    Design and development of a plastic scintillator based whole body β/γ contamination monitoring system

    2021-06-04 08:42:50VaishaliThakurAmitJainAshokkumarRekhaAnilkumarPravinSawantProbalChaudhuryChaudhari
    Nuclear Science and Techniques 2021年5期

    Vaishali M. Thakur· Amit Jain · P. Ashokkumar · Rekha Anilkumar ·Pravin Sawant · Probal Chaudhury · L. M. Chaudhari

    Abstract Plastic scintillation detectors based whole body β/γ contamination monitors are developed for use in radiation facilities. This microcontroller-based multi-detector system uses 13 plastic scintillator detectors, with minimized dead detection zones, monitoring the whole body,and conforming to the contamination limit prescribed by the regulatory authority. This system has the features for monitoring hands, feet, head, and face β/γ using contamination monitors and portal exit monitors. It can detect gamma sources at a dose rate of 10 nGyh-1.The system is calibrated using β sources 90Sr/90Y, 204Tl,and 36Cl,and the efficiency is found to be 29%,22%,and 18%,respectively.The minimum detectable β/γ contamination is 0.15 Bqcm-2,which is significantly less than the minimum detection objectives on head, face, hands, and feet.

    Keywords Plastic scintillation detector · Contamination monitor · Microcontroller · Minimum Detection Objective(MDO) · Minimum Detection Concentration (MDC)

    1 Introduction

    Monitoring radioactive contamination on the external body parts of a radiation worker is an integral part of radiological surveillance in a radiation facility. The monitoring technique and system should be adequate to properly assess the occupational radiation safety being maintained and compliance with the requirements of safety directives by regulatory authorities. Contamination monitors are essential for ensuring that the external radioactive contamination levels of an individual are within the regulatory limits. A properly designed contamination monitor is a regulatory requirement that controls radiation exposure,prevents the spread of radioactive contaminations,helps to provide information regarding the regular conditions of the facility,and ensures good working practices.Various types of radiation detectors, such as pancake GM detector, gas flow proportional counters, plastic scintillator and ZnS(Ag) scintillation detectors, are utilized for developing contamination-monitoring systems[1–3].An alpha surface contamination monitor based on a thick gaseous electron multiplier was proposed by Xiao et al. [4]. Qu et al. [5]reported a new method based on position-sensitive measurements of beta surface contamination, where the detector was a large-area plastic scintillator. The simulation and experimental results provided accurate information of the contamination distribution. Gas flow counterbased contamination monitors require a constant flow of the detector gas medium,and its maintenance is difficult if the entrance window foil ruptures. In the case of gas-flow proportional counters,a constant supply of filled gas at the required pressure must be ensured. Geiger Muller (GM)counter based systems are cheaper; however, despite the use of numerous GM tubes, they suffer from large dead zones,thereby reducing the reliability.NaI(Tl)scintillator detector-based quick scan whole body walkthrough monitors can detect internally deposited gamma-emitting radioisotopes and can be effectively utilized for quick decision making on countermeasures in case of an emergency [6]. An ideal detector used in body contamination monitoring should have a large surface area to cover the body part to be monitored, high beta detection sensitivity of beta energies of ~100 keV and above, and good gamma rejection,limiting the minimum detectable activity(MDA). Plastic scintillation detectors for beta and gamma radiation monitoring applications are extensively used mainly because of the easy availability of large-area detectors having the required specifications of shape and size, low price, good sensitivity, and good reliability[7–12]. In addition, they can be easily implemented for operations in rugged fields and require little maintenance.

    Occupational workers engaged in radiological operations must be monitored for external beta and gamma contamination and screened according to national regulatory specifications [13]. No transferable contamination on body surfaces is allowed. This entails regular personnel contamination monitoring as per the regulatory stipulations to ensure that no individual carries any radioactive contamination out of the facility. Radiation workers must use radiation monitoring instruments, protective equipment,clothing,and other devices appropriately provided to them to ensure personnel safety. They always use personnel protective wear such as lab coats or specific plant cloths,and the possibility of external β contamination on the front and back sides of the body is below average in normal operating conditions. However, the possibility of β contamination on other parts such as hands and the sole of shoes cannot be ignored, and it is mandatory to undergo a self-check before leaving the workplace/laboratory. Most of the radioactive materials handled are beta gammaemitting sources. Despite adequate precautions taken during handling of these sources, there is a possibility of radioactive contamination over the most vulnerable parts such as the hands and feet. Based on the ICRP ALARA principle,exposure of individuals to radiations should be as low as possible, cognizant of social and economic considerations. Therefore, a plastic scintillator detector-based whole body contamination monitoring system is designed and developed.Commercially available plastic scintillation detectors suitable for our application are procured for this purpose. The system is a combination of portal and contamination monitors, comprising cylindrical plastic scintillation detectors (CPDs) and large-area plastic scintillation detectors (LPDs) for the detection of gamma and beta/gamma contamination, respectively. This wholebody contamination monitor proves very effective in the estimation of personnel contamination at radiation facilities where beta/gamma radioactive isotopes are handled. The system is equipped with a user friendly interface and will help individuals to complete a self-check for potential radioactive contamination, thereby limiting the associated radiation risks. The design aspects, development methodology,and characterization results of a body contamination monitor based on plastic scintillation detectors are presented in this paper.

    2 System design and description

    The system comprises 13 plastic scintillator detectors based on combinations of two different shapes and sizes.It uses eight rectangular LPDs (350 mm × 150 mm × 0.5 mm) to monitor β and γ contamination on the hands, feet,face, and head and five cylindrical CPDs (1000 mm long and a diameter of 51 mm) to monitor γ contamination on the front and back sides of the body. Figure 1 shows the system after complete fabrication during testing and evaluation in our laboratory.

    Fig. 1 (Color online) Plastic scintillator detectors based whole body contamination monitoring system

    3 Detectors

    The LPD has a 350 mm × 150 mm × 4 mm light guide and an embedded 2π sensitive photomultiplier tube(PMT) provided with a 100 MΩ resistor divider having a maximum external dimension of 358 mm × 158 mm ×42 mm. It is procured as an integral assembly. The thin plastic scintillator detector has a thickness of 0.5 mm,which ensures the detection of beta particles with excellent gamma rejection.

