    Analysis of photostimulable phosphor image plate artifacts and their prevalence

    2022-01-24 09:24:28SaraMohamedElkhateebAshwagYagoubAloyounyMohamedMohamedSaeedOmerSoadMahmoudMansour
    World Journal of Clinical Cases 2022年2期

    INTRODUCTION

    Digital radiography has recently been used in dentistry as a substitute for conventional film radiography worldwide.Digital imaging provides new possibilities for recording and interpreting radiographic data in a user-friendly digital way for archiving and teleradiography[1].Digital imaging has many advantages over filmbased radiography,such as real-time imaging,not needing the use of darkroom chemicals,having image manipulation tools,better archiving,and decreased patient and operator radiation exposure[2-6].

    Dental radiography is provided by two main types of radiography:Conventional and digital radiography.Conventional radiography uses dental films that require chemical solutions for processing,more human resources,is time consuming,and with a higher radiation dose.This traditional type of radiography provides a permanent record of imaging without any possibility for archiving,and any error requires the retake of the radiograph,which exposes the patient to unnecessary additional radiation exposure.This is in contrast to digital radiography,which allows image manipulation to correct the visual characteristics of the image,such as contrast,brightness,and density,thus enhancing image quality without the need to retake the image.Digital radiography permits easy archiving of images and their electronic transfer between different specialties in the dental field.Digital radiographic technology uses electronic image receptors (digital sensors or plates),which are based on two main techniques in acquiring the image:A direct method using a chargecoupled device (CCD) or complementary metal oxide semiconductor (CMOS) sensors[7] that transmit the signal of the exposed plate directly over a wire with a real-time image on the monitor or indirect method using a photostimulable phosphor storage plate (PSP),which forms a latent image when exposed to radiation.The stored energy is then transported to a computer for display using a laser scanner[8-10].

    The scanning times of PSP plates vary from a few seconds to several minutes,depending on the type of laser scanner used and the spatial and contrast resolutions of the image[5].PSP plates are available in a variety of sizes in a way similar to conventional films,so they are vulnerable to bending and scratching during handling[2,11].

    Moreover,PSP plates must be handled more carefully than films because they are reusable after erasing the image[12].PSP plates are selected by most dental practitioners because of their easy intraoral placement with little patient discomfort,as well as them being cordless and resembling conventional films.This is in comparison to the more difficult intraoral placement of CCD plates,with more patient discomfort caused by the stiffness of these plates with a cord linking them to the computer even though an image can be obtained promptly by the practitioner after exposure of the plate in this system[11,12].

    But, answered he, let a man be ever so strong, he cannot carry more than a hundredweight, and what is that for a king s daughter? Well, do as you like; I have said my say

    Digital radiography,like any evolving technology,produces a new type of image pitfall that remains a problem for clinicians in that it can be overwhelming.To the best of our knowledge,a systematic review of dental digital radiography artifacts in clinical usage has not yet been reported[13].In addition,few studies have assessed PSP image errors using illustrative figures[14].

    MATERIALS AND METHODS

    This retrospective descriptive study was conducted in the oral and maxillofacial radiology unit of the dental clinics of the College of Dentistry,Princess Nourah University (PNU),where conventional film-based radiography was gradually replaced by digital radiography starting in 2018.All the intraoral digital radiographs were acquired using (Gendex Expert DC.,United States) an intraoral X-ray machine with 7-mA 65-kVP using a PSP system (Soredex DIGORA Optime imaging plate) and laser scanners (Soredex DIGORA Optime),which can house all sizes of reusable intraoral PSP sensor plates with image acquisition software (MIPACS Dental Enterprise viewer 3.2.2).

    The investigators retrieved all the digital intraoral periapical and bitewing radiographs that were taken and approved by the clinicians from all dental specialties in our college and with the consent of the patients from April 2018 to April 2020.We used the Medicor Imaging/MIPACS Toolkit software,which regularly detects and records all deleted radiographs due to retakes performed by all oral radiology technicians in the radiology unit with clarification of the image type.All the retakes were screened by two well-experienced oral and maxillofacial radiologists for evaluation of the present artifact type and cause,while the remaining images without errors were excluded from the study.Artifacts were classified into three categories:plate errors,scanning errors,and operator errors.First,two observers sat together in a collaboration session to determine the criteria and subtypes for each category of digital artifacts to unify the interpretation process.

    The observers independently evaluated and agreed on image artifacts.When disagreement existed among them,consensus was reached through discussion.Out of the selected cases,1000 images with errors were reevaluated by both investigators after 2 wk for the calculation of interobserver reliability.Our retrospective study was approved by the institutional review board (IRB) of PNU.

    The Princess heard his puffing18 and roaring, and growing frightened she cried: Oh dear! the bear is after us and will certainly catch us up! The tailor remained quite unmoved

    On the third31 morning after they had left their father s house they set about their wandering again, but only got deeper and deeper into the wood, and now they felt that if help did not come to them soon they must perish

    Statistical analysis

    Statistical analysis was performed using SPSS Statistics version 23 (IBM Corp.,New York,NY;formerly SPSS Inc.,Chicago,IL,United States),and the frequencies and percentages of PSP artifacts were calculated.Interobserver reliability was analyzed with kappa analyses,which was interpreted as follows:value<0 denoted less than chance agreement,0.01-0.20 denoted slight agreement,0.21-0.40 denoted fair agreement,0.41-0.60 denoted moderate agreement,0.61-0.80 indicated substantial agreement,and 0.81-0.99 was considered almost perfect agreement[13].

