    A joint computational-experimental study of intracranial aneurysms: Importance of the aspect ratio*

    2016-10-18 05:36:46SimonSuiManLAIAbrahamYikSauTANGAndersonChunOnTSANGGilbertoKaKitLEUNGAlfredCheukHangYUKwokWingCHOW

    Simon Sui-Man LAI, Abraham Yik-Sau TANG, Anderson Chun-On TSANG, Gilberto Ka-Kit LEUNG,Alfred Cheuk-Hang YU, Kwok Wing CHOW

    1. Department of Electrical and Electronic Engineering, University of Hong Kong, Hong Kong, China,

    E-mail: simonlai@hotmail.com.hk

    2. Department of Mechanical Engineering, University of Hong Kong, Hong Kong, China

    3. Department of Surgery, Li Ka Shing Faculty of Medicine, University of Hong Kong, Hong Kong, China

    4. Department of Electrical and Computer Engineering, University of Waterloo, Ontario, Canada

    ?

    A joint computational-experimental study of intracranial aneurysms: Importance of the aspect ratio*

    Simon Sui-Man LAI1, Abraham Yik-Sau TANG2, Anderson Chun-On TSANG3, Gilberto Ka-Kit LEUNG3,Alfred Cheuk-Hang YU4, Kwok Wing CHOW2

    1. Department of Electrical and Electronic Engineering, University of Hong Kong, Hong Kong, China,

    E-mail: simonlai@hotmail.com.hk

    2. Department of Mechanical Engineering, University of Hong Kong, Hong Kong, China

    3. Department of Surgery, Li Ka Shing Faculty of Medicine, University of Hong Kong, Hong Kong, China

    4. Department of Electrical and Computer Engineering, University of Waterloo, Ontario, Canada

    Rupture of a cerebral aneurysm (abnormal swelling of blood vessel in the brain) will cause subarachnoid hemorrhage, and will result in an alarming rate of mortality and morbidity. A joint computational-experimental study is conducted to assess the importance of the aspect ratio in the dynamics of blood flow. The aspect ratio is defined here to be the ratio of the height of the aneurysm to the linear dimension of the neck. Idealized models of such aneurysms located near a bifurcation point were investigated. Numerical simulations for hemodynamic properties like shear stress and flow rate were performed. The computational results were verified experimentally with specially fabricated phantoms, blood mimicking fluid and Doppler ultrasound imaging. Excellent agreements were obtained. Two features are highlighted, providing information in the intensely debated link between rupture risk and geometric factors. On increasing the aspect ratios, firstly, a jet impinging on the distal part of the neck can be observed, and secondly,a region of positive shear stress gradient can be found there. Furthermore, computational analyses for four patient-specific models were conducted to correlate with the results of idealized models and to provide further clinical insight.

    intracranial aneurysms, aspect ratio, Doppler ultrasound

    Introduction

    The applications of computational and experimental techniques of fluid mechanics to study the human cardiovascular system have received intensive attention recently[1]. Arterial aneurysms are abnormal dilatations of blood vessels. Mechanisms for their generation and development are still not fully understood[2],but vascular biology and hemodynamics are believed to play crucial roles. Besides large blood vessels like the abdominal aorta and the thoracic aorta, such aneurysms also occur frequently in smaller vessels, e.g. intracranial (or cerebral) aneurysms in the circulation system in the brain[3].

    With increasingly sophisticated imaging techniques, more such cerebral aneurysms are detected before catastrophic events occur. These abnormal swellings may occur in as much as a few percent of the total population[4]. The main risk of cerebral aneurysms is the phenomenon of rupture, which leads to subarachnoid hemorrhage. The rate of mortality is alarming, and any necessary subsequent medical procedure for surviving patients poses a heavy financial burden for the society. The traditional treatment for unruptured aneurysms is open surgery, but recently endovascular methods have proven to be effective and less traumatic[5]. However, the risk of rupture needs to be balanced against the risk of treatment.

    Scientifically, the hemodynamics and properties of the vortices and impingement zone associated with these aneurysms have been investigated[6]. Among all these geometric and dynamic factors, e.g., size and location of the aneurysms, pressure, flow rate and shear stress, we focus on the aspect ratio. The aspect ratio (AR) is commonly defined to be the ratio of theaneurysm depth to the dimension of the neck, and has been demonstrated clinically to be a statistically significant quantity in terms of assessing the risk of rupture[7-9]. Indeed for irregular shaped aneurysms, a volume to ostium area ratio has also been proposed[10]. The main idea behind these proposals is that slower flow in high AR sacs might lead to various processes in vascular biology, which will weaken the aneurysm wall[11]. This conjecture is difficult to prove explicitly with the present state of scientific knowledge. Hence,in clinical experiments, elastase induced aneurysms in rabbit had been investigated[11].

    The main goal of this work is to conduct a joint computational-experimental study on intracranial aneurysms. Such swellings usually occur along the sidewall of a parent vessel (fusiform aneurysms) or near a bifurcation point (saccular aneurysms).

    Computational investigations on bifurcation aneurysms have captured the attention of researchers. The effect of aneurysm geometry, the roles of low shear stress and high pressure in aneurysm rupture have been assessed. As the AR of the aneurysm increased, a reduction in the wall shear stress and an intensified wall pressure were observed, which could lead to a higher risk of rupture[12]. After the deployment of a high porosity stent, numerical simulations of an idealized model of a basilar tip aneurysm exhibited substantial reductions in the velocity field, pressure and shear stress[13]. Furthermore, computations under a non-Newtonian fluid assumption were performed on cerebral aneurysm models with the presence of daughter saccules[14]. Complex flow patterns were observed inside the aneurysms, which might imply a greater chance of rupture. These studies demonstrate the merit of computational fluid dynamics (CFD) in analyzing the highly complex flow patterns inside the aneurysmal sac.

    On the other hand, phantom studies and enhanced imaging techniques to improve the understanding of the dynamics of intracranial aneurysms had also been carried out. Particle image velocimetry measurements of spherical aneurysm models were compared with computer simulations. Small variations in the geometric shape might produce significant changes in the hemodynamic parameters like vorticity and shear stress[15]. High frame rate angiography and computer fluid dynamics simulations had been assessed jointly,and these two approaches produced very good agreements[16].

    The idea here is to employ still another measurement technique, namely, ultrasound imaging, to study models of cerebral aneurysms. Important merits of ultrasound technology include (1) the noninvasive nature, (2) affordability, and (3) no known harmful side effects. Indeed abdominal aortic aneurysms induced in mice had been measured by ultrasound with the additional advantages of rapid imaging speed, reproducibility, and high resolution[17]. In contrast, the usage of ultrasound imaging to blood flow issues in the brain has not been extensive. One direction is to assess if vasospasm in the middle cerebral artery will reduce the flow in the corresponding extracranial internal carotid artery[18]. Indeed this potential of contrastenhanced ultrasound imaging to visualize the flow dynamics in cerebral aneurysms during neurosurgical procedure is valuable[19], and would be examined in this work with the issue of varying the aspect ratio.

