    A new approach for accurate determination of particle sizes in microfluidic impedance cytometry

    2022-02-01 01:47:06PriyadarshiAbbasiKumaranandChowdhury
    納米技術(shù)與精密工程 2022年4期

    N.Priyadarshi, U.Abbasi, V.Kumaran, and P.Chowdhury

    ABSTRACT In microfluidi impedance cytometry,the change in impedance is recorded as an individual cell passes through a channel between electrodes deposited on its walls,and the particle size is inferred from the amplitude of the impedance signal using calibration.However,because the current density is nonuniform between electrodes of finit width,there could be an error in the particle size measurement because of uncertainty about the location of the particle in the channel cross section.Here,a correlation is developed relating the particle size to the signal amplitude and the velocity of the particle through the channel.The latter is inferred from the time interval between the two extrema in the impedance curve as the particle passes through a channel with cross-sectional dimensions of 50μm(width)×30μm(height)with two pairs of parallel facing electrodes.The change in impedance is predicted using 3D COMSOL finite-elemen simulations,and a theoretical correlation that is independent of particle size is formulated to correct the particle diameter for variations in the cross-sectional location.With this correlation,the standard deviation in the experimental data is reduced by a factor of two to close to the standard deviation reported in the manufacturer specifications

    KEYWORDS Impedance cytometry,Lab on chip,COMSOL simulation

    I.INTRODUCTION

    Single-cell impedance cytometry is becoming a widely used noninvasive method for determining the properties of cells,such as their sizes,shapes,and dielectric properties.1The size of a single cell was firs determined by Coulter2via DC resistance measurement.More recently,AC methods have been used to determine both the size and properties of the cell membrane,such as cytoplasm and subcellular components.3–8In general,impedance cytometry of a cell is carried out in a microfluidi channel based on the impedance signal when the cell—suspended in a conductive solution—passes through microelectrodes fabricated on the inner walls of the channel,and accurate determination of particle size has been explored with different microelectrode configurations9–12

    The literature contains two different microelectrode configura tions,i.e.,(i)coplanar,where the electrodes are etched on one wall of the channel with rectangular cross section,and(ii)parallel facing,where the electrodes are located on opposite walls of the channel.A microfluidi chip with a coplanar configuratio is a relatively cheap device that is easy to fabricate.However,as the cross-stream distance of the particle from the nearest electrode surface increases,the measured signal amplitude decreases,and this limits the range of particle size for a particular channel configuration in a channel with a height of tens of micrometers,the minimum particle size that can be detected is~2μm.By contrast,with parallel facing electrodes,the impedance depends on only the perpendicular distance of the particle from the nearest electrode surface,increasing as that distance decreases.However,the fabrication procedure is more complicated because it requires accurate alignment of the electrodes on opposite walls.13

    Recently,the coplanar configuratio was used to measure the diameters of submicrometer particles.1,14–16De Ninno et al.14demonstrated accurate size determination of particles with diameters of 5.2μm and above by using fiv electrodes in a microchannel with cross-sectional dimensions of 40μm(width)×21μm(height)and with the addition of floatin electrodes;the impedance signal involved an additional parameter known as relative prominence,which was used to correct the particle diameter with a resolution of 0.3μm,where the resolution is define as the difference in particle size corresponding to twice the standard deviation of the distribution.Using double differential electrodes in the coplanar configuration Zhong et al.15measured particle size with a resolution of 0.1μm,but the channel dimensions were restricted to 8μm(width)×10μm(height),and such small channels are susceptible to clogging.Several studies have used parallel electrodes to determine particle diameters,1,17–19and in most cases particle diameters of 3–10μm were studied with a resolution of 1μm.

    Irrespective of the electrode configuratio in the microchannel,the differential signal amplitude generated by a particle as it passes the electrodes depends on its perpendicular distance from them.Known as position dependency,this phenomenon results in significan errors in diameter measurements,and various approaches have been used to understand the position dependency of the signal amplitude and correct for it.14,16,17Spencer et al.17measured both transverse and oblique signal amplitudes simultaneously by using multiple pairs of electrodes to determine the transit times of particles,and the relative values of these transit times were used to correct for the position dependency in the direction perpendicular to the electrode surface.The signals measured using three pairs of coplanar electrodes were fitte using a bipolar Gaussian equation,and shape parameters obtained from this fi varied with both particle position and diameter,allowing correction for position dependency.However,because the coplanar configuratio is more sensitive to measurement inaccuracy due to position dependency,using it to quantify smaller particle diameters is considered infeasible.

    Herein,we propose a method involving three pairs of parallel facing electrodes to improve the estimation of particle diameter.By extracting both peak amplitude and travel time from the measured signal traces and the theoretically simulated curves,this method provides a simple strategy for correcting for particle position dependency.The novelty of the present work stems from the choice of travel time,for which a general strategy is introduced.17,20

    FIG.1.Complete setup for impedance measurements of beads suspended in phosphate-buffered saline solution.