    RESULTS

    A total of 50000 intraoral digital radiographs were acquired in the 2 year-period from April 2018 to April 2020.Of these,3550 (7.1%) retakes were performed due to the presence of image artifacts.Of these retakes,5% was related to operator errors and 2.1% was related to plate and scanning errors.The calculated kappa value for interobserver reliability was 0.99,indicating almost perfect interobserver agreement.

    Moreover,to reduce the radiation exposure of the patient,it is clinically important to avoid the occurrence of image artifacts in order to minimize repeated radiographs,which could happen through proper understanding of the reasons and solutions for incidence of image artifacts and thus lessen its frequency,especially in the recent commonly used dental digital imaging technology[14].

    Imaging artifacts were divided into three categories.Of the 3550 retakes,operator errors were the most common and were observed in 2500 images (70.4%),while plate errors were detected in 685 images (19.3%) and scanning errors in 365 images (10.3%).

    He could have made himself quite comfortable with a bed of soft moss2, but the fear of wild beasts disturbed his mind, and at last he determined3 to spend the night in a tree

    Operator errors

    In this study,the operator error category consisted of eight subtypes that were closely similar to the same error categories in conventional film radiography,as this type of artifact does not depend on the type of image receptor except for the reversed or mirror image,movement of the plate inside the packet,and double exposure in the plate either due to partial erasing of the previous image or failure of the image scanning.

    The cone cut was the most common error observed in 988 images (39.5%) out of 3550 images (Figure 1A),followed by the artifacts of improper PSP placement in the mouth (30.4%),projection geometry (22.4%),unexposed plate (4.5%),movement of phosphor plate in the disposable pocket (1.6%) (Figure 1A and B),reversed image(0.8%) (Figure 1c and d),overexposed (0.6%),and plate bending (0.3%).

    Plate errors

    Bite marks were the most common error observed in 276 images (40.3%),while image size reduction was detected in 174 images (25.4%).In addition,plate scratches were detected in 19.6% of cases,with other errors being contamination of PSP (6.5%),peeling of the coat (5.7%),plate damage (2.3%),and fingerprints (0.1%)(Figures 2,3 and 4).

    We sang all the way home from church. At lunch, Mom had a surprise for us. She had bought a dozen eggs, and we had boiled Easter eggs with our fried8 potatoes! Late that afternoon, the minister drove9 up in his car. Mom went to the door, talked with him for a moment, and then came back with an envelope in her hand. We asked what it was, but she didn’t say a word. She opened the envelope and out fell a bunch10 of money. There were three crisp11 twenty-dollar bills, one ten-dollar bill and seventeen one-dollar bills.

    Scanning errors

    It has been reported that the higher the light intensity and the longer the exposure time,the greater the loss of plate information[27].This was supported by the study of Ang DB in 2006,which reported that plates were not exposed to any light before scanning presented no change in image quality even after many days of storage[25,28].

    DISCUSSION

    The second most common errors in the current study were plate errors that were detected in 685 intraoral images (1.37% of total images and 19.3% of retakes),followed by scanning errors seen in 365 intraoral images (0.73% of total images and 10.3% of retakes).Chiu[23] 2008 reported that scanning errors were the second most common error,followed by plate errors.

    PSP and CCD digital systems are preferred over traditional films because of the lower radiation dose for both patients and operators,less time needed,ability to perform image manipulation as well as enhancement without the need for retake,better archiving,and environmental friendliness.

    One day the woman stood at the window3 overlooking the garden, and saw there a bed full of the finest rampion:4 the leaves looked so fresh and green that she longed to eat them.5 The desire grew day by day, and just because she knew she couldn t possibly get any, she pined away and became quite pale and wretched. Then her husband grew alarmed and said:

    And there are surprises. One time I came home to find a note on the front door that led me to another note, then another, until I reached the walk-in closet. I opened the door to find Scott holding a pot of gold (my cooking kettle) and the treasure of a gift package. Sometimes I leave him notes on the mirror and little presents under his pillow.

    PSP plates are more comfortable to the patient than CCD because they are cordless,more flexible,and thin,which resembles standard films.However,despite their superiority,PSP plates are more susceptible to bending and scratches,require more time for scanning,and subsequently develop more image artifacts,which consequently affects image quality[15].In previous studies,it has been stated that 95%of PSP plates used for 10 wk (used approximately 50 times) became non-diagnostic and needed to be substituted[16,17].

    Joyce Thomas comments on the trickery and imagery of the bone. The bone provides the imagery of deprivation112 and starvation, one of the primary themes of the tale. Also, the bone Hansel uses imitates the fate awaiting his flesh (the bone could well be the gnawed113 remains114 of the cage s previous occupant) (Thomas 1989).Return to place in story.

    Radiographic artifacts that occur with plain film radiography are well-identified and documented.However,to the best of our knowledge,few studies have reported and categorized the artifacts of PSP plates[17-21].Most of these studies investigated them in medical radiology;however,very few studies were correlated to the dental field[14,15,20,22].

    Every team requires unity. A team has to move as one unit, one force, with each person understanding and assisting the roles of his teammates. If the team doesn t do this, whatever the reason, it goes down in defeat. You win or lose as a team, as a family. A successful team walks onto the field with issues of race, religion and all societal pressures ratcheted down to inconsequential by the strength of common goals.