    The plan of the paper can now be described. The CFD formulation, the fabrication of the phantom and experimental setup will be explained. The matching between CFD and experiments will then be discussed. Finally, conclusions will be drawn.

    Fig.1(a) Aneurysm morphology of a real patient

    Fig.1(b) Schematic diagram of the generalized model utilized in this study, with the location of different points near and inside the aneurysm (b1) Computational mesh of the numerical model for

    1. Materials and methods

    1.1 Computational fluid dynamics simulations

    1.1.1 Aneurysm models

    Both idealized and patient specific models will be investigated. The computational process was started by building the models through specialized software and would be followed by solving the equations of fluid dynamics[20-22]. For the idealized case, a sequence of three-dimensional intracranial aneurysm models was created via the computer aided design (CAD)software SOLIDWORKS 2013 (Dassault Systèmes,Waltham, MA, USA). These idealized models were based on the computed tomography (CT) data of a patient with a cerebral aneurysm located near an arterial bifurcation (Fig.1(a)). The effect of the aspect ratio will be elucidated quantitatively by varying the linear dimensions over a range of realistic values typically found in patients (Fig.1(b)). The distance from the apex of the aneurysm to the neck was set to be equal to the maximum transverse dimension of the aneurysm. As such the AR becomes the sole geometric parameter,with a smaller neck width leading to a larger AR. Three sets of models were tested, namely,2.0 and 3.0, similar to the range of values studied earlier in the literature. The software Gambit 2.4.6(ANSYS, Canonsburg, Pennsylvania, USA) was used for generating high quality mesh. The numerical simulations were then performed by Fluent 6.3 (ANSYS,Canonsburg, Pennsylvania, USA).

    1.1.2 Fluid mechanics

    The blood plasma was assumed to be a Newtonian fluid of density 1 060 kg/m3and viscosity 0.0035 kg/(m?s). The continuity and Navier-Stokes equations were solved by an unsteady solver in Fluent,with a time step of 0.001 s, and a residual error of 10-6. Tetrahedral elements with a size of 0.3 mm were employed. The computational mesh was shown in Fig.1(b). As an illustrative example, the mesh independence test for the case ofwas presented. Three sets of simulation with different number of elements were performed. The flow velocities at Point A were: 0.43 m/s (3×105grids), 0.44 m/s (8×105grids),0.44 m/s (1.4×106grids). Consequently, the mesh with 8×105grids was taken for further analysis.

    1.1.3 Boundary conditions

    A pulsatile velocity profile and a pressure waveform were prescribed at the inlet and outlets respectively (Fig.2). The wall was taken as rigid, and no slip boundary conditions were applied. The Reynolds number at the peak systole was 531, and thus the flow was assumed to be laminar.

    In the actual numerical simulations, a cycle will refer to a complete period of a heartbeat pulse with systole and diastole. Typically two cycles of computations would generate a periodic output, and data for the third cycle would be reported in the subsequent discussion. Unless otherwise stated, the properties analyzed refer to those at the systole phase of the cardiac cycle.

    Fig.2 Boundary conditions for the velocity and pressure waveforms prescribed at the inlet and outlets of the CFD simulations (a) and the experimental models (b)

    1.1.4 Patient-specific models

    In addition to the idealized models, four patientspecific cases were also included to provide clinical insight. These aneurysms were located on the internal carotid artery. The information of the patients and the linear dimensions of the aneurysms were summarizedin Table 1. In general the aneurysms were located either at or in the vicinity of a bifurcation (for Patient I and Patients II, III and IV respectively). The ARs of these aneurysms ranged from 1.1 to 2.2. The threedimensional aneurysm models were reconstructed from the angiograms of the patients via the software Mimics (Materialise, Leuven, Belgium) for CFD study.

    Table 1 Details of the patient information and the aneurysm geometries of the four patient-specific models. The AR ranges from 1.1 (Patient I) to 2.2 (Patient IV)

    1.2 Experimental set-up and ultrasound flow measurements

    1.2.1 Preparation of the phantoms

    Aneurysm models were first created using CAD software SOLIDWORKS 2013, with wall thickness of 0.8 mm and female plugs added at the inlet and outlets. This CAD model was then printed on a stereolithography machine (Eden 350 V, Objet Geometries,Rahovot, Israel), which could fabricate three-dimensional models with 16 μm layer thickness resolution and 42 μm layer plane resolution. A special compliant photopolymer (FullCure 930, Objet Geometries) was used for stereolithography production. As a supporting cast, gel-like supporting photosensitive resin (FullCure 705, Objet Geometries) was deposited in the hollow regions of the model during fabrication. The actual choice of vessel wall thickness depends on many competing factors. Vessel walls being too thin will lack structural rigidity, while vessel walls being too thick will suffer from excessive ultrasound attenuation. An optimal range of wall thickness has been identified.

    After the raw stereolithography builds were printed, the “phantoms” were then soaked in deionized water for two hours to rinse away the supporting cast surrounding the aneurysm. Subsequently, the supporting cast within the vasculature was carefully removed under water using a curette. These finished models were stored in distilled water at room temperature in closed containers. Such procedures would permit the fabricated models to be operational over time without significant change in mechanical properties. As quantified via tensile testing, the average elastic modulus of the wall was measured to be 391 kPa[23]. To compare with the computational results, phantoms with1.5, 2.0 and 3.0 were fabricated and tested(Fig.3).

    Fig.3 Phantom models of various ARs created by rapid prototyping machine

    1.2.2 Flow systems

    A computer controlled gear pump system(AccuFlow-Q, Shelley Medical Imaging, London, ON,Canada) was established to generate a pulsatile inflow at 72 cycles per minute, and a peak flow rate of 4.5 ml per second. A blood mimicking fluid (BMF) (Shelley Medical Imaging) with the following properties was utilized in this experiment[24]: (1) density is 1 037 kg/m3, (2) kinematic viscosity is 3.95×10-6m2/s,(3) acoustic speed is 1 548 m/s, (4) Newtonian fluid,(5) nylon scatter with blood-matching ultrasound scattering power. A physiological monitor with capability to record intraluminal pressure in real-time (HP Agilent CMS-2001/M1006B, Agilent Technologies,B?blingen, Germany) was deployed as a pressure sensing catheter at the outlets of the phantom.

    The flow circuit consisted of the following components. It started with the gear pump, where the BMF entered the phantom through the inlet and exited at the outlets through the mounted phantom. The pressure sensor was linked to a variable flow restrictor to produce the effects of varying peripheral resistance downstream. Before the steady state was reached, air bubbles were allowed to rise and settle in the reservoir,and the baseline pressure could be adjusted by raising or lowering the reservoir level (Fig.4).

    In preparation for the experiment, the three phantoms of different AR values were mounted on the same water tank to facilitate parallel study of their flow dynamics. Care was taken to avoid buckling the phantoms. Distilled water was filled to 2 cm atop the height of the phantoms.