    II.EXPERIMENTAL METHODS

    The experimental setup is shown schematically in Fig.1(a).A microfluidi chip with dimensions of 5.4 mm×5.4 mm with parallel facing electrodes was fabricated on a two-inch wafer using our in-house facilities.The top and bottom walls for the electrodes were made of Si/SiO2and glass,respectively.As shown in Fig.1(b),the configuratio comprised a microfluidi channel with cross-sectional dimensions of 50μm(width)×30μm(height).On the top and bottom walls of the channel were three pairs of electrodes,each electrode with a width of 30μm and separated by an edge-to-edge distance of 30μm.The electrical circuit is shown in Fig.1(a),and the center electrodes on the top and bottom walls were grounded.

    FIG.2.(a)Channelconfiguration showing(i)particle traveling in channelwith cross-sectionaldimensions of 30μm×50μm,(ii)top and bottom electrodes,(iii)applied voltage,and(iv)outputcurrent.(b)Mesh used in typical3D COMSOL simulations for a particle with a diameter of3μm,highlighting the higher meshing density atthe electrodes.(c)and(d)Potentialand currentdensity distributions in xz plane of channelwith an applied voltage of1 V.

    The Cr/Pt/Au electrodes(20 nm/150 nm/100 nm)were deposited on both the Si/SiO2and glass substrates,and the desired electrode pattern as shown in Fig.1(b)was fabricated using optical photolithography and ion milling.The channel fabrication and bonding were done using a photoimageable resist(Perminex 2015,USA)and a bonding chuck as per the manufacturer data sheet.See our previous work21for the details of fabricating the sensor.Using an existing semiconductor facility,the electrical contacts were wirebonded to a printed circuit board and connected to male header pins for direct coupling with the female sockets;this was possible by an innovative change in our design,with the top electrode contacts accessible through the bottom electrodes by fillin the gap with fin solder paste(Chip Quik-SMD291SNL T7,USA).Side holes for flui access were injected with transparent flui connectors that were glued using UV glue(Bondic,USA)directly on the chip.A photograph of the fabricated lab-on-a-chip is shown in Fig.1(c).The inlet of the flui connector was then connected to a syringe pump via a Teflo tube(inner diameter:0.3 mm;outer diameter:0.8 mm).

    Polystyrene beads with a density of 1.05 g/cm3,a DC conductivity of 6.6×10?4S/m,and diameters of 2μm,3μm,4μm,5μm,and 6μm(Lab261,1%solid)were used for the measurements.They were suspended in phosphate-buffered saline(PBS)with a DC conductivity of 1.6 S/m and diluted to a concentration of~200 beads per microliter.The carrier signal applied to the top electrodes had a frequency of 1.8 MHz and a peak-to-peak amplitude of 7.5 V.The currents from the bottom left and right electrodes were converted to a voltage difference using transimpedance operational amplifier(THS 4303)in which the inverting inputs were grounded.The outputs of the two transimpedance amplifier were connected to the input of a differential amplifie(ADA4927),which provided the voltage difference between the two electrodes.At constant voltage,the current is inversely proportional to the impedance,and the results herein are expressed in terms of the impedance.As a cell passes sequentially past the left and right electrodes,the difference in impedance is positive when the particle is between the upstream pair of electrodes and then becomes negative when it passes through the downstream pair.

    The impedance measurements of the fast-moving cells suspended in the PBS solution required a high sampling rate.The fastest-moving cells in the channel with cross-sectional dimensions of 50μm×30μm at a flo rate of 30μl/min travelled at 0.7 m/s,and the time to cross the electrode zone of length 150μm was~0.2 ms.Therefore,a sampling rate of 350 000 samples per second provided~75 measurements within the time taken for a cell to cross the electrode zone,which was sufficien to capture the signal with no distortion.The current lock-in amplifie designed by MicroX could generate three different frequencies in a single channel ranging from 10 kHz to 15 MHz and also had the in-built feature of real-time data processing for event counting by a custom-made algorithm programmed in the digital signal processor integrated in the lock-in board.The demodulated data from the lock-in amplifie were transferred to computer memory through a universal serial bus and were recorded in binary format.The frequency generation,low-pass-filte cut-off frequency,and sampling rate were controlled using an application programming interface made for a Linux-based system.The front-end electronics shown in Fig.1 had precision resistors of 1 kΩ with a tolerance of less than 0.05%,and these were used to set the required gain of the differential amplifier

    FIG.3.Differential currentΔI versus particle position along x direction of microchannelfor different z positions perpendicular to electrodes.The insetshows the maximum impedance at different z locations scaled by the impedance at the center ofthe channelat z=0.

    FIG.4.(a)Velocity profile for a 30μm(height)×50μm(width)channelobtained from COMSOL simulation(verified using analyticalsolution).(b)Normalized impedance as a function of velocity normalized with maximum velocity at y=z=0.

    III.THEORETICAL ANALYSIS

    Electric-fiel simulations were carried out using the COMSOL 5.6 finite-elemen solver,and the analytical solution for a single pair of electrode was reported earlier.22The potential V in the flui and particle is governed by the Laplace equation?2V=0.The boundary condition of zero potential gradient was applied at the top and bottom surfaces of the channel,i.e.,n??V=0,because these surfaces are considered to be insulating;here,n is the unit normal to the surface.The boundary condition of constant potential was applied at the electrode surfaces.Because the boundary condition changes discontinuously at the junction between a bare channel wall and an electrode deposited thereon,the fiel lines have large curvature near the electrodes,and because of this,it was necessary to refin the mesh near the electrodes as shown in Fig.2(b).In total,2×106mesh points were used in a typical simulation configuratio of the type shown in Fig.2(b).Distributions of potential and current density across the channel with an applied voltage of 1 V are shown in Fig.2(c).