    Intraoral radiographs requiring retakes due to image artifacts included 2869 (80.8%)periapical (PA) images,518 (14.6%) bite wing images,and 163 blank images (4.6%).

    Therefore,the aim of this study was to detect the type,frequency,and reasons for the occurrence of intraoral image artifacts acquired by PSP plates in our dental clinics,and to propose probable methods to avoid these image artifacts.

    Although PSP plates can be exposed to regular light while being uncovered and imported into the scanner,it is recommended that exposure to regular light should not exceed 10 minutes,and that the scanning procedure should not be delayed more than that.Otherwise,the information in the plate will fade away[24],which will lead to increased signal loss and a reduction in the signal-to-noise ratio[25,26].It could appear as total image fading with too bright an image or non-uniform image density caused by partial exposure of PSP to excessive ambient light before scanning.In addition,fluctuating signal loss leads to a noisy image[24].

    In the present study,only 0.07% of 50000 intraoral images reported in the existing study presented with movement artifacts of the PSP plate inside the disposable packet.However,in Gulsahi and Secgin study in 2016,it was recorded frequently,whereas other studies did not report this type of artifact.It has been stated that this artifact is detected only with the Digora system,where both cardboard cover and disposable plastic envelopes should be used.While the cardboard cover is used to protect the PSP plates,they may unfortunately cause motion of the plates,resulting in these artifacts[24].

    Digital intraoral imaging systems have gradually replaced film-based imaging in recent years,as this technology has many advantages over conventional imaging.Two types of receptors are used for digital intraoral radiography:solid-state sensors of either CCD or CMOS,which are used with a wire and PSP plates that are cordless.Solid-state sensors have been used for more than two decades,while systems that use PSP plates have only recently been used in clinical practice.Each receptor type has its own advantages and disadvantages[2].

    In the current study,bite marks were the most common plate error subtypes observed in 276 images (40.3%).

    5. Wolf: The wolf has become a popular image in fairy tales thanks to this tale and The Tale of the Three Little Pigs. The wolf is a common predator16 in the forest and thus is a natural choice for the story unlike the witch, ogre or troll found in other tales. The wolf is often a metaphor17 for a sexually predatory man.

    The most probable cause of the increased incidence of bitemarks in our study is the disposable plastic packet.It is used for infection control and acts as a light barrier,but it does not provide proper safety against plate damage from bite marks,bending,or pressure.Another possible reason for plate bitemarks would be the patient unintentionally biting on the plate.Thus,the patient must be comprehensively instructed to avoid doing so before the exposure.

    That was 34 years ago. Today, the boy is a successful architect in New York City. His father’s cabin is still there on the island in the middle of the lake. He takes his own son and daughters fishing from the same dock.

    It was reported that pediatric periapical radiographs of primary incisors that were taken by a modified technique by making the child bite on a size 2 PSP plate to keep it in place during exposure simulating occlusal radiograph could be one of the reasons for PSP bite marks.Snap-A ray film holding devices with teeth-like edges were used.Roberts,Mol,2004 reported that the risk of teeth biting in pediatric primary incisors can be reduced by placing “adhesive backed sponge-like pads” on each side of the disposable sleeve that holds the reusable size 2 plate[2].

    Image size reduction artifact after scanning was reported in 174 images (25.4% of retakes) and 0.3% of all intraoral images,which can be related to scanning errors.This was also reported by various studies[20].

    In the present study,PSP plate scratches were detected in 19.6% of patients,while surface contamination of PSP was seen in 6.5%.To the best of our knowledge,the actual reasons for plate surface contamination and scratches have not been completely recognized and investigated.Kalathingal[17] 2010 stated that the possible cause of plate surface contamination could be the adhesive utilized in the barrier sheath,which could have affected the plate before scanning.The same study found that the plates that were used in dental colleges,as in our study,were more susceptible to scratches because of the increased number of people handling the plate,especially when they were discovered after scanning and before packaging the plate with a new protective sheath.Also,Kalathingal[17] 2010 reported that the hard rubbing of PSP plates with alcohol could be a cause for more scratches,and recommended that only the plates with visible surface contamination should be lightly wiped to remove any contaminants[17].

    Naturally a great scramble70 ensued, and at last the laughter and shouting awoke the Queen, who rang for her maids to ask the reason of such an unwonted hurry-burly

    In the current study,peeling of the coat was found in 5.7%,plate damage was recorded in 2.3%,and fingerprints were found in 0.1% of plates.These findings were in the line with those reported by Gulsahi and Secgin in 2016,wherein they found damaged plates to occur in only a few images[20].The most probable causes of plate damage are tough handling of PSP plates during their placement in the mouth,extensive bending,forceful placement of the plate into the scanner,or mechanical stresses exerted by the scanner roller during scanning[15].

    Scanning errors were detected in 365 images (0.73% of all intraoral images),where delayed scanning was the most common error in this type and was seen in 145 images(0.29%).Despite Gulsahi and Secgin[24] reporting 12.6% of all images exhibited nonuniform brightness in their 2016 study,Chiu[23] recorded 0.4% delayed scanning errors in their 2008 study.

