    Design, Characterization, and 3D Printing of Cardiovascular Stents with Zero Poisson’s Ratio in Longitudinal Deformation

    2021-12-25 10:03:00ChengjinWangLeiZhangYongcongFangWeiSun
    Engineering 2021年7期

    Chengjin Wang, Lei Zhang,*, Yongcong Fang, Wei Sun,d

    a Biomanufacturing Center, Department of Mechanical Engineering, Tsinghua University, Beijing 100084, China

    b Biomanufacturing and Rapid Forming Technology Key Laboratory of Beijing, Beijing 100084, China

    c‘‘Biomanufacturing and Engineering Living Systems” Innovation International Talents Base (111 Base), Beijing 100084, China

    d Department of Mechanical Engineering, Drexel University, Philadelphia, PA 19104, USA

    ABSTRACT

    Keywords:Additive manufacturing 3D printing Screw extrusion Cardiovascular stent Zero Poisson’s ratio Inherent drawbacks associated with drug-eluting stents have prompted the development of bioresorbable cardiovascular stents. Additive manufacturing (3-dimentional (3D) printing) has been widely applied in medical devices. In this study, we develop a novel screw extrusion-based 3D printing system with a new designed mini-screw extruder to fabricate stents. A stent with a zero Poisson’s ratio (ZPR)structure is designed, and a preliminary monofilament test is conducted to investigate appropriate fabrication parameters.3D-printed stents with different geometric structures are fabricated and analyzed by observation of the surface morphology.An evaluation of the mechanical properties and a preliminary biological evaluation of 3D-printed stents with different parameters are carried out.In conclusion,the screw extrusion-based 3D printing system shows potential for customizable stent fabrication.

    1. Introduction

    With the application of a polymer coating that contains and controls the release of antiproliferative drugs, drug-eluting stents(DES)have become the gold standard for interventional treatment of coronary artery disease to date [1–3]. However, the backbone materials of DES are non-degradable metal materials that will cage the vessel permanently,which may lead to several inherent drawbacks, such as late restenosis, stent thrombosis, and chronic inflammation, and may impede the restoration of physiological function [4,5].

    Problems associated with DES have prompted the development of bioresorbable stents(BRS),which provide transient support and then disappear;thus BRS can reduce the risk of late stent thrombosis and avoid prolonged dual antiplatelet therapy [6,7]. The conventional manufacturing process for metal stents is laser micro cutting, which has also been applied in the processing of BRS.Stepak et al. [8] investigated the influence of laser parameters on the quality and geometry of polylactide-based polymer stents.Guerra et al. [9] explored the feasibility of a 1.08 μm wavelength fibre laser to cut polycaprolactone (PCL) sheets, as a feasibility study for stents manufacturing. Nevertheless, laser processing is a thermal process that may lead to thermal damage, such as heat-affected zones, micro-cracks, tensile residuals, and dross deposition [10,11]. More importantly, laser cutting, as a kind of subtractive manufacturing method,is not suitable for personalized customization.

    It has been observed that additive manufacturing (also known as 3-dimentional (3D) printing) has evolved into a versatile technology and economical solution for medical applications and customizable fabrication. Kaesemeyer et al. [12] used a rapid stent fabrication system with an extrusion system and a four-axis motion system to fabricate bioresorbable polystatin stents comprising lactide, glycolide, caprolactone, and lovastatin(60:15:10:15 parts by weight).Park et al.[13]evaluated the properties of a 3D-printed drug-coated PCL stent that was implanted into the porcine femoral artery.Guerra et al.[10]presented a novel 3D additive manufacturing machine to produce BRS with PCL filament material,and investigated the effects of the process parameters on the stent’s features.Wu et al.[14]manufactured polylactic acid (PLA) stents with a negative Poisson’s ratio (NPR) structure using a commercial fused deposition modeling (FDM) printer and PLA filaments, and studied the radial compressive property of the PLA stent. Although the results of these studies were promising,limitations remain regarding the stent’s shape. The BRS fabricated in the aforementioned studies have a cylindrical shape and a consistent diameter.In addition,stents generally undergo longitudinal foreshortening(a contraction in the length of the stent as the stent dilates).Clinicians must select stents with a longer length than the length of the plaque geometry to be stented. However, the degree of radial expansion and the final length of stents vary from person to person. Thus, the selection of a stent relies on the experienced judgment of the physician, which may go wrong and lead to malposition of the stent.Even more seriously,too much foreshortening would generate friction between the stent end and the artery,which would injure the delicate endothelium [15].

    In this study,we propose a method for fabricating cardiovascular stents;we also develop a novel screw extrusion-based 3D printing system that includes a newly designed mini-screw extruder.A novel design of a stent with zero Poisson’s ratio (ZPR) structure is introduced. Preliminary monofilament testing was conducted to investigate the fabrication parameters of the home-made extruder.Subsequently, stents with different materials, diameters, and geometric parameters were fabricated by changing the diameter,length, and shape of the mandrel. The surface morphology and the effects of geometric and fabrication parameters on the mechanical properties of the 3D-printed PCL stents were studied.

    2. Design of stents with a zero Poisson’s ratio structure

    At present,the majority of commercially available stents have a Z-shaped sequential-rings structure.This kind of design consists of two parts: Z-shaped rings (known as ‘‘struts”) and connectors(known as‘‘bridges”)[16].The Z-shaped rings are meant to expand during deployment and to support the vessel from collapse after the withdrawal of the balloon catheter. The connectors give the stent the ability to travel through the curvature of the arteries.

    To solve the longitudinal foreshortening problem, we designed a stent with a ZPR structure,as depicted in Fig.1.The structure unit(Fig. 1(a)) is similar to the semi-re-entrant honeycomb structure reported by Attard et al. [17], which exhibits a ZPR property. The Poisson’s ratio (υ) is defined as the ratio of the transverse strain(εi) and the axial strain (εj) in the direction of the loading force.The Poisson’s ratio can be expressed as follows:

    The structural unit of stent geometry can be defined by the following parameters:l represents the length of the inclined rib of the Z-shaped element;θ represents the angle of the Z-shaped element;h represents the length of the connector unit;and t represents the diameter (or width) of the strut. With the grid shown in Fig. 1(b)rolled up into a tube, as shown in Fig. 1(c), D represents the diameter of the stent and H represents the total length of the connectors. Here we introduce N1and N2as the number of structural units in the circumferential direction and axial direction,respectively.

    When stents are dilated or compressed, the force is loaded in the radial direction. In Figs. 2(a) and (b), the angle θ changes with an increase of r (the radius of the stent), which may be written as follows:

    For loading in the radial direction (Oxr), which results in an infinitesimal change dθ in the angle θ,the strain in the Oxidirection can be expressed as follows:

    where δxiis the deflection in Xidirection and i = r, l.

    As can be seen from Fig.2(c),the change in the diameter of the stent when it is dilated mainly depends on the deformation of the inclined rib of the Z-shaped ring; therefore, we focus here on the deformation of the inclined rib under loading. An inclined rib of length l can be likened to a cantilever beam that is loaded and guided at one end and fixed at the other(Fig.2(d))[18].The deflection of the rib in the x direction (δx) can be decomposed into δ⊥xgenerated by the bending moment and δ||xproduced by the component of the force F in the direction parallel to the rib.The deflection δ⊥xof the guided end and the deflection δ||xof the rib [19,20] are given by the following:

    where E0is the Young’s module of the material, I=πt4/64 is the second moment of the rib, t is the diameter of the stent strut, and A=πt2/4 is the area of the cross-section of the rib.

