    Eff cient M ethods for Approaching Functional Safety to Hardware System s of M edical Devices

    2015-06-01 10:01:18KIMGi-young,WANGDa-wei,PARKHo-joon
    中國醫(yī)療設(shè)備 2015年12期

    Eff cient M ethods for Approaching Functional Safety to Hardware System s of M edical Devices

    KIM Gi-young1, WANG Da-wei1, PARK Ho-joon2,
    JANG Joong-soon1
    1.Department of Industrial Engineering, A jou University, Suwon Gyeonggi-do 443-749, South Korea;2.Medical Device Evaluation Engineering Center, Korea Testing Laboratory, Guro-gu Seoul 152-718, South Korea

    A medical device is an instrument that includes components, parts, or accessories to diagnose or treat patients.Since the complexity of medical devices has increased in recent years, functional safety and basic safety are required to ensure the overall device safety.Functional safety is part of the overall safety that relates to the equipment under control(EUC)and to the EUC control system that depends on the correct functionality of the electrical/electronic/programmable electronic(E/E/PE)safety-related systems.This study proposes approach methods to functional safety of medical devices for which it is important to correctly identify the safety functions and the safety integrity level(SIL).The relationship between the functional safety and essential performance is identi fi ed focusing on the safety function.The essential performance of E/E/PE systems is de fi ned as the safety function of the functional safety.The target SIL of the essential performance is determ ined according to the potential risk levels, based on the classi fi cation rules of medical devices.This approach is applied to the pulse oximeter as a case study.The target SIL for the functionality of the power-failure alarm condition is determined to be SIL1.The target SILs of other functions are determ ined as SIL2.

    functional safety;medical device;essential performance;classi fi cation rule;safety function;pulse oximeter

    0 INTRODUCTION

    A medical device is generally used in patients, and its use may be associated to potential injuries.Since the risk of use of a medical device may be associated w ith a severe outcome, it can result in patient injuries.It is important for medical devices to achieve the required safety by reducing potential risks.Risk management for medical equipment should be carried out in accordance to the ISO 14971 standard.This standard presents the process and framework applied to the risk management of medical devices[1].The process includes identifying hazards associated w ith medical devices, estimating and evaluating risks, controlling the risks, and monitoring the effectiveness of the applied controls.

    As part of the risk management activities, the safety of medical devices has to be evaluated according to the IEC 60601 standard.This standard deals w ith the requirements that ensure the basic safety and essential performance of medical electrical equipment and systems.The IEC 60601 standard is divided into the IEC 60601-1 and the IEC 60601-2/ISO 80601-2 series.The IEC 60601-1 standard includes the general requirements that should be applied to medical devices[2].It covers the basic safety requirements of medical electrical equipment, and serves to ensure that no single electrical, mechanical, or functional failure poses an unacceptable risk to patients and operators.The IEC 60601-2/ISO 80601-2 series account for the basic safety and essential performance requirements in accordance to the features of each medical device.

    A lthough risk management and safety assessment have been carried out for im proving medical device safety, an increased number of failures of medical devices have occurred due to the increase in their operation hours.The probability of failures is raised ow ing to the degradation of hardware components as a result of cumulative loading after continuous use[3].These failures perm it the patient to be exposed to intolerable risks.The methods that treat the reliability and safety of medical devices over time should be applied to the design and development phases.As one of the alternative approaches, functional safety has recently been required to beenforced to improve the safety of medical devices.

    Functional safety is required for equipment under control(EUC), where the control systems depend on electrical/electronic/programmable electronic(E/E/PE)safety-related systems[4].These control systems can be de fi ned as being a part of the overall safety system.Kim et al[5]emphasized that the key concept of functional safety is consisted of safety functions and safety integrity.A safety function is a specific function implemented to m itigate or elim inate the risks to acceptable levels.Safety integrity is the probability that the required safety function is successfully performed by E/E/PE safety-related systems.When a selected safety function is operated w ith the required safety integrity, functional safety of the designated system is implemented according to the required level.

    Prior research studies on functional safety have been mainly conducted using software available in the medical device field.The IEC 62304 standard includes the software development process to achieve functional safety of medical devices[6].Studies on the quality of the medical device software that was developed according to the IEC 62304 standard have already been conducted[7–8].Investigators have sought to develop schemes for the traceability of medical software to ensure the safety and the quality[9-10].When medical software is developed, the application of the IEC 61508 standard is inappropriate in the medical device sector[11].MDevSPICE-Adept process is thus utilized to evaluate the software safety of medical device[12].

    Several studies have been conducted to implement functional safety to medical devices at the system level.Risk analysis is conducted to enhance the safety for the wearable walking assistant robots[13].The effect of risk reduction is compared between traditional and functional safety methods.Even though healthcare systems are in place, safety veri fi cation has not been carried out suf fi ciently in the development and integration phases.In order to solve the problem, functional safety must be applied in the fi elds of medical information and devices[14].

    A lthough several studies related to functional safety of medical devices have been performed, previous research studies have been associated w ith lim itations.These research studies have mainly focused on the software development process of the medical devices.Conceptual studies have been partially conducted for functional safety.The needs for functional safety in the medical fi eld have only been emphasized in the published research studies.The specific scheme for the approach of functional safety is not provided in previous studies.Application targets and scopes of functional safety do not re fl ect the characteristics of the medical device sector.The subject of the functional safety is the E/E/PE system involved in safety functions.Nevertheless, previous studies are associated with errors in that the systems other than the E/E/PE systems are considered as the boundary of functional safety.The safety integrity level(SIL)has been determ ined not for the safety functions but for the E/E/PE systems.To solve this problem, studies related to the approach of functional safety have to be executed for medical devices.

    The purpose of this study is to present a method to determ ine the safety functions and SIL of functional safety based on the characteristics of the safety aspects for the medical device sector.The scope and subjects are defined in order to implement the functional safety of medical devices.Functional safety analysis is performed based on the relationship of safety functions, and on the essential performance specified in the standards related to the medical device safety.The SIL of the essential performance is determ ined according to the potential risk levels, based on the classi fi cation rules of medical devices.As a case study, the approach is applied to the pulse oximeter.