    The detector has an 8.8 mg cm-2entrance window as a reflector and a 20 μm polyamide (Kapton) window for scratch protection. The detector is fitted inside a stainless steel(SS)enclosure tray.Above the Kapton window,an SS honeycomb mesh of 0.1 mm thickness with 90% opening is fitted to the SS tray to protect it from any rough handling during the operation. The entire assembly is wrapped in a polyethylene sheet and placed in another 1 mm thick SS box having a lining of 3 mm lead sheets at the bottom and four sides of the detector for gamma shielding. The LPD has two flying leads RG-174 wires to have high voltage(HV) input and signal output connected to miniature high voltage (MHV) and Bayonet Neill Concelman (BNC)connectors, respectively. The connectors are fitted to a C clamp attached to the SS box.The LPD is light tight and its performance is appropriately tested before it is integrated into the system.

    The CPDs are assembled in our laboratory after procuring bare cylindrical plastic scintillators having a length of 1 m and diameter of 51 mm.The plastic detector is wrapped in a thin aluminum foil reflector and coupled to a 51 mm diameter PMT (9266KBS). The detector housing is a 1000 mm long cylindrical polyvinyl chloride (PVC)pipe with one end cap,and the other end is open,where the PMT is coupled to the detector using silica gel. The inner diameter of the PMT enclosure is 52 mm with a screwing arrangement housing to have a light tight assembly.Fourteen pin base B14A socket with divider networks interconnect the PMT and the preamplifier (PA) unit. The detector, PMT and PA are called CPD module and placed in the system with 180° covered (semicircle of backside detector) using a 3 mm lead sheet to provide gamma shielding. The assembled CPD module in our laboratory is rugged for normal handling and requires no further maintenance,and care is taken to make it completely light tight.

    4 System frame

    A SS material is used for the fabrication of the system to have a rigid body, to ensure ease of cleaning and decontamination and help to reduce the gamma background around the detection area. The SS frame has an external physical dimension of 100 cm width, 220 cm height, and 60 cm depth. The frame comprises two feet grill window slots below each installed LPD module. It has two frontal hand monitoring stations with two slots to monitor both hands.Slots are made available at a height of 120 cm with two LPDs facing each other arranged at a gap of 10 cm to monitor beta/gamma hand contamination on either side of the palm.Two infrared(IR)switches are fitted at a depth of 25 cm inside the slots to ensure appropriate positioning of the hands inside the slot. The head and face (HF) monitoring station has two LPDs mounted at right angles to each other. The HF LPDs are mounted on a moving frame to accommodate the likely height variation (150–185 cm)among occupational workers. The HF detectors are mechanically fitted at the top of the SS frame having a sliding pulley arrangement with a counter dead weight balance. The CPDs are arranged within the SS frame such that three detectors are on the backside and two detectors are placed in front to monitor back and chest side of the worker respectively. Perspex sheet (3 mm) covering the inner side of the frame where the person is being monitored. All other sides are covered with 1 mm thick SS sheet.The back CPDs are spaced at a 10 cm gap fixed at a height of 20 cm from the bottom, whereas the two front CPDs are at gap of 5 cm at the same elevation.

    5 Power supply

    The system operates on a 230 V AC, 50 Hz mains supply.A switch mode power supply is used to convert the AC mains to DC levels of ± 12 V and + 5 V, and is placed at the bottom of the back monitoring section of the frame. A 12 V DC is required to operate seven HV modules utilized to power the 13 detectors on a twin sharing basis. The detector processing electronic cards require a DC power supply of ± 12 V and 5 V. The connectors are placed on the power supply distributor cards to energize the respective electronic components of the entire system.

    6 Processing electronics

    Visible light is generated by the interaction of radiations with the detector material.The detector signal is converted to an electrical pulse by the PMT, which is coupled to the detector. The HV of 750 V required for the PMT is generated using commercially available DC to DC converter HV modules.Pulse processing electronics convert the PMT signals into digital pulses. Each detector module is interfaced with the PA, amplifier and discriminator (PAD)electronics to convert detector pulses to transistortransistor logic (TTL) output pulses. Figure 2 shows a functional block diagram for processing electronics of the system. The PA, placed near the detectors, provides an interface between the detector and pulse processing electronics.

    Its output is fed to an amplifier circuit using operational amplifier integrated circuits(ICs).The amplified pulses are discriminated from noise using a comparator IC. Pulses of amplitude less than 200 mV are rejected for noise discrimination, and the pulses of higher amplitude are further converted to TTL pulses using a monostable multivibrator IC. The typical PAD output TTL pulse width is 5 μs. The PAD cards are placed near the respective detectors, and TTL pulses are fed to the counting electronics through 50 Ω impedance coaxial BNC cables.

    Individual detector TTL pulses are counted by a 13 channel counter timer module (CTM) and processed using microcontroller-based control electronics called singleboard computer (SBC). The CTM is developed using Philips PCF8583/PCF8593 14 ICs. One of the ICs is used as a real-time clock (RTC) to maintain date/time information, and the rest are used in the event counter mode.Each IC communicates with the SBC using an interintegrated circuit (I2C) bus on a data pin (SDA) and clock pin(SCL).All the SCL pins of the 14 ICs are tied high with the microcontroller SCL and connected in parallel. Individual RTC/event counter IC SDAs are connected one at a time to the microcontroller SDA through two 8-to-1 decoder ICs (40C51).The selection of the event counter is performed using six select control lines mapped in the data memory. Each detector TTL pulse from the detector is connected to one of the inputs of the AND gate, and other inputs are controlled by the enable/disable signal of the counter from the SBC card.The output of the AND gate is connected to the respective counters.

    The SBC is based on Philips 80C552 [14] microcontroller, which is interfaced with a 32 kilo byte (kB) Erasable and Programmable Read Only Memory (EPROM)IC,32 kB Random Access Memory(RAM)IC,240 × 128 dots alpha-numeric graphic liquid crystal display (LCD),CTM,audio/visual high alarm indicator,personal scanning mode status indicator switches, RS-232 serial port interface, and panel PC, as depicted in Fig. 3. The distinct advantages of microcontroller-based control electronics are their simplicity, low power consumption, failure-proof continuous operation, and operational flexibility.