    Fig.1. Design of stent with ZPR structure.(a)The structural unit;(b)2D expanded drawing of stent;(c)3D model of stent.l0:the half of the distance between the adjacent ribs.

    Fig.2. Schematic diagram of deformation of Z-shaped ring and an inclined rib.(a)A Z-shaped ring;(b)an inclined rib;(c)deformation of a structural unit cell under loading in 2D model; (d) deformation of an inclined rib considering moments and axial force.

    In this study, l, t, and θ are usually set as 1.5 mm, 0.2 mm, and 60°, respectively. Thus, we can find that

    This indicates that in this study, δ||xis negligible relative to δ⊥x,which means that the deformation stretching of the rib can be viewed as negligible. Thus, the deflection of the rib in the x direction (δx) can be written as follows:

    The force F can be written in terms of the applied stress as F =σth. Using the standard definition of the Young’s modulus,and substituting for I, the equivalent elastic modulus of the rib in the x direction is

    Here,we substitute l,t,h,and θ as 1.5 mm,0.2 mm,1.5 mm,and 60°, and find that

    From Eqs.(1)–(13),we can infer that during the stent expansion process, only deformation of the stent structure (the ribs) occurs,and the stent material itself will not deform. Thus, the Poisson’s ratio of structure in Fig. 1(c) for loading in the radial direction can be written as follows:

    Because dXl/dθ=0, it is obvious that vrl=0.

    Thus, in this application, the direction of the loading force is parallel to the circumferential direction (Oxl) in the twodimensional (2D) expanded drawing of the stent (Figs. 1(b) and 2(c)) when the stent is dilated. Introducing the ZPR structure(v =0) into stent geometry provides an ideal situation in that the length of the stent does not change in the expansion process.

    3. A 3D printing system based on micro-extrusion technology

    3.1. System configuration

    Many of the polymers that are widely used in medical devices,such as poly(L-lactide) acid (PLLA) and PCL, are thermoplastic materials. FDM as a kind of extrusion-based 3D printing technology that has been widely applied in processing polymer materials.Conventional FDM equipment is based on filament extrusion and the three-axis printing technology, so it requires raw materials to be formed into filaments before being printed. Furthermore, frequent failures of filament buckling during the filament extrusion process result in interruption of the extrusion and necessitate numerous operator interventions [21,22]. In contrast, screw extrusion equipment offers better flexibility in the choice of raw materials. Wang et al. [23] developed a precision extruding deposition (PED) system consisting of a screw-extruder to fabricate PCL tissue scaffolds; the presented PED system directly extrudes materials in its granulated or pellet form without the filament preparation, and deposits freeform. This configuration opens up the opportunity for the use of a wider range of materials.Nonetheless,when the hollow structure and complicated pattern of a stent are taken into consideration,this kind of layer-by-layer deposition method has some limitations, as it is inevitably necessary to print numerous support structures. In addition, it is worth mentioning that the molecular weight of the polymer material decreases with increasing residence time of the melt inside the extruder,so a short and miniature-sized extruder is desirable in the process of biopolymer extrusion [24].

    To alleviate these problems, we proposed a screw extrusionbased 3D printing system with the addition of a rotation axis(the fourth axis) and designed a mini-screw extruder. The complete process of stent fabrication is presented in Fig. 3(a).The screw extrusion-based 3D printing system consists of three main components: an x–y–z motion system; a rotation axis(rotating mandrel); and a mini-screw extruder (Fig. 3(b)). The x–y–z motion system adopts a gantry structure, and Panasonic AC servo motors (Japan) are applied for the x-, y-, and z-axes.The shape and length of the rotating mandrel are flexible and can be changed accordingly. To ensure deposition accuracy, the movement of the x–y–z motion system, the rotational velocity of the mandrel, and the rotational velocity of the miniextruder are synchronized based on a programmable multiple axis controller (PMAC) provided by OMRON’s Delta Tau Data Systems (USA).

    To ensure the processing quality, the mini-extruder should be taken into consideration as a crucial component in the fabrication system. The structure and main components of the mini-extruder designed in this study are shown in Fig. 3(c). The heating element is attached to the outer surface of the barrel to provide heat for the material. Thus, when the mini-extruder operates, materials in the form of pellets or powder can be added into the extruder through the material inlet and will then be transported downward by the motor-driven screw.The chamber between the screw and the barrel is filled with molten material.Finally,the pressure generated in the chamber helps the molten material to be extruded through the nozzle tip as filaments.

    3.2. Research on printing parameters

    To determine a suitable temperature to fabricate 3D-printed stents, extrusion of PCL (material safety data sheet (MSDS)440744, Sigma-Aldrich, USA) monofilament was first conducted using the home-made mini-extruder in order to observe the forming quality of the extruded filaments.Considering that the width of the stent strut is 100–400 μm,the diameter of the nozzle tip outlet(φnozzle) was set as 250 μm. Differential scanning calorimetry(DSC) of the raw PCL material was performed using a DSC Q2000 instrument (TA Instruments, Italy) in the temperature range of-80–200°C,at a heating rate of 10°C?min-1,and under a nitrogen flow. Uniaxial tensile tests of the PCL monofilament samples were carried out with a WDW 3020 universal testing machine (Changchun Kexin Test Instrument, China), with a load cell of 50 N and a crosshead displacement of 1 mm?min-1at room temperature.

    The DSC experiment with the raw PCL material shows that PCL has a low melting temperature (Tm) of about 60 °C and a wide range of processing temperatures (Fig. 4(a)). Generally, for the material to melt sufficiently,the heating temperature of the extruder (Textruder) should be set above the melting point (Tm) of the material.As shown in Fig.4(b),PCL monofilaments were fabricated at five different temperatures between 75 and 85 °C (75, 77.5, 80,82.5, and 85 °C), with the extruder’s rotation speed set at 11.2 r?min-1. The samples produced by these temperatures were termed T75, T77.5, T80, T82.5, and T85, respectively. As shown in Fig. 4(c), the filament diameters (φfilament) of samples T75, T77.5,T80, T82.5, and T85were determined to be (319.33 ± 5.65),(318.03±6.69),(311.14±2.21),(312.18±3.19),and(315.10±9.48)μm,respectively.The tensile strength of samples T75,T77.5,T80,T82.5,and T85was determined to be(16.97±0.80),(15.53±1.73),(16.22±0.56),(15.04±0.96),and(14.87±1.55)MPa,respectively.

    Fig.3. The custom-made screw extrusion-based 3D printing system.(a)Configuration of stent fabrication system;(b)schematic diagram of 3D printing system;(c)structure and main components of the new-designed mini-extruder.

    Fig. 4. Investigation of the influence of heating temperature on the extrusion of PCL monofilament. (a) DSC curve of PCL; (b) PCL monofilaments fabricated at five different temperatures between 75 and 85°C;(c) the effect of temperature on the tensile strength and diameter of monofilaments;(d)scanning electron microscope(SEM)image of PCL filament extruded at 80 °C (SEM, FEI Quanta 200, the Netherlands). Error bars in (c) represent means ± standard deviations. HV: high voltage; Mag: magnitude.