    In the present study,the examined intraoral radiographic errors were categorized into three main categories:operator,plate,and scanning errors.Operator errors were the most common and were observed in 2500 out of 50000 intraoral images (5%) and in 70% of 3550 retakes,which is in accordance with the findings of Hui-Lin Chiu[23]2008.regarding the increased incidence of operator errors as seen in the current study.In addition,this reflects the need for broader training of radiology technicians.

    In the current study,the most common subtype of scanning error was delayed scanning (39.9% of retakes),which was comparable to the results of ?al??kan and Sumer in their study in 2017,who affirmed that the probable cause for this artifact is the elimination of the plate from its cover after radiation exposure to inhibit infection before scanning[24].

    Delayed scanning artifacts (non-uniform density or bright image) were the most common errors observed in 145 images (39.9%).White (radiopaque) lines (37.2%),blank images (13.4%),black (radiolucent) lines (5.5%),and double images (incomplete or partial erasing) (4.1%) were also seen.(Table 1,Figures 5 and 6).

    The second scanning error was the presence of white lines in 136 images (37.2%),which were reported to be due to dust or dirt particles on the slim scanning opening of the scanner.This caused them to remain fixed during the scanning process or dirt on the rollers,which acted as a blockage to the laser light.Thus,resulting in production of areas devoid of signals.Therefore,the mechanical scanner transport system should be checked and cleaned regularly,with additional replacement of the belt,if necessary[15].

    Aha! they both exclaimed3. You ARE from Iraq! Tara smiled and said yes. Then she apologized to both of them for lying the day before. She explained that she had not wanted to get into an Arabic conversation with them. It had been her experience that many ESL students continued to speak their native language in ESL class, and Tara had not come to ESL class to practice her Arabic. In her opinion, ESL students should try to speak English only.

    The black or radiolucent lines that were detected in 20 images (5.5%) were hypothesized by a previous study to have resulted from electromagnetic interfering artifacts that were caused by anything that interfered,interrupted,decreased,or limited efficient scanner performance[15,29].

    Moreover,?al??kan A and Sumer AP[26] denoted this artifact as a ridging artifact,and advocated that these black lines occurred due to fast variations in the intensity of the light of the stimulating lasers as well as the loss of harmonization between rapid scan cycles and image plate movement.It is recommended to add proper electromagnetic shielding,appropriate voltage supply,uninterrupted power supply,and regular maintenance of scanner performance to avoid the occurrence of artifacts in radiolucent lines[15].

    The previously mentioned errors observed in our study could be limited by proper regular orientation and enforcement by all operators,technicians,and students of the appropriate procedures of gentle handling of the PSP plates during exposure,scanning,and after scanning.This also includes focusing on the correct light rubbing of the PSP plates only with visible surface contamination to increase their longevity and usability.Moreover,regular checkups of PSP plates should be regularly performed to check the integrity of the plates.In addition,periodic maintenance,cleaning,and calibration of scanning devices would significantly reduce the number of scanning errors.Furthermore,additional comprehensive training of radiology technicians regarding application of proper radiographic techniques using the PSP system would be very beneficial in reducing operator errors.

    CONCLUSION

    The use of digital imaging in dental practice has recently made a revolution in image recording and analysis.Our institution made a significant modification by digitizing all dental images recently acquired using mainly the PSP system.Our study investigated the prevalence of intraoral image artifacts in our clinics,which were characteristic of PSP plates wherein the most common artifacts were bitemarks,image size reduction,scratches,and delayed scanning.Defining the causes of these artifacts and identifying methods for preventing them are of great clinical significance.Further research on PSP artifacts for further error identification and proper handling is needed and is a process that is essential to produce superior diagnostic images needed for instituting proper dental care.

    ARTICLE HIGHLIGHTS

    Research background

    Nowadays,digital imaging outweighs conventional imaging and has been used widely in dentistry.Digital radiography allows image manipulation to adjust the visual characteristics of the image,such as contrast,brightness,and density,thus enhancing image quality without the need to retake the image.

    Research motivation

    Digital imaging provides an easier,comfortable,and user-friendly way for recording and interpreting radiographic data for archiving and teleradiography.

    Research objectives

    To detect the frequency,type,and reasons behind the appearance of intraoral image artifacts acquired by photostimulable phosphor plates (PSP).

    Research methods

    This retrospective descriptive study.A total of 50000 intraoral radiographs were retrieved from the clinical database from April 2018 to April 2020 to evaluate the reason,type,and solutions to these image artifacts.All intraoral digital radiographs were acquired using an intraoral X-ray machine with 7-mA,65-kVP using a PSP system and laser scanners,which can house all sizes of reusable intraoral PSP sensor plates with image acquisition software.

    Research results

    Imaging artifacts were divided into three categories;operator,plate,and scanning errors.Out of 3550 retakes,5%,1.37%,and 0.73% were related to the operator,plate,and scanning errors,respectively.The cone cut was the most common operator error(988 images),Bite marks were the most common plate error (276 images),and delayed scanning artifacts were the most common scanning errors (145 images).

    Research conclusions

    Our study discussed intraoral image artifacts that were characteristic of PSP,where the most common artifacts were bitemarks,image size reduction,scratches,and delayed scanning.

    Research perspectives

    Thus,recognizing intraoral radiographic image errors and defining the causes and their trouble-shooting are crucial factors in making images possess great clinical impacts.