    Fig. 2 (Color online) Block diagram of the processing electronics for the whole body contamination monitoring system

    Using the master (microcontroller) slaves (RTC/event counter ICs) communication protocol, the SBC and CTM communicate with each other through the I2C bus and interface with each other using a 10 pin flat ribbon cable(FRC). The counters and RTC of the CTM are memory mapped with a microcontroller. The CTM can be controlled sequentially by selecting each successively by the SBC using select control lines, SDA and SCL. The event counter mode includes start, stop counting, and a reading counter. The RTC can be loaded with date/time and read such that data can be logged and stored.The global enable or disable signal from the SBC to the CTM is provided to have the same time slot counting of all detector pulses.

    The RAM and RTC have a lithium battery backup for data and RTC date/time retention,respectively,in the case of a power failure. The microcontroller has a watchdog timer, which automatically resets the system if it goes out of synchronization so that it has a 24 × 7 h system operation.The system acquires the detector pulses by CTM for a set acquisition time. If the registered counts exceed the set alarm level, a high alarm is initiated, and the data are stored in the memory with the date and time information in its RAM. The audio-visual alarm module is a self-contained unit designed to provide audio and high visual alarms.The 32 kB data memory is used to save high alarm data with date/time information and system parameters.If a high alarm is not set, the system continues to operate. The SBC is interfaced with personal scanning mode status indicator switches, S1 to S5, to have a close watch on the standing position of the scanned person within the system.S1 and S2 are light sensors that identify whether the monitoring person is positioned properly over the foot detector frames in the system.If the HF monitoring station is properly adjusted, switch S3 is activated. The IR switches, S4 and S5, are activated when the person to be monitored correctly sets their hands in the hand monitoring slots and presses it. A 240 × 128 dot alpha-numeric/graphics LCD(Oriole make model number:OGMY-24011SF) with backlight is used to provide clear visibility even in low ambient light conditions for displaying counts registered by the detectors. The LCD is memory mapped and selected using chip select (CS) and controlled using enable,register select,and eight data lines.The operational functionality of the system is governed by controller electronics with embedded software in the EPROM. The software is developed using the assembly language in a Keil environment.The SBC board is used as a development kit with a program and data memory overlapped in the upper 32 kB address (8000H to FFFF). The embedded monitor program allows hex file to be downloaded in the RAM to be executed for testing.The final tested program is burnt in the EPROM to have a system’s operational program.

    Fig. 3 Functional block diagram of a microcontroller-based single board computer (SBC)

    The system also includes a personal computer (PC) to facilitate data analysis using the graphical user interface(GUI) software. The SBC transfers the acquired detectors counts to the PC in the Windows operating system using a standard serial RS232 protocol. The data are displayed to communicate whether the contamination is more than the permissible level prescribed by the regulatory authority[13].

    The design of the system is rugged with SS body and may require very little corrective maintenance. Routine preventive maintenance of the system may include periodic cleaning of the displays and sensors using a soft cloth dampened with only water. The SS frame and lead shielding of the detectors make the system very heavy and fixed at a location.

    7 Methodology

    When the power is ON,the software initializes counters,memory, LCD, and serial interfaces. The system software acquires 13 detector data for the set data acquisition time(1–15 s) continuously to assess the background radiological status of the system located at the workplace. The detector counts are displayed on the LCD and sent to the PC along with the polling status of switches S1 to S5. The situation of uninterrupted switches leads to a background acquisition.This mode of operation is useful for generating baseline background data using various standard sources to ensure proper functionality and setting of the alarm level.It can also be utilized for the periodic evaluation of the efficiency of the system. The background count from each detector when the person is standing on the contamination monitor is less than the earlier registered counts in the absence of the person,which is inferred to be owing to selfshielding effects of individuals and compensated in the personal contamination scanning mode.

    When the worker walks in the frame and properly stands on the feet, the IR proximity switches S1 and S2 are interrupted. Thus, the system waits for the scan mode, and instructions are displayed on a PC monitor with audio instruction to pull and properly adjust the HF monitor with face and head, thereby interrupting switch S3 located on the face detector.Once S3 is interrupted,the instruction to place hands correctly into the slots and to press switches S4 and S5 is given. Once all the switches are activated (S1 to S5),the person is asked to wait until the data acquisition is complete. A count down from 10 to 0 is displayed per second.Once the acquisition time is‘‘0’’,the instruction to remove hands position back HF monitor module and to leave the monitoring area of the system is given. If the person does not position properly, depending upon the status of the associated switches,an appropriate message is displayed on the PC monitor to follow the instructions.For example, if the left hand switch is not interrupted, a message ‘‘insert left hand properly’’ will be displayed on the monitor after a beep sound. A graphical representation of the radiological status of a human figure is shown.If any of the detectors show higher counts than the set limit for a high alarm, it will be indicated by a green to red color change of the corresponding body part on the human figure and displayed on the monitor. The system captures the photograph and saves, along with data waits for proper countermeasures to be taken by the health physicist. An audio visual high alarm is generated and is ON until the acknowledgment signal is initiated. In a report on high alarm data with RTC timing data, all counts of CPDs converted to dose rate and that from LPDs converted to activity is updated in PC as a pdf file. The data are also displayed on the LCD and stored in the SBC’s memory,which can be retrieved.If the radiological status is normal,then green LED glows and ‘thank you and have a good day’ message with the radiological status on the human figure is displayed.

    When the contamination monitor is first powered ON,it takes approximately two minutes to complete the initiation of the system, which also includes the booting of the computer. Subsequently, the system automatically enters into the background count acquisition mode. The system will be ready to monitor any individual worker after the first set of background data is acquired.

    8 Results and di scussions

    Eight LPDs and five CPDs were assembled and integrated with the electronics inside the SS frame.The results of the performance evaluation of the detectors and calibration results of the system are described in the following sections. The specifications prescribed by the International Electro-technical Commission (IEC) [15] and similar guidelines by the Bureau of Indian Standards(BIS)[16]for hand and foot contamination monitors have been adopted during the development of this system.

    9 System performance evaluation

    The LPDs are used to assess the β/γ contamination status of the hands, feet, head, and face of the monitored individual. The combined entrance window thickness of the detector is 20 mg cm-2,and the corresponding cutoff β particles of energy are less than 100 keV[17].Experiments were conducted to study the variation in β detection efficiency over the surface of the detector and with respect to the source to detector distance.

