    Evaluating newer generation intraocular lens calculation formulas in manual versus femtosecond laser-assisted cataract surgery

    2021-08-16 09:56:12HarrishNithianandanSoumyaSharmaEricTamHannahChiuRajivMainiSohelSomani

    Harrish Nithianandan, Soumya Sharma, Eric S. Tam, Hannah Chiu, Rajiv Maini,Sohel Somani

    1Department of Ophthalmology & Visual Sciences, McGill University, Montreal H4A 3S5, Canada

    2Ryerson University, Toronto, Ontario M5B 2K3, Canada

    3Department of Ophthalmology, University of Toronto, Toronto M5T 3A9, Canada

    4Uptown Eye Specialists, Vaughan, Ontario L4K 2Z5, Canada

    5William Osler Health System, Brampton, Ontario L6R 3J7,Canada

    Abstract

    INTRODUCTION

    The predictability of refractive outcomes following uncomplicated cataract surgery has steadily improved over the years, and this is largely attributable to more accurate intraocular lens (IOL) power calculation formulas[1-2].Although more recent formulas generally outperform those of previous generations, there is considerable debate regarding the selection of the most accurate formula[3-4]. Among the more popular IOL formulas, the Haigis[5], and newer generation Barrett Universal II (Barrett)[6]and Hill-Radial Basis Function(Hill-RBF)[7]formulas have shown superiority to other formulas[8-11]. These formulas are felt to perform relatively consistently across the range of axial lengths (AL)[2]. And their use of multiple-variable vergence models has drawn the attention of cataract surgeons[12].

    Although some authors advocate for the use of different formulas for varied ocular parameters[2]. It is important to also consider the modality of cataract surgery being employed. Namely, there has been increasing integration of the femtosecond laser into phacoemulsification surgery given its ability to assist with many critical steps of the procedure[13].From a refractive standpoint, it has been suggested that the use of refractive femtosecond laser-assisted cataract surgery(ReLACS) may optimize the circularity and centration of the anterior capsulotomy[14]and ultimately the effective lens position when compared to conventional manual cataract surgery (MCS)[15]. The impact of these potential differences on the performance of modern IOL formulas has not been elucidated.

    Given this void in the literature, this study was designed to evaluate and compare the performance of the Haigis, Barrett,and Hill-RBF 2.0 IOL formulas within a single IOL platform across eyes treatedviaReLACS and MCS. This would further add to the literature on comparing the performance of each formula, but more importantly, would provide guidance on whether surgeons should select a particular formula based on the modality of cataract surgery.

    SUBJECTS AND METHODS

    Ethical ApprovalThis retrospective interventional comparative case series was approved by the Research Ethics Board of the William Osler Health System and was conducted in accordance with the tenets of the Declaration of Helsinki.Electronic medical records were reviewed at a private cataract surgery center (Vaughan, Ontario, Canada) between June 30th,2017 to August 30th, 2018 to identify patients who: 1) had uncomplicated phacoemulsification surgery with implantation of a Tecnis intraocular lens (AMO Tecnis ZCB00, Johnson &Johnson Vision, Santa Ana, CA, USA); 2) had preoperative biometric testing completed using the IOL Master 700 (Zeiss,Oberkochen, Germany); 3) had a lens thickness (LT) of at least 2.5 mm; 4) had cylinder less than 4.0 D; and 5) AL standard deviation (SD) of 15 μm or less, representing a good reliability index. Only one eye per patient was used, and the better of two eyes included with postoperative best corrected visual acuity (BCVA) of at least 20/40 or better[16]. Exclusion criteria included any history of refractive surgery or corneal disease, postoperative refractive error greater than 2.0 D, and postoperative BCVA worse than 20/40.

    Preoperative ExaminationPreoperative demographic variables of interest included age, gender, and laterality of the included eye. Baseline biometric data obtained from the IOL Master 700 included AL, anterior chamber depth(ACD), keratometry values, LT and horizontal white to white diameter (WTW). In an effort to increase the internal validity of the study, the older generation Hoffer Q formula[17]was included in this analysis to serve as a reference point against which the newer generation formulas would be compared.The postoperative refraction for a Tecnis IOL was predicted using the IOL Master 700 for the Haigis, Barrett, and Hoffer Q formulas. Hill-RBF (version 2.0) calculations were performed using the online web-based calculator[7].