    4. Results

    4.1. Stent fabrication

    As reported by Liu et al.[25],in the hot melt extrusion process,when the extruder motor is closed at the end of a line, the screw stops rotating immediately. However, material continues to be extruded from the nozzle tip because the pressure generated inside the extruder is not released.This phenomenon is called‘‘leakage of materials” and has an adverse effect on fabrication structure. Taking the design of the stent into consideration, the planning of the printing trajectory is crucial in the process of stent fabrication. If repeated starting and stopping of the extruder can be avoided—which also means that there is no need to suddenly move the extruder from one point to another during the printing process—then leakage of materials can be effectively eliminated.

    Accordingly, based on the design of the stent, auxiliary lines(gray lines in Fig. 5(a)) were introduced into the printing strategy.The 2D expanded drawings of the printing trajectory are shown in Fig. 5(a), where the direction of the arrows represents the moving direction of the extruder in the printing process. In this printing strategy, the molten filaments are deposited into a series of Zshaped lines (red lines) along the circumferential direction in the first layer. Adjacent Z-shaped lines (red lines) are connected with a ‘‘bridge” line (short black lines) at the end points. In the second layer,parallel straight lines(long black lines),which have a length of H, are printed along the axial direction, and adjacent straight lines (long black lines) are connected with auxiliary lines (gray lines). The materials fuse together at the overlap points of the Zshaped lines (red lines) and straight lines (long black lines).According to the forming method of intersection points,this printing method is called the ‘‘overlapping printing method” in this study.

    The designed screw extrusion-based 3D printing system with a mini-extruder was then employed to fabricate stents. The molten material filament was extruded from the nozzle and deposited onto the mandrel. With the synchronized movement of the x–y–z-axes and the rotation of the mandrel, the circumferential and axial filaments were collected successively to complete the fabrication of a stent. Stents were fabricated in a customized manner by changing the diameter and shape of the mandrel and the geometric structure of the stent model. A schematic of the stent fabrication process is shown in Fig. 5(b). The fabrication temperature was set at 80 °C. PLLA (RESOMER L210S, Evonik,Germany) stents and PCL (MSDS 440744, Sigma-Aldrich, USA)stents with different geometric structure were also fabricated(Figs. 5(c)–(e)).

    Table 1 The width of the stent strut based on different movement velocity and extruder rotation speed.

    Fig. 5. Fabrication of 3D-printed bioresorbable cardiovascular stents. (a) Printing trajectory of stents; (b) schematic of the stent fabrication process; (c) PLLA stents with different geometry structure; (d, e) tapered PCL stents; (f)–(j) images of stents listed in Table 2 (A1–E3).φ: the diameter of stents (unit: mm);ω: the angular velocity.

    To determine the influence of the stent diameter(φ),the length of the connector unit(h),and the extruder rotation speed(n)on the mechanical properties of the 3D-printed stents,five series(A–E)of 3D-printed PCL stents were fabricated (Figs. 5(f)–(j)). The experimental groups and parameters are shown in Table 2. For different groups of stents, l, θ, and H were constant. In this study, we set l=1.5 mm,l0=π/4-0.75,θ=60°,and H=15 mm.Thus,according to πφ=2N1(lcos(θ)+l0) and H =N2h (Fig. 1), when φ = 2, 3, and 4 mm, N1is 4, 6, and 8, respectively, and when h = 1.00, 1.25,and 1.50 mm, N2is 15, 12, and 10, respectively.

    4.2. Morphological analysis

    The surface morphology of the stents was observed by scanning electron microscopy (SEM, FEI Quanta 200, the Netherlands) at an accelerating voltage of 15 kV. Before SEM analysis, the samples were sputter-coated with gold for 60 s. As shown in Fig. 6(a), the stents developed with the‘‘overlapping printing method”generally showed good surface quality. The intersection points of the stents fabricated with the ‘‘overlapping printing method” are fused by two filaments. It can be seen from the SEM image (Fig. 6) that the PCL monofilaments melted and combined together at theintersection points,ensuring that the stent has good continuity and strength at these points.

    Table 2 Experimental groups and parameters of 3D-printed PCL stents.

    Fig. 6. SEM images of stent fabricated with ‘‘overlapping printing method.” (a) Mag 85×; (b, c) Mag 200×.

    In addition,a Keyence digital microscope(VHX-500,Japan)was used to observe the surface morphology and measure the width of the stent strut (t). The rotation speed of series D, A–C, and E was 8.4, 11.2, and 14 r?min-1. As depicted in Fig. 7, the average strut width of series D, A–C, and E was (220 ± 12), (362 ± 42), and(452 ± 35) μm. Specifically, the average width of with the stents in groups A1–A3 and groups B1–B3 was (349 ± 28) and(334 ± 23) μm, and the average width for the stents in groups C1–C3 was (404 ± 38)μm.

    4.3. Mechanical properties

    A three-point bending test and radial compression test were performed on the 3D-printed PCL stents listed in Table 2 (A1–E3)by installing the corresponding test apparatus (Figs. 8(a) and (b))on an INSTRON 3365 testing machine (USA) at room temperature according to American Society of Testing Materials (ASTM)F2606-08 and ASTM F3067-14, respectively [26,27]. In addition, a radial expansion test of the groups C1 and C2 stents was performed by means of a dilated balloon (Fig. 8(c)).

    4.3.1. Three-point bending test

    The three-point bending test is a common means of investigating stent’s bending flexibility, which affects the conformability of the stent (Fig. 9). A stent with good bending flexibility can easily be delivered through the curvature of the artery.Bending flexibility can be represented by bending stiffness: The higher the bending stiffness,the lower the flexibility.According to its definition,bending stiffness can be calculated by E ?I, where E and I are elastic modulus and area moment of inertia.E ?I is given by the following equation [26,28]:

    where F, L, and f are the bending force, bending span length, and deflection, respectively. In this test, L is set as 11 mm; thus, when the deflection is set at a certain value,stents with less bending load(F)have better bending flexibility.When the deflection is 2 mm,the loading force of the stents of groups A,B,and C is shown in Fig.9(d);the force of groups B, D, and E is shown in Fig. 9(h).

    In Figs. 9(a)–(d), the bending flexibility of different groups is B1 < A1 ≈C1, A2 < B2 < C2, and A3 < B3 < C3. As the length of the connector increases from 1 to 1.5 mm, for the φ2, φ3, and φ4stents, the ratio of the loading force between groups C1/A1, C2/A2, and C3/A3 is 1.01, 0.77, and 0.49, respectively. As shown in Figs. 9(e)–(h), the bending flexibility of different groups is E1 < B1 < D1, E2 < B2 < D2, and E3 < B3 < D3.

    Fig. 7. The width of stent strut of five series (A–E). (a)–(e) Photographs of stents with size marking taken by Keyence VHX-500 microscope; (f) the width of strut of different groups (A1–E3). Error bars in (f) represent means ± standard deviations.

    Fig. 8. Mechanical properties tests of stents. (a) Three-point bending test; (b) radial compression test; (c) radial expansion test.