    1 ESSENTIAL PERFORM ANCE OF M EDICAL DEV ICES AND APPL ICAT ION SCOPES OF FUNCTIONAL SAFETY

    Risk management in accordance to the ISO 14971 standard is the highest level of activity performed to achieve the safety of medical devices.The assessment of the basic safety and essential performance is a fraction of risk management activities for medical devices.The requirements for satisfying the basic safety of medical devices are speci fi ed in the general standard IEC 60601-1, and the standard series IEC 60601-2/ISO 80601-2.The IEC 60601-1 deals w ith the general requirements that should be applied to the medical devices.The IEC 60601-2/ISO 80601-2 series present the requirements of basic safety and essential performance re fl ecting the characteristics of each medical device.Figure 1 show s the hierarchical structure of standards related to the safety of medical devices.

    Figure 1 Hierarchical structure of m edical device safety standards

    Basic safety means freedom from an unacceptable risk directly caused by physical hazards, when the medical equipment is used under normal and single-fault conditions[1].Essential performance is the collective capability of a clinical function related to the safety.It is not included to the range of basic safety.Essential performance is easily understood by considering whether its absence or degradation would result in an unacceptable risk[15].The essential performance of each medical device used in the market has been selected based on the hazard analysis and risk assessment by medical device experts.The essential performance of several medical devices that are extensively used and potentially have a high potential risk is speci fi ed in the particular standard series IEC 60601-2/ISO 80601-2.Essential performance has relevance to functional safety.Therefore, an additional theoretical study should be conducted to investigate the two concepts.

    Essential performance is analogous to functional safety[16].Nevertheless, essential performance should be distinguished from functional safety[17].Essential performance is thought to correlate signi fi cantly w ith functional safety from the risk point of view.The application targets of these two concepts are the same as the safetyrelated areas that can be caused by the intolerable risk.Therefore, essential performance must be considered in the determination of the application scope of functional safety for medical devices.As shown in Figure 2, essential performance of medical devices is de fi ned as the safety function of functional safety.In accordance to the functional safety concept, the safety function is the important feature that must be executed to reduce or prevent the risk, based on the hazard analysis and risk assessment.If the safety function of functional safety is inappropriately executed, the intolerable risk results in harm and injuries to the patients.The essential performance of the medical device is also one of the safety-related factors.If essential performance is not properly executed, harm can be incurred when a hazardous situation occurs.Therefore, the essential performance can be regarded to be equally important to the safety function, and is one of the most signi fi cant elements of functional safety.

    Figure 2 Relationship of the key elements of functional safety and essential performance

    Essential performance of medical devices is grouped into two categories:functions w ith E/E/PE systems, and functions w ithout E/E/PE systems.The safety functions are categorized into two types:One is only executed to enhance safety, and the other is conducted to achieve better performance and safety simultaneously.Figure 3 shows the relevance between safety functions and essential performance.Among the variety of essential performance of medical devices, the essential performance using the E/E/PE system is only included in the safety function of functional safety.In contrast, the essential performance in cases where the E/E/PE system is not used is excluded from the application boundary of functional safety.For example, essential performance such as the energy output function of electrical surgical instruments, is included in the boundary of functional safety.However, essential performance, such as maintaining the sterilized electrode that does not use the E/E/PE system, has no relevance to the safety function of functional safety.

    Figure 3 Relationship of safety functions and essential performance

    If the essential performance is not specified in the particular standards it should be determ ined based on the ISO 14971 standard.The essential performance of a medical device is defined by the risk management activities.The essential performance should be checked depending on whether the E/E/PE systems are used or not.When the E/E/PE system is used for the implementation of the essential performance, it can be considered that it constitutes the safety functions of functional safety.If the E/E/PE system is not used, the essential performance is not included in the application boundary(Figure 4).

    Figu re 4 Flow d iagram ind icative of the selection o f sa fety functions considering essential perform ance.E/E/PE:Electrical/electronic/programmable electronic.

    2 SAFETY INTEGRITY LEVEL OF ESSENTIAL PERFORMANCE

    Safety integrity is a probability that an E/E/PE safetyrelated system properly performs the safety functions under speci fi ed conditions, and w ithin a stated period of time.SIL is the measure for evaluating at which degree of reliability the safety-related function operates, as it is implemented in EUC[18].As shown in Table 1, SIL is the discrete level corresponding to the range of safety integrity values in the IEC 61508 standard, which is the typical standard of functional safety.Speci fi cally, SIL4 is the highest level of safety integrity, and SIL1 is the lowest level[4].The higher the SIL is required, the lower the probability that the specified safety functions of the safetyrelated system w ill fail to operate.

    Table 1 Safety integrity levels based on mode of operations

    The mode of operations has to be determined for establishing the target SIL.The mode of operations is the way of operating safety functions, according to the demand rate.It is divided into three types:low-demand mode, high-demand mode, and continuous mode[19].If the demand rate of safety functions is no greater than one per year, the speci fi ed safety functions are considered to be in a low-demand mode.The parameter for the target SIL is selected as the average probability of a dangerous failure on demand of the safety function(PFDavg), in the lowdemand mode.If the demand rate of safety functions is greater than one per year, the speci fi ed safety functions are regarded to exist in the high-demand more or in the continuous mode.The parameter for the target SIL is selected as the average frequency of a dangerous failure of the safety function(PFH), in the highdemand more or in the continuous mode.When the demand rate is unknown, PFH is used as the parameter to determine the target SIL.

    Information on app lication areas of functional safety should be fully taken into account to conduct the risk assessment and determine the target SIL of the selected safety functions of functional safety.A lthough the same safety function is implemented, the SIL has different targets, depending on the application sector[20].SIL determination should be performed by taking into account the characteristics of the medical devices sector.Medical devices are classi fi ed in accordance to the potential risk levels.As shown in Table 2, the recommended target SIL is decided according to the detailed classi fi cation scheme.