    Surgical TechniqueStandard techniques were employed for cataract extraction using either MCS or ReLACS. MCS was completed using a 2.5 mm main clear corneal incision and the WhiteStar Signature Phacoemulsification System (Advanced Medical Optics, Santa, Ana, CA, USA) for cataract extraction.For eyes undergoing ReLACS, the Catalys Precision Laser System with Liquid Optics Interface (Abbott Medical Optics,Santa Ana, CA) was used to create the 2.5 mm main corneal incision, a 1.2 mm side port incision, a 5.0 mm diameter anterior capsulotomy and for lens fragmentation with grid softening. The eyes then proceeded with standard phacoemulsification surgery as described for the MCS group.All eyes were prescribed topical bromfenac one drop daily,topical besifloxacin one drop three times daily, and loteprednol gel applied three times daily for a total of 30d following the procedure.

    Postoperative ExaminationManifest refraction was measured at the one-month postoperative visit. The postoperative refractive error was calculated by subtracting the predicted spherical equivalence from the measured spherical equivalence. This calculation was computed for all formulas.The mean errors of all formulas were adjusted to zero by retrospectively optimizing the lens factor. For the Haigis formula, an optimized a0 lens constant of 1.712 was computed using 0.4 and 0.1 as the a1 and a2 constants, respectively[5]. For the Barrett formula, the optimized lens factor was computed to be 2.15. For the Hill-RBF and Hoffer Q formulas, the optimized lens constants were 119.5 and 5.85, respectively.Following optimization, the mean absolute error (AE) and median absolute error (MedAE) were computed.

    Outcome MeasuresThe primary outcome was to compare the one-month MedAE between the formulas among MCS and ReLACS eyes. Secondary outcomes included the proportion of eyes achieving an AE within 0.5 D, and the correlation between AE and AL among MCS and ReLACS eyes.

    Statistical AnalysisCategorical parameters were reported as proportions with differences compared using the Fisher exact test. Continuous variables were reported as mean±standard deviation (SD) with differences compared using thettest.Differences in MedAE were compared using the Friedman test. Post-hoc analyses were completed using the Wilcoxon signed-rank test, with multiple comparisons corrected for using the Bonferroni method. Pearson correlational analysis was completed to compute correlation coefficients. Statistical analysis was performed using SPSS (IBM Corp, Armonk, New York; software version 22).

    RESULTS

    A total of 158 eyes met the inclusion criteria with 64 eyes in the MCS group and 94 eyes in the ReLACS group.Demographic and preoperative characteristics of both groupsare illustrated in Table 1. Patients in the ReLACS group were younger (P=0.002) and had slightly thinner lenses (P=0.02)when compared to the MCS group. There were no differences between the groups when comparing AL, mean keratometry,cylinder, ACD, WTW, IOL power, and distributions of gender and laterality (allP>0.05).

    Table 1 Baseline characteristics of eyes that underwent ReLACS and MCS

    Table 2 One-month refractive outcomes of the intraocular lens formulas within MCS and ReLACS eyes

    Table 2 outlines the refractive outcomes at one-month postoperative within the MCS and ReLACS groups. Across all eyes, the Friedman test confirmed that there were statistically significant differences in the accuracy of the formulas(P=0.005). Post-hoc analysis revealed that the MedAE was significantly less when using the Barrett formula [MedAE=0.21 D,interquartile range (IQR): 0.11-0.38 D] compared to Haigis(MedAE=0.27 D, IQR: 0.12-0.50 D,P=0.005) and the Hoffer Q formula (MedAE=0.25 D, IQR: 0.10-0.46 D,P<0.001). The MedAE of the Hill-RBF formula (MedAE=0.25 D, IQR: 0.11-0.39) was similar to the Barrett formula (P=1.0).

    When comparing the MedAE among ReLACS eyes, the Friedman test confirmed that there were statistically significant differences in the accuracy of the formulas (P=0.001). Post-hoc analyses revealed that the Barrett formula (MedAE=0.21 D,IQR: 0.11-0.38 D) was more accurate than the Haigis formula(MedAE=0.27 D, IQR: 0.12-0.52 D,P=0.006) and the Hoffer Q formula (MedAE=0.28 D, IQR: 0.11-0.49 D,P<0.001). The distribution of the Hill-RBF formula (MedAE=0.23 D, IQR:0.11-0.40 D,P=1.0) was similar to that of the Barrett formula among ReLACS eyes.

    Αmong MCS eyes, the Friedman test confirmed that there were no statistically significant differences in the accuracy of the formulas (P=0.59). The MedAE (IQR) of the Haigis, Barrett,Hill‐RBF and Hoffer Q formulas were 0.23 D (0.08‐0.49 D),0.21 D (0.10-0.38 D), 0.26 D (0.11-0.40 D) and 0.20 D (0.10-0.46 D), respectively.