    Fig. 9. Force-versus-deflection curves for 3D-printed PCL stents with different geometric unit parameter and fabrication parameter: for (a), (b), and (c), the length of connector(h)is 1,1.25,and 1.5 mm,respectively;(d)the loading force in(a),(b),and(c)of stents with different diameter and length of connector(h)when the deflection is 2 mm;for(e),(f),and(g),the extruder’s rotation speed(n)is 8.4,11.2 and 14 r?min-1,respectively;(h)the loading force in(e),(f),and(g)of stents with different diameters and fabrication parameters (n) when the deflection is 2 mm. Error bars in (d) and (h) represent means ± standard deviations.

    Table 3 Bending stiffness values of 3D-printed stents.

    The bending stiffness values (average value) of the stents at a deflection of 2 mm are given in Table 3. In contrast, the A1, C1,C3, and D1–D3 stents have relatively better bending flexibility. In a previous publication [29], the expanded Abbott Absorb GT1 stents (diameter: 3.0 mm, length: 18 mm) were shown to have a bending stiffness value of 4.20 N?mm2,which is close to the bending stiffness values of the A1, C1, C3 and D1–D3 stents. This indicates that within these 15 groups of stents, the A1, C1, C3, and D1–D3 stents have applicable bending flexibility.

    4.3.2. Radial compression test

    The Radial compression test was applied to evaluate the radial strength, which determines the radial support force of the stent.Stents with good radial strength can provide better support.Radial strength can be represented by radial stiffness: The higher the radial stiffness, the lower the radial strength. Schmidt et al. [30]suggested that radial stiffness is determined from the profile decreasing in a physiologically relevant range from 0 to 0.2 bar(1 bar=105Pa)outer pressure(Eq.(16)).ASTM F3067-14[27]suggests that when using an area normalized force,the instantaneous stent diameter should be used (instead of the starting diameter),multiplied by the initial length (Eqs. (17)–(18)). Accordingly, the radial stiffness is the compression ratio when the instantaneous pressure acting on the stents is 0.2 bar.

    A stent’s radial strength is mainly related to the width (t) and number of Z-shaped rings. In Figs. 10(a)–(d), it can be seen that the radial stiffness of the different groups is Bi > Ci > Ai (i = 1, 2,3); thus, the radial strength of the different groups is Bi < Ci < Ai(i = 1, 2, 3). From Figs. 10(e)–(h), it is obvious that the radial stiffness of the different groups is Di > Bi > Ei (i = 1, 2, 3); thus, the radial strength of the different groups is Di < Bi < Ei (i = 1, 2, 3).

    In a previous publication [30], the radial stiffness varied from 0.51% to 3.99%, which is close to the radial stiffness values of the A2, A3, C2, C3, and E3 stents. This indicates that within these 15 groups of stents, the A2, A3, C2, C3, and E3 stents have applicable bending flexibility.

    4.3.3. Radial expansion test

    Samples(the C1 and C2 stents)were expanded from the original mode until the expansion ratio became 10% and 20%, respectively.Group C1 stents with a diameter of 2 mm were expanded into about 2.2 and 2.4 mm, respectively. Group C2 stents with a diameter of 3 mm were expanded into about 3.3 and 3.6 mm,respectively. After that, the radial forces were removed and the stents recoiled in the radial direction.The diameter of each sample in the original mode (Oφ), expansion maximum mode (Mφ) and recoil mode(Rφ)is shown in Fig.11(a).The radial recoil ratio after removing the radial force can be calculated by the following:

    The average radial recoil ratio of the C1 stents expanded to 110% and 120% was 69% and 41%, respectively, while that of the C2 stents was 77% and 71%. The extent of the radial recoil of the 3D-printed PCL stents was unsatisfactory, which is related to the excellent elongation rate of PCL material. To resolve this problem,other biomaterials will be explored in future studies.

    The length of each sample in the original mode (OL), expansion maximum mode (ML), and recoil mode (RL) is shown in Fig. 11(b).The longitudinal foreshortening ratio when the stents were expanded into the maximum mode can be calculated by the following:

    The longitudinal foreshortening ratio of the C1 stents expanded to 110%and 120%was 0.33%and 0.13%,respectively, while that of the C2 stents was 0.40%and 0.09%.In contrast,current commercial stents have a shortening rate of 3%–5%[31].Thus,it can be verified that the stent geometry designed in this study can achieve a substantially steady axial length after being expanded.

    4.4. Biocompatibility

    Hemolysis analysis of the 3D-printed stents was conducted in accordance with ASTM F756-17[32].Fresh human blood obtained from voluntary donors was legally supplied by Beijing Huaxin Hospital, the First Affiliated Hospital of Tsinghua University. The hemolysis rate was determined by measuring hemoglobin release under static conditions,as previously reported[33].The hemolysis rate was calculated according to Eq. (21):

    where ODtrepresents the optical density of the test group,and ODpand ODnrepresent the absorbance of the positive and negative groups,respectively.According to ASTM F756-17 [32], the hemolysis of materials is classified as non-hemolytic(0–2%),slightly hemolytic (2%–5%), or hemolytic (> 5%). The hemolysis rate of the 3Dprinted PCL was 0.07% ± 0.02%, which is non-hemolytic.

    A cytocompatibility evaluation of the 3D-printed stents was performed according to ISO 10993–5:2009 [34]. Human umbilical vein endothelial cells (HUVECs; ATCC, USA) were cultured in clonetics endothelial cell growth medium-2 (EGM-2; cc3162, Lonza,Switzerland). Extracts of 3D-printed PCL were prepared by immersing samples in EGM-2 at 37 °C for 24 h. Cell viability was detected by adding 10 μL of cell counting kit-8 (CCK-8)(Dojindo Laboratories, Japan) in each well after the HUVECs were cultured in extracts for 96 h; next, then the optical density (OD)value for each well was read at a wavelength of 450 nm in order to determine the cell viability on a microplate reader (Multiskan,Thermo, USA). The cell viability was calculated according to Eq. (22):

    where ODtrepresents the OD value of the test group, and ODpand ODnrepresent the OD values of the positive and negative groups,respectively. According to the cell viability value, the cytotoxicity reactivity of the materials was classified as Grade 0(≥100%),Grade 1(80%–99%),Grade 2(50%–79%),Grade 3(30%–49%),or Grade 4(0–29%). When the grade is more than 2, a material is considered to have cytotoxic effects. The cell viability of the 3D-printed PCL was 90% ± 5%, which is Grade 1.

    Fig. 10. Force-versus-compression ratio curves for PCL stents with different geometric unit parameter and rotation speed (n). For stents in (a), (b), and (c), the length of connector(h)is 1,1.25,and 1.5 mm,respectively;(d)the radial stiffness of stents in(a),(b),and(c);for stents in(e),(f),and(g),rotation speed(n)is 8.4,11.2,and 14 r?min-1,respectively; (h) the radial stiffness of stents in (e), (f), and (g). Error bars in (d) and (h) represent means ± standard deviations.

    Fig. 11. Radial expansion results: (a) the change of stent diameter; and (b) the change of stent length. Error bars represent means ± standard deviations.

    The results of the hemolysis rate and cell viability confirmed the potential application of 3D-printed stents in vascular implantation.