    Table 2 Recommended SIL based on medical devices classif cation

    The medical devices are classi fi ed into three classes in the United States:class Ⅰ, class Ⅱ, and class Ⅲ[21].In Europe, the medical devices are classi fi ed into four classes:class Ⅰ, classⅡ a, class Ⅱ b, and class Ⅲ[22].In South Korea, the regulation enacted by the Korean M inistry of Food and Drug Safety classifies medical devices into four classes, as in Europe:class 1, class 2, class 3, and class 4[23].In every country, the classi fi cation regulations of medical devices have a sim ilarity in that, when the risk of a medical device is high, the number assigned to each class is increased.The item list of medical devices belonging to each class is sim ilar among the United States, European Union, and South Korea.

    Based on the classification performed in South Korea, tongue depressors and manual exam tables are typical class 1 medical devices.Class 1 medical devices have a low risk and a simple structure.A lmost all class 1 medical devices do not carry out functions of measurement, diagnosis, or energy radiation.They do not commonly use the E/E/PE systems to operate safety functions.Therefore, class 1 medical devices are excluded from the boundary of functional safety.

    The medical devices that are primarily used for diagnostic purposes are included into the class 2 boundary.Diagnostic X-ray equipment and diagnostic ultrasound are included into class 2 medical devices.Failures of class 2 medical devices enable the patient to be directly or indirectly exposed to harm.For example, excessive doses of radiation or ultrasound result in causing harm and direct damage to the tissue.The diagnostic functions cannot be executed ow ing to the decrease of image quality.A misdiagnosis can result due to the poor image quality of the diagnostic medical devices.The patient is exposed to a potentially hazardous situation.In this case, risk reduction activities are required.SIL1 should be established for safety functions in order to reduce potential risk and to enhance safetyof class 2 medical devices.

    Class 3 mainly consists of medical devices used to treat the patient or the disease.Electro–surgical systems and surgical laser are contained in class 3.Medical devices for treatment require higher electrical or mechanical energy than diagnostic medical devices of class 2.They transfer the energy to patients through the applied parts that directly contact the body.Therefore, the severity of the risk applied to the person is higher than class 2 medical devices.SIL2 should be established for safety functions to reduce the potential risk and to enhance the safety of class 3 medical devices.

    Class 4 medical devices are inserted and implanted into the body to work sem i-permanently, and are associated w ith the heart and blood vessels.Among a variety of class 4 medical devices, cardiopulmonary bypass devices, pacemakers and intra-aortic balloon pumps are operated by the E/E/PE systems.These medical devices have a high risk in regard to infections and biological portability.The risks associated w ith functional safety issues also lead to fatal harm and may threaten the life of the patients.Therefore, the safety functions of medical devices that belong to class 4 should be targeted as SIL3.

    The SIL determ ination based on the classification of medical devices is not complete.The m inimum guidelines are presented to de fi ne the target SIL of essential performance that is included in the functional safety boundary.The effect of malfunctions or failures of essential performance should be con fi rmed.SIL1 is allocated to the safety functions whose failures result in m inor injuries.When the failures of the safety functions give rise to serious injuries in a few people or to curable injuries in many people, the safety function’s target is considered to be SIL2.SIL3 is targeted to the safety function for which the malfunction causes serious injuries to many people or death to a few people due to failures.The death of many people is rarely caused by the failures of the medical devices[24].Therefore, SIL1, SIL2, or SIL3, are usually enough to establish the target of safety functions of medical devices.

    Functional safety of the medical device is implemented for hardware systems by using determined safety functions and SILs.Information for the target hardware system is collected.The requirements for implementing safety functions and SILs are speci fi ed at the system level.The requirements are divided into two groups;hardware safety integrity requirements and systematic safety integrity requirements.These requirements are allocated to the subsystem level.

    3 CASE STUDY

    This research conducted a case study to de fi ne the safety functions of functional safety and to establish the target SIL of the safety functions.The pulse oximeter was selected as the target medical device for the case study.A pulse oximeter is the medical device used to measure blood oxygen saturation, that is, the percentage of the amount of hemoglobin bound to oxygen w ith respect to the total amount of hemoglobin[25].Oxygen is an essential element for metabolism in human beings.If the blood oxygen concentration decreases, people can experience an increase in their heart rate, headaches, and nausea, in m ild cases.When the symptoms are severe, convulsions, morbus ceruleus, and unconsciousness can occur.Finally, cardiac arrest or brain death may occur due to a severe decrease of blood oxygen concentration.Therefore, the pulse oximeter has to operate properly in order to prevent additional damages in patients.

    In accordance to the specified process of this research, the related standards o f the pu lse oxim eter have to be checked to define the safety functions of functional safety.The essential performance is confirmed to select the safety functions of the functional safety for the pulse oximeter.The essential performance of the pulse oximeter is specified by the paticular standard ISO 80601-2-61[26].The groups of essential performance include the accuracy of measurement, detection of error/alarms, and prevention of incorrect output.Since all essential performances of the pulse oximeter use the E/E/PE system, it is considered as a device w ith safety functions of functional safety.The accuracy of SpO2and heart rate is included in the safety functions related to the measurement accuracy.The measured data have to be quickly updated to disp lay the current measurement, and must be renewed w ithin time intervals of 30 s or less.This function is related to the prevention function of incorrect outputs.Error detection enables the users to know that the optical sensor is disconnected from the extension cable, or that the power supply is converted from an external power into an internal power.In regard to the alarm functions, the IEC 60601-1-8 standard, commonly applied to medical devices, speci fi es the priority of alarms considering the risk level[27].Every essential performance of the pulse oximeter is included in the scope of functional safety.Therefore, the relationship between safety functions of functional safety and essential performance of the pulse oximeter is shown in Figure 5.