    Simulation results for the change in current due to a particle with a diameter of 3μm are shown in Fig.3,where the impedance is shown as a function of downstream distance at different z locations in the direction perpendicular to the top and bottom walls.Even for the same particle size,the impedance curves differ significantly in shape depending on the z position of the particle.In each case,there are at least four extrema as the particle passes through the channel.When it passes through the center of the channel,the current changes relatively smoothly and can be fitte well by Gaussian curves.When the particle is close to a wall,there are two maxima in the current when the particle crosses the left and right edges of each electrode,and the amplitude deviates from single Gaussian behavior with two sharp peaks near the edges of the electrodes.17

    To understand the position dependency of the peak amplitude,the maximum from each set of curves for different z values was extracted and then normalized by the impedance at the center of the channel at z=0[Z/Z0∝ΔImax(z)/ΔImax(0)].Here,the maximum of the time trace of the impedance is used for bimodal curves,and Z0is the maximum impedance measured at z=0.The fitte curve in the inset of Fig.3(b)shows that the maximum amplitude changes exponentially as a function of vertical position.

    A series solution can be obtained for the flui velocity profil for the flo in a rectangular channel by using separation of variables.The velocity depends on the pressure gradient dp/dx along the channel:23

    where?h/2≤z≤h/2 is the direction perpendicular to the electrodes,and?w/2≤y≤w/2 is that perpendicular to the flo direction and parallel to the electrodes.The maximum velocity at the center of the channel at y=z=0 is

    and the average velocity vavis

    Although the velocity profil in 3D space is an inverted paraboloid(not shown here),for y=0 it becomes a parabola as shown in Fig.4(a),which can be fitte using

    At the experimental flo rate of 30μl/min,the value of vmaxat the center was 0.68 m/s,and for this maximum velocity,the coefficient a1,a2,and a3for the parabolic fi in Fig.4(a)are 0.68 m/s,?1.13×10?5s?1,and?0.003 m?1s?1,respectively.

    Because the particle position and size are unknown,the velocity is used as a parameter for correcting the particle size.The velocity is inferred from the time taken by the particle to travel from one electrode pair to the next.Although the maximum impedance is independent of the y position,the maximum velocity in a horizontal plane at a fixe z location does depend on the horizontal cross-stream coordinate y.Here,the maximum velocity vmaxat the channel center at y=z=0 is used as the normalizing factor.

    The variation of Z/Z0versus vx/vmaxat y=0 is shown in Fig.4(b),and the following provides a good fi for this curve:

    where d is the corrected diameter,D is the electrical diameter,and c1,c2,and c3are the fittin constants.The curve fittin was done using cftool in MATLAB,using the linear least-squares method known as the Levenberg–Marquardt algorithm to obtain a minimum R2=99%.

    For the present experimental system,the electrical diameter D is define as

    where G is the amplificatio factor to account for the currentto-voltage converter board and other electronic peripherals to be calibrated from experimental data.The value of G was estimated as 73.5μm(nA)(?1/3)using the assumption that the true diameter and the electrical diameter are equal for particles moving at the center of the channel with maximum velocity.Equation(5)allows the corrected diameter to be determined once the values of D and the particle velocity vxare known.

    FIG.6.(a)Scatter plotof velocity versus electricaldiameter for 2-μm particles from separate runs,showing clearly the instrument resolution of up to 1.7μm.(b)Histogram of corrected diameter of 2-μm particles.(c)and(d)Color density plots of velocity versus electricaland corrected diameter,respectively.

    IV.RESULTS AND DISCUSSION

    Figure 5 shows time traces of impedance measurements with sampling at intervals of 3.3μs for particles with a diameter of 3μm.Figure 5(a)shows that the peak amplitude of Z/Z0is in the range of 1%–1.75%,which corresponds to variation of the electrical diameter D by 20%.This agrees well with the theoretical prediction as shown in the inset of Fig.3.Figures 5(b)–5(e)show expanded time traces of the impedance as a single particle travels through the channel.The velocities inferred from the time interval between two extrema are 0.54 m/s,0.49 m/s,0.36 m/s,and 0.28 m/s,respectively,and the values of the electrical diameter D derived using Eq.(5)are 3.05μm,3.35μm,3.45μm,and 3.71μm,respectively.From these figures it is clear that the shapes of the impedance trajectories of particles moving with different velocities corresponding to different z positions agree qualitatively with the theoretical predictions shown in Fig.3.As the velocity decreases,the particles are located closer to the electrodes,and the impedance trajectory has a bimodal Gaussian shape.

    FIG.7.Separate runs with different particle sizes:(a)and(b)electrical-diameter histogram for particles ofdifferent sizes,and corresponding correction in histogram shape;(c)and(d)scatter plot of velocity versus electricaldiameter for different particle sizes,and corresponding change in plot due to size correction.

    TABLE I.Parameters used to fitdata in Figs.5 and 6.