    5. Discussion

    5.1. Effect of printing parameters on printing-ability and structural formability

    5.1.1. Heating temperature of the extruder

    Before stent fabrication, extrusion of PCL monofilaments was conducted to investigate the effect of the heating temperature of the extruder (Textruder) on the strength and dimensional stability of the extruded monofilament. By comparing the accuracy of the filament diameter of different groups, it was found that the PCL filaments had better dimensional stability when Textruderwas 80 and 82.5 °C (Fig. 4(c)), while samples extruded at 75 and 80 °C had the advantage in terms of tensile strength.Consequently,considering both strength and dimensional stability,the heating temperature of the extruder was set at 80 °C. This temperature was used in the subsequent stent fabrication process. As shown in Figs.4(b)and(d),the PCL filament extruded at 80°C showed good surface morphology.

    5.1.2. Movement velocity of the x-axis system, rotational velocity of the mandrel, and extruder rotation speed

    Table 1 shows that the width of the stent strut(t) can be changed by adjusting the movement velocity (VRx) and extruder rotation speed (n). As VRxincreases, the width of the stent strut becomes smaller and the printing stability decreases (the coefficient of variation increases). When extruder rotation speed (n)increases, the width of the stent strut increases and shows better precision (the coefficient of variation is lower). Therefore, for a comprehensive consideration of printing stability(e.g.,for a coefficient of variation less than 10%)and printing efficiency,the movement velocity(VRx)was set at 1 mm?s-1and the width of the strut was controlled by changing the extruder rotation speed (n).Accordingly, the movement velocity of the x-axis (Vx) is√---3/2 mm?s-1(0.866 mm?s-1) and the linear velocity of the rotating mandrel (VR) was 0.5 mm?s-1. Thus, the rotational velocity of the mandrel (nm, unit:r?min-1) was 30/(πφ) r?min-1(where φ represents the inner diameter of the stent). Extruder rotation speed (n)of 8.4, 11.2, and 14 r?min-1were then adopted in the subsequent experiments.

    5.1.3. Filament extrusion speed

    The filament extrusion speed depends on the rotation speed of the extruder. The higher the rotation speed is, the faster the filaments are extruded. The rotation speed of series D, A–C, and E was 8.4, 11.2, and 14 r?min-1, respectively. As depicted in Fig. 7,the stents of series A, B, and C show a middle value in the width of strut (t), while series D has a relatively lower t and series E has a greater t. It is clear that the width of the stent strut is positively related to the rotation speed of the extruder: The greater the rotation speed, the larger the width of the strut. The stents in the series A,B,and C were fabricated with the same rotation speed.However,compared with the stents in series A and B,the stents in series C had a relative larger dimension of the stent strut. Therefore, the stability of stent fabrication needs to be improved in the future studies.

    5.2. Effect of structural parameters on the mechanical properties of 3D-printed stents

    In the literature,various methods have been used to examine a stent’s mechanical properties.The bending flexibility property and radial strength are two important indicators of the mechanical properties of a stent, which are closely related to the geometric parameters of the stent[14,28].Based on the discussion in Sections

    4.3.1 and 4.3.2, it is noted that bending flexibility can be represented by bending stiffness,with a higher bending stiffness resulting in a lower flexibility. Radial strength can be represented by radial stiffness, with a higher radial stiffness resulting in a lower radial strength. In this study, the main parameters affecting the mechanical properties of stents were the length of the connector and the width of the stent strut.

    5.2.1. Length of connector (h)

    The difference between series A, B, and C is the length of the connector (h), which is 1, 1.25, and 1.5 mm, respectively. Because the length of the stent is constant, it is clear that the longer the connector h is, the fewer the number of Z-shaped rings will be,resulting in greater bending flexibility and lower radial strength.Thus, the theoretical result for bending flexibility and radial strength should be Ai < Bi < Ci and Ci < Bi < Ai (i = 1, 2, 3),respectively.

    In regard to bending flexibility,Figs. 9(a)–(d)and Table 3 show that the bending flexibility of the different groups is B1 Ci > Ai (i = 1, 2, 3);thus, the radial strength of different groups is Bi < Ci < Ai (i = 1,2,3),which is different from the theoretical result.This finding will be explained in Section 5.2.2.

    In conclusion,stents with a longer connector have an advantage in terms of bending flexibility.However,it is not conclusive in this part whether a longer connector is strictly unfavorable for radial strength.

    5.2.2. Width of stent strut (t)

    The difference between series D, B, and E is the extruder rotation speed (n), which is 8.4, 11.2, and 14 r?min-1, respectively.Because the width of the stent strut is positively related to the rotation speed of the extruder, the greater the rotation speed is,the larger the width of the strut will be,resulting in lower pending flexibility and greater radial strength. Thus, the theoretical result for bending flexibility and radial strength should be Ei < Bi < Di and Di < Bi < Ei (i = 1, 2, 3), respectively.

    As expected, Figs. 9(e)–(h), Table 2, and Figs. 10(e)–(h) show that the experimental results for bending flexibility and radial strength are consistent with the theoretical situation.These results confirm that the strut width(t)is a dominating factor for the radial strength of stents. Furthermore, as mentioned above, the radial strength of groups A–C is Bi < Ci < Ai (i = 1, 2, 3), which does not match with the theoretical result of Ci < Bi < Ai. This result can be explained in terms of the width of the stent strut (t). From Fig. 7(f), it can be determined that the strut width of the Ci stents is relatively larger than that of the Bi stents (i = 1, 2, 3).

    5.2.3. Brief summary

    Based on the discussion in Sections 5.2.1 and 5.2.2, it can be seen that bending flexibility and radial strength are a pair of contradictory properties. This is a result of a trade-off of stent structure parameters to realize integrated mechanical properties. For stents of the same diameter, the mechanical properties (bending flexibility and radial strength) of the stents are closely related to the length of the connector and the width of the stent strut. In regard to the bending flexibility of stents with the same diameter,it can be seen that stents with a longer connector and a smaller width of stent strut (t) have advantages; therefore, stents with a loose geometric structure tend to show better flexibility.In regard to the radial strength of stents with the same diameter, it can be inferred that the strut width(t)is a dominating factor.The answer can be obtained from Eq. (12), which indicates that the structural strength of the stent is proportional to t3and inversely proportional to h; this means that the properties of the stent are more susceptible to the strut width(t).It can be inferred that stents with a larger connector (h) and a moderate strut width are more conducive for a balance between bending flexibility and radial strength.

    6. Conclusions

    In this study, a home-made screw extrusion-based 3D printing system using a mini-screw extruder was developed to fabricate stents while adding a fourth rotation axis (the mandrel). A stent with ZPR structure was designed and was demonstrated to be steady in axial length after expansion.Preliminary testing was conducted to investigate appropriate fabrication parameters with monofilament experiments. Subsequently, stents (made of PLLA and PCL)with different shape and geometric structures were fabricated in a customized manner using the ‘‘overlapping printing method.”In addition,the mechanical properties of PCL stents with different parameters were determined.

    In conclusion, the screw extrusion-based 3D printing system reported here shows potential for customizable stent fabrication.PCL stents are well studied in terms of morphology and mechanical analysis. However, the insufficiency of stent flexibility indicates a need for more efforts in stent design in future research. Furthermore, in comparison with metal stents, PCL stents are not completely reliable in term of radial strength and radial recoil. PLLA or other biopolymers with greater strength and a lower elongation rate will be explored for stent fabrication.