    Across all eyes, the Barrett (87% of eyes) and Hill-RBF (85%)formulas were more likely to result in AE less than 0.5 D when compared to Haigis (75%) and Hoffer Q (77%,P<0.001).Among ReLACS eyes, Barrett (86%) and Hill-RBF (84%)were similarly more likely to achieve AE less than 0.5 D when compared to Haigis (74%) and Hoffer Q (74%,P<0.001). This trend was also observed among MCS eyes (P<0.001), with 89%, 89%, 77% and 81% of eyes achieving AE less than 0.5 D using the Barrett, Hill-RBF, Haigis and Hoffer Q formulas,respectively.

    Among ReLACS eyes, Pearson correlational analysis did not reveal a significant association between AL and AE when using the Barrett (r=0.076,P=0.47), Haigis (r=0.037,P=0.72),Hoffer Q (r=0.16,P=0.14) or Hill-RBF (r=-0.027,P=0.80)formulas. Among MCS, there was no significant correlation between AL and AE when using the Hoffer Q (r=-0.16,P=0.20) or the Haigis formula (r=-0.143,P=0.26). However,AL and AE were significantly correlated among MCS eyes when using the Barrett (r=-0.25,P=0.04) and Hill-RBF (r=-0.25,P=0.04) formulas.

    DISCUSSION

    To our knowledge, this study is among the first to evaluate the performance of the Haigis, Barrett, and Hill-RBF formulas among eyes treatedviaReLACS and MCS. Among MCS eyes, there was no significant difference in MedAE between the formulas, however among ReLACS eyes, there was a pronounced benefit of pursuing the newer generation Barrett and Hill-RBF formulas. Interestingly, our results suggest that AL was associated with the magnitude of AE when these newer generation formulas are used in MCS eyes, however AL did not influence AE among ReLACS eyes. Across all eyes,and especially in ReLΑCS eyes, the Hoffer Q was suboptimal when compared to the newer generation formulas.

    Recently, Whanget al[18]compared refractive outcomes between eyes treatedviafemtosecond laser-assisted and conventional cataract surgery using the Barrett and Haigis formulas. Their results indicated superior refractive outcomes among eyes treated using the femtosecond laser-assisted approach[18]. However, their study’s analysis was not designed to detect performance differences between formulas, and rather focused on cataract surgery modality as a modulator of refractive accuracy. Our study, therefore, complements their results and found that the Barrett formula was more accurate and was associated with a greater likelihood of refractive success across all eyes, and in eyes undergoing ReLACS. Therefore, these results suggest that surgeons may benefit from using the Barrett formula over Haigis for their ReLACS cases. However, for MCS cases, both formulas are likely to perform comparably.

    Although not statistically significant, our results did indicate a trend towards smaller MedAE achieved using the Barrett when compared to the Hill‐RBF. The Hill‐RBF is an artificial intelligence-based formula that uses pattern recognition and an algorithmic process to predict the optimal IOL power given the AL, ACD and corneal power based on a pre-existing database of patients. In their study of 127 eyes, Wanet al[19]did not detect differences in MedΑE when comparing the Barrett and Hill-RBF formulas. Similarly, Robertset al[8]found both the Hill-RBF and the Barrett formulas to reduce refractive error in short and long eyes, in their cohort of 400 patients, however they did detect fewer refractive surprises when using the Barrett formula. Kaneet al[4]found Barrett to be superior in the reduction of absolute error when compared to Hill-RBF in a large cohort of 3122 patients. The apparent superiority Barrett when compared to Hill-RBF may be attributable to its use of multiple parameters to calculate a better estimation of the effective lens position.

    ACD, LT, and WTW were included as parameter inputs into the Barrett formula calculation, as they may contribute to a better estimation of lens position and thus more accurate results[6]. The Haigis formula, however, did not use LT or WTW as parameter inputs. These differences may explain the more accurate results obtained by Barrett in the present study.A previous Meta-analysis has shown that when created by ReLACS, capsulotomies were more likely to be of the intended diameter, more horizontally centred, and more circular(depending on the measure of circularity used)[14]. Therefore,the benefit of the Barrett formula among ReLΑCS eyes may reflect some degree of synergy between the increased accuracy of the estimated lens position by Barrett and the theoretically more predictable and centred capsulorrhexis offered by ReLACS.

    The superiority of Barrett to other formulas, including Haigis,has been demonstrated by a number of studies. In their consecutive series of 18 501 cases, Melleset al[2]found that the Barrett provided a more accurate prediction of postoperative refraction when compared to Haigis and performed well across a range of ALs. Across all eyes in their series, the MedAE for the Barrett and Haigis formulas were 0.25 and 0.28 D,respectively, which were similar to the results of our study[2].Similarly, Kaneet al[4]found Barrett to be superior to Haigis when comparing refractive accuracy in their series of 3241 patients, especially in eyes with AL greater than 22.0 mm.