    Fig.4 The experimental set-up showing all the major components (a), The phantom models with different aspect ratios being mounted in parallel for comparison (b)

    1.2.3 Doppler ultrasound measurement: Quantitative flow investigation

    An ultrasound scanner (SonixTouch, Ultrasonix,Richmond, BC, Canada) equipped with a linear array transducer (L14-5W, Ultrasonix) was used for flow measurements. The transducer was vertically positioned 0.015 m above the apical point of the phantom,and the imaging plane was aligned with the coronal(frontal) plane of the aneurysm. Pulsed Doppler measurements with an angle corrected flow velocity field were then taken at the inlet, the left and right outlets. The axial (relative to transducer, or vertical direction, defined as theaxis here) flow velocity was also recorded at selected points along the aneurysm, namely,sites “A”, “B”, “C”, “D”, “E” and “F” (Fig.1(b)).

    For all the pulsed Doppler measurements, ultrasound beams of frequency 5 MHz and range gate of 2.0 mm were employed. For the fast flow sites, the pulse repetition frequency (PRF) was set to be 5 kHz with a wall filter cut-off of 100 Hz, and hence a flow velocity of up to 0.6 m/s was captured. Conversely,for the slow-flow sites, the PRF was chosen to be 1.3 kHz with a wall filter cut-off of 25 Hz, and thus a flow velocity of up to 0.15 m/s was recorded. For data reported in the next section, six consecutive pulse cycles were measured for each phantom and each measuring site.

    1.2.4 Duplex color Doppler ultrasonography: Qualitative flow visualization

    Using the same ultrasound scanner, duplex color Doppler ultrasonography was performed to acquire images of the flow patterns inside the aneurysm. The color box was set to cover the entire aneurysm and the three branches. The frequency parameter was kept at 5 MHz, while PRF and wall cut-off were set to 1.3 kHz and 6 Hz respectively. Doppler ensemble size would use ten samples, and no beam steering nor Doppler angle correction was performed, resulting in a velocity range of. Furthermore, in a color Doppler image, the red/blue color denoted the flow directions toward/away from the transducer along the vertical direction respectively.

    1.2.5 Flow visualization by ultrasound imaging

    At the inlet of the phantom, the input flow rate was set to have two peaks in the velocity field, namely,at 0.44 m/s at the systolic peak, followed by another maximum at 0.18 m/s. At the outlet the peak flow rates corresponding to a velocity of about 0.35 m/s for both the left and right outlet branches were implemented. The outlet pressure was calibrated to attain a physiologically realistic value of 120/80 mmHg. Theseprofiles were selected as they were used frequently in many studies of carotid arteries earlier in the literature[23].

    Fig.5 CFD results and ultrasound Doppler signals showing the velocity field at various locations near and inside the aneurysm for the case of, Sites “A” through “F” and points “2” and “3” are defined according to Fig.1(b)

    For all the phantoms studied, aliasing occurred at the entry of the aneurysm due to the high velocity jet,and one single vortex was observed inside the aneurysm. Finally, the wall of the phantom was observed to display a slight movement due to the compliance of the elastic material, but these small amplitude vibrations did not substantially affect the agreement with CFD results presented in the next section. Shadowing effect below the aneurysm was also noted, due to the inherently high ultrasound signal attenuation of the compliant wall material[23].

    2. Results and discussions

    2.1 Fixed aspect ratio

    The flow dynamics in the aneurismal sac was analyzed. Inside the aneurysm, at the inflow region(Sites “A”, “B”, “C”, Fig.1(b)), as blood moved up from the neck to the apex, the axial flow velocity was reduced, and the waveform broadened. At the outflow region (Sites “D”, “E”, “F”), as blood moved from the apex down to the neck, the axial flow speed increased and the waveform narrowed. The peak velocity at the outflow region was lower than that of the inflow region. However, the flow rate in and out of the aneurysm should have a net sum of zero, as the aneurysm was a blind sac, and this balance was maintained by the difference in areas of the influx and outgoing flows. This flow pattern indeed compared favorably with the computations and the experiments conducted earlier in the literature[25]. Without stenting, blood flow entered the aneurysm near the distal neck and created an anticlockwise vortex. The flow would then leave the aneurysm near the proximal region[25].

    Basically, our CFD simulations and the experimental results showed extremely good agreements. As an illustrative example, the case ofwas highlighted. The velocity fields obtained computationally at the entrance of the aneurysm (site “A”), ascending portion of the stream (sites “B”, “C”), apex (site“D”), descending portion (sites “E”, “F”) were reported (Fig.5). The corresponding data of the velocity from experiments were displayed too (Fig.5). The shapes of the curves were almost identical. Furthermore, the maxima of the velocity field of the computational and experimental data agreed well, within just a few percent of each other.

    2.2 Varying aspect ratio

    The flow pattern near the neck region and inside the aneurysm was determined both computationally and experimentally as depicted in Fig.6. Fora significant portion of the fluid went into the aneurysm instead of flowing to the downstream vessel directly. For, however, strong flow impingement was observed near the neck. About half of the flow entered the aneurysm. In general, for higherAR, the area of the interface at the neck which exhibited velocity influx into the aneurysm would be smaller than that of the outflow region.

    Fig.7 CFD results and ultrasound Doppler signals showing the velocity at Point “B” (a) and “Point” D (d) for different aspect ratios:1.5, 2.0 and 3.0

    Another illuminating perspective was to examine the properties and extent of the jet impingement. The size of the jet impinging on the aneurysm and the location of impact were also obtained both by computations and experiments (Fig.6). Again excellent agreements were observed. In general, for lower aspect ratio, the jet was wider, and did not hit the wall directly as this fluid stream crossed into the aneurysmal sac and eventually reached the distal wall. However, for higher aspect ratio, the jet of fluid entering the aneurysm was narrower and impinged directly on the aneurysm wall.

    These trends were further confirmed by both the computational and experimental velocity fields recorded at the point of impact (Point “B”) and the apex(Point “D”) of the aneurysm (Fig.7). With increasing ARs, the maximum fluid velocity was lower at the apex region, as the main blood stream encountered difficulties in entering the sac in view of the smaller size of the entrance.

    2.3 Shear stress gradient

    Given the promising agreements between these two approaches, it would be instructive to use CFD to assess the shear stress at the aneurysm, where experimental measurements might be most difficult. Earlier works in the literature[26,27]had suggested that an abnormally low shear stress or an extraordinarily steep shear stress gradient might damage the normal functions of the endothelial cells. We shall demonstrate that aneurysms with higher aspect ratios do give rise to sharper shear stress gradients, thus lending support to the clinical beliefs that the risk of rupture grows with an increasing AR.

    The wall shear stress at the distal aneurysmal wall for different ARs was first calculated (Fig.8). The horizontal axis was the distance measured from the aneurysm neck. A local maximum was observed in the case ofand. This phenomenon was probably due to the flow impingement onto the aneurysmal wall, resulting in a high shear stress locally. The shear stress, however, just dropped gradually for the case of

    Fig.8 Shear stress at the distal aneurysmal wall for different ARs

    Fig.9 Shear stress gradients at the distal aneurysmal wall for different ARs

    The shear stress gradient for various aspect ratios was then obtained by numerical differentiation (Fig.9). All three curves displayed a numerically large, but algebraically negative, slope near the wall. However, the curves forandchanged direction and exhibited a maximum, thus necessarily incurring regions of positive gradients[28]. Such sharp changes in gradients may have adverse consequences in terms of vascular biology of the endothelial cells,and will be pursued quantitatively in the future.