    Figure 6(a)is a scatter plot of velocity versus electrical diameter D for particles with a diameter of 2μm.This shows clearly that the instrument resolution was~1.7μm,and there was electrical noise for electrical diameters below this value.The maximum measured particle velocity vmaxwas~0.67 m/s,which is very close to the theoretically predicted value of~0.68 m/s for the aforementioned channel dimensions and flo rate.The increase of D signifie the position dependency of the particle in the z direction as shown in Fig.6(c).Using Eq.(5),the corrected particle diameter d was derived for the whole population of data;the corresponding histogram plot is shown in Fig.6(b),and a color density plot of velocity versus d is shown in Fig.6(d).These show a significan narrowing of the distribution and a reduction in the coefficien of variation,with the particle size now independent of velocity.The values of the coeffi cients c1,c2,and c3used in Eq.(5)were obtained from the theoretical results given in Table I.

    The density of sampling points is shown in the electricaldiameter–velocity plane in Fig.7 for particles with average diameters of~3.0μm,4.0μm,5.0μm,and 6.0μm.For all diameters,the density plot in particle-diameter–velocity space[Fig.7(a)]shows a systematic variation of the electrical diameter with velocity,which is a consequence of the position dependence of the impedance in the z direction.Applying Eq.(5)to all the individual populations reduces the standard deviation in the distributions for all sizes,and the particle size is now independent of velocity.Figure 7(b)shows that the histogram of the uncorrected diameter is an asymmetric Gaussian distribution with a relatively large standard deviation.When Eq.(5)is applied,the distribution shown in Fig.7(d)is closer to a symmetric Gaussian with a smaller standard deviation.From these plots,it is inferred that the correlation used to correct the electrical diameter works effectively for all particle sizes with the same fittin constants.

    Figure 8 shows similar plots for experiments with a mixture of beads.Figure 8(a)shows that regardless of particle size,the ratio D/d varies in a similar manner with velocity.Figure 8(b)shows that the histogram of the uncorrected diameter is an asymmetric Gaussian distribution and for each particle diameter,the distribution width is large.The electrical diameter was corrected using Eq.(5)with the same values for c1,c2,and c3as those in the simulations for all particle sizes.There is a significan reduction in the spread of the histograms for particles of different sizes as shown in Figs.8(c)and 8(d),and the comparative results are provided in Table II for different particle sizes.From Table II,for the separate runs it is clear that there was a reduction in the standard deviation by a factor of two irrespective of particle diameter,which indicates the validity of the empirical relationship in Eq.(5).The experimental data from the mixed runs show again the applicability of Eq.(5),where both the standard deviation and coefficien of variation(CV)were found to be of the same orders of magnitude as those estimated from the separate runs.The estimated CV for each particle diameter was found to be very close to that from the manufacturer;the exception was that for the diameter of 3μm,for which the CV provided by the manufacturer is very low compared to those for the other diameters.

    FIG.8.Mixed runs with different particle sizes:(a)and(b)electric-diameter histogram,and corresponding correction in histogram shape;(c)and(d)scatter plot of velocity versus electricaldiameter for mixed runs,and corresponding plot with correction.

    TABLE II.Comparison between statisticaldata provided by manufacturer and those obtained before and after corrections were applied to the experimentaldata.

    V.CONCLUSION

    An important difficult in determining the diameter of a particle of micrometer size by using impedance measurements in a microchannel is the uncertainty about its location in the channel cross section.For a particle of given size,the measured impedance is higher if the particle is closer to the electrodes and lower when it is in the center of the channel.Because of this,the electrical diameter inferred from impedance measurements has a significan error when compared to the true diameter.Herein,a procedure was formulated for correcting the electrical diameter to obtain the true diameter.The particle velocity is used as a single parameter to represent the variation in position in the cross section,and the ratio of the electrical and true diameters is expressed as a three-parameter fi[Eq.(5)].Although the velocity depends on the two cross-stream coordinates whereas the impedance correction depends on only the coordinate perpendicular to the electrodes,we found a very good correlation between the velocity and the correction to the diameter using the three-parameter fit There is a reduction by a factor of two in the standard deviation of the measured particle diameter by using this empirical correlation.

    Surprisingly,we also found that the coefficient in Eq.(5)are independent of the particle diameter.The same coefficient were found to minimize the standard deviation in particle size for particles with diameters of 2–8μm.This indicates that the coefficient are functions of only the channel and electrode configuration and do not depend on the size of the particle in the channel.Therefore,once the coefficient have been calibrated for one particle size for a given channel configuration they can be used for other particle sizes.This provides a robust method for correcting the particle size for variations in cross-sectional location for samples in which the particle size distribution is discrete,so that the spread in the size distribution is smaller than the difference in the discrete size ranges.A more sophisticated procedure is being developed for size correction when there is a continuous size distribution.

    ACKNOWLEDGMENTS

    The authors would like to thank(i)the Director of CSIR–NAL for supporting this activity and(ii)the Polish grant committee for funding the“Bridge Alpha”grant for this project.

    AUTHOR DECLARATIONS

    Conflict of Interest

    The authors have no conflict to disclose.

    DATA AVAILABILITY

    The data that support the finding of this study are available from the corresponding author upon reasonable request.