    Given that IOL formulas tend to be less accurate among eyes of very short or long ALs, the present study aimed to evaluate the performance of formulas across the full range of ALs of the included participants. A recent Meta-analysis found that the Barrett and Haigis formulas performed comparably in long eyes[11]. In short eyes, studies did not detect a significant difference in refractive performance between the two formulas[1,4,9]. Our results indicated that AL was not significantly associated with ΑE in the ReLΑCS groups when using the Barrett or Hill-RBF, however AL was negatively correlated with AE among MCS eyes when using the Barrett and Hill-RBF formulas. In other words, shorter eyes were more likely to have greater AE among MCS eyes, and this further supports the potential benefit of pursuing ReLACS together with a newer generation formula.

    The retrospective design of this study was one of the limitations of this study, as the authors were not able to prospectively match the baseline characteristics of the MCS and ReLACS groups. The sample size was relatively small and not equal between the groups, which warrants investigations in future larger cohorts and of other modern IOL formulas to provide further clarity regarding the ideal choice of IOL formula for ReLACS and MCS, respectively. The strengths of this study include its use of a consistent IOL design with lens constant optimization to allow for sound statistical testing, and an adequate follow up period of one month for all patients, at which time postoperative refraction is typically known to have stabilized.

    In conclusion, our study supports the superiority of the newer generation Barrett and Hill-RBF formulas to Haigis overall in patients undergoing phacoemulsification with implantation of a Tecnis IOL. Surgeons employing ReLACS may have increased refractive accuracy especially when using the Barrett or Hill-RBF formulas, as there is a strong opportunity to predict and implement an accurate effective lens position. Further studies are warranted to investigate whether one of these newer generation formulas outperforms the other.

    ACKNOWLEDGEMENTS

    Presentations:The data presented in this manuscript has been presented in part at: 1) The American Society of Cataract and Refractive Surgery Annual Meeting, May 2019, San Diego,California, USA; 2) The Canadian Ophthalmological Society Annual Meeting, June 2019, Quebec City, Quebec, Canada.

    Conflicts of Interest:Nithianandan H,None;Sharma S,None;Tam ES,None;Chiu H,None;Maini R,None;Somani S,None.