    日韩伦理黄色片| 日本色播在线视频| 男人添女人高潮全过程视频| 韩国高清视频一区二区三区| 亚洲色图综合在线观看| 美女国产视频在线观看| 久久99蜜桃精品久久| 精品国产露脸久久av麻豆| 亚洲在线观看片| 蜜桃亚洲精品一区二区三区| 一区二区三区免费毛片| 国产精品精品国产色婷婷| 自拍偷自拍亚洲精品老妇| 亚洲av二区三区四区| 99re6热这里在线精品视频| 人妻 亚洲 视频| 免费看日本二区| 国产老妇伦熟女老妇高清| 亚洲aⅴ乱码一区二区在线播放| 久久精品综合一区二区三区| av播播在线观看一区| 国产亚洲5aaaaa淫片| 全区人妻精品视频| 国产在线男女| 亚洲精品一区蜜桃| 蜜桃亚洲精品一区二区三区| 水蜜桃什么品种好| 亚洲av一区综合| 一级a做视频免费观看| 3wmmmm亚洲av在线观看| 午夜福利高清视频| 国产男女超爽视频在线观看| 国产伦精品一区二区三区四那| 蜜桃久久精品国产亚洲av| 99久久精品国产国产毛片| 国产精品久久久久久av不卡| 久久影院123| 日本一二三区视频观看| 亚洲精品国产色婷婷电影| 成人国产av品久久久| 麻豆精品久久久久久蜜桃| 国产国拍精品亚洲av在线观看| 亚洲精品亚洲一区二区| 最近中文字幕2019免费版| 夫妻性生交免费视频一级片| 亚洲人成网站在线观看播放| 别揉我奶头 嗯啊视频| 久久精品夜色国产| 男插女下体视频免费在线播放| 国产黄片视频在线免费观看| 一本—道久久a久久精品蜜桃钙片 精品乱码久久久久久99久播 | 欧美精品一区二区大全| 亚洲欧美日韩东京热| 欧美激情国产日韩精品一区| 久热这里只有精品99| 美女被艹到高潮喷水动态| 国产精品国产三级国产专区5o| 高清欧美精品videossex| 午夜激情久久久久久久| 亚洲伊人久久精品综合| av国产免费在线观看| 91午夜精品亚洲一区二区三区| 国产精品.久久久| 欧美老熟妇乱子伦牲交| 中国美白少妇内射xxxbb| 中文字幕久久专区| 又爽又黄a免费视频| 最近的中文字幕免费完整| 天天躁日日操中文字幕| 欧美激情在线99| 午夜免费鲁丝| 精品久久国产蜜桃| 有码 亚洲区| 色吧在线观看| 国产精品av视频在线免费观看| 深夜a级毛片| 哪个播放器可以免费观看大片| 亚洲精品亚洲一区二区| 亚洲精品久久久久久婷婷小说| 国产高潮美女av| 天堂中文最新版在线下载 | 美女脱内裤让男人舔精品视频| 一个人看视频在线观看www免费| 又黄又爽又刺激的免费视频.| 97在线人人人人妻| 内地一区二区视频在线| 亚洲精品成人av观看孕妇| 午夜福利视频1000在线观看| av国产久精品久网站免费入址| 国产成人免费无遮挡视频| 日韩强制内射视频| 精品久久久久久久久av| 国产成人精品婷婷| 精品久久久久久久久亚洲| 黄片无遮挡物在线观看| 午夜福利在线在线| 男女边摸边吃奶| 在线看a的网站| 插逼视频在线观看| 在线a可以看的网站| 夜夜看夜夜爽夜夜摸| 亚洲无线观看免费| 国产毛片在线视频| 国产一区有黄有色的免费视频| 如何舔出高潮| 婷婷色av中文字幕| 国产 精品1| 男人爽女人下面视频在线观看| 国产片特级美女逼逼视频| 一区二区三区乱码不卡18| 嫩草影院新地址| 青青草视频在线视频观看| 中国国产av一级| 嫩草影院新地址| 毛片一级片免费看久久久久| 欧美激情在线99| 嫩草影院入口| 一二三四中文在线观看免费高清| 少妇裸体淫交视频免费看高清| 99九九线精品视频在线观看视频| 噜噜噜噜噜久久久久久91| av一本久久久久| 99九九线精品视频在线观看视频| 久久综合国产亚洲精品| 久久ye,这里只有精品| 99久国产av精品国产电影| 一二三四中文在线观看免费高清| 久久97久久精品| 综合色丁香网| 久久久久久久精品精品| 插阴视频在线观看视频| 欧美成人a在线观看| av卡一久久| 久久午夜福利片| 最近手机中文字幕大全| 一区二区三区免费毛片| 久久久久九九精品影院| 老司机影院成人| 最近手机中文字幕大全| 简卡轻食公司| 丝袜脚勾引网站| 国产亚洲精品久久久com| 国产精品99久久99久久久不卡 | 久久久欧美国产精品| 天堂俺去俺来也www色官网| 国模一区二区三区四区视频| 在线观看美女被高潮喷水网站| 欧美另类一区| 日韩av免费高清视频| 黑人高潮一二区| 爱豆传媒免费全集在线观看| 国产精品av视频在线免费观看| 高清日韩中文字幕在线| 美女国产视频在线观看| av又黄又爽大尺度在线免费看| 