    2.4 Patient-specific cases

    To provide further clinical insight, several patient-specific models were tested computationally and compared with the idealized models. The flow pattern near the neck region and inside the aneurysm was portrayed in Fig.10. For patient I with lower1.1), the jet was wider, and entered the aneurysm without hitting onto the distal aneurysmal wall. However, as AR increases, a narrower fluid jet was observed. For patient IV with higher AR, the jet of fluid impinged directly on the distal aneurysm wall. This phenomenon compared favorably with the pattern observed in idealized models (Fig.6).

    Fig.10 Computational velocity contour plots for the four patient-specific models with AR ranging from 1.1 (Patient I)to 2.2 (Patient IV)

    3. Conclusions

    The present paper conducts a joint computational-experimental study on the importance of the aspect ratio, defined here as the ratio of the aneurysm depth to the linear dimension of the neck. A higher aspect ratio has usually been associated with a greater risk of rupture statistically through clinical data[7-10]. However,the underlying flow dynamics is still not yet fully understood. Here the accuracy of computational simulations is verified by comparing with the flow measurements through phantoms fabricated from specialized materials with idealized geometries. Excellent agreements between the two approaches are obtained. Similar flow patterns are also observed in patient-specific models. This paper extends our previous works in twoaspects. Firstly, a more realistic boundary condition used frequently in the literature, where the velocity profile attains a nonzero value at the close of each cardiac cycle[29], is adopted here, and experimental verification is incorporated[30]. Secondly, our earlier work concentrated on such a joint computationalexperimental study for bifurcation and sidewall aneurysms with a fixed AR[31]. We have now extended the consideration to varying aspect ratios in this work.

    With different aspect ratios, two significant changes are observed from these joint investigations. Firstly, jets of blood stream entering an aneurysmal sac of higher aspect ratio tend to be narrower, and may impinge on the weakened wall directly. Secondly, the shear stress at the distal portion of the neck exhibits sharp variations and regimes of positive gradient are found for aneurysms with larger aspect ratios, providing a platform for the potential damage of the endothelial cells[28].

    In this study, the flow visualization in the phantoms is performed by ultrasonography. This is indeed a useful tool in medical engineering, and can be applied to study different blood vessel diseases, e.g. stenosis and ulcers[23]. The merits of ultrasound imaging include (1) noninvasive in nature, (2) relatively low cost,and (3) no known harmful side effects. Furthermore,particle injection into the fluid is not necessary. Consequently, this technique has great potential to be developed into an intraoperative tool in neurosurgical operations in the future.

    There are various possible extensions for the present work. In terms of geometric configuration, the length of the parent artery upstream employed in the simulation might affect the flow dynamics and the wall tension[32]. A similar issue can be examined here. Real blood vessels in patients are elastic but there is no universal agreement in the literature regarding the effects of elasticity on the hydrodynamics, with some studies showing effects on the velocity field[33], while other works might indicate similar results for both rigid and elastic models[34]. Nevertheless in future studies on rupture risk, wall thickness and elasticity must be restored[35]. Post-treatment wall shear stress and displacement distribution were investigated by utilizing patient specific basilar tip model with fluidstructure interaction simulation[36]. A joint computational-experimental study using patient-specific configuration will be fruitful. The boundary conditions could also be varied to investigate the effect of the peripheral resistance of the downstream branches[22].

    In conclusion, many clinical questions are still not fully understood and many challenges remain ahead[37-41]. Besides the aspect ratio, geometric factors like the orientation of the side branch relative to the aneurysm and the curvature of the parent vessel might be critical criteria. These issues should be addressed through such computational-experimental approaches in the near future.

    Acknowledgements

    Partial financial support has been provided by the Seed Funding Program for Basic Research of The University of Hong Kong, funding program from the Research Grants Council (Grant No. GRF 785113)and Innovation and Technology Support Program through contracts ITS/011/13 and ITS/150/15 of the Government of the Hong Kong Special Administrative Region.

    References

    [1] TAYLOR C. A., DRANEY M. T. Experimental and computational methods in cardiovascular fluid mechanics[J]. Annual Review of Fluid Mechanics, 2004, 36: 197-231.

    [2] LASHERAS J. C. The biomechanics of arterial aneurysms[J]. Annual Review of Fluid Mechanics, 2007, 39: 293-319.

    [3] HUMPHREY J. D., TAYLOR C. A. Intracranial and abdominal aortic aneurysms: Similarities, differences and need for a new class of computational models[J]. Annual Review of Biomedical Engineering, 2008, 10: 221-246.

    [4] CEBRAL J. R., RASCHI M. Suggested connections between risk factors of intracranial aneurysms: A review[J]. Annals of Biomedical Engineering, 2013, 41(7): 1366-1383.

    [5] D’URSO P. I., LANZINO G. and CLOFT H. J. et al. Flow diversion for intracranial aneurysms: A review[J]. Stroke,2011, 42(8): 2363-2368.

    [6] SFORZA D. M., PUTMAN C. M. and CEBRAL J. R. Hemodynamics of cerebral aneurysms[J]. Annual Review of Fluid Mechanics, 2009, 41: 91-107.

    [7] UJIIE H., TAMANO Y. and SASAKI K. et al. Is the aspect ratio a reliable index for predicting the rupture of a saccular aneurysm?[J]. Neurosurgery, 2001, 48(3): 495-503.

    [8] WEIR B., AMIDEI C. and KONGABLE G. et al. The aspect ratio (dome/neck) of ruptured and unruptured aneurysms[J]. Journal of Neurosurgery, 2003, 99(3): 447-4 51.

    [9] NADER-SEPAHI A., CASIMIRO M. and SEN J. et al. Is aspect ratio a reliable predictor of intracranial aneurysm rupture?[J]. Neurosurgery, 2004, 54(6): 1343-1348.

    [10] YASUDA R., STROTHER C. M. and TAKI W. et al. Aneurysm volume-to-ostium area ratio: A parameter useful for discriminating the rupture status of intracranial aneurysms[J]. Neurosurgery, 2011, 68(2): 310-318.

    [11] ZENG Z., DURKA M. J. and KALLMES D. F. et al. Can aspect ratio be used to categorize intra-aneurysmal hemodynamics? -A study of elastase induced aneurysms in rabbit[J]. Journal of Biomechanics, 2011, 44(16): 2809-2816.

    [12] WANG Qing, WANG Wei-zhe and FEI Zhi-min et al. Simulation of blood flow in intracranial ICA-PComA aneurysm via computational fluid dynamics modeling[J]. Journal of Hydrodynamics, 2009, 21(5): 583-590.

    [13] QIU Xiao-ning, FEI Zhi-min and ZHANG Jue et al. Influence of high-porosity mesh stent on hemodynamics of intracranial aneurysm: A computational study[J]. Journal of Hydrodynamics, 2013, 25(6): 848-855.

