产成人精品久久二区二区91| 日韩一卡2卡3卡4卡2021年| 亚洲 欧美一区二区三区| 每晚都被弄得嗷嗷叫到高潮| av在线播放免费不卡| 免费女性裸体啪啪无遮挡网站| 亚洲第一欧美日韩一区二区三区 | 一区二区三区乱码不卡18| 999久久久国产精品视频| 国产精品免费一区二区三区在线 | 久久久欧美国产精品| 老鸭窝网址在线观看| 国产精品免费大片| 一级黄色大片毛片| 国产在线视频一区二区| 日本一区二区免费在线视频| 精品国内亚洲2022精品成人 | 制服人妻中文乱码| 国产无遮挡羞羞视频在线观看| 亚洲视频免费观看视频| 成人手机av| 国产成人av激情在线播放| 亚洲少妇的诱惑av| 亚洲av国产av综合av卡| 最近最新中文字幕大全电影3 | 久久亚洲精品不卡| 国产免费福利视频在线观看| 91麻豆av在线| 亚洲五月色婷婷综合| 99精品欧美一区二区三区四区| 波多野结衣一区麻豆| 王馨瑶露胸无遮挡在线观看| 国产亚洲精品第一综合不卡| 视频区欧美日本亚洲| 美女主播在线视频| 免费日韩欧美在线观看| 国产又色又爽无遮挡免费看| 久久免费观看电影| 淫妇啪啪啪对白视频| 99国产极品粉嫩在线观看| 丝袜美腿诱惑在线| 2018国产大陆天天弄谢| 黄色成人免费大全| 99在线人妻在线中文字幕 | 黄片大片在线免费观看| 久久青草综合色| 男女午夜视频在线观看| cao死你这个sao货| 高清在线国产一区| 亚洲 国产 在线| videosex国产| 中文字幕人妻熟女乱码| 桃花免费在线播放| 成年人免费黄色播放视频| 国产精品久久久久久精品电影小说| 精品久久蜜臀av无| 国产亚洲精品久久久久5区| 最近最新免费中文字幕在线| 50天的宝宝边吃奶边哭怎么回事| 日本五十路高清| 亚洲精华国产精华精| 一区二区三区国产精品乱码| 成年动漫av网址| 后天国语完整版免费观看| 亚洲专区字幕在线| av欧美777| 无人区码免费观看不卡 | 欧美激情 高清一区二区三区| 久久性视频一级片| e午夜精品久久久久久久| 看免费av毛片| 黄色 视频免费看| 久久久国产精品麻豆| 首页视频小说图片口味搜索| 精品福利观看| 中文字幕人妻丝袜一区二区| 国产成人欧美在线观看 | 热re99久久国产66热| 悠悠久久av| 日本a在线网址| 国产日韩一区二区三区精品不卡| www.自偷自拍.com| 国产精品av久久久久免费| 一级片'在线观看视频| 一区二区三区国产精品乱码| 国产深夜福利视频在线观看| 亚洲av国产av综合av卡| 国产精品久久久久成人av| 精品少妇黑人巨大在线播放| 亚洲综合色网址| 999久久久精品免费观看国产| 亚洲国产精品一区二区三区在线| 脱女人内裤的视频| 久久精品国产亚洲av香蕉五月 | 亚洲一码二码三码区别大吗| 亚洲国产欧美在线一区| 蜜桃国产av成人99| 欧美精品一区二区大全| 久久 成人 亚洲| 色精品久久人妻99蜜桃| 国产aⅴ精品一区二区三区波| 精品人妻1区二区| 久久午夜亚洲精品久久| 老司机在亚洲福利影院| 日韩视频一区二区在线观看| 久久久久久久久免费视频了| 欧美av亚洲av综合av国产av| 制服诱惑二区| 国产精品98久久久久久宅男小说| 色94色欧美一区二区| 两个人看的免费小视频| videos熟女内射| 中文字幕人妻熟女乱码| 精品少妇一区二区三区视频日本电影| 天堂中文最新版在线下载| 免费一级毛片在线播放高清视频 | 亚洲,欧美精品.| 两人在一起打扑克的视频| 热re99久久精品国产66热6| 国产成人精品无人区| 99久久99久久久精品蜜桃| 宅男免费午夜| www.