    To estimate the percentage variation of beta detection efficiency over the surface of a large-area plastic scintillator, a study was conducted by recording the efficiency along the length of the detector using a204Tl beta source.Figure 4 shows the variation in beta detection efficiency over the surface of the LPD. The scintillator exhibited a slightly lower efficiency at the borders of the detector on both sides. The profile was observed to be slightly asymmetric,and the maximum efficiency was at the center.The efficiency varied between 19.6% at the sides to 22% at the middle along the central line and lengthwise.At the corners of the detector, the efficiency decreased to 17–18%. With gamma-emitting standard sources137Cs and60Co at the center of the detector, the γ response of the LPD was observed to be 400 cps μGy-1h-1.

    The detector response with the variation of distance between the pure beta-emitting source204Tl beta source(742 Bq,endpoint energy 763 keV)was observed.Because the detector response was the maximum at the center, the source was kept at the center and moved away from the center.Figure 5 presents the distance vs.efficiency plot for the204Tl beta source. The counting efficiency of the LPD on its surface at distances of 5 mm, 10 mm, 20 mm,30 mm, and 50 mm at the center of the detector was recorded and plotted. A maximum efficiency of 22% was observed on the surface. It varied from 15 to 4% at distances of 5 and 50 mm. The geometry of the LPD and status indicator switches were arranged such that the source of the contamination (hands, feet, head and face) was not more than 10 cm from the respective detector, for which the efficiency was ~13%.

    After installing the detectors in the system, the detector performance was evaluated using a204Tl source placed on contact at nine points, such as four corners, four center points of surface edges,and a central point.The percentage efficiency variation was ± 10% for all eight detectors.

    Fig. 4 Efficiency variation over the surface along the length of LPD

    Fig. 5 Distance versus efficiency plot along the central axis of LPD

    The LPD response to the activity variation for the137Cs source was performed to determine the linearity in an activity range of approximately 100–2500 Bq.The sources were prepared by depositing known activity on an aluminum planchette using a standard137Cs solution.

    These sources were kept 3 mm from the center of the detector,and readings in CPS were recorded.The observed CPS versus activity with standard deviation for different standard activity of the source used is shown in Fig. 6.Very good linearity with beta activity response was observed, and the efficiency was 11% with a sensitivity of 0.638 Bq/CPS.

    For the detector response evaluation, test point sources were used. In an actual scenario of personnel contamination monitoring,the contaminant source may not be a point source.Therefore,the efficiency variation from the point to the90Sr/90Y spread-out source (15 cm × 10 cm) was 29%and 33%, respectively, when placed at the center of the detector in contact. The spread-out source was kept at six different locations in three columns and two rows. The efficiency varied from 33 to 35%, and the variation was less than that of the point source when both moved along the surface of the detector.

    Fig. 6 LPD response to varying activities of 137Cs source

    Fig. 7 Gamma response of CPD for 60 μCi, 137Cs disk source

    The gamma response of the CPD was evaluated using a 60 μCi137Cs disk source. The source to detector distance was varied up to 40 cm along the middle of a vertically mounted CPD. Counts were recorded for 10 s. Figure 7 shows a plot of counts (for 10 s) versus the source to detector distance from the center of the detector. At a monitoring distance of approximately 5 cm, the response was 58 cps μCi-1.

    The count rate variation along the length of the detector was within 10%, and the radial response was nearly uniform. A set of data were recorded to study the dose rate response of the CPD for gamma radiation exposure. Figure 8 shows the linear response of the detector up to 60 μGy h-1with a sensitivity of 1000 cps μGy-1h-1,and the pulses turned to saturate beyond a dose rate of 100 μGy h-1.

    The detector energy response was evaluated using241Am,137Cs, and60Co to cover the energy range of 60 keV to 2 MeV. The response of the detector was independent of the energy,and the variation in the detector response was observed to be within ± 20%.

    The average background counts of the detectors, when tested in our laboratory, were observed as approximately 20 cps for the LPD and approximately 50 cps for the CPD.The data acquisition of the detectors in the background mode was studied after keeping a known137Cs test source near the system for a brief period. Figure 9 shows the combined responses of all detectors on the introduction and removal of the test source.In addition,it was observed that the CPDs yielded a higher response than the LPDs conforming to their position and gamma sensitivity.

    Fig. 8 Linearity test of CPD for different gamma dose rates

    Fig. 9 (Color online) Responses of the detectors on the introduction and removal of a γ source

    The detector geometry and their placement were carefully optimized for probable β, γ contamination detection on the hands,feet,head and face.This arrangement was set to ensure minimal dead detection zones in such a system.Five sensors were used to monitor the appropriate positioning of the hands, feet, and face/head of the personnel.

    This, in turn, ensured the accuracy of the beta/gamma contamination measurement by the system. Such position sensors are to be incorporated into contamination monitors based on the requirements of IEC standards.Moreover,the main operating system is provided with an uninterruptible power supply. Key features of the contamination-monitoring system are listed in Table 1.

    A reduction in recorded counts with respect to the prevailing background counts was observed for all the detectors, whereas non-contaminated personnel stood in the contamination monitor for scanning. The observed reduction in the count rate can be attributed to self-shielding.Human subjects monitored can vary in height from 145 to 185 cm and weight from 45 to 100 kg,so the effect of selfshielding.The average percentage reduction in background counts was maximum for the HF detectors (approximately 12%) and approximately 6% for the hand and foot detectors.

    It should be noted that the pair of hand detectors is covered and shielded from all sides except the slot for hand entry, whereas HF detectors are not shielded along one surface. The CPDs showed a reduction of 2–4%. This observation was incorporated into the software to account for the corrected background counts during the process of personnel contamination monitoring.

    Table 1 Key features of whole body contamination monitor

    10 Calibration and alarm setting of the system

    The LPDs were calibrated using standard radioactive disk sources.The efficiencies(on contact)were 29%,22%,18%, and 11% for90Sr/90Y,204Tl,36Cl and137Cs sources,respectively.