    在线免费观看的www视频| 老司机午夜福利在线观看视频| 很黄的视频免费| 蜜桃久久精品国产亚洲av| 国产在视频线在精品| 在线播放国产精品三级| 看黄色毛片网站| 麻豆一二三区av精品| 成人高潮视频无遮挡免费网站| 少妇熟女aⅴ在线视频| 国产伦在线观看视频一区| 黄色一级大片看看| 国内少妇人妻偷人精品xxx网站| 三级国产精品欧美在线观看| 一级黄色大片毛片| 亚洲专区国产一区二区| 老熟妇乱子伦视频在线观看| 欧美激情久久久久久爽电影| 两个人的视频大全免费| 国产爱豆传媒在线观看| 欧美日本亚洲视频在线播放| 亚洲片人在线观看| 十八禁网站免费在线| 国产精品日韩av在线免费观看| 听说在线观看完整版免费高清| 免费搜索国产男女视频| a在线观看视频网站| 51国产日韩欧美| 国产精品久久久久久久久免 | 欧洲精品卡2卡3卡4卡5卡区| 看免费av毛片| 在线观看av片永久免费下载| 国产一区二区三区在线臀色熟女| 久久草成人影院| 欧美日韩国产亚洲二区| www日本黄色视频网| a级毛片免费高清观看在线播放| 在现免费观看毛片| 国产伦精品一区二区三区四那| 国产一区二区三区视频了| 欧美午夜高清在线| 美女cb高潮喷水在线观看| 99久久成人亚洲精品观看| 少妇高潮的动态图| 国内揄拍国产精品人妻在线| 免费观看人在逋| 男人的好看免费观看在线视频| 欧美日韩福利视频一区二区| 色综合欧美亚洲国产小说| 午夜视频国产福利| 日本与韩国留学比较| 欧美性猛交黑人性爽| 99精品久久久久人妻精品| 淫妇啪啪啪对白视频| 亚洲人成网站在线播放欧美日韩| 久久中文看片网| 97热精品久久久久久| 别揉我奶头 嗯啊视频| 国产精品国产高清国产av| 亚洲欧美日韩无卡精品| 观看美女的网站| 亚洲一区高清亚洲精品| 黄色配什么色好看| 一进一出好大好爽视频| 好男人在线观看高清免费视频| 真实男女啪啪啪动态图| 亚洲中文字幕日韩| 国产一区二区激情短视频| 成人av在线播放网站| 成人欧美大片| 国产av在哪里看| 听说在线观看完整版免费高清| 国产日本99.免费观看| 国产精品人妻久久久久久| 成人国产一区最新在线观看| 观看美女的网站| 亚洲精品456在线播放app | 精品人妻1区二区| 最近视频中文字幕2019在线8| 中文字幕精品亚洲无线码一区| 真人做人爱边吃奶动态| 精品久久久久久,| 搡老妇女老女人老熟妇| 精品人妻一区二区三区麻豆 | 国内久久婷婷六月综合欲色啪| 国产乱人伦免费视频| 精品国产三级普通话版| 免费观看的影片在线观看| 精品久久久久久久久久免费视频| 精品国产亚洲在线| 久久久久久大精品| 欧美色欧美亚洲另类二区| 99久久99久久久精品蜜桃| 精品人妻1区二区| 日本免费a在线| 男人狂女人下面高潮的视频| 成年版毛片免费区| 男女床上黄色一级片免费看| 精品无人区乱码1区二区| 在线播放无遮挡| 精品人妻1区二区| 超碰av人人做人人爽久久| 一级av片app| 伊人久久精品亚洲午夜| 人妻制服诱惑在线中文字幕| 国产三级中文精品| 女人十人毛片免费观看3o分钟| 蜜桃亚洲精品一区二区三区| 亚洲中文字幕日韩| 国产精品av视频在线免费观看| 日本 欧美在线| 国产在视频线在精品| 永久网站在线| 国产一区二区三区在线臀色熟女| 一卡2卡三卡四卡精品乱码亚洲| 久久久久久久久久成人| 国产不卡一卡二| 91久久精品国产一区二区成人| 欧洲精品卡2卡3卡4卡5卡区| 日韩欧美 国产精品| 国产一区二区在线av高清观看| 别揉我奶头~嗯~啊~动态视频| 小蜜桃在线观看免费完整版高清| 欧美在线黄色| 国产亚洲av嫩草精品影院| 啦啦啦观看免费观看视频高清| 国产午夜福利久久久久久| 一二三四社区在线视频社区8| 99在线视频只有这里精品首页| 久久久久久久亚洲中文字幕 | 亚洲成a人片在线一区二区| 中文资源天堂在线| av国产免费在线观看| 女人十人毛片免费观看3o分钟| 色尼玛亚洲综合影院| 国产视频一区二区在线看| 男女床上黄色一级片免费看| 色视频www国产| 国产精品99久久久久久久久| 国产高清激情床上av| 欧美成人一区二区免费高清观看| 日日夜夜操网爽| 精品人妻偷拍中文字幕| 免费大片18禁| 久久国产精品人妻蜜桃| 日本免费一区二区三区高清不卡| 一个人免费在线观看电影| 99热这里只有精品一区| 一进一出抽搐gif免费好疼| 中文字幕熟女人妻在线| 久久久久精品国产欧美久久久| 