    Acknowledgements

    The authors acknowledge funding support from the Beijing Municipal Natural Science Foundation,China(Z150001).This work was also supported by Beijing Anzhen Hospital.

    Compliance with ethics guidelines

    Chengjin Wang,Lei Zhang,Yongcong Fang,and Wei Sun declare that they have no conflict of interest or financial conflicts to disclose.

    在线播放无遮挡| 国内精品美女久久久久久| 国产精品永久免费网站| 69av精品久久久久久| 成人三级黄色视频| 高清视频免费观看一区二区 | 免费大片18禁| 3wmmmm亚洲av在线观看| 看片在线看免费视频| 欧美性猛交╳xxx乱大交人| av天堂中文字幕网| 长腿黑丝高跟| h日本视频在线播放| 一边亲一边摸免费视频| 欧美变态另类bdsm刘玥| 国产麻豆成人av免费视频| 免费不卡的大黄色大毛片视频在线观看 | 中文字幕久久专区| 久久精品夜夜夜夜夜久久蜜豆| 久久精品国产亚洲网站| 久久草成人影院| 欧美不卡视频在线免费观看| 欧美激情在线99| 黄色日韩在线| 精品99又大又爽又粗少妇毛片| 久久韩国三级中文字幕| 99在线人妻在线中文字幕| 国产真实伦视频高清在线观看| 两性午夜刺激爽爽歪歪视频在线观看| 久久久久久久久久久丰满| 国产精品久久久久久精品电影小说 | 国产精品一区二区性色av| 国产免费又黄又爽又色| 美女高潮的动态| 一区二区三区乱码不卡18| 精品久久久久久电影网 | 亚洲四区av| 国产一区亚洲一区在线观看| 精华霜和精华液先用哪个| 欧美区成人在线视频| 搞女人的毛片| 亚洲av成人精品一区久久| 好男人在线观看高清免费视频| 岛国在线免费视频观看| 九色成人免费人妻av| 免费播放大片免费观看视频在线观看 | 大香蕉久久网| 国产伦精品一区二区三区四那| 亚洲国产精品国产精品| 六月丁香七月| 毛片一级片免费看久久久久| 18禁裸乳无遮挡免费网站照片| 99在线视频只有这里精品首页| 少妇人妻精品综合一区二区| av视频在线观看入口| 精品午夜福利在线看| 色哟哟·www| 成人毛片a级毛片在线播放| 三级经典国产精品| 免费av观看视频| 国产色婷婷99| 亚洲人成网站在线播| 欧美极品一区二区三区四区| 国产91av在线免费观看| 精品久久久久久久末码| 亚洲国产欧美人成| 亚洲欧洲国产日韩| 亚洲真实伦在线观看| 麻豆久久精品国产亚洲av| 亚洲成人av在线免费| 亚洲中文字幕日韩| 汤姆久久久久久久影院中文字幕 | 非洲黑人性xxxx精品又粗又长| 淫秽高清视频在线观看| 亚洲真实伦在线观看| 欧美最新免费一区二区三区| 日韩欧美在线乱码| 一级毛片aaaaaa免费看小| 波多野结衣巨乳人妻| 精品久久久久久久人妻蜜臀av| 久久久久免费精品人妻一区二区| 久久久久免费精品人妻一区二区| 色综合色国产| 欧美日韩综合久久久久久| 一级黄色大片毛片| 人人妻人人澡人人爽人人夜夜 | 老师上课跳d突然被开到最大视频| 韩国av在线不卡| 国产精品蜜桃在线观看| 国产色婷婷99| 亚洲欧美成人精品一区二区| 国产精品一区二区在线观看99 | 久久久久免费精品人妻一区二区| 网址你懂的国产日韩在线| 在线观看av片永久免费下载| 久久久久久久久大av| 国产精品精品国产色婷婷| 狠狠狠狠99中文字幕| 一区二区三区四区激情视频| 99久国产av精品国产电影| 偷拍熟女少妇极品色| 一区二区三区乱码不卡18| 精品人妻熟女av久视频| 国产精品嫩草影院av在线观看| 亚洲人与动物交配视频| 啦啦啦观看免费观看视频高清| 色尼玛亚洲综合影院| a级一级毛片免费在线观看| 国产免费男女视频| 九九在线视频观看精品| 啦啦啦观看免费观看视频高清| 精品99又大又爽又粗少妇毛片| av国产久精品久网站免费入址| 一本一本综合久久| 欧美另类亚洲清纯唯美| 色尼玛亚洲综合影院| 欧美激情国产日韩精品一区| 白带黄色成豆腐渣| 欧美潮喷喷水| 综合色丁香网| 日本-黄色视频高清免费观看| 天天一区二区日本电影三级| 少妇的逼水好多| 国产色爽女视频免费观看| 如何舔出高潮| 能在线免费看毛片的网站| 观看美女的网站| 成人特级av手机在线观看| 国产一区亚洲一区在线观看| 国产精品av视频在线免费观看| 2022亚洲国产成人精品| 国产一级毛片七仙女欲春2| 亚洲成人av在线免费| 黄色一级大片看看| 欧美日韩国产亚洲二区| 秋霞伦理黄片| 中文字幕人妻熟人妻熟丝袜美| 嘟嘟电影网在线观看| 成人二区视频| 日韩一区二区三区影片| 久久久久网色| 99久久精品国产国产毛片| 国产亚洲精品久久久com| 亚洲欧美清纯卡通| 成人无遮挡网站| 日本av手机在线免费观看| 欧美性感艳星| av又黄又爽大尺度在线免费看 | 我的老师免费观看完整版| 日韩,欧美,国产一区二区三区 | 七月丁香在线播放| 亚洲怡红院男人天堂| 国产精品一及| 草草在线视频免费看| 日韩精品有码人妻一区| 少妇人妻一区二区三区视频| 亚洲精品自拍成人| 亚洲精品亚洲一区二区| 精品一区二区三区人妻视频| 日本免费在线观看一区| 熟女电影av网| 在线播放无遮挡| 国产午夜精品一二区理论片| 青春草视频在线免费观看| 99久久人妻综合| 国产久久久一区二区三区| 亚洲欧美一区二区三区国产| 在线免费十八禁| 成人av在线播放网站| 国产麻豆成人av免费视频| 国产麻豆成人av免费视频| 久久久国产成人免费| 国产成人aa在线观看| 最近的中文字幕免费完整| 国产中年淑女户外野战色| 精品欧美国产一区二区三| 最近2019中文字幕mv第一页| 亚洲欧美日韩高清专用| 日本欧美国产在线视频| 日日干狠狠操夜夜爽| 国产麻豆成人av免费视频| 亚洲人成网站在线播| 中文资源天堂在线| 人人妻人人澡欧美一区二区| 久久人人爽人人爽人人片va| 