    Figu re 5 C lassification of safety functions and per form ancerelated functions

    The essential performance related to the accuracy of measurement and prevention of incorrect outputs constitutes fundamental functions that are continuously perform ed.Therefore, the mode of operations is selected to be in continuous mode.In regard to the essential performance related to the detection of error, alarms are only activated to ensure the safety of the pulse oximeter.These functions have relatively lower demand rate compared to the primary functions.A ll of the essential performance operates more than at least once per year.The operating mode is determ ined to be in a high-demand mode or in a continuous mode of operations for the essential performance.PFH is selected as the quantitative parameter for the target SIL.

    The pulse oximeter is a class 2 medical device based on item classi fi cation regulations.SIL1 should basically be assigned to the essential performance of the pulse oximeter in accordance to SIL determ ination rules, as shown in Table 2.Additional risk assessment is performed by considering the characteristics of the essential performance.The functions that m easure the pulse rate and blood oxygen levels are closely related to human life support.If these functions do not operate properly, this may lead to serious patient injuries.The probability of patient deaths due to the malfunction of the pulse oximeter is very low.A lthough the alarms of the pulse oximeter are not operated normally, hospital staff visually con fi rms the symptoms and takes action on patients in order to avoid incidents that may lead to death.

    The assigned SILs are modified by the results of risk managem ent.The function for the pow er-failure alarm condition is considered to be SIL1.Even though the speci fi ed essential performance is out of the order state, the hazardous situation does not occur immediately in this case.This essential performance has a low probability of causing harm to patients.Therefore, the target of the power-failure alarm condition is established to be SIL1.In contrast, SIL2 is assigned to other functions, such as the SpO2accuracy, pulse rate accuracy, protection against hazardous output, detection of probe and probe cable extender fault, and alarms condition priority.

    When fi ve types of essential performance are in failure occurrence situations, this can directly result in serious injuries for the patients, such as convulsions and morbus ceruleus.Therefore, the target for this essential performance is selected to be SIL 2, as shown in Table 3.

    4 CONCLUSION

    This study has p resented the app roach m ethod o f functional safety for medical devices.The analysis comparing the concept of safety function and the essential performance has been performed to approach functional safety.Essential performance of the medical device is sim ilar to the safety function of functional safety.Therefore, the lists on the essential performance should be considered to define safety functions of functional safety.Essential performance based on E/E/PE systems is selected to de fi ne the safety function of functional safety.Information on the essential performance that is presented in particular standards perm its designers and developers to efficiently define safety functions of functionalsafety.When essential performance of medical devices is used for selection of the safety function, the cost and time can be reduced, and the validity and adequacy can be ensured.

    Table 3 Target safety integrity levels for essential perform ance of the pu lse oximeter

    SIL is determ ined in accordance to the characteristics of the medical devices sector.Medical devices are classi fi ed according to the potential risk level.Information on classification regulation and risk management of medical devices is used to determ ine the target SIL of essential performance.SIL1, SIL2, or SIL3 are universally applicable for the target of safety functions of medical devices.SIL3 is allocated to a high-risk essential performance of medical devices and SIL1 is targeted to a low-risk essential performance.

    The case study is carried out for the pulse oximeter by using the presented approach methods.The ISO 80601-2-37 is a particular standard for the pulse oximeter and presents six types of essential performance.All the essential performance of the pulse oximeter is based on an E/E/PE system.Therefore, all essential performance is included in the boundary of functional safety.Among the essential performances, SIL1 is allocated to a power-failure alarm condition.Other essential performance targets are determ ined to be SIL2.

    [REFERENCES]

    [1]International Organization for Standardization,ISO 14971 Medical devices-Application of risk management to medical devices, 2007[S].

    [2]International Electrotechnical Commission,IEC 60601-1, Medical electrical equipment-Part 1:General requirements for basic safety and essential performance,2005[S].

    [3]Ohring M.Reliability and failure of electronic materials and devices[M].Academ ic Press,1998.

    [4]International Electrotechnical Commission,IEC 61508-1,Functional safety of Electric/Electronic/Programmable Electronic safetyrelated systems- Part 1:General requirements, 2010[S].

    [5]Kim GY,Ko BG,Chan SI,et al.Assessment Procedure of Safety Integrity Level(SIL)Based on Flow chart[J].J App l Re, 2010,10(2):107-122.

    [6]International Electrotechnical Commission,IEC 62304 Medical device software–Software life cycle processes, 2006[S].

    [7]Huhn M,Zechner A.Arguing for software quality in an IEC 62304 compliant development process[C].Proceeding ISoLA'10 Proceedings of the 4th international conference on Leveraging applications of formal methods, verification, and validation-Volume Part II,2010:296-311.

    [8]Burton J,McCaffery F,Richardson I.A risk management capability model for use in medical device companies[C].Proceedings of the 2006 international workshop on Software quality,2006:3-8.

    [9]Regan G,McCaffery F,M cDaid K,et al.Medical device standards' requirements for traceability during the software development lifecycle and imp lementation of a traceability assessment model[J].Comp Stand Inter,2013,36(1):3-9.

    [10]McCaffery F,Casey V,Sivakumar MS,et al.Medical device software traceability[C].In:Software and Systems Traceability,London:Springer,2012:321-339.

    [11]Jordan P.Standard IEC 62304-M edical device software-Software lifecycle processes[C].In:Software for Medical Devices 2006,The Institution of Engineering and Technology Sem inar,IET,2006:41-47.

    [12]M cHugh M,M cCaffery F,Casey V.How amendments to the medical device directive affect the development of medical device software[C].European Systems and Software Process Improvement and Innovation Conference, EuroSPI,2011.

    [13]Nabeshima C,Kawamoto H,Sankai Y.Typical risks and protective measures of wearable walking assistant robots[C].In System Integration(SII),2011 IEEE/SICE International Symposium on,2011:914-919.

    [14]Chadwick L,Fallon EF,van der Putten WJ,et al.Functional safety of health information technology[J].Health Inform J,2012, 18(1):36-49.

    [15]Schm idt M.Essential Performance in Medical Equipment[J].MDDI,2007,29(8):92-99.