    The limit of β/γ fixed contamination, which is the derived working level (DWL) prescribed by the national regulatory authority, is 350 Bq for the hands and 0.37 Bq cm-2for personal shoes, whereas 1.5 Bq cm-2for skin contamination and 2 Bq cm-2for personal clothing [13]. Therefore, our minimum detection objective(MDO)is set by a regulator that is equal to the DWL value.In the case of an average 300 cm2measurement area of both palms having four LPDs for hand contamination monitoring, the corresponding MDO is 90 Bq. In the case of an average 300 cm2foot measurement area, the MDO for the detector was assumed to be 100 Bq.

    The minimum detection concentration(MDC)values for the system are calculated using the Currie equation given in Eq. (1) [18].

    where NDis the minimum number of counts needed from the source to ensure that the false-negative rate is not larger than 5% and σNBis the square root of the background counts. Equation (2) provides the corresponding MDC value.where T is the counting time, which is set to 10 s and E is the efficiency of the LPD.

    The MDC values were estimated using an average background count rate of 20 cps and an efficiency of 15%at a distance of 5 mm for the204Tl β source and found to be 45 Bq for the LPD.The corresponding contamination level was estimated as 0.15 Bq cm-2. The minimum detection level (MDL) is less if either the source is in contact with the detector or if the source endpoint energy is greater than 763 keV. Thus, the alarm level can be set anywhere between the MDC and MDO values.

    The average background count rate for the CPD was approximately 50 cps. It corresponded to nearly 1000 cps per μGy h-1for a prevalent laboratory background radiation dose rate 50 nGy h-1. The MDC for the CPD was also calculated using Eqs. (1) and (2). The NDvalue was obtained 11 cps with the Currie equation for the counting time of 10 s. Therefore, the contamination-monitoring system can detect a gamma source with a minimum dose rate of 10 nGy h-1, which corresponds to an activity of 280 Bq of137Cs and 76 Bq of60Co at 5 cm from the CPD surface for a counting duration of 10 s.These activity levels are much below the exemption levels set by international bodies[19].To identify the inadvertent movement of radioactive sources, an alarm level of 1000 cps was set for the CPDs when counted in the background mode.Generally,this instrument is placed at the exit of the white zone area of a facility where the radiation dose rate should not exceed 1 μGy h-1[13].

    11 Conclusion

    A whole-body β/γ contamination monitor was developed by employing 13 plastic scintillator detectors. The hands, feet, face, and head were monitored independently using appropriate position sensors as per IEC and BIS standards. The MDC values of the detectors employed in the system were found to be significantly lower than those prescribed by the national regulatory authority.The MDLs for hand and foot contamination detectors(LPDs)for204Tl beta-emitting source at 5 mm detector to source distance were found to be 45 Bq and 0.15 Bq·cm-2, respectively.This corresponded to 50% and 40% of the respective MDOs. This system is likely to serve as a combination of portal exit and contamination monitors,and it can function as a dual system. Any gamma radiation source movement through the system can be alerted with an MDL of 10 nGy h-1. The system is appropriately designed to satisfy the regulatory requirements and will help to complement a series of upstream controls in the facility to ensure radiological safety.

    AcknowledgementsWe gratefully acknowledge the guidance and encouragement given by Shri Suresh Babu, Director, HS&E Group,BARC during this study. We thank our colleagues Smt. Yogita Nivalkar, Shri Lal Bahadur Shastri, Smt. Megha Alat, and Shri C Andrew for their support and cooperation during the course of this development work. We also appreciate the help rendered by our colleagues Shri Kailash Gharat and Shri Chintaman Gaikwad in finalizing the concept. We thank our colleagues at the RSSD workshop Kailas Belwalkar, Shri Nitin Ghaitadake, Shri D. N. Patil, and Shri Pratap Patil, for the fabrication of the mechanical assembly.

    Author contributionsAll authors VMT, AJ, PA, RA, PS, PC and LMC contributed to the study conception and design. Material preparation by VMT, AJ, PA and PS. Data collection by VMT and PS.Analysis were performed by VMT,PA and RA.The first draft of the manuscript was written by VMT. All authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.