    国产日韩欧美亚洲二区| 国产成人欧美在线观看 | 啦啦啦在线观看免费高清www| 精品少妇久久久久久888优播| 亚洲一区中文字幕在线| 搡老岳熟女国产| 午夜福利在线免费观看网站| 色婷婷av一区二区三区视频| 99久久99久久久精品蜜桃| 国产一级毛片在线| 国产欧美亚洲国产| 一区二区av电影网| 咕卡用的链子| 午夜影院在线不卡| 午夜福利网站1000一区二区三区| 操美女的视频在线观看| 中文字幕精品免费在线观看视频| 精品人妻一区二区三区麻豆| 免费观看av网站的网址| 视频区图区小说| 免费久久久久久久精品成人欧美视频| 免费av中文字幕在线| 国产精品亚洲av一区麻豆 | 亚洲成人一二三区av| 日韩一区二区三区影片| 妹子高潮喷水视频| 嫩草影视91久久| 美女午夜性视频免费| 国产精品久久久人人做人人爽| 国产精品秋霞免费鲁丝片| 国产精品人妻久久久影院| 国产人伦9x9x在线观看| 99久国产av精品国产电影| 亚洲精品成人av观看孕妇| av在线老鸭窝| 捣出白浆h1v1| 久久国产精品大桥未久av| 大香蕉久久成人网| 少妇人妻久久综合中文| av片东京热男人的天堂| 亚洲av成人精品一二三区| 久久久久视频综合| 桃花免费在线播放| 最新在线观看一区二区三区 | 欧美激情 高清一区二区三区| 国产男女超爽视频在线观看| 国产成人精品久久二区二区91 | 亚洲,一卡二卡三卡| 高清欧美精品videossex| 精品人妻一区二区三区麻豆| 亚洲国产精品一区二区三区在线| av又黄又爽大尺度在线免费看| 午夜福利影视在线免费观看| 波多野结衣av一区二区av| 国产精品av久久久久免费| 亚洲国产av新网站| 成年人午夜在线观看视频| 国产一区有黄有色的免费视频| 国产av码专区亚洲av| 十分钟在线观看高清视频www| 国产免费一区二区三区四区乱码| 欧美av亚洲av综合av国产av | 精品少妇一区二区三区视频日本电影 | 人人妻人人爽人人添夜夜欢视频| 最黄视频免费看| 国产1区2区3区精品| 中文字幕制服av| 女人久久www免费人成看片| 亚洲天堂av无毛| 久久精品久久精品一区二区三区| 国产在线一区二区三区精| 欧美最新免费一区二区三区| 啦啦啦在线观看免费高清www| 成年人午夜在线观看视频| 日韩精品免费视频一区二区三区| 亚洲精品久久久久久婷婷小说| 国产亚洲最大av| 国产日韩欧美视频二区| www.av在线官网国产| 美女主播在线视频| 美国免费a级毛片| 丝袜喷水一区| 亚洲一区二区三区欧美精品| 日本爱情动作片www.在线观看| 欧美 日韩 精品 国产| 啦啦啦 在线观看视频| 久久久久精品久久久久真实原创| 我要看黄色一级片免费的| 如何舔出高潮| 日韩精品免费视频一区二区三区| 在线精品无人区一区二区三| 精品国产乱码久久久久久男人| 久久天躁狠狠躁夜夜2o2o | 99热国产这里只有精品6| 女人精品久久久久毛片| 国产精品欧美亚洲77777| 中文字幕av电影在线播放| 99九九在线精品视频| 中文天堂在线官网| 汤姆久久久久久久影院中文字幕| 久久这里只有精品19| 制服人妻中文乱码| 2021少妇久久久久久久久久久| 日韩 欧美 亚洲 中文字幕| 欧美精品人与动牲交sv欧美| 校园人妻丝袜中文字幕| 在线观看人妻少妇| 亚洲精品视频女| 亚洲av电影在线观看一区二区三区| 国产精品一二三区在线看| 久久亚洲国产成人精品v| 亚洲一卡2卡3卡4卡5卡精品中文| 日韩免费高清中文字幕av| 国产色婷婷99| 制服诱惑二区| 无限看片的www在线观看| 亚洲国产看品久久| 亚洲色图 男人天堂 中文字幕| av在线播放精品| 嫩草影视91久久| 免费观看人在逋| 啦啦啦在线观看免费高清www| 最近的中文字幕免费完整| 黄片小视频在线播放| 日本午夜av视频| 亚洲成人手机| 亚洲精品久久成人aⅴ小说| 丝袜人妻中文字幕| 99久久综合免费| 国产成人系列免费观看| 街头女战士在线观看网站| 国产1区2区3区精品| 午夜福利视频在线观看免费| 一区二区三区精品91| 亚洲,一卡二卡三卡| 亚洲成人免费av在线播放| 午夜福利网站1000一区二区三区| 国精品久久久久久国模美| 日日啪夜夜爽| 十八禁人妻一区二区| 久久99热这里只频精品6学生| 亚洲美女搞黄在线观看| 精品免费久久久久久久清纯 | 欧美成人午夜精品| 亚洲成人av在线免费| 国产男人的电影天堂91| 老司机靠b影院| 黑丝袜美女国产一区| 在线观看三级黄色| 如何舔出高潮| 