    [14] WANG Sheng-zhang, CHEN Jia-liang and DING Guanghong. Non-newtonian computational hemodynamics in two patient-specific cerebral aneurysms with daughter saccules[J]. Journal of Hydrodynamics, 2010, 22(5): 639-646.

    [15] HOI Y., WOODWARD S. H. and KIM M. et al. Validation of CFD simulations of cerebral aneurysms with implication of geometric variations[J]. ASME Journal of Biomechanical Engineering, 2006, 128(6): 844-851.

    [16] JOU L. D., MAWAD M. E. Analysis of intra-aneurysmal flow for cerebral aneurysms with cerebral angiography[J]. American Journal of Neuroradiology, 2012, 33(9): 1679-1684.

    [17] MARTIN-MCNULTY B., VINCELETTE J. and VERGONA R. et al. Noninvasive measurement of abdominal aortic aneurysms in intact mice by a high-frequency ultrasound imaging system[J]. Ultrasound in Medicine and Biology, 2005, 31(6): 745-749.

    [18] SCH?NING M., SCHEEL P. and WITTIBSCHL?GER J. et al. The effect of vasospasm on cerebral perfusion: A color duplex study of the extra- and intracranial cerebral arteries[J]. Ultrasound in Medicine and Biology, 2012,38(3): 360-367.

    [19] H?LSCHER T., RODRIGUEZ-RODRIGUEZ J. and WILKENING W. G. et al. Intraoperative brain ultrasound: A new approach to study flow dynamics in intracranial aneurysms[J]. Ultrasound in Medicine and Biology,2006, 32(9): 1307-1313.

    [20] LAM S. K., FUNG G. S. K. and CHENG S. W. K. et al. A computational study on the biomechanical factors related to stent-graft models in the thoracic aorta[J]. Medical and Biological Engineering and Computing, 2008, 46(11): 1129-1138.

    [21] TANG A. Y. S., FAN Y. and CHENG S. W. K. et al. Biomechanical factors influencing type B thoracic aortic dissection: Computational fluid dynamics study[J]. Engineering Applications of Computational Fluid Mechanics, 2012, 6(4): 622-632.

    [22] TANG A. Y. S., CHUNG W. C. and LIU E. T. Y. et al. Computational fluid dynamics study of bifurcation aneurysms treated with pipeline embolization device: Side branch diameter study[J]. Journal of Medical and Biological Engineering, 2015, 35(3): 293-304.

    [23] LAI S. S. M., YIU B. Y. S. and POON A. K. K. et al. Design of anthropomorphic flow phantoms based on rapid prototyping of compliant vessel geometries[J]. Ultrasound in Medicine and Biology, 2013, 39(9): 1654-1664.

    [24] RAMNARINE K. V., NASSIRI D. K. and HOSKINS P. R. et al. Validation of a new blood-mimicking fluid for use in Doppler flow test objects[J]. Ultrasound in Medicine and Biology, 1998, 24(3): 451-459.

    [25] LIOU T. M., LI Y. C. Effects of stent porosity on hemodynamics in a sidewall aneurysm model[J]. Journal of Biomechanics, 2008, 41(6): 1174-1183.

    [26] MALEK A. M., ALPER S. L. and IZUMO S. Hemodynamic shear stress and its role in atherosclerosis[J]. JAMA The Journal of the American Medical Association,1999, 282(21): 2035-2042.

    [27] SHOJIMA M., OSHIMA M. and TAKAGI K. et al. Magnitude and role of wall shear stress on cerebral aneurysm: Computational fluid dynamic study of 20 middle cerebral artery aneurysms[J]. Stroke, 2004, 35(11): 2500-2505.

    [28] DOLAN J. M., MENG H. and SINGH S. et al. High fluid shear stress and spatial shear stress gradients affect endothelial proliferation, survival and alignment[J]. Annals of Biomedical Engineering, 2011, 39(6): 1620-1631.

    [29] KU D. N., GIDDENS D. P. and ZARINS C. K. et al. Pulsatile flow and atherosclerosis in the human carotid bifurcation-Positive correlation between plaque location and low and oscillating shear-stress[J]. Arteriosclerosis,1985, 5(3): 293-302.

    [30] TANG A. Y. S., LAI S. K. and LEUNG K. M. et al. Influence of the aspect ratio on the endovascular treatment of intracranial aneurysms: A computational investigation[J]. Journal of Biomedical Science and Engineering,2012, 5(8): 422-431.

    [31] TSANG A. C. O., LAI S. S. M. and CHUNG W. C. et al. Blood flow in intracranial aneurysms treated with pipeline embolization devices: Computational simulation and verification with Doppler ultrasonography on phantom models[J]. Ultrasonography, 2015, 34(2): 98-108.

    [32] CHEN Jia-liang, DING Guang-hong and YANG Xin-jian et al. Effects of parent artery segmentation and aneurismal-wall elasticity on patient-specific hemodynamic simulations[J]. Journal of Hydrodynamics, 2011, 23(5): 660-668.

    [33] ZHAO Jun-wei, YIN Wen-yi and DING Guang-hong et al. Numerical simulation and analysis on the hemodynamics of an elastic aneurysm[J]. Journal of Hydrodynamics,2008, 20(2): 216-224.

    [34] BAZILEVS Y., HSU M. C. and ZHANG Y. et al. A fullycoupled fluid-structure interaction simulation of cerebral aneurysms[J]. Computational Mechanics, 2010, 46(1): 3-16.

    [35] CEBRAL J. R., VAZQUEZ M. and SFORZA D. M. et al. Analysis of hemodynamics and wall mechanics at sites of cerebral aneurysm rupture[J]. Journal of NeuroInterventional Surgery, 2015, 7(7): 530-536.

    [36] CHEN Jia-liang, WANG Sheng-zhang and DING Guanghong et al. Patient-specific blood dynamic simulations in assessing endovascular occlusion of intracranial aneurysms[J]. Journal of Hydrodynamics, 2009, 21(2): 271-276.

    [37] HAN Xu, LIU Xiao-sheng and LIANG Fu-you. The influence of outflow boundary conditions on blood flow patterns in an AcoA aneurysm[J] Chinese Journal of Hydrodynamics, 2015, 30(6): 692-700(in Chinese).

    [38] MU Li-zhong, HE Ying and CHEN Jun-yuan et al. Study of blood flow regulation in a patient-specific model for the circle of Willis with an aneurysm by computational and in-vitro simulation[J] Chinese Journal of Hydrodynamics, 2015, 30(6): 707-715(in Chinese).

    [39] KAN Xiao-xin, BAI Yi-fan and MA Tao et al. Fluid-structure interaction of aortic dissection based on the computed tomography reconstructed model[J]. Chinese Journal of Hydrodynamics, 2015, 30(6): 701-706(in Chinese).

    [40] LIU Hai-bin, QIN Kai-rong and YUAN Wen-xue et al. Acute effect of moderate intensity aerobic cycling exercise on carotid arterial stiffness and hemodynamics in sedentary controls[J]. Chinese Journal of Hydrodynamics,2015, 30(6): 721-728(in Chinese).