精华液| 国产精品麻豆人妻色哟哟久久| 亚洲熟女精品中文字幕| 午夜激情av网站| 中文字幕另类日韩欧美亚洲嫩草| 亚洲色图综合在线观看| 国产精品久久久人人做人人爽| 国产单亲对白刺激| 热re99久久精品国产66热6| 亚洲精品中文字幕一二三四区 | 亚洲熟妇熟女久久| 夫妻午夜视频| 免费少妇av软件| 国产成人一区二区三区免费视频网站| 成人特级黄色片久久久久久久 | 精品午夜福利视频在线观看一区 | av有码第一页| 国产一区二区 视频在线| 国产高清激情床上av| 99热网站在线观看| 亚洲av成人不卡在线观看播放网| 男人操女人黄网站| 亚洲熟女毛片儿| 亚洲黑人精品在线| 免费在线观看视频国产中文字幕亚洲| 欧美激情极品国产一区二区三区| 国产在线精品亚洲第一网站| h视频一区二区三区| e午夜精品久久久久久久| 另类精品久久| 午夜福利在线观看吧| 下体分泌物呈黄色| 男女免费视频国产| 精品久久久精品久久久| 女人久久www免费人成看片| 久久久久久久久免费视频了| 国产淫语在线视频| 777米奇影视久久| 国产一区有黄有色的免费视频| 日韩一卡2卡3卡4卡2021年| 999久久久国产精品视频| 后天国语完整版免费观看| 成人特级黄色片久久久久久久 | 桃花免费在线播放| 久久精品国产a三级三级三级| 操出白浆在线播放| 老熟妇仑乱视频hdxx| 亚洲五月婷婷丁香| 麻豆乱淫一区二区| 久久久久国内视频| 亚洲欧美精品综合一区二区三区| 极品教师在线免费播放| 窝窝影院91人妻| 日本黄色日本黄色录像| 国产成人av激情在线播放| 另类亚洲欧美激情| 熟女少妇亚洲综合色aaa.| 国产xxxxx性猛交| 777米奇影视久久| 巨乳人妻的诱惑在线观看| 菩萨蛮人人尽说江南好唐韦庄| 大型黄色视频在线免费观看| 欧美日韩福利视频一区二区| 精品国产亚洲在线| 日本av免费视频播放| 亚洲欧美激情在线| 中国美女看黄片| 欧美 亚洲 国产 日韩一| 免费观看av网站的网址| av电影中文网址| 免费观看av网站的网址| 国产精品久久久久成人av| 男女之事视频高清在线观看| 香蕉久久夜色| 欧美日韩亚洲综合一区二区三区_| 亚洲专区字幕在线| 久久人妻福利社区极品人妻图片| 久久99一区二区三区| 咕卡用的链子| 成在线人永久免费视频| 天天躁日日躁夜夜躁夜夜| 中文字幕人妻丝袜制服| 欧美乱码精品一区二区三区| 国产亚洲精品久久久久5区| av电影中文网址| 免费女性裸体啪啪无遮挡网站| 亚洲自偷自拍图片 自拍| 日韩欧美免费精品| 一本大道久久a久久精品| 亚洲九九香蕉| 一本久久精品| 91成年电影在线观看| 国产免费视频播放在线视频| 性高湖久久久久久久久免费观看| 大片免费播放器 马上看| 性高湖久久久久久久久免费观看| 国产成人精品久久二区二区免费| 精品久久久久久久毛片微露脸| 精品卡一卡二卡四卡免费| 成在线人永久免费视频| 一本综合久久免费| 男女之事视频高清在线观看| 岛国在线观看网站| 欧美在线一区亚洲| 一二三四在线观看免费中文在| 男女边摸边吃奶| 成人特级黄色片久久久久久久 | 美女扒开内裤让男人捅视频| 久久狼人影院| 在线播放国产精品三级| 久久毛片免费看一区二区三区| svipshipincom国产片| 亚洲精品久久成人aⅴ小说| 中文字幕最新亚洲高清| 男女床上黄色一级片免费看| 一级毛片女人18水好多| 五月开心婷婷网| 久久精品国产综合久久久| 日韩欧美免费精品| 国产精品国产高清国产av | cao死你这个sao货| 在线观看66精品国产| 嫁个100分男人电影在线观看| 不卡一级毛片| 精品一区二区三区四区五区乱码| 丁香六月天网| 免费久久久久久久精品成人欧美视频| 一区二区日韩欧美中文字幕| 久久ye,这里只有精品| 亚洲精品美女久久av网站| 看免费av毛片| 亚洲性夜色夜夜综合| 手机成人av网站| 交换朋友夫妻互换小说| 国产日韩欧美视频二区| 美女主播在线视频| 国产深夜福利视频在线观看| 十分钟在线观看高清视频www| avwww免费| 欧美精品av麻豆av| 成年人免费黄色播放视频| 大片免费播放器 马上看| 黄频高清免费视频| 激情在线观看视频在线高清 |