    Design of a personnel safety interlock system for proton therapy

    2021-05-06 03:21:12ShouYuanWangYunTaoSongHanShengFengShiLiHaiLinCaoJingZhangOuWenHuangZhuLi
    Nuclear Science and Techniques 2021年4期

    Shou-Yuan Wang· Yun-Tao Song · Han-Sheng Feng · Shi Li ·Hai-Lin Cao · Jing Zhang · Ou-Wen Huang · Zhu Li

    Abstract Proton therapy is the most advanced radiotherapy approach in the world, and causes less damage to normal human tissue than traditional radiotherapy.Because the treatment process produces a high-energy proton beam,the personnel safety interlock system mainly considers measures to protect personnel from radiation hazards during beam preparation and the beam release process.Unlike other safety interlock systems, the personnel safety interlock system designed in this study focuses on the safety and stability of the system itself. The hardware and software of important interlock control loops are designed and developed according to the requirements of Safety Integrity Level 3 specified by IEC61508. A set of redundant ring networks was developed to ensure that damage to a certain network line does not affect the normal operation of the system. A set of friendly operation interfaces and data storage systems were developed to ensure that the operator can monitor the data in real time and trace the data. The personnel safety interlock system mainly includes a beam enabling function, clearance function, and emergency stop function. The system was put into actual use and successfully ensured personnel safety.

    Keywords Personnel safety · Radiation protection ·Interlock system · Safety integrity levels

    1 Introduction

    Compared with other radiotherapy methods, proton therapy has the advantages of a better dose distribution and higher local dose. Clinical treatment data show that the 5-year survival rate is as high as 70% [1]. The proton therapy system includes an accelerator system, an energy selection system,a beamline system,and a treatment room.A high-energy proton beam is generated by the accelerator,adjusted to a pre-set energy level by the energy selection system, and then transmitted to the treatment room by the beam line system to treat the patient tumor.

    A superconducting cyclotron was developed and built by the Hefei proton therapy project with an energy level of 200 MeV and a beam current of 400 nA.During the system operation, along with high radiation fields, measures need to be taken to ensure the safety of on-site workers. Based on radiation safety and the reliability requirements of the interlock system,in this paper,the design and development of a personnel safety interlock system(PSIS)depending on the safety requirements of IEC 61508 is introduced. The safety integrity level (SIL), a key concept in functional safety standards, is a performance metric used to measure the reliability of safety systems that conduct specific safety functions [2]. The SIL is widely used in high-risk industries, such as the nuclear, petroleum, chemical, and steel industries.In the system design stage, the SIL design must be carried out at the hardware and software levels to ensure that the entire control loop meets the safety requirements.The system safety that satisfies the SIL is mainly embodied in two aspects. On the one hand, it has a lower probability of failure, and on the other, it has a series of failure protection measures in the event of failure. SIL1, SIL2, and SIL3 are commonly used SIL levels. The higher the level is, the higher the safety factor [3]. In this project, because proton therapy is a high-risk industry, its personnel safety interlock system needs to meet the application requirements of SIL3. Some of the terms used to describe the safety level of the control loop are listed below, and Table 1 shows the criteria correspondence between PFDSIF and SIL:

    PFDS: average failure probability on sensor demand.

    PFDL: average failure probability on logic controller demand.

    PFDA: average failure probability on actuator demand.

    PFDSIF: average failure probability on safety interlock function demand.

    PFDSIF = PFDS + PFDL + PFDA.

    In PSIS, the main functions are the beam enable function, clearance function, and emergency stop function [4].The key control loop of the interlock function, such as the beam enable function and emergency stop function, meet the requirements of SIL3. In addition, a friendly graphical user interface (GUI) was developed for operators to remotely monitor the safety system.

    There are two treatment rooms in the project, and the workplace depending on the radiation intensity is divided into a control area and a supervision area to meet the requirements of the China National Standard GB18871-2002 (basic international standards for protection against ionizing radiation and for the safety of the radiation sources) [5]. The control area is shown in the dotted area in Fig. 1, and focus needs to be on radiation safety management.

    2 System requirements

    According to the requirements ment ioned in the introduction to ensure the radiation safety of personnel in the control area, the safety interlock system of proton therapy personnel mainly needs to consider the safety requirementsof the system from two aspects:readying the beam release and the actual beam release itself. During the preparation of the beam release,it is necessary to consider the possible radiation risks of workers and take necessary measures to avoid these risks, and in the process of beam release,measures should be taken to shut off the beam in time to prevent or reduce the harm caused by radiation.

    Table 1 Correspondence between PFDSIF and SIL

    Fig. 1 (Color online) Control and supervision areas

    Therefore, it is necessary to arrange a large number of interlock devices to collect field data and use the controller to process the data. To realize the safety function, it is necessary to select the hardware according to the requirements of the safety interlock system first,and then arrange the devices in various areas of the site.Next,it is necessary to carry out safety programming in the controller and collect real-time data of the devices through the controller.In addition, a GUI was developed to remotely monitor the key parameters of the interlock system [6].

    2.1 Beam enable requirements

    The interlock system requires a series of measures to confirm the beam-enabled conditions before the beam is released, and attempts to avoid the risk of radiation accidents. Therefore, the requirements for enabling a beam release are as follows:

    1. A series of measures are needed to confirm that no personnel are within the radiation area before the beam is released to prevent injury in such personnel from the radiation [7, 8].

    2. It is necessary to confirm the states of all emergency stop interlocks on site, and that all emergency stop interlocks have been reset.

    3. Radiation monitoring equipment must be ready to confirm the radiation status of the site.

    4. To avoid an accidental operation of the beam, a key switch should be set in the control room to confirm the beam-enabled condition.

    5. It is necessary to confirm the states of the interlock system itself, and the interlock system should be opened and operated normally.