在线观看一区二区三区| 男女下面进入的视频免费午夜| 又大又黄又爽视频免费| 男女下面进入的视频免费午夜| 草草在线视频免费看| 美女国产视频在线观看| 亚洲av成人精品一二三区| 街头女战士在线观看网站| 一本色道久久久久久精品综合| 亚洲国产欧美在线一区| 亚洲内射少妇av| 日韩成人av中文字幕在线观看| 午夜福利视频精品| 最近手机中文字幕大全| 51国产日韩欧美| 日日啪夜夜爽| 80岁老熟妇乱子伦牲交| 欧美激情久久久久久爽电影| 波野结衣二区三区在线| 纵有疾风起免费观看全集完整版| 成人黄色视频免费在线看| 日韩大片免费观看网站| 91久久精品国产一区二区三区| 久久影院123| 免费观看无遮挡的男女| 一本久久精品| 亚洲人与动物交配视频| 99热这里只有精品一区| 国产乱人偷精品视频| 国产精品国产av在线观看| 亚洲色图av天堂| 欧美三级亚洲精品| 日产精品乱码卡一卡2卡三| 26uuu在线亚洲综合色| 欧美精品国产亚洲| 在线观看美女被高潮喷水网站| 亚洲国产最新在线播放| 国产中年淑女户外野战色| 欧美日韩亚洲高清精品| 亚洲三级黄色毛片| 色5月婷婷丁香| 在线天堂最新版资源| 2022亚洲国产成人精品| 亚洲av免费在线观看| 国产免费一区二区三区四区乱码| 久久久久久久精品精品| 亚洲国产最新在线播放| 亚洲精品成人av观看孕妇| 国产大屁股一区二区在线视频| 色5月婷婷丁香| 国产爱豆传媒在线观看| 久久久久精品久久久久真实原创| 日韩三级伦理在线观看| 日本爱情动作片www.在线观看| 成年人午夜在线观看视频| 日本黄色片子视频| 精品亚洲乱码少妇综合久久| 久热这里只有精品99| 国产白丝娇喘喷水9色精品| 午夜福利高清视频| 男女下面进入的视频免费午夜| 国产亚洲最大av| 亚洲va在线va天堂va国产| 免费在线观看成人毛片| 亚洲天堂国产精品一区在线| 国产又色又爽无遮挡免| 大陆偷拍与自拍| av在线观看视频网站免费| 久久精品久久久久久噜噜老黄| 激情五月婷婷亚洲| 免费观看的影片在线观看| 国产一级毛片在线| 久久精品熟女亚洲av麻豆精品| 欧美激情在线99| 简卡轻食公司| 中文字幕免费在线视频6| 美女被艹到高潮喷水动态| 大陆偷拍与自拍| 欧美日韩视频精品一区| 高清av免费在线| 欧美xxxx黑人xx丫x性爽| a级毛色黄片| 97在线人人人人妻| 国产精品99久久久久久久久| 永久免费av网站大全| eeuss影院久久| 欧美性猛交╳xxx乱大交人| 亚洲欧美清纯卡通| 国产一级毛片在线| 日韩在线高清观看一区二区三区| 夜夜看夜夜爽夜夜摸| 精品国产乱码久久久久久小说| 国产视频首页在线观看| 亚洲av免费高清在线观看| 午夜精品国产一区二区电影 | 国产欧美亚洲国产| 亚洲精品乱码久久久久久按摩| 一级a做视频免费观看| av卡一久久| 亚洲精品成人av观看孕妇| 日本一本二区三区精品| 国产精品.久久久| 男人舔奶头视频| 3wmmmm亚洲av在线观看| 中国国产av一级| 亚洲四区av| 99热这里只有精品一区| 久久久久国产网址| 亚洲av欧美aⅴ国产| 日本色播在线视频| 国产亚洲午夜精品一区二区久久 | 中文资源天堂在线| 夜夜看夜夜爽夜夜摸| 久久国产乱子免费精品| 国产精品三级大全| 午夜福利视频1000在线观看| 亚洲人成网站在线播| 精品少妇黑人巨大在线播放| 国产乱人视频| 一本色道久久久久久精品综合| 一二三四中文在线观看免费高清| 婷婷色av中文字幕| 看黄色毛片网站| 亚洲欧美精品自产自拍| 国产一区二区在线观看日韩| 日韩强制内射视频| 狂野欧美激情性bbbbbb| 国产精品麻豆人妻色哟哟久久| 国产熟女欧美一区二区| 欧美日韩精品成人综合77777| 精品一区在线观看国产| 国模一区二区三区四区视频| 免费看日本二区| kizo精华| 婷婷色麻豆天堂久久| 成人特级av手机在线观看| 久久久久精品性色| 久久久久久久亚洲中文字幕| 全区人妻精品视频| 免费av毛片视频| 女人十人毛片免费观看3o分钟| 国产亚洲精品久久久com| 国产av国产精品国产| 看非洲黑人一级黄片| 亚洲av不卡在线观看| 神马国产精品三级电影在线观看| 又黄又爽又刺激的免费视频.