    全区人妻精品视频| 亚洲第一av免费看| 亚洲欧洲日产国产| 亚洲av日韩在线播放| 久久久久久久久久成人| 午夜免费观看性视频| 伊人久久国产一区二区| 国产日韩欧美视频二区| 日本爱情动作片www.在线观看| 亚洲国产欧美在线一区| 国产精品人妻久久久影院| 九九久久精品国产亚洲av麻豆| 插逼视频在线观看| 狂野欧美白嫩少妇大欣赏| 国产精品女同一区二区软件| 免费久久久久久久精品成人欧美视频 | 伦精品一区二区三区| 又黄又爽又刺激的免费视频.| 国产乱来视频区| 在现免费观看毛片| 18禁在线无遮挡免费观看视频| 久久精品久久精品一区二区三区| 人人妻人人澡人人看| 久久精品国产亚洲av涩爱| 五月天丁香电影| 美女主播在线视频| 国产白丝娇喘喷水9色精品| 免费黄频网站在线观看国产| 狂野欧美白嫩少妇大欣赏| 性色avwww在线观看| 曰老女人黄片| 欧美日韩av久久| 国产成人免费无遮挡视频| 色哟哟·www| 最新中文字幕久久久久| 欧美精品一区二区大全| 岛国毛片在线播放| 久久久久国产网址| 国产精品国产av在线观看| 免费看av在线观看网站| 亚洲国产最新在线播放| 欧美国产精品一级二级三级| 最黄视频免费看| 高清午夜精品一区二区三区| 日韩成人伦理影院| av黄色大香蕉| 久久99热6这里只有精品| av国产久精品久网站免费入址| av免费在线看不卡| 国产精品国产三级专区第一集| 国产高清不卡午夜福利| 一区二区日韩欧美中文字幕 | 亚洲国产av影院在线观看| 熟女电影av网| 岛国毛片在线播放| .国产精品久久| 久久久久久久久久成人| 精品99又大又爽又粗少妇毛片| 大片免费播放器 马上看| 美女福利国产在线| 一本久久精品| 春色校园在线视频观看| videossex国产| 国精品久久久久久国模美| 综合色丁香网| 91精品国产国语对白视频| 精品国产国语对白av| 一二三四中文在线观看免费高清| 国产成人免费观看mmmm| 免费高清在线观看视频在线观看| 亚洲av欧美aⅴ国产| 亚洲精品乱码久久久久久按摩| 国产黄色免费在线视频| 久久热精品热| 黑丝袜美女国产一区| 亚洲无线观看免费| 久久久久久久亚洲中文字幕| 99热全是精品| 蜜桃国产av成人99| 精品99又大又爽又粗少妇毛片| 热re99久久精品国产66热6| 欧美bdsm另类| 久久久午夜欧美精品| 男女边摸边吃奶| 免费观看a级毛片全部| .国产精品久久| 一区二区日韩欧美中文字幕 | 男女高潮啪啪啪动态图| 日韩不卡一区二区三区视频在线| 男女边摸边吃奶| 欧美三级亚洲精品| av网站免费在线观看视频| 黑丝袜美女国产一区| 亚洲精品美女久久av网站| 亚洲人成网站在线观看播放| 日本91视频免费播放| 亚洲国产成人一精品久久久| 亚洲精品第二区| 中文乱码字字幕精品一区二区三区| 美女内射精品一级片tv| 色视频在线一区二区三区| 国产精品三级大全| 成人国产av品久久久| 制服人妻中文乱码| 久久精品久久久久久噜噜老黄| 人成视频在线观看免费观看| 黄色一级大片看看| 国产精品一区二区三区四区免费观看| 老司机影院毛片| 日本黄大片高清| 美女大奶头黄色视频| 91久久精品电影网| 国产一区二区三区av在线| www.色视频.com| √禁漫天堂资源中文www| 精品一区在线观看国产| 国产片内射在线| 久久99热这里只频精品6学生| 青春草亚洲视频在线观看| 国产成人精品婷婷| 亚洲精品色激情综合| av免费在线看不卡| 黑人猛操日本美女一级片| 青春草亚洲视频在线观看| 日日摸夜夜添夜夜爱| 日韩欧美精品免费久久| a 毛片基地| 999精品在线视频| 天堂俺去俺来也www色官网| a级毛色黄片| 亚洲少妇的诱惑av| 久热这里只有精品99| 国产精品一区二区在线观看99| 亚洲国产日韩一区二区| 午夜视频国产福利| 一区二区日韩欧美中文字幕 | 久久97久久精品| 亚洲av男天堂| 欧美日韩视频精品一区| 午夜av观看不卡| 成人黄色视频免费在线看| 69精品国产乱码久久久| 日本午夜av视频| 亚洲av成人精品一二三区| 日韩制服骚丝袜av| 国产男女内射视频| 又粗又硬又长又爽又黄的视频| 国产伦精品一区二区三区视频9| 亚洲天堂av无毛| 亚洲一区二区三区欧美精品| 成人毛片60女人毛片免费| 你懂的网址亚洲精品在线观看| 久久久久久久久久人人人人人人| 久久久国产精品麻豆| 国产精品一区二区三区四区免费观看| 99国产精品免费福利视频| 熟女av电影| 久久毛片免费看一区二区三区| 国产精品人妻久久久久久| 日韩电影二区| 国产av一区二区精品久久| 日韩,欧美,国产一区二区三区| 日本-黄色视频高清免费观看| 黑丝袜美女国产一区| 看十八女毛片水多多多| 日本免费在线观看一区| 国产 一区精品| 中文精品一卡2卡3卡4更新| 天美传媒精品一区二区| 嫩草影院入口| 美女国产高潮福利片在线看| 在线免费观看不下载黄p国产| 久久精品国产亚洲av天美| av专区在线播放| 亚洲国产日韩一区二区| 久久精品国产亚洲av涩爱| 80岁老熟妇乱子伦牲交| 男人添女人高潮全过程视频| 日韩成人伦理影院| 你懂的网址亚洲精品在线观看| 日本av免费视频播放| 99热网站在线观看| 国产精品人妻久久久影院| 久久影院123| 亚洲婷婷狠狠爱综合网| 亚洲国产色片| 久久久久久人妻| 国产一区有黄有色的免费视频| 亚洲国产日韩一区二区| 亚洲精品自拍成人| 免费不卡的大黄色大毛片视频在线观看| av一本久久久久| 黑人欧美特级aaaaaa片| 久久综合国产亚洲精品| 最近手机中文字幕大全| 国产69精品久久久久777片| 极品少妇高潮喷水抽搐| 麻豆成人av视频| 不卡视频在线观看欧美| 日本黄色日本黄色录像| 国产精品.