    2.2 Interlock requirements of emergency stop

    During the process of beam release, it is necessary to consider the risk of workers staying in the radiation area or entering the radiation area by mistake, and a series of measures should be taken by the safety interlock system to stop the beam. The requirements of the emergency stop interlock are as follows:

    1. If an emergency occurs during the beam-out process,the personnel in the control or supervision areas can press the emergency stop button to stop the beam.

    2. During the beam release process,a beam stop interlock is required to detect whether the clearance door is open.

    3. During the beam release process, if the interlock system fails, a beam-stop interlock is required.

    2.3 Device requirements

    Hardware is needed to realize the software function of the beam enabling and emergency stop interlock. A controller is needed to realize the interlock logic, and a field sensor is required to collect the field signals.

    1. A controller is needed to implement the safety interlock program, and a series of input and output modules are required to monitor and control the safety equipment [9].

    2. The safety system needs to know whether the area dose has been exceeded, and it is therefore possible to arrange the radiation monitoring equipment in the control and supervision areas to monitor the radiation dose.

    3. To prevent people from staying in the site when the accelerator is running,clearance buttons and clearance doors must be arranged in the control area. The interlock system needs to monitor the clearance button and the position switch signal of the clearance door for the personnel in the control area to clear the site before leaving the field. Sound and light alarms are required during the process to remind us that the site is currently being cleared.

    4. Emergency stop buttons need to be arranged in the control and supervision areas for interlock protection under emergency conditions, and the personnel safety interlock system needs to monitor whether the emergency stop button has been pressed.

    5. To prevent the accelerator from being turned on accidentally, a key switch for the accelerator beam needs to be set in the supervision area, and the personnel safety interlock system needs to collect the signal of the key switch for the interlock logic.

    2.4 System safety requirement

    The system includes two parts: hardware and software.To meet the safety requirements of the safety interlock system and enable the system to operate safely and stably[10],it is necessary to select safety devices and ensure safe communication. It is also necessary to select appropriate programming software for the safe programming of the key program blocks.

    1. In the hardware selection, it is necessary to fully consider the system safety.For the key safety interlock function, it is necessary to select the controller and sensor meeting the SIL3 level according to the requirements of the safety standard. For a general logic function, a normal device is simply needed.

    2. In software programming, it is necessary to apply a program according to different security levels. The security program needs to use the program block of the security program library,and the general program uses the program block of the general program library.

    2.5 Interface requirement

    Users need to obtain the safety interlock parameters conveniently and quickly, and judge whether the current beam output conditions have been met or whether there is an interlock trigger. Therefore, a friendly interlock interface is required.The system states can be monitored at the interface,and the state history can be queried.The specific requirements are as follows:

    1. It is necessary to remotely monitor the main safety signals, such as the radiation value of the area, area clearance states, emergency stop states, and beam enable states.

    2. It is necessary to trace the historical information.In the case of an emergency stop,it is necessary to determine the cause of the fault for maintenance.

    3 Design principles

    Considering the safety and reliability,the overall design of PSIS adopts a control loop that meets the SIL in the design of an important interlock control loop [11], as shown in Fig. 2.

    The interlock control loop represents a complete control link, including the sensor used to monitor the field status,the signal acquisition module,the logic processing module,the signal beam output module,and the actuator to execute the interlock command. In the SIL3 control loop, the device selected, hardware wiring, and communication methods all need to meet SIL3.

    Programs with low safety levels are called general programs, and programs with high safety are called safety programs. Their programming is relatively independent,each of which having its own set of program libraries that meet the safety requirements.

    General procedures typically apply general sensors and actuators, and safety equipment uses safety sensors and actuators to meet the SIL requirements. The general program and the safety program operate together in the safety controller, although the controller distinguishes two programs and runs them in different processing layers [12].

    The difference between safety hardware and general hardware lies mainly in two aspects:failure probability and failure protection. Manufacturers will test the failure probability of the safety hardware in a specific environment. The lower the failure probability, the higher the reliability and the higher the SIL level. When the safety sensor fails, it can send a fail-safe signal to the controller,and the controller can generate an interlock to protect the safety of the personnel. When the safety controller fails, it can send a fail-safe signal to the actuator, and the actuator can generate interlocking actions to protect the personnel.When the safety actuator fails, it interlocks to protect the safety of the personnel. General hardware only needs to perform routine inspections and does not need to test the failure probability. It is generally only used for process control and does not trigger an interlock action when it fails.

    The safety software was programmed in the safety logic controller to match the safety sensors and safety actuators.Hardware devices that meet SIL3 are generally redundant.The safety program can determine whether the interlock protection program will be triggered according to whether the redundant circuit is abnormal. However, safety programs are fail-safe. When there is an abnormality in the controller device or communication, or when a power failure occurs, a fail-safe signal is sent to the actuator to generate interlocking actions to protect the safety of the personnel.

    General software is only used for process control,with a lower security level, such as field clearance control and alarm light control, and when such software fails, there is no effect on personnel safety, and thus no corresponding failure protection measure is required.

    Consider an emergency stop when the area is cleared as an example. After the area is cleared, the door position switch monitors whether the interlock door is opened.If the interlock door is opened,it sends an interlock signal to the controller, and the controller executes the beam-off command. If the cable of the door position switch is disconnected, the controller will detect the abnormality of the door position switch and execute the beam-off command.When the controller fails,the fail-safe is activated,and the relay receives the fail-safe signal and executes the beamoff command.When the relay fails,it will also activate the fail-safe system to shut off the beam.

    4 Architecture design

    To realize a personnel safety interlock system, the system architecture was first designed. This system is divided into four layers according to their functional characteristics: the display, data service, function, and device layers[13],as shown in Fig. 3.In the architecture,the safety and general parts are distinguished. The yellow background in Fig. 3 is the safety part, which is designed based on SIL3.