| 成年女人在线观看亚洲视频 | 亚洲精品国产色婷婷电影| 老司机影院成人| 国产免费又黄又爽又色| 免费观看av网站的网址| 免费播放大片免费观看视频在线观看| 99九九线精品视频在线观看视频| 青春草国产在线视频| 久久亚洲国产成人精品v| 国产黄色免费在线视频| 久久久精品94久久精品| 国产中年淑女户外野战色| 亚洲国产精品999| 国产女主播在线喷水免费视频网站| 亚洲国产av新网站| 六月丁香七月| 欧美高清性xxxxhd video| 白带黄色成豆腐渣| 3wmmmm亚洲av在线观看| 欧美zozozo另类| 狂野欧美白嫩少妇大欣赏| 国产成人a区在线观看| 欧美日韩一区二区视频在线观看视频在线 | 国产伦在线观看视频一区| av在线观看视频网站免费| 激情五月婷婷亚洲| 国产精品久久久久久精品电影小说| videos熟女内射| 久久精品人人爽人人爽视色| 少妇人妻 视频| 99九九在线精品视频| 国产深夜福利视频在线观看| 国产成人精品在线电影| 色吧在线观看| 日韩免费高清中文字幕av| 亚洲国产看品久久| 国产女主播在线喷水免费视频网站| 美女扒开内裤让男人捅视频| 久久久久网色| 在线观看免费视频网站a站| 高清黄色对白视频在线免费看| 各种免费的搞黄视频| 国产午夜精品一二区理论片| 女性生殖器流出的白浆| 久久久久精品久久久久真实原创| 在线观看免费高清a一片| 三上悠亚av全集在线观看| 久久女婷五月综合色啪小说| 精品国产一区二区久久| 亚洲,一卡二卡三卡| 美女视频免费永久观看网站| 精品人妻熟女毛片av久久网站| 久久99一区二区三区| 免费av中文字幕在线| 亚洲五月色婷婷综合| kizo精华| 国产熟女午夜一区二区三区| 黑人欧美特级aaaaaa片| 亚洲精品国产区一区二| 亚洲精品美女久久av网站| 看非洲黑人一级黄片| 精品卡一卡二卡四卡免费| 18禁观看日本| 国产精品av久久久久免费| 日韩av免费高清视频| 国产乱来视频区| 成人亚洲精品一区在线观看| 别揉我奶头~嗯~啊~动态视频 | 亚洲人成网站在线观看播放| 国产黄色免费在线视频| 搡老乐熟女国产| 女的被弄到高潮叫床怎么办| 叶爱在线成人免费视频播放| 国产成人欧美在线观看 | 叶爱在线成人免费视频播放| 国产成人欧美在线观看 | 久久99精品国语久久久| 亚洲精品av麻豆狂野| 免费黄网站久久成人精品| 丝袜美腿诱惑在线| 亚洲国产中文字幕在线视频| 国产黄色视频一区二区在线观看| 久久亚洲国产成人精品v| 欧美日韩福利视频一区二区| 男男h啪啪无遮挡| 国产成人系列免费观看| 丝袜人妻中文字幕| 丝袜美足系列| 国产成人系列免费观看| 亚洲激情五月婷婷啪啪| 成人影院久久| 在线观看www视频免费| 日韩,欧美,国产一区二区三区| 菩萨蛮人人尽说江南好唐韦庄| 在线 av 中文字幕| 亚洲人成网站在线观看播放| 制服诱惑二区| 国产成人免费无遮挡视频| 丝袜喷水一区| 纵有疾风起免费观看全集完整版| 精品第一国产精品| 99久久人妻综合| 亚洲七黄色美女视频| 男女国产视频网站| 999精品在线视频| 国产精品一区二区在线观看99| 爱豆传媒免费全集在线观看| 97精品久久久久久久久久精品| 久久av网站| 精品国产乱码久久久久久小说| 最近手机中文字幕大全| 考比视频在线观看| 看非洲黑人一级黄片| 日本欧美国产在线视频| 国产不卡av网站在线观看| 69精品国产乱码久久久| 国语对白做爰xxxⅹ性视频网站| av视频免费观看在线观看| 亚洲人成电影观看| 一区二区三区四区激情视频| 亚洲第一av免费看| a级毛片在线看网站| 丰满迷人的少妇在线观看| 亚洲国产精品国产精品| 成年美女黄网站色视频大全免费| 国产亚洲午夜精品一区二区久久| 秋霞在线观看毛片| 在线天堂中文资源库| 性少妇av在线| 国产 精品1| 19禁男女啪啪无遮挡网站| 亚洲一级一片aⅴ在线观看| 女人爽到高潮嗷嗷叫在线视频| 国产又爽黄色视频| 欧美日韩视频精品一区| 免费日韩欧美在线观看| 亚洲男人天堂网一区| 国产欧美日韩综合在线一区二区| 亚洲欧美中文字幕日韩二区| 国产成人一区二区在线| 日本vs欧美在线观看视频| 少妇人妻精品综合一区二区| 深夜精品福利| 熟女少妇亚洲综合色aaa.| 国产精品久久久久久精品古装| 欧美xxⅹ黑人| 亚洲精品久久午夜乱码| 无遮挡黄片免费观看| 亚洲精品久久久久久婷婷小说| 天美传媒精品一区二区| 国产xxxxx性猛交| 免费在线观看黄色视频的| av女优亚洲男人天堂| 九草在线视频观看| 国产激情久久老熟女| 嫩草影视91久久| 国产极品天堂在线| 美国免费a级毛片| 久久久久精品人妻al黑| 又粗又硬又长又爽又黄的视频| 丝袜喷水一区| 亚洲精品美女久久av网站| 国产一区亚洲一区在线观看| 国产一级毛片在线| 天天躁夜夜躁狠狠久久av| av电影中文网址| 1024香蕉在线观看| av片东京热男人的天堂| 亚洲精品在线美女| 老司机靠b影院| 大话2 男鬼变身卡| 亚洲,一卡二卡三卡| 亚洲,欧美精品.