久久久| 两个人的视频大全免费| 日韩av不卡免费在线播放| 69精品国产乱码久久久| 一本—道久久a久久精品蜜桃钙片| 亚洲欧洲精品一区二区精品久久久 | 我要看黄色一级片免费的| 最后的刺客免费高清国语| 欧美日韩成人在线一区二区| 观看av在线不卡| 一级毛片 在线播放| 有码 亚洲区| 色视频在线一区二区三区| 久久国产亚洲av麻豆专区| 美女cb高潮喷水在线观看| 国产精品一区www在线观看| 老女人水多毛片| 久久久久久久久久成人| a级毛片在线看网站| 日本av免费视频播放| 九九久久精品国产亚洲av麻豆| 国产成人精品婷婷| 黑人欧美特级aaaaaa片| 精品久久久久久久久av| 亚洲精品乱久久久久久| 国产男女内射视频| 在线观看三级黄色| 亚洲av成人精品一二三区| 91国产中文字幕| 又大又黄又爽视频免费| 免费久久久久久久精品成人欧美视频 | 国产精品嫩草影院av在线观看| 久热久热在线精品观看| 亚洲欧美成人精品一区二区| 国产色婷婷99| 亚洲人成网站在线观看播放| 伦理电影大哥的女人| 免费观看av网站的网址| 亚洲国产精品专区欧美| 日本爱情动作片www.在线观看| 午夜激情av网站| 特大巨黑吊av在线直播| 黑人巨大精品欧美一区二区蜜桃 | 在线观看一区二区三区激情| 女性生殖器流出的白浆| 哪个播放器可以免费观看大片| 日韩不卡一区二区三区视频在线| 亚洲美女黄色视频免费看| 少妇丰满av| 国产成人精品婷婷| 国产免费视频播放在线视频| 亚洲成人一二三区av| 两个人免费观看高清视频| 国产av精品麻豆| 日本与韩国留学比较| 777米奇影视久久| 久久午夜福利片| 下体分泌物呈黄色| 亚洲精品国产av蜜桃| 久久久久久久久久成人| 亚洲精品亚洲一区二区| 国产精品人妻久久久影院| 日韩成人av中文字幕在线观看| 青春草亚洲视频在线观看| 久久国产亚洲av麻豆专区| 国产黄色视频一区二区在线观看| 国产色爽女视频免费观看| 黄色欧美视频在线观看| 亚洲欧美成人精品一区二区| 国产熟女欧美一区二区| 国产 一区精品| 日韩av免费高清视频| 丝袜美足系列| 99热全是精品| 蜜桃久久精品国产亚洲av| 国产成人freesex在线| 我的老师免费观看完整版| 在线观看免费日韩欧美大片 | 亚洲欧美中文字幕日韩二区| 中文字幕亚洲精品专区| 精品视频人人做人人爽| 国产在线免费精品| 日韩人妻高清精品专区| 草草在线视频免费看| 久久久久人妻精品一区果冻| 一区二区三区乱码不卡18| 欧美3d第一页| 天天影视国产精品| 欧美人与善性xxx| 大话2 男鬼变身卡| 人人妻人人澡人人看| 久久99热这里只频精品6学生| 另类精品久久| 国产色爽女视频免费观看| 成年女人在线观看亚洲视频| 欧美激情极品国产一区二区三区 | 在线天堂最新版资源| 精品久久久久久电影网| 91久久精品国产一区二区成人| 国产黄色视频一区二区在线观看| 91成人精品电影| 成年美女黄网站色视频大全免费 | 18禁观看日本| 看免费成人av毛片| 欧美亚洲 丝袜 人妻 在线| 看非洲黑人一级黄片| 国产男人的电影天堂91| 国产成人免费无遮挡视频| 啦啦啦啦在线视频资源| 亚洲精品一二三| 久久久午夜欧美精品| 亚洲欧洲精品一区二区精品久久久 | 欧美97在线视频| 国产av码专区亚洲av| 只有这里有精品99| 色视频在线一区二区三区| 国产欧美日韩一区二区三区在线 | 一级二级三级毛片免费看| 高清欧美精品videossex| 久久精品久久久久久久性| 两个人免费观看高清视频| 22中文网久久字幕| 久久影院123| 在线 av 中文字幕| 国产淫语在线视频| 日本黄大片高清| 欧美国产精品一级二级三级| 日日摸夜夜添夜夜爱| 亚洲精品日本国产第一区| 国产色爽女视频免费观看| .国产精品久久| 搡女人真爽免费视频火全软件| 日本av手机在线免费观看| 韩国高清视频一区二区三区| 精品人妻在线不人妻| 久久99一区二区三区| 十分钟在线观看高清视频www| 男女国产视频网站| 人妻系列 视频| 欧美老熟妇乱子伦牲交| 欧美三级亚洲精品| 国产精品秋霞免费鲁丝片| 亚洲色图综合在线观看| 国产精品一二三区在线看| 日韩精品免费视频一区二区三区 | 成年人午夜在线观看视频| 又黄又爽又刺激的免费视频.| 国产精品一区二区在线观看99| 3wmmmm亚洲av在线观看| 青青草视频在线视频观看| 麻豆乱淫一区二区| 两个人免费观看高清视频| 亚洲av欧美aⅴ国产| 欧美精品国产亚洲| 2021少妇久久久久久久久久久| 新久久久久国产一级毛片| 国产白丝娇喘喷水9色精品| 五月伊人婷婷丁香| 亚洲欧美日韩卡通动漫| 欧美日韩亚洲高清精品| 成人免费观看视频高清| 美女国产视频在线观看| av不卡在线播放| 亚洲人成网站在线观看播放| 在线天堂最新版资源| 亚洲伊人久久精品综合| 热re99久久国产66热| 午夜精品国产一区二区电影| 欧美日韩综合久久久久久| 免费久久久久久久精品成人欧美视频 | 熟女电影av网| 女人精品久久久久毛片| 日韩欧美精品免费久久| 成年女人在线观看亚洲视频| 中文欧美无线码| 精品卡一卡二卡四卡免费| 欧美97在线视频| 满18在线观看网站| 国产男女内射视频| 热99国产精品久久久久久7| 精品亚洲成国产av| 精品国产一区二区三区久久久樱花| 五月玫瑰六月丁香| 韩国av在线不卡| 国产精品一区二区三区四区免费观看| 中国三级夫妇交换| 成年人午夜在线观看视频| 日韩一区二区视频免费看| 欧美精品人与动牲交sv欧美| 丰满迷人的少妇在线观看| 国产精品一二三区在线看| 97超碰精品成人国产| 国产精品一国产av| 嫩草影院入口| 一级毛片aaaaaa免费看小| 亚洲熟女精品中文字幕| 卡戴珊不雅视频在线播放| 国产一级毛片在线| 亚洲精品一二三| 欧美日韩在线观看h| 在线观看免费高清a一片| 午夜福利视频在线观看免费| 边亲边吃奶的免费视频| 久久久国产一区二区| 午夜激情福利司机影院| 亚洲无线观看免费| 国产一区二区在线观看日韩| 99久久综合免费| 日韩视频在线欧美| 国产极品天堂在线| 