    [16]Pinheiro P.Functional Safety Today's Medical Developments Magazine[EB/OL].http://www.onlinetmd.com/TMD-0911-functional-safety.aspx#.VQDt7k39mUk.

    [17]Silberberg JL.FDA/CDRH Experience w ith EMC Problem Reports,Standards,and Guidance,Center for Devices and Radiological Health,US Food and D rug Adm inistration, 2010[S].

    [18]International Electrotechnical Comm ission.IEC 61508-4, Functional safety of Electric/Electronic/Programmable Electronic safety-related systems-Part 4:Definitions and abbreviations, 2010[S].

    [19]Jin H,Lundteigen MA,Rausand M.Reliability performance of safety instrumented system s:A common approach for both low-and high-demand mode of operation[J].Reliab Eng Syst Safe,2011,96(3):365-373.

    [20]Smith DJ,Simpson KGL.Safety Critical Systems Handbook:A STRAIGHTFOWARD GUIDE TO FUNCTIONAL SAFETY, IEC 61508(2010 EDITION)AND RELATED STANDARDS, INCLUDING PROCESS IEC 61511 AND MACHINERY IEC62061 AND ISO 13849[M].The Netherlands:Elsevier,2010.

    [21]US Departm ent of Health and Hum an Services,Food and Drug Adm inistration,Center for Devices and Radiological Health,Center for Biologics Evaluation and Research.The 510(k)Program:Evaluating Substantial Equivalence in Premarket Noti fi cations, Food and Drug Administration[S].2011:1-42.

    [22]EUROPEAN COMM ISSION DG HEALTH AND CONSUMER.MEDICAL DEVICES:Guidance Document-Classification of Medical Devices,2010[S].

    [23]Department of Medical Equipment Policy.Regulations on The M edical Device Classification,M inistry of Food and Drug Safety,2015[S].

    [24]Marszal EM.Tolerable risk guidelines[J].Isa T,2001,40:391-399.

    [25]Runcim an WB,Webb RK,Barker L,et al.The Australian Incident Monitoring Study.The pulse oximeter:applications and limitations-an analysis of 2000 incident reports[J].Anaesth Intensive Care,1993,21(5)543-550.

    [26]International Organization for Standardization,ISO 80601-2-61, Medical electrical equipment-Part 2-61:Particular requirements for basic safety and essential performance of pulse oximeter equipment[S].

    [27]International Electrotechnical Comm ission,IEC 60601-1-8 Medical electrical equipment-Part 1-8:General requirements for basic safety and essential performance-Collateral Standard:General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems, 2006[S].

    R197.39 [Document code]A

    10.3969/j.issn.1674-1633.2015.12.008

    1674-1633(2015)12-0027-07

    Received:2015-06-15 Revised:2015-08-11

    JANG Joong-soon, Professor.