    The display layer is the GUI of the PSIS, which runs independently in the control room, and is only used todisplay the state.When the GUI fails,the program will still be executed in the underlying controller without affecting the safety program. The GUI includes the clearance information, radiation monitor information, emergency stop information,and beam ready/release information.The operator can easily obtain the system operation information from the GUI and proceed to the next step.

    Fig. 2 Control loop

    The main function of the data service layer is to store historical data, including manager information, historical data information, operation records, alarms, and interlock information. The data service layer of the safety interlock system archives and records all information generated during operation, such as radiation monitoring system operation records, emergency stop button information, site clearing information, site clearing door switch states, and enabled key switch states. When needing to retrieve data for analysis,operators can extract storage information from the database.

    The logic of the personnel safety interlock system is realized in the functional layer, and the PSIS consists of four main functional modules:the communication module,input signal processing module, main logic processing module, and output processing module [14]. In each function module,the PSIS distinguishes between the safety functions and general functions. The safety program uses the safety function block, which can self-check the input signal, internal communication, and output signal of the program. When the self-inspection is abnormal, the interlock command is executed. The safety functions include PROFIsafe communication, safety signal processing, an emergency stop function, and a beam enable function. A safety library that satisfies SIL3 with a series self-check functions is included in the safety controller programming platform.The safety program was developed using programs of the safety software library. The general function is the general action flow because the auxiliary program of the safety system does not directly participate in the interlocking process, and sound and light alarms or necessary logic signals for the safety program are provided on-site. General functions include general communication,general signal processing, a clearance function, and an alarm lamp function.

    The device layer also distinguishes general and safety devices according to the actual application requirements.SIL3 devices were selected as the key interlock control loop hardware. As the main difference between the SIL3 device and the general device, the failure probability is lower; however, when the equipment fails, it can send safety signals to trigger an interlock. The safety devices include an interlock controller,emergency stop button,and door position switch to conduct an interlock function with a high level of safety. General equipment includes a clearance button, alarm lamp, low security level logic,or alarm function.

    The average failure probability on the demand of the safety control loops is shown in Table 2, and thecalculation indicates that the designed safety control loops meet the requirements of SIL3.

    Table 2 Safety interlock function PFD

    5 Hardware design

    As shown in Fig. 4, the field interlock devices are relatively scattered, although the interlock functions are relatively centralized. One main controller and three remote modules are selected to realize the interlock function in the personnel safety interlock system.The remote module was used to monitor and control the personnel safety interlock device on-site. The personnel safety interlock equipment can be divided into SIL3 safety devices and general devices according to the safety functions conducted.Safety devices include emergency stop buttons, door position switches,key switches, relays, radiation monitors, and accelerator beam-off devices. Common equipment includes a clearance button and an alarm lamp. The remote modules connect with the device through a hard wire to ensure the signal transmission rate and reliability. For safety devices,the remote module adopts the SIL3 level redundant wiringmode, which can identify a series of line faults, such as a short circuit and open circuit,and prevent a safety function failure [15]. For the general module, the remote module adopts single-channel wiring mode to meet the functional requirements. The main controller collects the signals uploaded by all remote modules as the judgment condition of the interlock logic and executes the safety interlock program.The main controller identifies the safety program and the general program, and then runs them separately. If the general program is abnormal, the operation of the safety program will not be affected. The communication protocol between the main controller and the remote module uses a PROFIsafe protocol meeting the SIL3 level.In addition, PSIS designs and develops a set of ring networks for data communication. The PROFIsafe protocol and ring network meet the failure requirements. The controller and remote module of the PSIS include two network ports,which are connected in series individually,forming a closed-loop network. When a line in the ring network is damaged, it will not affect the network communication of the system, thereby significantly improving the reliability of the network communication. In addition, when the ring network fails, it will be detected by the safety controller,and the safety system will trigger the failsafe interlock to shut off the beam.

    The PSIS is connected to the ion source (IS), radio frequency (RF), and vertical deflection (VD). These devices are accelerator subsystems for producing a beam.When the beam-off logic is active,the HSIS interlocks the IS,RF,and VD to shut off the beam [16, 17].

    The interface of the PSIS runs on a workstation used to monitor the states of the site. It has a low security level,and when the GUI fails,the personnel safety interlock logic continues running in the controller.The GUI uses the TCP/IP protocol, which is the general protocol for communication [18].

    The data of the personnel safety interlock is stored on the server,and the operator can reference the historical data stored on the server for research through the system software. As a part of the auxiliary function of interlock system, the data storage function has a low safety level requirement, and uses a general network protocol for communication.

    6 Logic design

    The control logic of the PSIS was carried out in the main controller. The personnel safety interlock receives the signals of the clearance button, door position switch,emergency stop button, radiation monitoring, key switch,and other field devices arranged in the control area. The layout of the local interlock system device is shown in Fig. 5. The signals are logic inputs, and the controller applies the logic through the program. The logic output devices are IS, RF, and VD, which are used for beam-off logic. The logic function can be divided into the clearance function, enable function, emergency stop function, and a series of signal processing functions.

    The clearance function is realized by the clearance button and clearance door. As shown in Fig. 5, the clearance function of the proton therapy system is divided into five routes, and each route is an independent area. The main purpose of the clearance is to ensure that no people are in the control area before the accelerator is released.Therefore,it is necessary to establish a clearance route that can cover the control area.According to the characteristics of this project,a clearance route is set,which can guide the worker in clearing the control area.As shown in Fig. 5,the accelerator and transportation line areas consist of routes 1,2, and 3. When routes 1, 2, and 3 are all cleared, the interlock system judges that the accelerator and transportation line areas are cleared. The clearing of the treatment room 1 area is route 4,and when the clearance of route 4 is completed, treatment room 1 is cleared. Route 5 is used for the clearance of treatment room 2. When the route 5 area is cleared,the treatment room 2 area is cleared.

    A series of area-clearance logic approaches were designed. Each area is relatively independent, and the destruction of the clearance area does not destroy the clearance of the other areas. When clearing the radiation areas,the staff only need to clear the uncleared area,which significantly improves the efficiency of the site clearance.