国产黄片视频在线免费观看| 精品99又大又爽又粗少妇毛片| 99视频精品全部免费 在线| 亚洲av福利一区| 久久 成人 亚洲| 国产乱人偷精品视频| 亚洲激情五月婷婷啪啪| 99九九线精品视频在线观看视频| 亚洲精品第二区| 一个人看视频在线观看www免费| 免费播放大片免费观看视频在线观看| 热re99久久国产66热| 国产在线一区二区三区精| 国产有黄有色有爽视频| 九色成人免费人妻av| 三级国产精品片| 黄色一级大片看看| 中国三级夫妇交换| 狂野欧美白嫩少妇大欣赏| 国产探花极品一区二区| 极品人妻少妇av视频| 在现免费观看毛片| 亚洲欧美成人精品一区二区| 美女主播在线视频| 国产在视频线精品| 精品一品国产午夜福利视频| 日本与韩国留学比较| 亚洲高清免费不卡视频| 国产色婷婷99| 日本与韩国留学比较| av卡一久久| 亚洲av成人精品一区久久| 久久久午夜欧美精品| 久久国产亚洲av麻豆专区| 一本—道久久a久久精品蜜桃钙片| av电影中文网址| 少妇被粗大的猛进出69影院 | 午夜免费观看性视频| 国产黄频视频在线观看| 亚洲精品视频女| 人妻系列 视频| 高清午夜精品一区二区三区| 一级a做视频免费观看| 久久久久久人妻| 久久久久久久久久人人人人人人| 国产精品免费大片| 老熟女久久久| 久久99蜜桃精品久久| av不卡在线播放| 日韩,欧美,国产一区二区三区| 免费不卡的大黄色大毛片视频在线观看| 人妻少妇偷人精品九色| 看免费成人av毛片| 少妇的逼好多水| 免费黄网站久久成人精品| 婷婷色麻豆天堂久久| 一区二区三区乱码不卡18| 欧美+日韩+精品| 国产欧美日韩综合在线一区二区| 亚洲,一卡二卡三卡| 国产日韩欧美在线精品| 制服诱惑二区| 日韩亚洲欧美综合| 国产欧美亚洲国产| 男女边吃奶边做爰视频| 亚洲精品乱码久久久v下载方式| 国产精品 国内视频| 亚洲高清免费不卡视频| 久久狼人影院| 久久精品久久久久久噜噜老黄| a级毛片黄视频| 成人亚洲精品一区在线观看| 国产日韩一区二区三区精品不卡 | 亚洲三级黄色毛片| 日本av免费视频播放| 久久国内精品自在自线图片| 一本一本久久a久久精品综合妖精 国产伦在线观看视频一区 | 欧美国产精品一级二级三级| 中文字幕免费在线视频6| 欧美精品人与动牲交sv欧美| 国精品久久久久久国模美| 久久国产精品男人的天堂亚洲 | 成人毛片60女人毛片免费| 久久精品久久精品一区二区三区| 久久久久久人妻| 亚洲av国产av综合av卡| av国产久精品久网站免费入址| 久久久精品免费免费高清| 久久久久精品久久久久真实原创| 丰满少妇做爰视频| 伦精品一区二区三区| 亚洲精品乱久久久久久| 国产精品久久久久久久久免| 人妻夜夜爽99麻豆av| 久久韩国三级中文字幕| 免费看不卡的av| 亚州av有码| 亚洲人成网站在线播| 久久久亚洲精品成人影院| videossex国产| 日韩成人伦理影院| 国产精品无大码| 亚洲综合精品二区| 人人妻人人澡人人爽人人夜夜| 久久久精品94久久精品| 午夜影院在线不卡| 3wmmmm亚洲av在线观看| 亚洲,一卡二卡三卡| 两个人免费观看高清视频| 国产极品粉嫩免费观看在线 | 高清在线视频一区二区三区| av在线观看视频网站免费| 亚洲欧美中文字幕日韩二区| 一级毛片黄色毛片免费观看视频| 日日爽夜夜爽网站| 少妇的逼水好多| 成人国产麻豆网| 久久久精品区二区三区| 22中文网久久字幕| 最近手机中文字幕大全| 岛国毛片在线播放| 免费大片18禁| 国产成人精品一,二区| 亚洲精品日韩av片在线观看| 熟女人妻精品中文字幕| 日本黄色日本黄色录像| 有码 亚洲区| 久久久精品区二区三区| 亚洲精品视频女| 狂野欧美激情性bbbbbb| 在线天堂最新版资源| 亚洲国产最新在线播放| 亚洲中文av在线| 91在线精品国自产拍蜜月| 日韩欧美一区视频在线观看| 极品人妻少妇av视频| 妹子高潮喷水视频| 国产精品成人在线| 51国产日韩欧美| 黑丝袜美女国产一区| 亚洲精品国产色婷婷电影| 男女无遮挡免费网站观看| xxx大片免费视频| 久久久久久久亚洲中文字幕| 最近最新中文字幕免费大全7| 97在线人人人人妻| 一级毛片 在线播放| av在线老鸭窝| 国产欧美日韩综合在线一区二区| 亚洲av国产av综合av卡| 啦啦啦中文免费视频观看日本| 欧美97在线视频| 国产一区亚洲一区在线观看| 久久久久久久亚洲中文字幕| 日韩中文字幕视频在线看片| 涩涩av久久男人的天堂| 国产成人免费观看mmmm| 91成人精品电影| 国产午夜精品久久久久久一区二区三区| 老司机亚洲免费影院| 波野结衣二区三区在线| 国产成人免费无遮挡视频| 欧美激情极品国产一区二区三区 | 男女国产视频网站| 国产精品不卡视频一区二区| 中文字幕最新亚洲高清| 久久这里有精品视频免费| 男女边吃奶边做爰视频| 午夜免费观看性视频| 九色成人免费人妻av| h视频一区二区三区| 97在线视频观看| 免费看不卡的av| 制服人妻中文乱码| 在线播放无遮挡| 欧美性感艳星| 精品人妻熟女毛片av久久网站| 狂野欧美激情性xxxx在线观看| 考比视频在线观看| 国产精品一区二区在线不卡| 久久亚洲国产成人精品v| 亚洲欧洲精品一区二区精品久久久 | 美女主播在线视频| 99精国产麻豆久久婷婷| 在线观看三级黄色| 精品一区二区三卡| 国产国拍精品亚洲av在线观看| 黄色欧美视频在线观看| 99热网站在线观看| 自线自在国产av| 草草在线视频免费看| 亚洲不卡免费看| 91久久精品国产一区二区成人| 一本一本久久a久久精品综合妖精 国产伦在线观看视频一区 | 一边亲一边摸免费视频| 男女无遮挡免费网站观看| 亚洲内射少妇av| 久久久欧美国产精品| 色视频在线一区二区三区| 国产精品.久久久| 一区二区三区乱码不卡18| 高清午夜精品一区二区三区| 日韩免费高清中文字幕av| 亚洲欧美一区二区三区黑人 | 波野结衣二区三区在线|