    [41] WANG Yan-xia, GAO Zheng-ming and LIU Bo et al. A flow chamber device for simulating oscillatory shear stress in the common carotid artery after exercise training[J]. Chinese Journal of Hydrodynamics, 2015, 30(6): 650-656(in Chinese).

    September 20, 2015, Revised April 1, 2016)

    * Biography: Simon Sui-Man LAI (1985-), Male,

    Master, Research Associate

    Abraham Yik-Sau TANG,

    E-mail: aystang@hku.hk

    青春草视频在线免费观看| 国产蜜桃级精品一区二区三区| 欧美激情国产日韩精品一区| 国产精品一区二区免费欧美| 欧美xxxx性猛交bbbb| 一区二区三区四区激情视频 | 免费在线观看影片大全网站| 神马国产精品三级电影在线观看| 赤兔流量卡办理| 国产精品永久免费网站| 成人鲁丝片一二三区免费| 51国产日韩欧美| 1024手机看黄色片| 丝袜美腿在线中文| 色综合亚洲欧美另类图片| 少妇人妻一区二区三区视频| 两个人的视频大全免费| 亚洲在线观看片| 免费黄网站久久成人精品| 99热6这里只有精品| 97热精品久久久久久| 亚洲精品粉嫩美女一区| 波多野结衣高清作品| 丰满的人妻完整版| 日日撸夜夜添| 搡女人真爽免费视频火全软件 | 国产不卡一卡二| 亚洲国产精品sss在线观看| 午夜福利18| 精品福利观看| 亚洲人成网站在线观看播放| 成人美女网站在线观看视频| 精品一区二区三区av网在线观看| 99久国产av精品| 日韩亚洲欧美综合| 欧美成人a在线观看| 久久6这里有精品| 丰满乱子伦码专区| 色在线成人网| 人妻夜夜爽99麻豆av| 久久久久国产网址| 91在线观看av| 91狼人影院| 欧美人与善性xxx| 可以在线观看毛片的网站| 别揉我奶头 嗯啊视频| 精品一区二区三区视频在线| 色播亚洲综合网| 欧美zozozo另类| 夜夜爽天天搞| 亚洲欧美成人综合另类久久久 | 一本久久中文字幕| 国产成人a∨麻豆精品| 床上黄色一级片| 日韩制服骚丝袜av| 美女免费视频网站| 神马国产精品三级电影在线观看| 99久久精品一区二区三区| 老女人水多毛片| 亚洲人成网站高清观看| 久久草成人影院| 人妻久久中文字幕网| 免费黄网站久久成人精品| 欧美成人精品欧美一级黄| 国产精品久久电影中文字幕| 晚上一个人看的免费电影| 九九爱精品视频在线观看| 直男gayav资源| 中文字幕av在线有码专区| 国产精品乱码一区二三区的特点| 精品熟女少妇av免费看| 久久久午夜欧美精品| 成人国产麻豆网| 国产精品一区二区三区四区久久| 亚洲无线在线观看| 九九热线精品视视频播放| 亚洲熟妇中文字幕五十中出| a级毛片免费高清观看在线播放| 一级毛片电影观看 | 日本a在线网址| 69av精品久久久久久| 成人无遮挡网站| 久久天躁狠狠躁夜夜2o2o| 欧美zozozo另类| 亚洲性夜色夜夜综合| 亚洲精品久久国产高清桃花| 国产人妻一区二区三区在| а√天堂www在线а√下载| 我要看日韩黄色一级片| 噜噜噜噜噜久久久久久91| 91久久精品国产一区二区三区| 成年女人永久免费观看视频| 欧美一区二区国产精品久久精品| 男女边吃奶边做爰视频| 亚洲成人中文字幕在线播放| 国产午夜精品久久久久久一区二区三区 | 成人亚洲精品av一区二区| 久99久视频精品免费| 最近在线观看免费完整版| 亚洲国产精品成人综合色| 少妇猛男粗大的猛烈进出视频 | 欧美最黄视频在线播放免费| 一本精品99久久精品77| 国产在线精品亚洲第一网站| 成人精品一区二区免费| 天堂av国产一区二区熟女人妻| 看片在线看免费视频| 观看免费一级毛片| 亚洲国产精品久久男人天堂| 精品99又大又爽又粗少妇毛片| 麻豆乱淫一区二区| 亚洲人与动物交配视频| 麻豆av噜噜一区二区三区| 色尼玛亚洲综合影院| 国产人妻一区二区三区在| 久久久久九九精品影院| 国产成年人精品一区二区| 女生性感内裤真人,穿戴方法视频| 美女xxoo啪啪120秒动态图| 国产高清有码在线观看视频| 精品久久久久久久久av| 亚洲精品国产成人久久av| 丰满乱子伦码专区| 日本爱情动作片www.在线观看 | 黄色欧美视频在线观看| 在线播放国产精品三级| 国产乱人偷精品视频| 亚洲av成人精品一区久久| 97超碰精品成人国产| 成人永久免费在线观看视频| 日韩一区二区视频免费看| 97超碰精品成人国产| 不卡视频在线观看欧美| 国产伦一二天堂av在线观看| 熟女电影av网| 插逼视频在线观看| 欧美绝顶高潮抽搐喷水| 久久久久久久久久成人| 九九在线视频观看精品| 欧美又色又爽又黄视频| 精品久久久久久久久亚洲| 久久精品国产亚洲av香蕉五月| 国产成年人精品一区二区| 亚洲av一区综合| 亚洲性夜色夜夜综合| 国产精品久久久久久av不卡| 国产淫片久久久久久久久| 成人二区视频| 国产91av在线免费观看| 热99re8久久精品国产| 亚洲高清免费不卡视频| 男人的好看免费观看在线视频| 在线看三级毛片| 毛片一级片免费看久久久久| 久久精品综合一区二区三区| 99九九线精品视频在线观看视频| videossex国产| 在现免费观看毛片| 久久久色成人| 美女内射精品一级片tv| 毛片女人毛片| 久久草成人影院| 可以在线观看毛片的网站| 午夜免费男女啪啪视频观看 | 久久久久久久久大av| 免费电影在线观看免费观看| 舔av片在线| 我要搜黄色片| www.