国产精品久久视频播放| 99热全是精品| 午夜久久久久精精品| 亚洲三级黄色毛片| 在线播放无遮挡| 国产白丝娇喘喷水9色精品| 国内精品美女久久久久久| 亚洲精品亚洲一区二区| av免费在线看不卡| 久久久久久久午夜电影| 亚洲婷婷狠狠爱综合网| 在线免费十八禁| 又爽又黄无遮挡网站| 亚洲三级黄色毛片| 亚洲在久久综合| 高清午夜精品一区二区三区| 成人av在线播放网站| 日韩 亚洲 欧美在线| 精品人妻一区二区三区麻豆| 午夜精品一区二区三区免费看| 国产精品永久免费网站| 国产又色又爽无遮挡免| 看免费成人av毛片| 亚洲色图av天堂| 欧美人与善性xxx| 欧美日韩国产亚洲二区| 中文资源天堂在线| 亚洲成av人片在线播放无| 精品久久久噜噜| 成人综合一区亚洲| av在线老鸭窝| 久久久久久久亚洲中文字幕| 天堂网av新在线| 欧美高清成人免费视频www| av在线亚洲专区| 久久久久国产网址| h日本视频在线播放| 国产高清不卡午夜福利| 最近中文字幕2019免费版| 国产一级毛片七仙女欲春2| 亚洲真实伦在线观看| 欧美日韩在线观看h| 久久99热这里只有精品18| 日韩制服骚丝袜av| 97在线视频观看| 一级毛片电影观看 | 国产成年人精品一区二区| АⅤ资源中文在线天堂| 亚洲欧洲日产国产| 欧美性猛交╳xxx乱大交人| 亚洲国产精品成人综合色| 18禁在线无遮挡免费观看视频| 亚洲va在线va天堂va国产| 亚洲av熟女| 大香蕉97超碰在线| 国产精品一二三区在线看| 久久精品国产99精品国产亚洲性色| 又黄又爽又刺激的免费视频.| videos熟女内射| 日本熟妇午夜| 在线天堂最新版资源| 神马国产精品三级电影在线观看| 国产一区二区在线观看日韩| 菩萨蛮人人尽说江南好唐韦庄 | 一个人免费在线观看电影| 一个人观看的视频www高清免费观看| 久久久久网色| 日本黄色片子视频| 观看免费一级毛片| 国产色爽女视频免费观看| 午夜视频国产福利| 久久久欧美国产精品| 久久精品国产亚洲av涩爱| 两性午夜刺激爽爽歪歪视频在线观看| 国产伦一二天堂av在线观看| 成年版毛片免费区| 国产成人精品久久久久久| av播播在线观看一区| 搡女人真爽免费视频火全软件| av免费在线看不卡| 亚洲自偷自拍三级| 一级二级三级毛片免费看| 一个人免费在线观看电影| 日本免费a在线| 国产亚洲av嫩草精品影院| 国产美女午夜福利| 99热全是精品| 尤物成人国产欧美一区二区三区| 精品久久久久久久人妻蜜臀av| 国内精品一区二区在线观看| 午夜福利在线观看免费完整高清在| 日韩精品有码人妻一区| 黑人高潮一二区| 噜噜噜噜噜久久久久久91| 亚洲国产精品成人综合色| 国产黄色视频一区二区在线观看 | 国语自产精品视频在线第100页| 精品酒店卫生间| 精品国产一区二区三区久久久樱花 | 中文字幕av成人在线电影| 日产精品乱码卡一卡2卡三| 国产欧美另类精品又又久久亚洲欧美| 最近视频中文字幕2019在线8| 少妇猛男粗大的猛烈进出视频 | 三级经典国产精品| 最近中文字幕2019免费版| 午夜福利在线在线| 久久精品人妻少妇| 久久国产乱子免费精品| av又黄又爽大尺度在线免费看 | 国产片特级美女逼逼视频| 久久精品夜色国产| 亚洲av中文字字幕乱码综合| 亚洲自拍偷在线| 久久精品夜夜夜夜夜久久蜜豆| 国产成年人精品一区二区| 久久6这里有精品| 国产麻豆成人av免费视频| 国产久久久一区二区三区| 啦啦啦韩国在线观看视频| 岛国毛片在线播放| 亚洲第一区二区三区不卡| av在线观看视频网站免费| 3wmmmm亚洲av在线观看| 国产亚洲91精品色在线| 免费观看性生交大片5| 日本一二三区视频观看| 非洲黑人性xxxx精品又粗又长| 高清毛片免费看| 国产精品电影一区二区三区| 亚洲在久久综合| 欧美精品一区二区大全| 日韩精品有码人妻一区| 一级二级三级毛片免费看| 国产在视频线精品| 特大巨黑吊av在线直播| 久久久a久久爽久久v久久| 十八禁国产超污无遮挡网站| kizo精华| 2021天堂中文幕一二区在线观| 亚洲av免费在线观看| 我要看日韩黄色一级片| 日韩高清综合在线| a级毛色黄片| 天堂网av新在线| 人妻制服诱惑在线中文字幕| 国产一级毛片七仙女欲春2| 只有这里有精品99| 亚洲熟妇中文字幕五十中出| 少妇猛男粗大的猛烈进出视频 | 欧美激情国产日韩精品一区| 最近中文字幕2019免费版| 国产精品国产三级专区第一集| 国产精品人妻久久久影院| 国产成人a∨麻豆精品| a级毛色黄片| 联通29元200g的流量卡| 大香蕉久久网| 边亲边吃奶的免费视频| 97人妻精品一区二区三区麻豆| АⅤ资源中文在线天堂| 一区二区三区四区激情视频| 午夜久久久久精精品| 大话2 男鬼变身卡| 嫩草影院精品99| 国产69精品久久久久777片| av专区在线播放| 欧美三级亚洲精品| 久久99蜜桃精品久久| 日本猛色少妇xxxxx猛交久久| 国内精品宾馆在线| av专区在线播放| 亚洲欧美精品自产自拍| 色综合亚洲欧美另类图片| 亚洲aⅴ乱码一区二区在线播放| 一级黄片播放器| 99热精品在线国产| 精品熟女少妇av免费看| 精品一区二区免费观看| 91精品伊人久久大香线蕉| 18禁动态无遮挡网站| 91狼人影院| 91aial.com中文字幕在线观看| 国产精品精品国产色婷婷| 久久精品国产亚洲网站| 中文资源天堂在线| 国产极品天堂在线| 国产视频首页在线观看| 51国产日韩欧美| 久久久久久久久久成人| 2021天堂中文幕一二区在线观| 久久久久久久午夜电影| 欧美xxxx性猛交bbbb| 亚洲在久久综合| 国产午夜精品久久久久久一区二区三区| 国产精品爽爽va在线观看网站| 免费黄色在线免费观看| 联通29元200g的流量卡| 亚洲中文字幕一区二区三区有码在线看| 菩萨蛮人人尽说江南好唐韦庄 | 性色avwww在线观看| 又爽又黄a免费视频| av天堂中文字幕网| 在线观看一区二区三区| 少妇猛男粗大的猛烈进出视频 | 亚洲精品自拍成人| 国产一区二区在线av高清观看| 成人av在线播放网站| 亚洲三级黄色毛片| 两个人的视频大全免费| 九色成人免费人妻av| 人妻少妇偷人精品九色| 国产在线男女| 日本五十路高清| 