    E-mail:jsjang@ajou.ac.kr

    国产乱人偷精品视频| 亚洲精品日韩在线中文字幕| 九九久久精品国产亚洲av麻豆| 精品一区在线观看国产| 亚洲精品成人av观看孕妇| 欧美一级a爱片免费观看看| 久久久久国产网址| 亚洲熟女精品中文字幕| 免费观看性生交大片5| 亚洲欧美一区二区三区国产| 欧美变态另类bdsm刘玥| 国产极品天堂在线| 国产精品精品国产色婷婷| 国产黄色视频一区二区在线观看| 一级爰片在线观看| av在线蜜桃| 国产精品一及| 欧美日韩国产mv在线观看视频 | 尤物成人国产欧美一区二区三区| av天堂中文字幕网| 国产 一区精品| 一级毛片电影观看| 下体分泌物呈黄色| 天堂网av新在线| 美女脱内裤让男人舔精品视频| 国内揄拍国产精品人妻在线| 亚洲高清免费不卡视频| 内地一区二区视频在线| 国产男女超爽视频在线观看| 最近最新中文字幕免费大全7| 一个人观看的视频www高清免费观看| 婷婷色av中文字幕| 黄色配什么色好看| 99久久精品热视频| 亚洲欧美清纯卡通| 国产探花在线观看一区二区| 男人舔奶头视频| 日本猛色少妇xxxxx猛交久久| 99久久精品国产国产毛片| 日日啪夜夜爽| 国产午夜福利久久久久久| 亚洲综合色惰| 日韩免费高清中文字幕av| 亚洲av在线观看美女高潮| 五月开心婷婷网| 亚洲熟女精品中文字幕| 日日摸夜夜添夜夜爱| 亚洲av免费高清在线观看| 日韩一区二区三区影片| 人妻 亚洲 视频| 亚洲丝袜综合中文字幕| 天天躁夜夜躁狠狠久久av| 国产成人福利小说| 日韩一区二区三区影片| 免费大片黄手机在线观看| 婷婷色综合大香蕉| 亚洲精品影视一区二区三区av| 国产91av在线免费观看| 精品国产一区二区三区久久久樱花 | 欧美日本视频| 2021天堂中文幕一二区在线观| 日韩成人伦理影院| 婷婷色综合大香蕉| 美女xxoo啪啪120秒动态图| 久久人人爽人人片av| 97超碰精品成人国产| 日韩av在线免费看完整版不卡| 国产一区二区亚洲精品在线观看| 国产视频内射| 国产伦理片在线播放av一区| 看非洲黑人一级黄片| 国精品久久久久久国模美| 特大巨黑吊av在线直播| av福利片在线观看| av又黄又爽大尺度在线免费看| 国内精品宾馆在线| 看黄色毛片网站| 91久久精品电影网| 久久影院123| 三级国产精品片| av在线亚洲专区| 日韩欧美精品v在线| 午夜免费鲁丝| 亚洲精品一区蜜桃| 国产熟女欧美一区二区| 久久久国产一区二区| 国产老妇伦熟女老妇高清| 国产欧美日韩精品一区二区| 亚洲成人精品中文字幕电影| 欧美成人一区二区免费高清观看| 观看免费一级毛片| 日韩av在线免费看完整版不卡| 我要看日韩黄色一级片| 欧美极品一区二区三区四区| 国产老妇女一区| 我的女老师完整版在线观看| 久久久久精品性色| 免费看不卡的av| 一区二区三区精品91| 视频中文字幕在线观看| 在线亚洲精品国产二区图片欧美 | 最后的刺客免费高清国语| 最近最新中文字幕大全电影3| 国产精品久久久久久久电影| 激情五月婷婷亚洲| 日本黄色片子视频| 最近中文字幕2019免费版| 一本色道久久久久久精品综合| 男人和女人高潮做爰伦理| 国产成人精品一,二区| 两个人的视频大全免费| 天美传媒精品一区二区| 97在线人人人人妻| 在线看a的网站| 国产一区二区亚洲精品在线观看| 国产欧美日韩一区二区三区在线 | 老司机影院成人| 秋霞伦理黄片| 男女啪啪激烈高潮av片| 免费观看无遮挡的男女| 中文字幕久久专区| 亚洲自拍偷在线| 国产免费视频播放在线视频| 午夜精品一区二区三区免费看| 纵有疾风起免费观看全集完整版| 欧美日韩视频高清一区二区三区二| 晚上一个人看的免费电影| 久久久久久久久久久免费av| 国产男女内射视频| 国产91av在线免费观看| 久久热精品热| 青春草亚洲视频在线观看| 免费看日本二区| 成年版毛片免费区| 一边亲一边摸免费视频| 嘟嘟电影网在线观看| 亚洲av二区三区四区| 免费电影在线观看免费观看| 韩国av在线不卡| 岛国毛片在线播放| 一边亲一边摸免费视频| 精品99又大又爽又粗少妇毛片| 成人亚洲欧美一区二区av| 亚洲自偷自拍三级| 国产精品久久久久久精品古装| 白带黄色成豆腐渣| 精品久久久久久久久av| 九九爱精品视频在线观看| 一级毛片黄色毛片免费观看视频| 成人漫画全彩无遮挡| 亚洲国产成人一精品久久久| 纵有疾风起免费观看全集完整版| 亚洲aⅴ乱码一区二区在线播放| 真实男女啪啪啪动态图| 亚洲自拍偷在线| 大片免费播放器 马上看| 老女人水多毛片| 日韩欧美 国产精品| 噜噜噜噜噜久久久久久91| 欧美成人精品欧美一级黄| 美女脱内裤让男人舔精品视频| 一边亲一边摸免费视频| 国产精品蜜桃在线观看| 1000部很黄的大片| 久久综合国产亚洲精品| 免费少妇av软件| 国产精品偷伦视频观看了| 精品少妇久久久久久888优播| 精品亚洲乱码少妇综合久久| 22中文网久久字幕| 欧美成人a在线观看| 大又大粗又爽又黄少妇毛片口| 亚洲精品自拍成人| 亚洲欧美清纯卡通| 人妻制服诱惑在线中文字幕| 久久久精品94久久精品| 国产日韩欧美亚洲二区| 黄片无遮挡物在线观看| 亚洲四区av| 中文资源天堂在线| 国产精品秋霞免费鲁丝片| 丝袜脚勾引网站| 国产成人freesex在线| 色吧在线观看| 免费看av在线观看网站| 成人毛片60女人毛片免费| 偷拍熟女少妇极品色| 内地一区二区视频在线| 男女边摸边吃奶| 国产av国产精品国产| 国产精品麻豆人妻色哟哟久久| 日韩一本色道免费dvd| 丝袜美腿在线中文| 人人妻人人看人人澡| 99热国产这里只有精品6| 亚洲av免费高清在线观看| 国产伦在线观看视频一区| 久久久成人免费电影| 麻豆成人午夜福利视频| 国产黄片美女视频| 国产视频首页在线观看| 亚洲人成网站高清观看| 成人国产麻豆网| 黄色视频在线播放观看不卡| 