色视频.com| 日本精品一区二区三区蜜桃| 淫秽高清视频在线观看| 男女之事视频高清在线观看| 日韩高清综合在线| 别揉我奶头~嗯~啊~动态视频| 国产在视频线在精品| 久久人妻av系列| 亚洲最大成人中文| 国产亚洲精品久久久久久毛片| 两个人的视频大全免费| 久久精品夜夜夜夜夜久久蜜豆| 亚洲内射少妇av| 成人午夜高清在线视频| 国产成人影院久久av| 久久久久性生活片| 99久国产av精品| 啦啦啦韩国在线观看视频| 国产精品亚洲一级av第二区| 无遮挡黄片免费观看| 激情 狠狠 欧美| 午夜激情福利司机影院| 午夜精品一区二区三区免费看| 赤兔流量卡办理| 99热这里只有是精品50| 中文字幕av在线有码专区| 久久久久久久久久成人| 麻豆久久精品国产亚洲av| 97人妻精品一区二区三区麻豆| 成人特级av手机在线观看| 日韩欧美免费精品| 一级毛片我不卡| 免费看av在线观看网站| 亚洲欧美日韩高清在线视频| 亚洲av中文字字幕乱码综合| 日本成人三级电影网站| 日产精品乱码卡一卡2卡三| 国产成人91sexporn| 听说在线观看完整版免费高清| 99久久精品国产国产毛片| 色视频www国产| 日韩欧美国产在线观看| 如何舔出高潮| 亚洲人成网站在线观看播放| 老熟妇乱子伦视频在线观看| 国产精品一区二区性色av| 在线天堂最新版资源| 亚洲一区二区三区色噜噜| 美女免费视频网站| 国国产精品蜜臀av免费| 在线观看午夜福利视频| 丰满乱子伦码专区| 日本一二三区视频观看| 一进一出抽搐动态| 亚洲精品乱码久久久v下载方式| 国产精品不卡视频一区二区| 永久网站在线| 99热只有精品国产| а√天堂www在线а√下载| 99久久精品国产国产毛片| 免费看av在线观看网站| 99久久精品一区二区三区| 国产黄片美女视频| 亚洲精品亚洲一区二区| 久久99热这里只有精品18| 97热精品久久久久久| 中出人妻视频一区二区| 别揉我奶头 嗯啊视频| 免费av观看视频| 全区人妻精品视频| 国产 一区精品| 国产一区二区在线av高清观看| 男人的好看免费观看在线视频| 国内精品宾馆在线| 国产精品久久久久久久久免| 精品久久久久久久人妻蜜臀av| 亚洲精品一卡2卡三卡4卡5卡| 色在线成人网| 亚洲熟妇熟女久久| 99riav亚洲国产免费| 中出人妻视频一区二区| 免费一级毛片在线播放高清视频| 在线观看66精品国产| 亚洲成人中文字幕在线播放| 久久久久久久午夜电影| 好男人在线观看高清免费视频| 欧美成人精品欧美一级黄| 级片在线观看| 白带黄色成豆腐渣| 色在线成人网| 露出奶头的视频| 非洲黑人性xxxx精品又粗又长| 联通29元200g的流量卡| 国产亚洲精品av在线| 日韩人妻高清精品专区| 3wmmmm亚洲av在线观看| 中文字幕熟女人妻在线| 亚洲四区av| 午夜久久久久精精品| 一级黄色大片毛片| 婷婷精品国产亚洲av在线| 丰满的人妻完整版| 在线观看美女被高潮喷水网站| 国产精品一区www在线观看| 天堂网av新在线| 插逼视频在线观看| 国产伦精品一区二区三区视频9| 97在线视频观看| 能在线免费观看的黄片| 十八禁国产超污无遮挡网站| 网址你懂的国产日韩在线| 国产真实伦视频高清在线观看| 欧美性感艳星| 国产黄色视频一区二区在线观看 | 一级毛片aaaaaa免费看小| 国产精品久久久久久久电影| 99热网站在线观看| 久久天躁狠狠躁夜夜2o2o| 日本撒尿小便嘘嘘汇集6| 国产精品电影一区二区三区| 亚洲国产高清在线一区二区三| 亚洲美女视频黄频| 免费搜索国产男女视频| 蜜桃亚洲精品一区二区三区| 色5月婷婷丁香| 2021天堂中文幕一二区在线观| 少妇的逼好多水| 欧美色视频一区免费| 乱码一卡2卡4卡精品| 在线a可以看的网站| 丰满的人妻完整版| 久久久久国内视频| 亚洲人成网站在线播| 国产av麻豆久久久久久久| 亚洲人成网站在线播放欧美日韩| 久久99热这里只有精品18| 亚洲精品久久国产高清桃花| 午夜久久久久精精品| 久久久久国产精品人妻aⅴ院| 精品国产三级普通话版| 超碰av人人做人人爽久久| 99热全是精品| 在线播放无遮挡| 亚洲成人精品中文字幕电影| 久久久精品欧美日韩精品| 亚洲人与动物交配视频| 18+在线观看网站| 人妻夜夜爽99麻豆av| 免费不卡的大黄色大毛片视频在线观看 | av中文乱码字幕在线| 99在线人妻在线中文字幕| 久久午夜亚洲精品久久| 99久久成人亚洲精品观看| h日本视频在线播放| 99久久精品热视频| 毛片一级片免费看久久久久| 亚洲精品久久国产高清桃花| 在线播放国产精品三级| 免费av不卡在线播放| 色噜噜av男人的天堂激情| 日韩高清综合在线| 亚洲美女搞黄在线观看 | av免费在线看不卡| 一本一本综合久久| 国产精品电影一区二区三区| 成人毛片a级毛片在线播放| 久久久久久久午夜电影| 久久午夜亚洲精品久久| 国产aⅴ精品一区二区三区波| 亚洲人成网站在线播放欧美日韩| 精品午夜福利在线看| 国产欧美日韩一区二区精品| 亚洲丝袜综合中文字幕| 亚洲国产高清在线一区二区三| 久久精品影院6| 成人av在线播放网站| 少妇的逼水好多| 日本五十路高清| 99九九线精品视频在线观看视频| 久久精品国产99精品国产亚洲性色| 黄色一级大片看看| 久久婷婷人人爽人人干人人爱| 亚洲精品影视一区二区三区av| 高清毛片免费看| .国产精品久久| 久久精品国产清高在天天线| 亚洲四区av| 看黄色毛片网站| 久久人人精品亚洲av| 久久久久久国产a免费观看| 51国产日韩欧美| 波多野结衣巨乳人妻| 亚洲精品一区av在线观看| 日本免费a在线| 欧美另类亚洲清纯唯美| 99视频精品全部免费 在线| 99久久中文字幕三级久久日本| 日韩高清综合在线| 亚洲国产精品国产精品| 国产亚洲精品久久久久久毛片| 亚洲自偷自拍三级| 三级毛片av免费| 两性午夜刺激爽爽歪歪视频在线观看| 国产午夜精品久久久久久一区二区三区 | 国产午夜精品久久久久久一区二区三区 | 日日撸夜夜添| 国产成人影院久久av| av专区在线播放| 国产午夜精品久久久久久一区二区三区 | 97在线视频观看| 人妻久久中文字幕网| 久久热精品热| 18禁在线无遮挡免费观看视频 | 联通29元200g的流量卡| 人妻夜夜爽99麻豆av| 亚州av有码| 午夜精品在线福利| 日韩成人伦理影院| 少妇丰满av| 久久午夜亚洲精品久久| 波多野结衣巨乳人妻| 看非洲黑人一级黄片| 久久九九热精品免费| 午夜视频国产福利| 国产av在哪里看| av在线播放精品| 乱码一卡2卡4卡精品| 22中文网久久字幕| 