一个人免费在线观看电影| 国产大屁股一区二区在线视频| 日本熟妇午夜| 美女大奶头视频| 免费大片18禁| 内地一区二区视频在线| 99久久九九国产精品国产免费| 国内揄拍国产精品人妻在线| 亚洲五月天丁香| 亚洲人成网站高清观看| 亚洲av日韩在线播放| 欧美精品国产亚洲| 能在线免费观看的黄片| 高清视频免费观看一区二区 | 熟女电影av网| 亚洲av成人av| 热99re8久久精品国产| 岛国毛片在线播放| 亚洲欧美成人综合另类久久久 | 天堂网av新在线| 欧美日韩一区二区视频在线观看视频在线 | 波野结衣二区三区在线| 美女国产视频在线观看| 最近的中文字幕免费完整| 97超碰精品成人国产| 汤姆久久久久久久影院中文字幕 | 亚洲中文字幕一区二区三区有码在线看| 国产精品人妻久久久影院| 国语对白做爰xxxⅹ性视频网站| 高清毛片免费看| 在线天堂最新版资源| 亚洲国产欧洲综合997久久,| 国产精品久久久久久精品电影小说 | 久久久久久久午夜电影| 韩国av在线不卡| 久久久久国产网址| 亚洲av.av天堂| 在现免费观看毛片| 欧美97在线视频| 午夜福利在线在线| 亚洲av熟女| 精品人妻一区二区三区麻豆| 高清午夜精品一区二区三区| 亚洲国产欧美在线一区| 国产中年淑女户外野战色| 成人高潮视频无遮挡免费网站| 九九久久精品国产亚洲av麻豆| 男插女下体视频免费在线播放| a级一级毛片免费在线观看| 深夜a级毛片| 2021少妇久久久久久久久久久| 麻豆一二三区av精品| 亚洲欧美精品专区久久| 亚洲aⅴ乱码一区二区在线播放| av黄色大香蕉| 三级国产精品欧美在线观看| 欧美性感艳星| 国产av一区在线观看免费| av在线蜜桃| 国产精品爽爽va在线观看网站| 久久草成人影院| 亚洲中文字幕日韩| 视频中文字幕在线观看| 免费黄网站久久成人精品| 成人二区视频| 99热这里只有是精品50| 久久久久国产网址| 美女黄网站色视频| 少妇熟女欧美另类| 欧美最新免费一区二区三区| 精品久久国产蜜桃| 亚洲国产精品久久男人天堂| av.在线天堂| 成人漫画全彩无遮挡| 国产av一区在线观看免费| 22中文网久久字幕| 日韩成人av中文字幕在线观看| 成人亚洲精品av一区二区| 精品一区二区三区人妻视频| 综合色丁香网| 久久这里有精品视频免费| 亚洲成人中文字幕在线播放| 久久久久九九精品影院| 免费av观看视频| 亚洲伊人久久精品综合 | 国产91av在线免费观看| 永久免费av网站大全| 免费观看性生交大片5| 大又大粗又爽又黄少妇毛片口| 91aial.com中文字幕在线观看| 午夜福利在线观看吧| 赤兔流量卡办理| 亚洲精品色激情综合| 桃色一区二区三区在线观看| 欧美区成人在线视频| 激情 狠狠 欧美| 乱码一卡2卡4卡精品| 久久精品久久久久久噜噜老黄 | 三级经典国产精品| 床上黄色一级片| 亚洲av电影不卡..在线观看| 国产精品国产三级专区第一集| 毛片一级片免费看久久久久| 少妇熟女aⅴ在线视频| 国产国拍精品亚洲av在线观看| 色尼玛亚洲综合影院| 精品久久久久久电影网 | 麻豆久久精品国产亚洲av| 精品免费久久久久久久清纯| 啦啦啦观看免费观看视频高清| 久久精品国产自在天天线| 国产成人aa在线观看| 在线a可以看的网站| 亚洲av电影在线观看一区二区三区 | 人人妻人人看人人澡| 亚洲av男天堂| 国产精品永久免费网站| 久久99蜜桃精品久久| 国产精品熟女久久久久浪| 国产毛片a区久久久久| 国产精品日韩av在线免费观看| 日韩视频在线欧美| 免费av毛片视频| 久久久国产成人精品二区| 国内精品美女久久久久久| 亚洲欧美日韩卡通动漫| 亚洲av中文字字幕乱码综合| 七月丁香在线播放| 嘟嘟电影网在线观看| 日本免费一区二区三区高清不卡| 最近中文字幕高清免费大全6| 麻豆一二三区av精品| 黑人高潮一二区| 老司机福利观看| 男人舔奶头视频| 国产精华一区二区三区| 91久久精品电影网| 男女国产视频网站| 真实男女啪啪啪动态图| 黄片wwwwww| 国产乱来视频区| 边亲边吃奶的免费视频| 欧美性猛交黑人性爽| 欧美高清性xxxxhd video| 中文字幕av成人在线电影| 丰满少妇做爰视频| 人人妻人人澡人人爽人人夜夜 | 国产精品人妻久久久久久| 国内精品美女久久久久久| 噜噜噜噜噜久久久久久91| 亚洲在线观看片| 亚洲美女视频黄频| a级毛色黄片| 91午夜精品亚洲一区二区三区| 午夜免费激情av| 白带黄色成豆腐渣| 午夜亚洲福利在线播放| 欧美变态另类bdsm刘玥| 丰满乱子伦码专区| 夜夜爽夜夜爽视频| 日日撸夜夜添| 国产久久久一区二区三区| 97人妻精品一区二区三区麻豆| 51国产日韩欧美| 日韩av在线免费看完整版不卡| 亚洲最大成人手机在线| 成年版毛片免费区| 天美传媒精品一区二区| 日韩在线高清观看一区二区三区| 91久久精品国产一区二区三区| 青春草视频在线免费观看| 亚洲一级一片aⅴ在线观看| 免费黄网站久久成人精品| 少妇人妻一区二区三区视频| 99热这里只有是精品在线观看| 永久免费av网站大全| 男人的好看免费观看在线视频| h日本视频在线播放| av视频在线观看入口| 高清视频免费观看一区二区 | 日韩大片免费观看网站 | 色5月婷婷丁香| 午夜亚洲福利在线播放| 在线播放无遮挡| 欧美性猛交黑人性爽| 欧美潮喷喷水| АⅤ资源中文在线天堂| 日本一本二区三区精品| 亚洲av免费在线观看| 国产又黄又爽又无遮挡在线| 精品一区二区免费观看| 蜜桃久久精品国产亚洲av| 少妇被粗大猛烈的视频| 久热久热在线精品观看| 日本与韩国留学比较| 三级国产精品片| 久久欧美精品欧美久久欧美| 乱系列少妇在线播放| 欧美精品一区二区大全| av女优亚洲男人天堂| 日本免费一区二区三区高清不卡| 99在线视频只有这里精品首页| 国产一级毛片七仙女欲春2| 禁无遮挡网站| 国产午夜福利久久久久久| av女优亚洲男人天堂| 国产高潮美女av| 成人性生交大片免费视频hd| 国产女主播在线喷水免费视频网站 | 久久欧美精品欧美久久欧美| 最近2019中文字幕mv第一页| 久久精品人妻少妇| 最近的中文字幕免费完整| 91午夜精品亚洲一区二区三区| 麻豆乱淫一区二区| 日本五十路高清|