大陆偷拍与自拍| 欧美国产精品一级二级三级 | 蜜桃久久精品国产亚洲av| 国产免费一级a男人的天堂| 一级毛片久久久久久久久女| 联通29元200g的流量卡| 精品一区二区免费观看| 啦啦啦啦在线视频资源| 永久免费av网站大全| 久久精品久久精品一区二区三区| 久久久精品免费免费高清| 欧美日韩精品成人综合77777| 国产亚洲av嫩草精品影院| 国产精品人妻久久久久久| 亚洲怡红院男人天堂| 黄色视频在线播放观看不卡| 日本免费在线观看一区| 丝袜美腿在线中文| 人人妻人人爽人人添夜夜欢视频 | 肉色欧美久久久久久久蜜桃 | 欧美区成人在线视频| 在线观看国产h片| 日日啪夜夜爽| 日韩伦理黄色片| 国产 精品1| 成年版毛片免费区| 国产综合精华液| 精品久久久噜噜| 日本色播在线视频| 久久人人爽人人片av| 99视频精品全部免费 在线| 激情 狠狠 欧美| 乱码一卡2卡4卡精品| 亚洲精品,欧美精品| 国产亚洲av嫩草精品影院| 日韩国内少妇激情av| 国产一区亚洲一区在线观看| 精品亚洲乱码少妇综合久久| 国产一区二区在线观看日韩| 欧美丝袜亚洲另类| 国产精品一区二区在线观看99| 成人特级av手机在线观看| 日日啪夜夜爽| 精品国产三级普通话版| 国产69精品久久久久777片| 在线天堂最新版资源| 在线观看美女被高潮喷水网站| 一级爰片在线观看| www.色视频.com| 麻豆精品久久久久久蜜桃| 国产午夜福利久久久久久| 国产高清有码在线观看视频| 亚洲欧美成人精品一区二区| 夫妻午夜视频| 一级爰片在线观看| av免费在线看不卡| 国产久久久一区二区三区| 国产亚洲5aaaaa淫片| 男人和女人高潮做爰伦理| 99久久九九国产精品国产免费| 国产精品爽爽va在线观看网站| 丝袜喷水一区| 亚洲国产精品专区欧美| 99热这里只有精品一区| 国产精品成人在线| 韩国av在线不卡| 国内精品美女久久久久久| 日日摸夜夜添夜夜爱| 国产精品人妻久久久久久| 亚洲自偷自拍三级| 老女人水多毛片| 久久精品国产自在天天线| 免费在线观看成人毛片| 亚洲一区二区三区欧美精品 | 免费看a级黄色片| 日本欧美国产在线视频| 国产精品久久久久久久电影| av国产久精品久网站免费入址| 亚洲国产精品999| 日韩av免费高清视频| av在线app专区| 哪个播放器可以免费观看大片| 日韩亚洲欧美综合| 麻豆成人av视频| 精品一区二区免费观看| 国产爽快片一区二区三区| 九九在线视频观看精品| 欧美性感艳星| 日韩在线高清观看一区二区三区| 精品一区在线观看国产| 赤兔流量卡办理| 在线观看三级黄色| 国产成人精品一,二区| 九九久久精品国产亚洲av麻豆| 一级av片app| 日韩av在线免费看完整版不卡| 热99国产精品久久久久久7| .国产精品久久| 亚洲精品久久午夜乱码| av在线亚洲专区| 国产一区二区亚洲精品在线观看| 美女被艹到高潮喷水动态| 一本色道久久久久久精品综合| 九九爱精品视频在线观看| 少妇人妻一区二区三区视频| 久久99热6这里只有精品| 中文天堂在线官网| 老司机影院成人| 欧美区成人在线视频| 亚洲,一卡二卡三卡| 欧美老熟妇乱子伦牲交| 亚洲av电影在线观看一区二区三区 | 精品国产三级普通话版| 亚洲久久久久久中文字幕| 亚洲欧美日韩无卡精品| 久久99热6这里只有精品| 日韩一本色道免费dvd| 亚洲最大成人手机在线| 亚洲精品一二三| 亚洲一区二区三区欧美精品 | 国产真实伦视频高清在线观看| 91久久精品国产一区二区三区| 男人狂女人下面高潮的视频| 一个人看的www免费观看视频| 日日啪夜夜撸| 深爱激情五月婷婷| 成人亚洲精品av一区二区| 亚洲电影在线观看av| 亚洲国产精品专区欧美| 99热全是精品| 大陆偷拍与自拍| 久久ye,这里只有精品| 人妻系列 视频| 久久亚洲国产成人精品v| 亚洲精品乱码久久久久久按摩| 国产综合精华液| 中文在线观看免费www的网站| 国产高潮美女av| 成人漫画全彩无遮挡| 久久精品国产亚洲av天美| 我的老师免费观看完整版| 中国国产av一级| 直男gayav资源| 三级国产精品欧美在线观看| 在线观看国产h片| 午夜精品国产一区二区电影 | 国产极品天堂在线| 国产高清有码在线观看视频| 国产亚洲av片在线观看秒播厂| 永久免费av网站大全| 我要看日韩黄色一级片| 久久久久久久久久久丰满| 亚洲av日韩在线播放| 欧美日韩视频高清一区二区三区二| 午夜视频国产福利| 国产成人福利小说| 国产午夜精品一二区理论片| 青春草亚洲视频在线观看| 免费av观看视频| 成人特级av手机在线观看| 人妻 亚洲 视频| 国产欧美另类精品又又久久亚洲欧美| 久久久久久久亚洲中文字幕| 国产欧美另类精品又又久久亚洲欧美| 大又大粗又爽又黄少妇毛片口| 自拍欧美九色日韩亚洲蝌蚪91 | 少妇熟女欧美另类| 久久久久久久久久人人人人人人| 色吧在线观看| 在线a可以看的网站| 亚洲精品国产色婷婷电影| 日韩一区二区三区影片| 欧美最新免费一区二区三区| 久久久精品欧美日韩精品| 欧美日韩亚洲高清精品| 久久久久久九九精品二区国产| 成人无遮挡网站| 亚洲无线观看免费| 亚洲精品色激情综合| 97精品久久久久久久久久精品| 波野结衣二区三区在线| 一区二区三区四区激情视频| 亚洲人成网站在线观看播放| 久久99热这里只有精品18| 国产亚洲一区二区精品| 国产老妇女一区| 听说在线观看完整版免费高清| 涩涩av久久男人的天堂| 日产精品乱码卡一卡2卡三| 亚洲高清免费不卡视频| 80岁老熟妇乱子伦牲交| 国产精品秋霞免费鲁丝片| 国产成人aa在线观看| 波野结衣二区三区在线| 亚洲欧洲日产国产| 极品少妇高潮喷水抽搐| 汤姆久久久久久久影院中文字幕| 亚洲婷婷狠狠爱综合网| 国产午夜精品一二区理论片| 99久久精品热视频| 99热这里只有是精品在线观看| 免费观看在线日韩| 欧美激情在线99| 久久久久久久久久久免费av| 亚洲精品国产av成人精品| 亚洲第一区二区三区不卡| 在线 av 中文字幕| 亚洲一区二区三区欧美精品 | 中文字幕av成人在线电影| 亚洲精品成人av观看孕妇| 2021天堂中文幕一二区在线观| 又爽又黄无遮挡网站| 亚洲精品成人久久久久久| 亚洲美女搞黄在线观看| 七月丁香在线播放| 美女被艹到高潮喷水动态| 亚洲精品国产色婷婷电影| 麻豆成人av视频| 亚洲,一卡二卡三卡| 欧美最新免费一区二区三区| 老司机影院成人| 91午夜精品亚洲一区二区三区| 三级国产精品片| 哪个播放器可以免费观看大片| 高清毛片免费看| 美女国产视频在线观看| 成人综合一区亚洲| 听说在线观看完整版免费高清| 亚洲在久久综合| 91精品伊人久久大香线蕉| 在线天堂最新版资源| 国产精品久久久久久精品电影小说 | av.