    Inside the control and supervision areas,a large number of emergency stop buttons are arranged for beam-off operations during emergency situations to protect the safety of the personnel. The personnel safety system monitors the states of the emergency stop buttons in real time. When an emergency stop button is pressed, the interlock action is activated.

    Radiation monitoring equipment is used to monitor the on-site radiation value to prevent radiation from exceeding the standard. The PSIS needs to determine whether the radiation monitoring equipment operates normally before the beam is released. If it is abnormal, the beam is not allowed. During the beam output, it is also necessary to determine whether the radiation on site exceeds the threshold value.If the threshold value is exceeded,it means that the personnel may be at risk of radiation exposure,and the beam flow needs to be interrupted.

    Fig. 5 (Color online) PSIS local device layout

    6.1 Clearance function

    The staff must clear the site according to the path indicated by the arrow.First,the operator needs to go to the starting point of the path to photograph the clearance button, and the clearance door needs to be closed within the time specified by the system.The route-clearance logic for the program is shown in Fig. 6. The interlock system monitors in real time whether the clearance button and clearance door are pressed or closed within the specified time.When the logic is correct,the interlock system judges that the route clearance has been completed. When all routes are cleared, the PSIS determines that all areas are cleared.

    6.2 Beam enable function

    To prevent human radiation hazards caused by the accelerator beam under unexpected situations, the safety interlock system needs to determine a series of prerequisites before releasing the beam,including the following:all control areas have been cleared,the emergency stop button has not been pressed in any areas, and the radiation monitoring system works normally without an alarm sounding.

    However, on-site staff members require management methods to confirm whether the external environment is normal, such as some necessary signing procedures, confirming whether the personnel in the control room are ready, and whether the server and GUI are working normally.When all external conditions are met,the staff needs to turn on the key switch to allow the beam to be released.

    The safety interlock system activates the interlocks of IS, RF, and VD when the beam release conditions are not met. At this time, the IS, RF, and VD will be interlocked,and the proton therapy system cannot release the beam.When the safety interlock system detects that all beam conditions are met, it releases the IS, RF, and VD interlocks,and the proton therapy system can release the beam.

    Fig. 6 Clearance logic

    6.3 Emergency stop function

    The emergency stop function means that, during the beam release process,the personnel safety system executes the beam-off emergency stop after detecting the beam-off condition. During the beam release process, the PSIS monitors the states of the emergency stop button,the states of the radiation monitor equipment, and whether the clearance door is open in real time. If the emergency stop button is pressed, it means that the staff has found an abnormality at the scene, and the emergency stop will be manually interrupted. To ensure the safety of the on-site staff,the proton therapy system arranges a large number of emergency stop buttons in the control area and the supervision area, and the staff can press any emergency stop button for beam stop logic. Radiation monitoring equipment is used to monitor the radiation situation in a specific location.If the radiation exceeds the limit,the equipment is deemed abnormal, or the radiation shielding fails [19], an emergency beam-off stop button is required. During the beam release process, if the clearance door is opened, it means that someone might have entered the control area by mistake. To ensure the safety of the personnel, the personnel safety interlock system needs to shut off the beam[20].

    The execution devices of the beam-stop interlock are the IS,RF,and VD.When the beam stop interlock is triggered,the interlock system sends the beam stop signal to these three devices simultaneously through a hard wire and redundantly shuts the beam flow down.

    7 Software development

    The system block is used for software development.The system block mainly includes the function block and data block.The system blocks are divided into general function blocks and safety function blocks,and safety programs and general programs are stored. The enable function and emergency stop function belong to the safety program and are programmed in the safety function block.The clearance function and the alarm light function belong to the general function and are programmed in the general function block.Data blocks are also divided into safety data blocks and general data blocks, which store the safety variables and variables, respectively, and the safety program calls the data in the safety data block for a logical execution. In the case of a system failure, the safety program automatically activates the fail-safe interlock function to interrupt the beam. As shown in Fig. 7, the yellow background color belongs to the safety system block, and the other colors belong to the general function block.

    8 GUI development

    In proton therapy systems, the interface of the PSIS plays an important role.The staff cannot directly obtain the information they want from the controller, and thus a human–computer interaction interface is necessary. The GUI was developed according to the requirements of the PSIS (see Fig. 8).

    In the GUI, the operator can directly obtain detailed information on the area clearance. The situation of each clearance route on the GUI interface is clearly displayed.The clearance button is green when pressed and orange when it is not pressed.The clearance door is green when it is closed and orange when it is not closed. The clearance route indicates the clearance state of the current area:clearance is complete when it is green, and incomplete when it is orange.Before the beam is enabled,the operator can determine the areas that need to be cleared on the interface and move to the corresponding area to conduct the clearing process.

    The operator can obtain emergency stop information for each area on the interface. Many emergency stop buttons are arranged in each area.When an emergency stop button is not pressed,it is displayed in green.When an emergency stop button is pressed, the display of the emergency stop button in this area is changed to red. An emergency stop button area is mainly divided into the accelerator area,beam line area, treatment room 1, treatment room 2,treatment control room 1, and treatment control room 2.The operator can determine the reason for the pressing of the emergency stop button according to the area where the emergency stop button is pressed,and reset the emergency stop button in this area after confirming whether the system is functioning normally.

    The system state represents the key signals of the safety interlock system.These key signals are summarized signals processed by the controller. When the beam enable indicator is green,it means that the external conditions of beam release have been met and a beam release is allowed, and when it is red,it means that the external conditions of beam release have not been met. When the emergency stop indicator is red,it means that the system has an emergency stop triggered; when it is green, it means there is no emergency stop. If the radiation alarm indicator is red, it means that the radiation is abnormal,and when it is green,it means that the radiation is normal.