| 中文字幕色久视频| 男人舔女人的私密视频| 麻豆av在线久日| 亚洲欧美清纯卡通| 久久久久精品人妻al黑| 欧美中文综合在线视频| 极品少妇高潮喷水抽搐| 国产熟女午夜一区二区三区| 狂野欧美激情性bbbbbb| 国产高清不卡午夜福利| 色视频在线一区二区三区| h视频一区二区三区| 国产成人91sexporn| 婷婷色av中文字幕| 国产极品天堂在线| 在现免费观看毛片| 夫妻性生交免费视频一级片| 亚洲精品av麻豆狂野| 国产毛片在线视频| 一边亲一边摸免费视频| 日日啪夜夜爽| 少妇被粗大猛烈的视频| 亚洲精品成人av观看孕妇| 男女下面插进去视频免费观看| 国产免费一区二区三区四区乱码| 在线精品无人区一区二区三| 男女高潮啪啪啪动态图| 巨乳人妻的诱惑在线观看| 色婷婷av一区二区三区视频| 熟妇人妻不卡中文字幕| 街头女战士在线观看网站| 一二三四在线观看免费中文在| 制服丝袜香蕉在线| av国产久精品久网站免费入址| 欧美日本中文国产一区发布| 亚洲情色 制服丝袜| avwww免费| 中文字幕色久视频| 乱人伦中国视频| 亚洲色图 男人天堂 中文字幕| 久久久久国产精品人妻一区二区| 国产免费福利视频在线观看| 亚洲欧美清纯卡通| 久久精品人人爽人人爽视色| 人人澡人人妻人| 考比视频在线观看| 免费久久久久久久精品成人欧美视频| 国产精品麻豆人妻色哟哟久久| 日本vs欧美在线观看视频| 999久久久国产精品视频| 国产一区二区三区av在线| 亚洲欧美日韩另类电影网站| 国产女主播在线喷水免费视频网站| av国产久精品久网站免费入址| 成人免费观看视频高清| 亚洲 欧美一区二区三区| 人妻一区二区av| 搡老乐熟女国产| 亚洲精华国产精华液的使用体验| 一级片'在线观看视频| 午夜影院在线不卡| 免费不卡黄色视频| 韩国精品一区二区三区| 精品国产乱码久久久久久小说| 侵犯人妻中文字幕一二三四区| 在线看a的网站| 国产精品嫩草影院av在线观看| 精品国产一区二区三区久久久樱花| 啦啦啦视频在线资源免费观看| 人妻一区二区av| 欧美精品av麻豆av| 精品一品国产午夜福利视频| 国产成人欧美在线观看 | 777米奇影视久久| 熟妇人妻不卡中文字幕| av福利片在线| 晚上一个人看的免费电影| 成人手机av| 日韩中文字幕欧美一区二区 | 日韩人妻精品一区2区三区| 国产精品一区二区精品视频观看| 久久99一区二区三区| av女优亚洲男人天堂| 国产女主播在线喷水免费视频网站| 女性被躁到高潮视频| 男女下面插进去视频免费观看| a级毛片黄视频| 免费观看人在逋| 欧美人与性动交α欧美软件| 国产伦人伦偷精品视频| 亚洲精品国产av成人精品| 叶爱在线成人免费视频播放| 亚洲欧美清纯卡通| 欧美日韩亚洲综合一区二区三区_| 午夜福利影视在线免费观看| 丝袜美腿诱惑在线| 超碰成人久久| 国产女主播在线喷水免费视频网站| 国产精品久久久久久精品电影小说| 肉色欧美久久久久久久蜜桃| 五月天丁香电影| 亚洲国产精品一区二区三区在线| 国产老妇伦熟女老妇高清| 肉色欧美久久久久久久蜜桃| 不卡av一区二区三区| 亚洲欧美一区二区三区黑人| 日本欧美国产在线视频| 美女大奶头黄色视频| 97精品久久久久久久久久精品| av在线观看视频网站免费| 丰满饥渴人妻一区二区三| 制服诱惑二区| 国产成人欧美| 久久久精品国产亚洲av高清涩受| videos熟女内射| 中文精品一卡2卡3卡4更新| 国产又爽黄色视频| 亚洲精品乱久久久久久| 美女中出高潮动态图| 亚洲欧洲日产国产| 久热这里只有精品99| 国产日韩一区二区三区精品不卡| 国产探花极品一区二区| 91精品国产国语对白视频| 最近中文字幕2019免费版| 不卡av一区二区三区| 国产精品久久久久久人妻精品电影 | 国产成人免费观看mmmm| 新久久久久国产一级毛片| 自拍欧美九色日韩亚洲蝌蚪91| 午夜免费男女啪啪视频观看| www.熟女人妻精品国产| 国产成人一区二区在线| 最近中文字幕高清免费大全6| 人人妻人人澡人人爽人人夜夜| 女人爽到高潮嗷嗷叫在线视频| 亚洲一码二码三码区别大吗| 少妇人妻精品综合一区二区| 亚洲精品,欧美精品| 午夜精品国产一区二区电影|