女人十人毛片免费观看3o分钟| 在线播放国产精品三级| 免费看光身美女| 我的女老师完整版在线观看| 亚洲国产精品成人久久小说 | 精品久久久久久久久久免费视频| 天天躁夜夜躁狠狠久久av| 精品免费久久久久久久清纯| 国产亚洲欧美98| 变态另类成人亚洲欧美熟女| 国产精品精品国产色婷婷| 国产色爽女视频免费观看| av天堂在线播放| av视频在线观看入口| 国产精品一二三区在线看| 免费无遮挡裸体视频| 久久久久久久午夜电影| 国产精品永久免费网站| 波多野结衣高清无吗| 女人十人毛片免费观看3o分钟| 亚洲国产精品国产精品| 91久久精品电影网| 最近中文字幕高清免费大全6| 久久久a久久爽久久v久久| 国产精品无大码| 欧美色视频一区免费| 人人妻人人澡人人爽人人夜夜 | 伦精品一区二区三区| 亚洲国产精品sss在线观看| 日韩高清综合在线| 91在线观看av| 亚洲综合色惰| 国产黄色视频一区二区在线观看 | 欧美性感艳星| 国产在视频线在精品| av天堂在线播放| 男人的好看免费观看在线视频| 国产伦一二天堂av在线观看| 久久久久久伊人网av| 国产日本99.免费观看| 国产亚洲精品av在线| 丰满人妻一区二区三区视频av| 国产视频内射| 三级男女做爰猛烈吃奶摸视频| 99在线人妻在线中文字幕| 校园春色视频在线观看| 在线观看av片永久免费下载| 欧美日韩乱码在线| 性欧美人与动物交配| 亚洲欧美精品自产自拍| 国产精品女同一区二区软件| 中文在线观看免费www的网站| 男人舔奶头视频| 一级毛片我不卡| 久久亚洲国产成人精品v| www.色视频.com| 日本黄色片子视频| 中文亚洲av片在线观看爽| 国内精品宾馆在线| 校园春色视频在线观看| 我的老师免费观看完整版| 俄罗斯特黄特色一大片| 赤兔流量卡办理| 97在线视频观看| 国产av麻豆久久久久久久| 天堂√8在线中文| 亚洲av不卡在线观看| 国产乱人视频| 中国美女看黄片| 少妇高潮的动态图| 一夜夜www| 久久久欧美国产精品| 高清日韩中文字幕在线| 国产69精品久久久久777片| 国产精品一区二区性色av| 激情 狠狠 欧美| 国产在视频线在精品| 免费在线观看影片大全网站| 天天躁日日操中文字幕| 国产亚洲精品av在线| 欧美精品国产亚洲| 国产91av在线免费观看| 99国产极品粉嫩在线观看| 国产精品嫩草影院av在线观看| h日本视频在线播放| 99九九线精品视频在线观看视频| 免费人成在线观看视频色| 亚洲国产欧洲综合997久久,| 国产精品永久免费网站| 精品久久久久久久久亚洲| 免费看av在线观看网站| 国产男靠女视频免费网站| 欧美色视频一区免费| 亚洲av中文字字幕乱码综合| 97超视频在线观看视频| 一卡2卡三卡四卡精品乱码亚洲| 亚洲国产高清在线一区二区三| 午夜亚洲福利在线播放| 成人欧美大片| 丰满人妻一区二区三区视频av| 校园春色视频在线观看| 99国产极品粉嫩在线观看| 亚洲欧美清纯卡通| 国内精品一区二区在线观看| 欧美性猛交黑人性爽| 国产大屁股一区二区在线视频| 欧美日韩在线观看h| 熟女人妻精品中文字幕| 成年女人看的毛片在线观看| 亚洲aⅴ乱码一区二区在线播放| 亚洲在线自拍视频| 午夜a级毛片| 亚洲一区高清亚洲精品| 国产老妇女一区| 最近中文字幕高清免费大全6| 国产亚洲精品综合一区在线观看| 免费高清视频大片| 国内久久婷婷六月综合欲色啪| 亚洲av中文av极速乱| 久久久久九九精品影院| 又黄又爽又免费观看的视频| 色哟哟·www| 亚洲人成网站高清观看| 久久99热这里只有精品18| 亚洲av电影不卡..在线观看| 青春草视频在线免费观看| 内射极品少妇av片p| a级毛片免费高清观看在线播放| 色av中文字幕| 亚洲精品在线观看二区| 熟女人妻精品中文字幕| 麻豆久久精品国产亚洲av| 久久久精品欧美日韩精品| 99久久精品一区二区三区| 有码 亚洲区| 日本三级黄在线观看| 日本色播在线视频| 在线看三级毛片| 日日啪夜夜撸| 热99re8久久精品国产| 日韩人妻高清精品专区| 国产亚洲av嫩草精品影院| 内射极品少妇av片p| 欧美一区二区亚洲| 看黄色毛片网站| 九九热线精品视视频播放| 国产欧美日韩精品一区二区| 日本精品一区二区三区蜜桃| 国产 一区 欧美 日韩| 99久久精品热视频| 蜜桃久久精品国产亚洲av| 国产伦精品一区二区三区视频9| 极品教师在线视频| 国产 一区精品| 免费看av在线观看网站| 亚洲精品在线观看二区| 能在线免费观看的黄片| av在线蜜桃| 国产欧美日韩一区二区精品| 日本黄大片高清| 亚洲欧美日韩高清专用| 久久热精品热| 国产麻豆成人av免费视频| 夜夜看夜夜爽夜夜摸| 男插女下体视频免费在线播放| .国产精品久久| 一个人观看的视频www高清免费观看| 国产探花极品一区二区| 欧美性猛交╳xxx乱大交人| 国产精品电影一区二区三区| 亚洲内射少妇av| 国产精品野战在线观看| 男女边吃奶边做爰视频| 联通29元200g的流量卡| 亚州av有码| 亚洲欧美日韩高清在线视频| 少妇被粗大猛烈的视频| 春色校园在线视频观看| 国产黄片美女视频| 两个人视频免费观看高清| 日韩大尺度精品在线看网址| 在线观看一区二区三区| 99久久成人亚洲精品观看| 久久精品国产清高在天天线| 国产在视频线在精品| 美女黄网站色视频| 丝袜喷水一区| 亚洲av免费在线观看| 亚洲在线观看片| 亚洲av美国av| 国产色婷婷99| 欧美一区二区国产精品久久精品| 国产aⅴ精品一区二区三区波| 国产黄a三级三级三级人| 日韩一区二区视频免费看| 国产精品三级大全| 日韩欧美在线乱码| 日韩三级伦理在线观看| 乱码一卡2卡4卡精品| 久久久久久久久久黄片| 国产成人a区在线观看| 变态另类丝袜制服| 小蜜桃在线观看免费完整版高清| 国产一区二区亚洲精品在线观看| 欧美+日韩+精品| 久久韩国三级中文字幕| 搡老熟女国产l中国老女人| 两个人视频免费观看高清| 三级经典国产精品| 天天躁日日操中文字幕| 小说图片视频综合网站| 国产亚洲精品久久久com| 寂寞人妻少妇视频99o| 能在线免费观看的黄片| 国产av在哪里看| 男人狂女人下面高潮的视频| 菩萨蛮人人尽说江南好唐韦庄 | 成人综合一区亚洲| 少妇的逼好多水|