    The operator can obtain the beam information on the GUI,check which conditions have not been met, and view the beam logic in the GUI. When all areas have been cleared,no emergency stop button has been pressed in any area, and the radiation monitoring equipment is not in an abnormal state, the beam key switch is turned on, and all red lines of the beam logic in the GUI will change to green lines. When the beam enable signal is green, it means that the beam-release conditions have been met and the beam release is allowed.

    The operator clicks the radiation monitor on the operation interface to view the specific data of the radiation monitor in each area, and clicks the History Record button to view the historical data of the interlock system, such as the time when the emergency stop button was pressed, the time of the clearance completion or failure, and the time record of the beginning and end of the beam.

    Fig. 7 (Color online) Software development

    Fig. 8 (Color online) Personnel safety interlock system GUI

    9 Conclusion

    To protect personnel from the radiation field during the operation of the proton therapy system, a personnel safety interlock system was designed and developed. The safety interlock system of this project provides a wealth of safety interlock functions to ensure the safety of the personnel within the radiation area. However, the safety interlock system of this project fully considers the safety requirements of the system,and the key interlock control loops are designed based on SIL3, which meets the requirements of the IEC61508 standard. To obtain the SIL3 control loop,SIL3 bottom equipment and a controller are selected, the SIL3 wiring mode is used, the communication protocol applies the SIL3 PROFIsafe protocol, and a set of safety ring network systems designed for the bottom equipment communication are used. As an advantage of the SIL3 design it reduces the failure probability of the system and provides fail-safe protection; in the case of equipment failure, interlock actions can be applied to protect the personnel.

    At present, the designed safety interlock system has passed a fail-safe test and has been used in the project.The safety interlock system can trigger a fail-safe protection in the case of a network failure or disconnection of the equipment signal line,effectively ensuring the safety of the personnel.Although the system was developed for specific facilities and used to protect on-site personnel during proton therapy, the data collection, data processing, site cleaning, emergency stop, and activation functions in the system can still be widely used in most radiological experimental sites. Because the PSIS is extremely important for protecting field personnel in the radiation industry,it is necessary to continue recording the problems that occur during the operation of the existing equipment,optimize the safety interlock system, and improve the overall stability of the safety interlock system operation.

    Author contributionsAll authors contributed to the study conception and design. Material preparation, data collection and analysis were performed by Shou-Yuan Wang, Jing Zhang, Ou-Wen Huang,Zhu Li. The first draft of the manuscript was written by Shou-Yuan Wang and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.