午夜福利在线观看免费完整高清在 | 久久亚洲真实| 麻豆一二三区av精品| 禁无遮挡网站| 亚洲不卡免费看| 色视频www国产| 老司机午夜十八禁免费视频| 国产精品一区二区免费欧美| 国产毛片a区久久久久| 91午夜精品亚洲一区二区三区 | 亚洲成人中文字幕在线播放| 国产国拍精品亚洲av在线观看| 一夜夜www| 亚洲成av人片免费观看| 免费黄网站久久成人精品 | 最近视频中文字幕2019在线8| 1024手机看黄色片| 国产在视频线在精品| 亚洲精品成人久久久久久| 宅男免费午夜| 夜夜爽天天搞| 成年女人毛片免费观看观看9| 国产av麻豆久久久久久久| 999久久久精品免费观看国产| 久久久成人免费电影| 国产欧美日韩一区二区精品| 99热这里只有是精品50| 亚洲经典国产精华液单 | 高清毛片免费观看视频网站| av在线天堂中文字幕| 亚洲五月婷婷丁香| 日本三级黄在线观看| 免费一级毛片在线播放高清视频| 免费黄网站久久成人精品 | www日本黄色视频网| 国产高清视频在线观看网站| 国产黄色小视频在线观看| 免费在线观看亚洲国产| 中文字幕高清在线视频| 成年女人看的毛片在线观看| 精品人妻视频免费看| 性色av乱码一区二区三区2| 桃色一区二区三区在线观看| 真人做人爱边吃奶动态| 中文字幕av成人在线电影| 免费高清视频大片| 99久久无色码亚洲精品果冻| 又爽又黄无遮挡网站| 国产欧美日韩精品一区二区| 亚洲成人久久性| 99热只有精品国产| 久久久久九九精品影院| 精品人妻一区二区三区麻豆 | 午夜亚洲福利在线播放| 国产色婷婷99| 精品国产三级普通话版| 欧美3d第一页| 国内少妇人妻偷人精品xxx网站| 亚洲av熟女| 天堂网av新在线| 精品人妻视频免费看| 久久午夜亚洲精品久久| 国产精品98久久久久久宅男小说| 国产精品自产拍在线观看55亚洲| 十八禁国产超污无遮挡网站| 国产真实伦视频高清在线观看 | 亚洲人成网站高清观看| 精品国产三级普通话版| 一进一出抽搐动态| 动漫黄色视频在线观看| 观看美女的网站| 国产色婷婷99| 高清日韩中文字幕在线| x7x7x7水蜜桃| 午夜福利在线观看免费完整高清在 | 蜜桃亚洲精品一区二区三区| 97碰自拍视频| 麻豆国产av国片精品| 亚洲最大成人中文| 国产一区二区三区在线臀色熟女| 啦啦啦观看免费观看视频高清| 性插视频无遮挡在线免费观看| www.熟女人妻精品国产| 欧美激情久久久久久爽电影| 欧美日本视频| 成年女人看的毛片在线观看| 亚洲无线在线观看| 色综合欧美亚洲国产小说| 午夜两性在线视频| 宅男免费午夜| 三级毛片av免费| 亚洲,欧美,日韩| 日韩中字成人| 中亚洲国语对白在线视频| 亚洲av成人精品一区久久| 2021天堂中文幕一二区在线观| 色播亚洲综合网| 最近最新中文字幕大全电影3| 日韩人妻高清精品专区| 精品午夜福利在线看| www.www免费av| 午夜视频国产福利| 亚洲国产精品sss在线观看| 国模一区二区三区四区视频| 午夜视频国产福利| 亚洲自拍偷在线| 亚洲av熟女| 中文在线观看免费www的网站| 欧美日韩综合久久久久久 | 麻豆成人av在线观看| 欧美黑人欧美精品刺激| aaaaa片日本免费| 亚洲欧美日韩东京热| 午夜免费成人在线视频| 欧美成人性av电影在线观看| 亚洲男人的天堂狠狠| 熟女电影av网| 成人特级av手机在线观看| 精品无人区乱码1区二区| 亚洲午夜理论影院| 久久精品久久久久久噜噜老黄 | 有码 亚洲区| 欧美3d第一页| 白带黄色成豆腐渣| 色吧在线观看| av女优亚洲男人天堂| 中文资源天堂在线| 久久精品久久久久久噜噜老黄 | a级毛片a级免费在线| 久久草成人影院| 女同久久另类99精品国产91| 国产国拍精品亚洲av在线观看| 国产人妻一区二区三区在| 午夜免费男女啪啪视频观看 | 国产亚洲精品综合一区在线观看| 欧美又色又爽又黄视频| 天美传媒精品一区二区| www.