午夜福利网站1000一区二区三区| 在线观看免费高清a一片| 一级爰片在线观看| 婷婷色综合大香蕉| 超碰成人久久| kizo精华| 看非洲黑人一级黄片| 午夜福利影视在线免费观看| 99国产精品免费福利视频| 精品一区二区免费观看| 啦啦啦啦在线视频资源| 人人妻人人澡人人爽人人夜夜| xxx大片免费视频| 亚洲欧美日韩另类电影网站| 黄色视频不卡| 精品午夜福利在线看| 黄网站色视频无遮挡免费观看| 久久久国产精品麻豆| 精品久久蜜臀av无| 国产欧美日韩一区二区三区在线| 中文天堂在线官网| 午夜91福利影院| 国产精品香港三级国产av潘金莲 | 久久青草综合色| 新久久久久国产一级毛片| 久久天躁狠狠躁夜夜2o2o | 亚洲国产欧美网| 国产野战对白在线观看| 90打野战视频偷拍视频| 国产极品天堂在线| 国产精品国产三级专区第一集| 成人黄色视频免费在线看| 在线观看免费午夜福利视频| 熟女少妇亚洲综合色aaa.| 黄色怎么调成土黄色| 老司机深夜福利视频在线观看 | 国产一区二区三区综合在线观看| 中文天堂在线官网| 精品国产乱码久久久久久小说| 国产男女内射视频| 中文字幕精品免费在线观看视频| 午夜影院在线不卡| 国产精品无大码| 日日摸夜夜添夜夜爱| 亚洲精品一区蜜桃| 亚洲激情五月婷婷啪啪| netflix在线观看网站| 一级爰片在线观看| 电影成人av| www.熟女人妻精品国产| www.自偷自拍.com| 日韩av不卡免费在线播放| 亚洲av电影在线观看一区二区三区| 热re99久久精品国产66热6| 波野结衣二区三区在线| 亚洲欧洲日产国产| 男人舔女人的私密视频| 日韩av不卡免费在线播放| 高清不卡的av网站| 亚洲第一av免费看| 日韩不卡一区二区三区视频在线| 国产乱来视频区| 成人亚洲欧美一区二区av| 国产精品 欧美亚洲| 亚洲精品国产av蜜桃| 纵有疾风起免费观看全集完整版| 王馨瑶露胸无遮挡在线观看| 男女床上黄色一级片免费看| 国产亚洲午夜精品一区二区久久| videos熟女内射| 国产精品国产三级专区第一集| 咕卡用的链子| 大香蕉久久网| 亚洲国产精品一区三区| 午夜福利乱码中文字幕| 成年女人毛片免费观看观看9 | 国产免费又黄又爽又色| 一区在线观看完整版| 五月天丁香电影| 国产成人精品福利久久| 在线观看国产h片| av在线app专区| 国产精品成人在线| a 毛片基地| 久久久国产精品麻豆| 欧美中文综合在线视频| 亚洲美女搞黄在线观看| 91国产中文字幕| 欧美国产精品va在线观看不卡| 99re6热这里在线精品视频| av网站在线播放免费| 成年av动漫网址| 婷婷色麻豆天堂久久| 夜夜骑夜夜射夜夜干| 99国产精品免费福利视频| 一级片'在线观看视频| 久久人人97超碰香蕉20202| a级毛片在线看网站| 天天躁日日躁夜夜躁夜夜| 欧美激情 高清一区二区三区| 亚洲成国产人片在线观看| 亚洲精品国产区一区二| 亚洲一区二区三区欧美精品| 久久久久精品性色| 亚洲人成77777在线视频| 久久热在线av| 另类亚洲欧美激情| 2018国产大陆天天弄谢| 少妇精品久久久久久久| 亚洲国产av影院在线观看| 日韩一区二区三区影片| 久久天躁狠狠躁夜夜2o2o | 国产xxxxx性猛交| av福利片在线| 国产精品久久久久久精品古装| av网站免费在线观看视频| 日韩欧美精品免费久久| 亚洲,欧美,日韩| 日本欧美视频一区| 亚洲av成人精品一二三区| 热99国产精品久久久久久7| 久久久精品免费免费高清| 人体艺术视频欧美日本| 精品一区二区三区四区五区乱码 | 久久国产亚洲av麻豆专区| 午夜福利一区二区在线看| 亚洲美女视频黄频| 亚洲成人一二三区av| 一级片免费观看大全| 亚洲国产最新在线播放| 国产精品一区二区在线观看99| 欧美久久黑人一区二区| 热99久久久久精品小说推荐| 国产成人啪精品午夜网站| 女人高潮潮喷娇喘18禁视频| 亚洲欧美成人综合另类久久久| 亚洲欧美激情在线| 久久久亚洲精品成人影院| 一本大道久久a久久精品| 九草在线视频观看| 人人妻人人添人人爽欧美一区卜| 午夜免费男女啪啪视频观看| 夫妻午夜视频| 国产淫语在线视频| av网站在线播放免费| 男女午夜视频在线观看| e午夜精品久久久久久久| 晚上一个人看的免费电影| 1024视频免费在线观看| 狂野欧美激情性bbbbbb| 美女大奶头黄色视频| 无限看片的www在线观看| 成人亚洲精品一区在线观看| 黄色毛片三级朝国网站| 国产在线免费精品| 国产成人免费无遮挡视频| 