    免费搜索国产男女视频| 国产精品久久久久久久电影| 久久这里只有精品中国| 日韩欧美三级三区| 色综合站精品国产| 一夜夜www| 日韩av在线大香蕉| 69人妻影院| 国产一区亚洲一区在线观看| 亚洲人与动物交配视频| 国产精品国产三级国产av玫瑰| 色在线成人网| 欧美精品国产亚洲| 亚洲无线在线观看| 黄色视频,在线免费观看| 91久久精品国产一区二区三区| 国产乱人视频| 国产乱人视频| 国产高潮美女av| 亚洲成人久久性| 91久久精品电影网| 国产精品无大码| 亚洲av五月六月丁香网| 国产高清不卡午夜福利| 亚洲中文字幕日韩| 亚洲中文日韩欧美视频| 18禁裸乳无遮挡免费网站照片| 久久久国产成人精品二区| 听说在线观看完整版免费高清| 亚洲av一区综合| 国产高清视频在线观看网站| a级毛片a级免费在线| 久久久久久久久久黄片| 久久中文看片网| 1000部很黄的大片| 午夜老司机福利剧场| 一进一出好大好爽视频| 五月玫瑰六月丁香| 噜噜噜噜噜久久久久久91| 人妻夜夜爽99麻豆av| 99热这里只有是精品在线观看| av在线天堂中文字幕| 91久久精品国产一区二区三区| 亚洲美女视频黄频| 国产黄a三级三级三级人| 久久人人爽人人爽人人片va| 嫩草影院精品99| 久久精品国产亚洲av香蕉五月| 性色avwww在线观看| 深夜精品福利| 97碰自拍视频| 中文字幕免费在线视频6| 午夜福利在线观看吧| 免费人成视频x8x8入口观看| 久久国内精品自在自线图片| 啦啦啦啦在线视频资源| 国产色婷婷99| 久久久欧美国产精品| 可以在线观看毛片的网站| 亚洲七黄色美女视频| 精华霜和精华液先用哪个| 日本在线视频免费播放| 九九热线精品视视频播放| 免费看av在线观看网站| 国产亚洲91精品色在线| av天堂在线播放| 此物有八面人人有两片| 免费无遮挡裸体视频| 18+在线观看网站| 午夜久久久久精精品| 男人的好看免费观看在线视频| 久久久久久大精品| 精品乱码久久久久久99久播| 久久精品国产自在天天线| 午夜福利在线在线| 蜜臀久久99精品久久宅男| 国产免费一级a男人的天堂| 亚洲激情五月婷婷啪啪| 日韩在线高清观看一区二区三区| 亚洲不卡免费看| 久久久精品大字幕| 97热精品久久久久久| 最近手机中文字幕大全| 18禁裸乳无遮挡免费网站照片| 亚洲人成网站在线播放欧美日韩| 亚洲精品一卡2卡三卡4卡5卡| 99精品在免费线老司机午夜| 久久久国产成人精品二区| 国产精品美女特级片免费视频播放器| 给我免费播放毛片高清在线观看| 九九久久精品国产亚洲av麻豆| 成人性生交大片免费视频hd| 久久久久久伊人网av| 少妇的逼水好多| 女同久久另类99精品国产91| 久久国产乱子免费精品| 俺也久久电影网| 简卡轻食公司| 国语自产精品视频在线第100页| 深爱激情五月婷婷| 99久久久亚洲精品蜜臀av| 99久国产av精品| 欧美性感艳星| 成人av一区二区三区在线看| 久久久久国产精品人妻aⅴ院| 99久久精品一区二区三区| 日本欧美国产在线视频| 日韩人妻高清精品专区| 国产成人aa在线观看| 精品午夜福利在线看| 亚洲av电影不卡..在线观看| 看黄色毛片网站| 国产男人的电影天堂91| 国产在线免费精品| 国产成人a∨麻豆精品| 69精品国产乱码久久久| 九九爱精品视频在线观看| 免费高清在线观看视频在线观看| 一级av片app| 久久综合国产亚洲精品| 少妇被粗大的猛进出69影院 | 国内少妇人妻偷人精品xxx网站| 免费黄频网站在线观看国产| 狂野欧美激情性bbbbbb| 亚洲综合色惰| 色94色欧美一区二区| 国产爽快片一区二区三区| 国产精品女同一区二区软件| 精品国产露脸久久av麻豆| 国产高清三级在线| 亚洲国产欧美日韩在线播放 | 18禁在线播放成人免费| 不卡视频在线观看欧美| 亚洲欧美清纯卡通| 伦理电影免费视频| 国产白丝娇喘喷水9色精品| 国产成人aa在线观看| 久久 成人 亚洲| 3wmmmm亚洲av在线观看| 三上悠亚av全集在线观看 | 国产亚洲5aaaaa淫片| 狂野欧美白嫩少妇大欣赏| 国产成人精品福利久久| 久久久久国产精品人妻一区二区| 亚洲精品国产色婷婷电影| 欧美日韩在线观看h| 十八禁网站网址无遮挡 | 美女内射精品一级片tv| 夜夜看夜夜爽夜夜摸| 国产乱人偷精品视频| 国产视频首页在线观看| 久久免费观看电影| 视频中文字幕在线观看| 欧美日韩精品成人综合77777| 国产在线视频一区二区| 国产色爽女视频免费观看| 在线观看免费高清a一片| 欧美激情国产日韩精品一区| 国产亚洲午夜精品一区二区久久| 性高湖久久久久久久久免费观看| 校园人妻丝袜中文字幕| 在线观看免费高清a一片| 午夜精品国产一区二区电影| 超碰97精品在线观看| 精品国产露脸久久av麻豆| 人妻人人澡人人爽人人| 国产一区二区三区av在线| 免费大片黄手机在线观看| 婷婷色av中文字幕| 国产免费视频播放在线视频| 成年美女黄网站色视频大全免费 | 亚洲,一卡二卡三卡| 天堂中文最新版在线下载| 两个人免费观看高清视频 | 亚洲精华国产精华液的使用体验| 寂寞人妻少妇视频99o| 香蕉精品网在线| 亚洲欧洲日产国产| 99久久精品热视频| 美女主播在线视频| av免费在线看不卡| av福利片在线观看| 特大巨黑吊av在线直播| 欧美3d第一页| 卡戴珊不雅视频在线播放| 亚洲国产av新网站| 女人精品久久久久毛片| 国产成人免费观看mmmm| 男人舔奶头视频| 亚洲久久久国产精品| 特大巨黑吊av在线直播| 国产欧美亚洲国产| 又粗又硬又长又爽又黄的视频| 十八禁高潮呻吟视频 | av视频免费观看在线观看| 老熟女久久久| 国产在线男女| 国产成人精品久久久久久| 纵有疾风起免费观看全集完整版| 日韩精品有码人妻一区| 国产男女内射视频| 久久午夜福利片| 麻豆精品久久久久久蜜桃| tube8黄色片| 国产乱人偷精品视频| 