熟女人妻精品国产| 三级国产精品欧美在线观看| 噜噜噜噜噜久久久久久91| aaaaa片日本免费| 欧美日韩黄片免| 男女视频在线观看网站免费| 99久久九九国产精品国产免费| 舔av片在线| 99精品在免费线老司机午夜| 十八禁网站免费在线| 精品久久久久久久久久久久久| 美女免费视频网站| 女人被狂操c到高潮| 亚洲精品亚洲一区二区| 无人区码免费观看不卡| 亚洲精品在线美女| 俺也久久电影网| 免费无遮挡裸体视频| 国内毛片毛片毛片毛片毛片| 国产亚洲精品av在线| 亚洲国产欧美人成| 国产成人a区在线观看| 婷婷六月久久综合丁香| 99热这里只有精品一区| 精品久久久久久久久亚洲 | 搡老熟女国产l中国老女人| 国产一区二区亚洲精品在线观看| 日日夜夜操网爽| 亚洲av免费在线观看| 久久精品国产亚洲av天美| 亚洲乱码一区二区免费版| 日韩有码中文字幕| 亚洲成人久久性| 毛片女人毛片| 亚洲中文字幕日韩| 免费观看的影片在线观看| .国产精品久久| 亚洲国产精品成人综合色| 亚洲一区二区三区不卡视频| 偷拍熟女少妇极品色| 在线国产一区二区在线| 亚洲狠狠婷婷综合久久图片| 亚洲成av人片在线播放无| 婷婷六月久久综合丁香| x7x7x7水蜜桃| 午夜福利高清视频| 亚洲欧美日韩高清专用| 少妇熟女aⅴ在线视频| 有码 亚洲区| 国产探花在线观看一区二区| 精品久久国产蜜桃| 久久久精品大字幕| 亚洲av成人av| 欧美最黄视频在线播放免费| 精品久久国产蜜桃| 免费观看精品视频网站| 十八禁网站免费在线| 国产成人a区在线观看| 在线免费观看的www视频| 级片在线观看| 国产精品免费一区二区三区在线| 观看免费一级毛片| 国产精品一区二区三区四区久久| 国产色婷婷99| 久久久久久国产a免费观看| 美女被艹到高潮喷水动态| 十八禁网站免费在线| 亚洲欧美日韩高清在线视频| 波多野结衣高清作品| 亚洲人成网站高清观看| 国产av不卡久久| 午夜精品一区二区三区免费看| 深夜精品福利| 人妻丰满熟妇av一区二区三区| 美女黄网站色视频| 久久国产精品人妻蜜桃| 亚洲国产欧洲综合997久久,| 欧美潮喷喷水| 久久久久久大精品| 国产伦一二天堂av在线观看| 亚洲七黄色美女视频| 成人高潮视频无遮挡免费网站| 日本a在线网址| 国内精品一区二区在线观看| 99国产精品一区二区蜜桃av| h日本视频在线播放| 亚洲七黄色美女视频| 久久精品综合一区二区三区| 国产精品电影一区二区三区| a级毛片免费高清观看在线播放| 丁香六月欧美| av在线天堂中文字幕| 中文资源天堂在线| 日日干狠狠操夜夜爽| 757午夜福利合集在线观看| 国产黄色小视频在线观看| 国产精品乱码一区二三区的特点| 69av精品久久久久久| 国产精品国产高清国产av| 啪啪无遮挡十八禁网站| 欧美区成人在线视频| 十八禁人妻一区二区| 亚洲精品在线美女| 久99久视频精品免费| 又粗又爽又猛毛片免费看| 日韩 亚洲 欧美在线| 亚洲精品亚洲一区二区| 久久久国产成人精品二区| 欧美在线黄色| 级片在线观看| 国产国拍精品亚洲av在线观看| 国产欧美日韩精品一区二区| 听说在线观看完整版免费高清| 国产成+人综合+亚洲专区| 国产黄色小视频在线观看| 又黄又爽又免费观看的视频| 日韩欧美精品免费久久 | 国产精品美女特级片免费视频播放器| av天堂在线播放| 欧美丝袜亚洲另类 | 深爱激情五月婷婷| 在线十欧美十亚洲十日本专区| 91久久精品国产一区二区成人| 亚洲精品一卡2卡三卡4卡5卡| 91久久精品国产一区二区成人| 三级男女做爰猛烈吃奶摸视频| 国产成人影院久久av| 永久网站在线| 国内精品久久久久久久电影| 美女免费视频网站| 午夜久久久久精精品| 波多野结衣高清作品| 真人做人爱边吃奶动态| 天美传媒精品一区二区| 在线观看66精品国产| 午夜免费激情av| 99久久久亚洲精品蜜臀av| 三级毛片av免费| 色在线成人网| 蜜桃久久精品国产亚洲av| 国产精品一及| 真人做人爱边吃奶动态| 观看美女的网站| 白带黄色成豆腐渣| 成年人黄色毛片网站| 欧美一区二区精品小视频在线| 老司机午夜十八禁免费视频| 精华霜和精华液先用哪个| 亚洲男人的天堂狠狠| 国产精品,欧美在线| 成人一区二区视频在线观看| 我要搜黄色片| 又紧又爽又黄一区二区| 国产人妻一区二区三区在| 国产男靠女视频免费网站| 精品日产1卡2卡| 国产高清激情床上av| 久久久久久久亚洲中文字幕 | 白带黄色成豆腐渣| 亚洲成av人片免费观看| 国产精品久久久久久亚洲av鲁大| 久久草成人影院| 久久久成人免费电影| 夜夜躁狠狠躁天天躁| 淫秽高清视频在线观看| 男女下面进入的视频免费午夜| 国产成人aa在线观看| 免费看a级黄色片| 欧美日韩亚洲国产一区二区在线观看| 99热只有精品国产| 99久久99久久久精品蜜桃| 一区二区三区四区激情视频 | 99久久精品国产亚洲精品| 精华霜和精华液先用哪个| 级片在线观看| 麻豆国产97在线/欧美| 亚洲一区二区三区不卡视频| 亚洲精华国产精华精| 永久网站在线| 国产亚洲欧美98| 亚洲美女视频黄频| 亚洲在线自拍视频| 黄色视频,在线免费观看| 成年版毛片免费区| 亚洲国产色片| 亚洲国产精品sss在线观看| 高清在线国产一区| 精品久久久久久久久av| 久久精品国产亚洲av涩爱 | 欧美日本视频| 18禁在线播放成人免费| 好看av亚洲va欧美ⅴa在| 超碰av人人做人人爽久久| 国产日本99.