亚洲欧美成人精品一区二区| av天堂久久9| 在线观看免费视频网站a站| 爱豆传媒免费全集在线观看| 亚洲av成人不卡在线观看播放网 | 亚洲精品日韩在线中文字幕| 一二三四在线观看免费中文在| 51午夜福利影视在线观看| 毛片一级片免费看久久久久| 午夜福利免费观看在线| 好男人视频免费观看在线| 日本黄色日本黄色录像| 天美传媒精品一区二区| 亚洲,欧美,日韩| 搡老乐熟女国产| 精品人妻在线不人妻| 777米奇影视久久| 在线观看人妻少妇| 狂野欧美激情性xxxx| 亚洲精品一区蜜桃| 桃花免费在线播放| av在线app专区| 久久这里只有精品19| 纯流量卡能插随身wifi吗| 中文欧美无线码| 日韩一区二区三区影片| 91国产中文字幕| 99热全是精品| 91国产中文字幕| 亚洲欧美激情在线| 亚洲美女黄色视频免费看| 99热全是精品| 欧美精品一区二区免费开放| 婷婷色av中文字幕| 欧美人与性动交α欧美精品济南到| 精品国产超薄肉色丝袜足j| 日韩大码丰满熟妇| 精品国产超薄肉色丝袜足j| 十八禁人妻一区二区| 久久人人爽人人片av| 男女边摸边吃奶| 国产黄色视频一区二区在线观看| 色精品久久人妻99蜜桃| 18在线观看网站| 丁香六月欧美| 久久精品人人爽人人爽视色| 99九九在线精品视频| 亚洲av电影在线进入| 国精品久久久久久国模美| 高清视频免费观看一区二区| 日本av手机在线免费观看| 高清视频免费观看一区二区| 国产99久久九九免费精品| 深夜精品福利| 亚洲精品av麻豆狂野| 高清视频免费观看一区二区| 亚洲 欧美一区二区三区| 91精品三级在线观看| 9色porny在线观看| 欧美日本中文国产一区发布| 久久 成人 亚洲| 19禁男女啪啪无遮挡网站| 久久精品久久精品一区二区三区| 性高湖久久久久久久久免费观看| 好男人视频免费观看在线| 免费高清在线观看日韩| 我要看黄色一级片免费的| 免费女性裸体啪啪无遮挡网站| 高清欧美精品videossex| 肉色欧美久久久久久久蜜桃| 久久精品亚洲熟妇少妇任你| 天天操日日干夜夜撸| 老司机影院成人| 亚洲七黄色美女视频| 在线 av 中文字幕| 色精品久久人妻99蜜桃| 国产亚洲一区二区精品| 日本av免费视频播放| 黄色 视频免费看| 欧美人与性动交α欧美精品济南到| 女人被躁到高潮嗷嗷叫费观| 又大又黄又爽视频免费| 日日啪夜夜爽| 国产精品 国内视频| 黄色毛片三级朝国网站| 国产亚洲av片在线观看秒播厂| 欧美激情高清一区二区三区 | 久久久精品免费免费高清| 免费不卡黄色视频| 青春草国产在线视频| 久久人妻熟女aⅴ| 亚洲,一卡二卡三卡| 国产精品国产av在线观看| 亚洲精品乱久久久久久| 在线亚洲精品国产二区图片欧美| 久久天躁狠狠躁夜夜2o2o | 一级a爱视频在线免费观看| 精品国产露脸久久av麻豆| 老司机亚洲免费影院| 啦啦啦 在线观看视频| 多毛熟女@视频| 亚洲国产毛片av蜜桃av| 欧美中文综合在线视频| 久久av网站| 一区二区三区四区激情视频| 久久女婷五月综合色啪小说| 国产精品熟女久久久久浪| 美女大奶头黄色视频| 日韩一区二区三区影片| 在线观看免费午夜福利视频| 丁香六月天网| 熟女少妇亚洲综合色aaa.| 久久毛片免费看一区二区三区| 欧美精品av麻豆av| 捣出白浆h1v1| 美国免费a级毛片| 久久99一区二区三区| 一二三四在线观看免费中文在| 欧美亚洲 丝袜 人妻 在线| 99热全是精品| 男人操女人黄网站| 高清在线视频一区二区三区| 中文精品一卡2卡3卡4更新| 久久婷婷青草| 精品一区二区三区av网在线观看 | 9色porny在线观看| 久久精品人人爽人人爽视色| 日本欧美国产在线视频| 国产精品国产三级专区第一集| 国产一级毛片在线| 桃花免费在线播放| 人妻 亚洲 视频| 天堂8中文在线网| 亚洲中文av在线| 欧美日韩国产mv在线观看视频| 亚洲第一区二区三区不卡| 亚洲欧美精品综合一区二区三区| 中文欧美无线码| 熟妇人妻不卡中文字幕| 色综合欧美亚洲国产小说| 久久精品aⅴ一区二区三区四区| 亚洲欧洲精品一区二区精品久久久 | av国产精品久久久久影院| 国产一区二区三区综合在线观看| 欧美乱码精品一区二区三区| 久久99一区二区三区| 不卡av一区二区三区| 国产男女超爽视频在线观看| 日韩一卡2卡3卡4卡2021年| svipshipincom国产片| 丝袜喷水一区| 成人国语在线视频| 日日啪夜夜爽| 成人国产麻豆网| 国产精品久久久久久精品古装| 