欧美精品高潮呻吟av久久| a级片在线免费高清观看视频| 精品国产一区二区三区久久久樱花| 日日啪夜夜撸| 精华霜和精华液先用哪个| 亚洲精品国产成人久久av| 亚洲欧美清纯卡通| 午夜精品国产一区二区电影| 精品视频人人做人人爽| 久久久久久久久久久丰满| 精品亚洲乱码少妇综合久久| av线在线观看网站| 一区在线观看完整版| 2018国产大陆天天弄谢| 日本91视频免费播放| 久久6这里有精品| 精华霜和精华液先用哪个| 久久久久久久久大av| av又黄又爽大尺度在线免费看| 欧美日韩视频高清一区二区三区二| 一区二区三区四区激情视频| 99久国产av精品国产电影| 在线亚洲精品国产二区图片欧美 | 麻豆乱淫一区二区| 精品亚洲成国产av| 大香蕉久久网| 精品国产乱码久久久久久小说| 69精品国产乱码久久久| 国产伦精品一区二区三区四那| 国产在线男女| h日本视频在线播放| 五月天丁香电影| 在线观看www视频免费| 中文字幕人妻熟人妻熟丝袜美| 国产日韩欧美视频二区| 九色成人免费人妻av| 亚洲国产精品成人久久小说| 免费播放大片免费观看视频在线观看| 国产亚洲最大av| 啦啦啦在线观看免费高清www| 777米奇影视久久| 噜噜噜噜噜久久久久久91| 最近手机中文字幕大全| 晚上一个人看的免费电影| 国产又色又爽无遮挡免| 人人妻人人爽人人添夜夜欢视频 | 国产亚洲午夜精品一区二区久久| av卡一久久| 精品国产国语对白av| 欧美少妇被猛烈插入视频| 亚洲国产精品一区三区| 国产亚洲一区二区精品| 欧美人与善性xxx| 国产成人精品无人区| 精品熟女少妇av免费看| 卡戴珊不雅视频在线播放| 两个人的视频大全免费| 久久精品国产鲁丝片午夜精品| 久久韩国三级中文字幕| 水蜜桃什么品种好| 男人狂女人下面高潮的视频| 嫩草影院新地址| 丰满迷人的少妇在线观看| 秋霞在线观看毛片| 亚洲无线观看免费| 老司机影院毛片| 视频区图区小说| 国产男女超爽视频在线观看| 国产成人91sexporn| 国产男女内射视频| 免费高清在线观看视频在线观看| 边亲边吃奶的免费视频| 秋霞在线观看毛片| 国产精品国产av在线观看| 晚上一个人看的免费电影| av女优亚洲男人天堂| 少妇熟女欧美另类| 午夜福利影视在线免费观看| 亚洲无线观看免费| 国产黄频视频在线观看| 国产一区二区在线观看日韩| 国产成人免费无遮挡视频| 一本久久精品| 国产精品熟女久久久久浪| 精品亚洲成a人片在线观看| 国产伦理片在线播放av一区| 国产伦在线观看视频一区| 纵有疾风起免费观看全集完整版| av福利片在线| 亚洲精品自拍成人| 内射极品少妇av片p| 狂野欧美激情性bbbbbb| 亚洲四区av| 国产精品.久久久| 亚洲精品一区蜜桃| 女性生殖器流出的白浆| 免费人妻精品一区二区三区视频| 高清av免费在线| 中文乱码字字幕精品一区二区三区| 亚洲不卡免费看| 亚洲精品日本国产第一区| 中文在线观看免费www的网站| 国产成人freesex在线| 五月开心婷婷网| 久久久久久久久久成人| 汤姆久久久久久久影院中文字幕| 各种免费的搞黄视频| 精品国产国语对白av| 精品卡一卡二卡四卡免费| 亚洲国产成人一精品久久久| 精品一品国产午夜福利视频| 91精品一卡2卡3卡4卡| 熟妇人妻不卡中文字幕| 亚洲欧美成人精品一区二区| 久久99一区二区三区| 亚洲综合色惰| 亚洲自偷自拍三级| av不卡在线播放| 97精品久久久久久久久久精品| 国产极品粉嫩免费观看在线 | 大片免费播放器 马上看| 乱系列少妇在线播放| 91午夜精品亚洲一区二区三区| www.色视频.com| 青青草视频在线视频观看| 天堂俺去俺来也www色官网| 亚洲,欧美,日韩| 国产白丝娇喘喷水9色精品| 日本wwww免费看| 精品久久国产蜜桃| 美女内射精品一级片tv| 亚洲国产精品国产精品| 老司机影院成人| 中文字幕免费在线视频6| 亚洲国产精品999| 亚洲电影在线观看av| 久久久久视频综合| 夜夜骑夜夜射夜夜干| 少妇精品久久久久久久| av在线老鸭窝| 夜夜看夜夜爽夜夜摸| 色婷婷av一区二区三区视频| 五月天丁香电影| 欧美激情国产日韩精品一区| 亚洲一区二区三区欧美精品| 天美传媒精品一区二区| 国产美女午夜福利| 日韩视频在线欧美| 久久免费观看电影| 欧美性感艳星| 久久久久国产精品人妻一区二区| 高清视频免费观看一区二区| 免费高清在线观看视频在线观看| 最近中文字幕高清免费大全6| 亚洲欧美一区二区三区国产| 久久久国产欧美日韩av| 永久网站在线| 精品久久久久久久久av| 欧美另类一区| 久热久热在线精品观看| 国内精品宾馆在线| 免费观看在线日韩| 3wmmmm亚洲av在线观看| 久久精品久久久久久噜噜老黄| 久久99蜜桃精品久久| 十分钟在线观看高清视频www | 成人免费观看视频高清| 我的老师免费观看完整版| 97在线人人人人妻| 尾随美女入室| 欧美一级a爱片免费观看看| 99精国产麻豆久久婷婷| 亚洲av福利一区| 一本—道久久a久久精品蜜桃钙片| 精品卡一卡二卡四卡免费| 久久久久久久国产电影| 久久久久久久久久人人人人人人| 久久久久久久久久久久大奶| 免费久久久久久久精品成人欧美视频 | 日韩不卡一区二区三区视频在线| 熟女人妻精品中文字幕| av有码第一页| 久久 成人 亚洲| 亚洲,欧美,日韩| 亚洲怡红院男人天堂| 日韩制服骚丝袜av| 精品久久久精品久久久| 日日摸夜夜添夜夜添av毛片| 你懂的网址亚洲精品在线观看| 九草在线视频观看| 亚洲欧美成人精品一区二区| 精品人妻一区二区三区麻豆| 日本欧美视频一区| 在线亚洲精品国产二区图片欧美 | 亚洲伊人久久精品综合| 国产在视频线精品| 涩涩av久久男人的天堂| 久久久久久人妻| 国产精品.