免费观看| 日韩精品中文字幕看吧| av视频在线观看入口| 国产伦人伦偷精品视频| 欧美激情久久久久久爽电影| 午夜福利在线观看吧| 日本成人三级电影网站| 精品久久国产蜜桃| 人人妻人人看人人澡| 亚洲国产欧美人成| 亚洲午夜理论影院| 男女之事视频高清在线观看| 白带黄色成豆腐渣| 欧美成人一区二区免费高清观看| 日本成人三级电影网站| 禁无遮挡网站| 免费在线观看影片大全网站| 欧美色视频一区免费| 精品不卡国产一区二区三区| 国产精华一区二区三区| 亚洲中文字幕一区二区三区有码在线看| 亚洲成av人片免费观看| 12—13女人毛片做爰片一| 在现免费观看毛片| 欧美日韩黄片免| 久久久色成人| 日本成人三级电影网站| 97热精品久久久久久| 国产精品98久久久久久宅男小说| 毛片一级片免费看久久久久 | 99久国产av精品| 国产真实伦视频高清在线观看 | 观看免费一级毛片| 免费电影在线观看免费观看| 欧美一级a爱片免费观看看| 欧美精品国产亚洲| 免费在线观看日本一区| 精品久久国产蜜桃| 少妇的逼水好多| 美女xxoo啪啪120秒动态图 | 在线观看美女被高潮喷水网站 | 男女下面进入的视频免费午夜| 亚洲人成网站在线播| 婷婷六月久久综合丁香| 精品久久久久久久久久久久久| 亚洲激情在线av| 尤物成人国产欧美一区二区三区| 国产私拍福利视频在线观看| 国产高清三级在线| АⅤ资源中文在线天堂| 国产欧美日韩精品一区二区| 色综合站精品国产| 色综合婷婷激情| 欧美性猛交╳xxx乱大交人| 亚洲一区二区三区不卡视频| 亚洲欧美日韩卡通动漫| 在线a可以看的网站| 午夜福利视频1000在线观看| 我的老师免费观看完整版| 无人区码免费观看不卡| 村上凉子中文字幕在线| 校园春色视频在线观看| 国产精品久久久久久精品电影| 亚洲成人久久爱视频| 又黄又爽又免费观看的视频| 亚洲欧美日韩卡通动漫| 精品不卡国产一区二区三区| 久久欧美精品欧美久久欧美| 1024手机看黄色片| 国语自产精品视频在线第100页| 亚洲aⅴ乱码一区二区在线播放| 99久久九九国产精品国产免费| 丝袜美腿在线中文| 中文亚洲av片在线观看爽| 国产亚洲精品久久久com| 国产免费一级a男人的天堂| 99在线视频只有这里精品首页| av在线蜜桃| 非洲黑人性xxxx精品又粗又长| 亚洲真实伦在线观看| 老司机福利观看| 男女做爰动态图高潮gif福利片| 国产aⅴ精品一区二区三区波| 久久人人爽人人爽人人片va | 99视频精品全部免费 在线| 国产精品嫩草影院av在线观看 | h日本视频在线播放| 国产精品久久视频播放| 精品一区二区三区视频在线观看免费| 简卡轻食公司| 国产午夜精品久久久久久一区二区三区 | 夜夜躁狠狠躁天天躁| 久久精品综合一区二区三区| 国产伦一二天堂av在线观看| 欧美zozozo另类| 国产老妇女一区| 别揉我奶头~嗯~啊~动态视频| 免费看a级黄色片| 内地一区二区视频在线| 亚洲av美国av| 波多野结衣巨乳人妻| 日本免费a在线| 免费无遮挡裸体视频| 久久性视频一级片| 日韩欧美国产一区二区入口| 女人被狂操c到高潮| 香蕉av资源在线| 在线国产一区二区在线| 一区二区三区四区激情视频 | 成人性生交大片免费视频hd| 搡女人真爽免费视频火全软件 | 久久精品国产99精品国产亚洲性色| 亚洲av成人av| .国产精品久久| 久久久国产成人精品二区| 亚洲内射少妇av| 免费av不卡在线播放| 日本黄色片子视频| АⅤ资源中文在线天堂| av天堂中文字幕网| 性欧美人与动物交配| 国产三级黄色录像| 全区人妻精品视频| 成人亚洲精品av一区二区| a级毛片a级免费在线| 精品国产亚洲在线| 欧美高清性xxxxhd video| 俄罗斯特黄特色一大片| 亚洲成人免费电影在线观看| 欧美成人免费av一区二区三区| a级毛片a级免费在线| 日本黄大片高清| 亚洲人成网站在线播| 欧美一区二区亚洲|