婷婷色麻豆天堂久久| 日韩 亚洲 欧美在线| 日韩一本色道免费dvd| 国产日韩欧美视频二区| 1024视频免费在线观看| 精品少妇一区二区三区视频日本电影 | 久久国产精品大桥未久av| av卡一久久| 五月开心婷婷网| 两个人看的免费小视频| 人妻 亚洲 视频| 亚洲国产中文字幕在线视频| 精品人妻熟女毛片av久久网站| 精品少妇内射三级| 成人毛片60女人毛片免费| 久久久精品区二区三区| 亚洲av男天堂| 考比视频在线观看| 91精品国产国语对白视频| 免费观看人在逋| 在线免费观看不下载黄p国产| 中文字幕人妻熟女乱码| 亚洲av成人不卡在线观看播放网 | 日韩电影二区| 菩萨蛮人人尽说江南好唐韦庄| 丝袜在线中文字幕| 一区二区三区精品91| 在线免费观看不下载黄p国产| 国产福利在线免费观看视频| 亚洲欧美清纯卡通| 国产熟女欧美一区二区| 午夜日韩欧美国产| 亚洲av日韩精品久久久久久密 | 性少妇av在线| 亚洲av在线观看美女高潮| 亚洲美女视频黄频| 男人舔女人的私密视频| 国产国语露脸激情在线看| 狠狠精品人妻久久久久久综合| 亚洲熟女精品中文字幕| 国产精品久久久久久精品古装| 国产精品成人在线| 另类精品久久| 成人黄色视频免费在线看| 美女视频免费永久观看网站| 久久99精品国语久久久| 国产乱人偷精品视频| 久久久精品94久久精品| 亚洲中文av在线| 曰老女人黄片| 国产极品粉嫩免费观看在线| 一区福利在线观看| 中文字幕av电影在线播放| 国产福利在线免费观看视频| 如何舔出高潮| 高清av免费在线| 日韩,欧美,国产一区二区三区| 一本大道久久a久久精品| 免费人妻精品一区二区三区视频| 精品亚洲成a人片在线观看| 国产人伦9x9x在线观看| 色网站视频免费| 亚洲精品成人av观看孕妇| 亚洲国产av新网站| 亚洲视频免费观看视频| 天天躁日日躁夜夜躁夜夜| 日韩大码丰满熟妇| 国产野战对白在线观看| 9色porny在线观看| 天堂中文最新版在线下载| 国产无遮挡羞羞视频在线观看| 一级毛片 在线播放| 国产不卡av网站在线观看| 国产成人a∨麻豆精品| 日韩人妻精品一区2区三区| 熟妇人妻不卡中文字幕| 免费av中文字幕在线| 人人妻人人澡人人爽人人夜夜| 狂野欧美激情性bbbbbb| 日韩一卡2卡3卡4卡2021年| kizo精华| 一本大道久久a久久精品| 黄片无遮挡物在线观看| 国产又爽黄色视频| 久久久久久久精品精品| 纵有疾风起免费观看全集完整版| 国产欧美日韩一区二区三区在线| 国产精品三级大全| 夫妻性生交免费视频一级片| 一级a爱视频在线免费观看| 菩萨蛮人人尽说江南好唐韦庄| 免费看不卡的av| 国语对白做爰xxxⅹ性视频网站| 超色免费av| 国产野战对白在线观看| 我的亚洲天堂| 亚洲精品aⅴ在线观看| 久久人人97超碰香蕉20202| 男女高潮啪啪啪动态图| 大陆偷拍与自拍| 精品国产超薄肉色丝袜足j| 精品午夜福利在线看| 九草在线视频观看| 亚洲欧美精品综合一区二区三区| 日本vs欧美在线观看视频| 久久av网站| 色婷婷久久久亚洲欧美| 别揉我奶头~嗯~啊~动态视频 | 97在线人人人人妻| 亚洲情色 制服丝袜| 亚洲国产精品一区二区三区在线| 中文字幕制服av| 激情五月婷婷亚洲| 欧美激情高清一区二区三区 | 可以免费在线观看a视频的电影网站 | 精品午夜福利在线看| 19禁男女啪啪无遮挡网站| 一级a爱视频在线免费观看| e午夜精品久久久久久久| 在线精品无人区一区二区三| 国产片特级美女逼逼视频| 久久久精品国产亚洲av高清涩受| 熟女av电影| 老司机影院成人| 亚洲欧美成人综合另类久久久| 宅男免费午夜| www日本在线高清视频| 欧美日韩福利视频一区二区| 一区二区日韩欧美中文字幕| 日韩欧美精品免费久久| av免费观看日本| 少妇被粗大的猛进出69影院| 婷婷色综合www| 日韩伦理黄色片| 国产av一区二区精品久久| av电影中文网址| 别揉我奶头~嗯~啊~动态视频 | 日韩 欧美 亚洲 中文字幕| 丝袜美足系列| 91aial.com中文字幕在线观看| 日韩精品免费视频一区二区三区| 精品国产国语对白av| 国产精品偷伦视频观看了| 男女无遮挡免费网站观看| 午夜91福利影院| 日本午夜av视频| 免费看不卡的av| 色婷婷久久久亚洲欧美| 极品人妻少妇av视频| 老汉色av国产亚洲站长工具| 97精品久久久久久久久久精品| 一二三四中文在线观看免费高清| 亚洲精品日韩在线中文字幕| 久久国产精品男人的天堂亚洲|