久久久| 精华霜和精华液先用哪个| 婷婷色综合www| 中国国产av一级| 少妇人妻一区二区三区视频| 亚洲欧洲精品一区二区精品久久久 | 99久久精品热视频| 国产精品女同一区二区软件| 水蜜桃什么品种好| 亚洲av不卡在线观看| 美女脱内裤让男人舔精品视频| 又大又黄又爽视频免费| 97精品久久久久久久久久精品| 熟女电影av网| 日日爽夜夜爽网站| 国产成人一区二区在线| 日韩强制内射视频| 国产高清三级在线| 深夜a级毛片| 国产精品一区www在线观看| 熟女av电影| 最新的欧美精品一区二区| 国产成人a∨麻豆精品| 亚洲精华国产精华液的使用体验| 久久久久久久久大av| 婷婷色综合www| 国产在线男女| 成年人免费黄色播放视频 | 国产欧美日韩一区二区三区在线 | 日本免费在线观看一区| 国产一区二区三区综合在线观看 | 人妻夜夜爽99麻豆av| 国产精品国产三级专区第一集| 久久久欧美国产精品| 蜜臀久久99精品久久宅男| 国产精品成人在线| 久久人人爽人人爽人人片va| 国产一区二区在线观看日韩| 亚洲欧洲日产国产| 嫩草影院入口| 亚洲天堂av无毛| 久久97久久精品| 大香蕉97超碰在线| 男人爽女人下面视频在线观看| 国产乱人偷精品视频| 少妇被粗大的猛进出69影院 | 男人爽女人下面视频在线观看| 99国产精品免费福利视频| 国产极品粉嫩免费观看在线 | 亚洲经典国产精华液单| 又爽又黄a免费视频| 亚洲精品一二三| 不卡视频在线观看欧美| 99热网站在线观看| 日韩,欧美,国产一区二区三区| 欧美精品人与动牲交sv欧美| 校园人妻丝袜中文字幕| 大片免费播放器 马上看| 欧美成人精品欧美一级黄| 亚洲第一av免费看| 99热这里只有是精品在线观看| 成人免费观看视频高清| av福利片在线观看| 日本午夜av视频| 日韩强制内射视频| 嫩草影院新地址| 国产爽快片一区二区三区| 一级a做视频免费观看| 国产成人免费观看mmmm| 日本免费在线观看一区| 久久久久久久精品精品| 国产精品蜜桃在线观看| 三级经典国产精品| 五月玫瑰六月丁香| 秋霞在线观看毛片| 成人亚洲欧美一区二区av| 亚洲av电影在线观看一区二区三区| 伊人久久国产一区二区| 久久免费观看电影| 大片免费播放器 马上看| 亚洲精品第二区| 亚洲第一av免费看| 又大又黄又爽视频免费| 午夜激情福利司机影院| 欧美老熟妇乱子伦牲交| 99热国产这里只有精品6| 国内精品宾馆在线| 午夜免费男女啪啪视频观看| 亚州av有码| 嘟嘟电影网在线观看| 免费观看在线日韩| 亚洲欧洲精品一区二区精品久久久 | 另类精品久久| 久久精品国产亚洲av涩爱| 久久人妻熟女aⅴ| 久久久久久伊人网av| 中文乱码字字幕精品一区二区三区| av黄色大香蕉| 日韩av不卡免费在线播放| 精品酒店卫生间| 黄色怎么调成土黄色| 日本爱情动作片www.在线观看| 久久久国产一区二区| 成人无遮挡网站| 欧美 日韩 精品 国产| 亚洲欧美精品自产自拍| 一区二区av电影网| 久久午夜福利片| 在线观看一区二区三区激情| 国产一级毛片在线| 在线观看www视频免费| 成人毛片60女人毛片免费| 国产精品国产三级国产av玫瑰| 日韩av在线免费看完整版不卡| 寂寞人妻少妇视频99o| 老女人水多毛片| 欧美人与善性xxx| 黄色一级大片看看| 欧美成人精品欧美一级黄| 欧美老熟妇乱子伦牲交| 国产精品国产三级国产专区5o| 人妻制服诱惑在线中文字幕| 国模一区二区三区四区视频| 搡老乐熟女国产| 六月丁香七月| 91精品一卡2卡3卡4卡| 永久免费av网站大全| 久久99热这里只频精品6学生| 久久久久久久亚洲中文字幕| 亚洲精品亚洲一区二区| 能在线免费看毛片的网站| 狠狠精品人妻久久久久久综合| 国产成人freesex在线| 免费大片黄手机在线观看| 亚洲精品国产成人久久av| 久久精品国产a三级三级三级| 九九久久精品国产亚洲av麻豆| 曰老女人黄片| 国产精品一区二区性色av| 人人澡人人妻人| 内射极品少妇av片p| 久久影院123| 中文资源天堂在线| av免费在线看不卡| 国产av一区二区精品久久| 亚洲内射少妇av| 精品一品国产午夜福利视频| 草草在线视频免费看| 日韩人妻高清精品专区| 人人妻人人看人人澡| 国产精品国产三级专区第一集| 国精品久久久久久国模美| 一级毛片aaaaaa免费看小| 日韩制服骚丝袜av| 建设人人有责人人尽责人人享有的| 久久久久久久久久久免费av| 国产精品欧美亚洲77777| 自线自在国产av| 国产熟女午夜一区二区三区 | 亚洲美女黄色视频免费看| 黑人巨大精品欧美一区二区蜜桃 | 亚洲欧美一区二区三区国产| 在线天堂最新版资源| 欧美精品亚洲一区二区| 天天躁夜夜躁狠狠久久av| 国产片特级美女逼逼视频| 欧美三级亚洲精品| 久久国产亚洲av麻豆专区| 黄色一级大片看看| 国模一区二区三区四区视频| 伦精品一区二区三区| 亚洲国产最新在线播放| 国产精品一区二区三区四区免费观看| 国精品久久久久久国模美| 欧美最新免费一区二区三区| 精品国产国语对白av| av福利片在线观看| 男男h啪啪无遮挡| 日本黄大片高清| 国产亚洲精品久久久com| 免费人妻精品一区二区三区视频| 亚洲国产成人一精品久久久| 亚洲欧美成人综合另类久久久| 日韩视频在线欧美| 搡老乐熟女国产| 99re6热这里在线精品视频| 欧美性感艳星| 免费大片黄手机在线观看| 欧美日韩国产mv在线观看视频| 制服丝袜香蕉在线| 亚洲国产精品成人久久小说| 波野结衣二区三区在线| 一级片'在线观看视频| 午夜视频国产福利| 搡女人真爽免费视频火全软件| 亚洲图色成人| 少妇被粗大猛烈的视频| 久久久久久久国产电影| 91aial.com中文字幕在线观看| 91精品国产九色| 五月伊人婷婷丁香| 久久久久久久国产电影| 国产真实伦视频高清在线观看| 成年av动漫网址| 少妇人妻一区二区三区视频| 精品亚洲成a人片在线观看| 五月玫瑰六月丁香| av在线app专区| 亚洲精品久久久久久婷婷小说| 男人和女人高潮做爰伦理| 肉色欧美久久久久久久蜜桃| 99热全是精品|