在线天堂| 国产老妇伦熟女老妇高清| 男人舔奶头视频| 超碰97精品在线观看| 久久精品久久久久久噜噜老黄| 精品亚洲乱码少妇综合久久| 一边亲一边摸免费视频| 伦精品一区二区三区| 久久韩国三级中文字幕| 日本黄大片高清| 狂野欧美激情性bbbbbb| 美女主播在线视频| 少妇裸体淫交视频免费看高清| 亚洲精品一二三| 午夜亚洲福利在线播放| 亚洲成人av在线免费| 男人和女人高潮做爰伦理| 日韩av在线免费看完整版不卡| 又爽又黄无遮挡网站| 国产伦精品一区二区三区视频9| 亚洲人成网站高清观看| 久久精品综合一区二区三区| 亚洲最大成人中文| 亚洲欧美清纯卡通| 男人舔奶头视频| 18禁动态无遮挡网站| 日韩大片免费观看网站| 成人一区二区视频在线观看| 超碰av人人做人人爽久久| 白带黄色成豆腐渣| 老司机影院成人| 亚洲真实伦在线观看| 精品99又大又爽又粗少妇毛片| 中国三级夫妇交换| 人妻 亚洲 视频| 黄色怎么调成土黄色| 成人亚洲精品一区在线观看 | 成年免费大片在线观看| 精品久久久久久电影网| a级一级毛片免费在线观看| 99九九线精品视频在线观看视频| av天堂中文字幕网| 水蜜桃什么品种好| 国产一区二区在线观看日韩| 免费少妇av软件| 欧美性猛交╳xxx乱大交人| 国产精品.久久久| 午夜福利视频1000在线观看| 亚洲精品视频女| 国产精品久久久久久久久免| 一本色道久久久久久精品综合| 亚洲人成网站在线观看播放| 午夜福利视频精品| 两个人的视频大全免费| 色视频www国产| 少妇人妻 视频| 国产色婷婷99| 精品少妇久久久久久888优播| 九九在线视频观看精品| 麻豆乱淫一区二区| www.色视频.com| 人妻 亚洲 视频| 女人十人毛片免费观看3o分钟| 精品熟女少妇av免费看| 久久精品国产亚洲网站| 又大又黄又爽视频免费| 精品亚洲乱码少妇综合久久| 亚洲av成人精品一二三区| 欧美成人a在线观看| 天堂俺去俺来也www色官网| 亚洲欧美一区二区三区黑人 | 搞女人的毛片| 日韩伦理黄色片| 国产精品一区www在线观看| 精品国产露脸久久av麻豆| 成人无遮挡网站| 高清午夜精品一区二区三区| 亚洲国产精品成人综合色| 九草在线视频观看| 久久韩国三级中文字幕| 色综合色国产| 六月丁香七月| 大陆偷拍与自拍| 插阴视频在线观看视频| av.在线天堂| 国产精品熟女久久久久浪| 在线看a的网站| av女优亚洲男人天堂| 天堂中文最新版在线下载 | 国产大屁股一区二区在线视频| 九九久久精品国产亚洲av麻豆| 欧美极品一区二区三区四区| 99热国产这里只有精品6| 欧美成人精品欧美一级黄| 国产真实伦视频高清在线观看| 国产91av在线免费观看| 久久久成人免费电影| 国产精品国产三级国产专区5o| 国产真实伦视频高清在线观看| 91精品国产九色| 国产乱人偷精品视频| 婷婷色麻豆天堂久久| 人妻系列 视频| 神马国产精品三级电影在线观看| 国产精品国产三级国产专区5o| 免费观看a级毛片全部| 亚洲精品乱码久久久v下载方式| av免费在线看不卡| 人妻 亚洲 视频| 日本av手机在线免费观看| 国产成人一区二区在线| 久久久久久久精品精品| 人妻制服诱惑在线中文字幕| 亚洲婷婷狠狠爱综合网| 精品国产露脸久久av麻豆| 99久国产av精品国产电影| 交换朋友夫妻互换小说| 97精品久久久久久久久久精品| 亚洲欧美日韩无卡精品| av线在线观看网站| 国产精品av视频在线免费观看| 91精品国产九色| 搡女人真爽免费视频火全软件| 涩涩av久久男人的天堂| 日韩大片免费观看网站| 日韩电影二区| 国产精品三级大全| 亚洲性久久影院| 六月丁香七月| 麻豆久久精品国产亚洲av| 在线免费观看不下载黄p国产| 国产 一区精品| 97在线人人人人妻| 有码 亚洲区| 人人妻人人看人人澡| 久久久久久久精品精品| 亚洲成人av在线免费| 亚洲人与动物交配视频| 3wmmmm亚洲av在线观看| 蜜桃久久精品国产亚洲av| 看非洲黑人一级黄片| 精品亚洲乱码少妇综合久久| 国内揄拍国产精品人妻在线| 日日啪夜夜爽| 国产男人的电影天堂91| 亚洲性久久影院| 18禁裸乳无遮挡免费网站照片| 欧美三级亚洲精品| 国产 精品1| 欧美日韩视频精品一区| 免费黄网站久久成人精品| 亚洲av男天堂| 80岁老熟妇乱子伦牲交| 啦啦啦中文免费视频观看日本| 一个人看的www免费观看视频| 男插女下体视频免费在线播放| 22中文网久久字幕| 91精品国产九色| 亚洲av男天堂| a级一级毛片免费在线观看| 色视频在线一区二区三区| 亚洲成人av在线免费| 亚洲四区av| 亚洲av国产av综合av卡| 天堂俺去俺来也www色官网| 久久久成人免费电影| 成人欧美大片| 亚洲av电影在线观看一区二区三区 | 成人亚洲精品一区在线观看 | 亚洲精品国产色婷婷电影| 人妻一区二区av| a级毛片免费高清观看在线播放| 午夜免费男女啪啪视频观看| 一二三四中文在线观看免费高清| 女人久久www免费人成看片| 国产在视频线精品| 国产乱人视频| 久久精品久久久久久久性| 少妇人妻 视频|