    Diagnostic problems in two-dimensional shear wave elastography of the liver

    2020-06-15 06:29:18HirokoNaganumaHideakiIshidaAtsushiUnoHiroshiNagaiHidekatsuKurodaMasahiroOgawa
    World Journal of Radiology 2020年5期

    Hiroko Naganuma,Hideaki Ishida,Atsushi Uno,Hiroshi Nagai,Hidekatsu Kuroda,Masahiro Ogawa

    Abstract

    Key words: Two-dimensional shear wave elastography;Ultrasound;Artifacts;Liver cirrhosis;Liver tumor;Computer simulation model

    INTRODUCTION

    Although liver biopsy has long been the standard method for diagnosing chronic liver disease,it is associated with many problems,such as postbiopsy complications[1],sampling error[2],and interobserver variability[3]. In a positive aspect,these problems motivated the development of noninvasive techniques to assess liver fibrosis,including magnetic resonance elastography[4-6]and ultrasound (US)-based elastography[7-10]. The latter has attained preferential use because of its lower cost and easier manipulation of the instrument[7-10]. Currently,there are four kinds of US-based elastography,namely strain elastography[11],transient elastography[12],point shear wave elastography[13],and two-dimensional shear wave elastography (2D-SWE)[9,14,15].Among these methods,2D-SWE has emerged as the most frequently used diagnostic US tool for hepatic fibrosis quantification;this is due to its ability to sample a large area in the liver,changing the sampling area quickly under B-mode observation and displaying a color mapping of shear wave (SW) values over the B-mode image[9,15].Measurement of SW values can be easily performed but it is also well known that the obtained SW values change according to the instrument itself and the level of expertise of the operator;as such,the certitude of those SW results is not always satisfactory[6,9,15]. The other reasons for this instability have not been clarified.

    This review reexamines the variability,using a method that is different from those in the commonly reported studies. US artifacts are routinely encountered in clinical practice and include refraction artifacts[16-18],attenuation artifacts[16,19,20],reflection artifact[16,21],reverberation artifacts[22]and others[16,23,24]. As use of 2D-SWE becomes more widespread,the problems related to US artifacts in it will be more important[25-27]. The aim of this review was,thus,to present a wide spectrum of US artifacts in hepatic 2D-SWE and to provide the simplest way to minimize the effects of such.

    THEORETICAL BASIS OF TWO-DIMENSIONAL SHEAR WAVE ELASTOGRAPHY

    The push-pulse produces small tissue movements in the plane of the push-pulse.These tissue movements,in turn,produce SWs that propagate and produce other minimal tissue movements in the horizontal plane of the push-pulse. These movements in the horizontal plane further propagate through the tissue in a sideway direction,away from the push-pulse. These SW movements (strictly speaking,tissue movements produced by SWs) are tracked by the regular-interval tracking US pulses,which are used to measure the arrival time of SWs. The simple formula is arrival time of SW/distance from the push-pulse,used to determine the speed of SWs. Such a measurement is possible because the speed of SWs is very slow (< 1-10 m/s)compared to the speed of US pulses (around 1540 m/s) (Table 1)[28,29]. There are two ways to quantify the relative tissue stiffness: SW speed expressed as m/s;and,kilopascal (kPa). Mostly,the SW speed is converted automatically to kPa,using the equation 1 kPa = 3 × P × (SW speed m/s)2,with the assumption that the examined tissue is always homogeneous,and where P is the tissue density and defined as 1.00 kg/m3.

    Table1 Acoustic characteristics of human tissues[28,29]

    The 2D-SWE method uses emission of multiple acoustic push-pulse to generate SWs from multiple lines in the tissue (Figure 1). As the SWs propagate,they cause tissue displacement,and this very minimal displacement is tracked by conventional pulse-echo US beams[6,30,31]. The US machine measures the propagation speed of SWs as they travel in tissue,and the measured SW speed at each point is displayed in color[6,32]. The reconstructed 2D color mapping of SWs is called “2D-SWE” and facilitates recognition of global distribution of tissue elasticity of the area.

    FACTORS THAT DEGRADE QUALITY OF TWODIMENSIONAL SHEAR WAVE ELASTOGRAPHY IMAGES

    For all,these US artifact-related problems are more clearly shown upon use of propagation mode,which can determine the causal artifact.

    Reverberation artifact

    Acoustic reverberation of the push-pulse at the liver capsule results in a false increase in SW speed. Reverberation artifacts arise when the US signals reflects repeatedly between two interfaces of largely different acoustic impedance,resulting in delayed echoes returning to the transducer at regular intervals. This can lead to reverberation artifacts on grayscale US image[33]. The same phenomenon of the push pulse is considered to occur between the liver capsule and the transducer. This phenomenon usually appears in the upper part of the cursor,presenting as a band of red (i.e. very increased SW speeds) (Figure 2). Thus,it is highly recommended to avoid this area when measuring SW values with SWE;if not,the results will be overestimated.

    Attenuation artifact

    Attenuation frequently occurs in a clinical setting and most commonly in obese or overweight patients. In these patients,a strongly attenuated push-pulse passing through the tissue results in poor creation of SWs at depth. This phenomenon usually appears at the distal part of the cursor,presenting as a colorless area (i.e. poor acquisition of SW data) (Figure 3). Thus,it is highly recommended to avoid this area when measuring SW values in obese or overweight patients.

    Probe compression artifact

    Excessive probe compression during image acquisition causes artificial compression of tissues just under the probe. This phenomenon is seen clearly when the liver is evaluated through subcostal scanning (Figure 4). SW measurement in the compressed area leads to a false elevation in SW values. A probe compression artifact appears near the probe,presenting as a reddish band. Thus,it is recommended not to compress the liver too much. The simplest way to avoid this artifact is to perform measurements through intercostal spaces only,where the liver tissue is less likely to be compressed by the probe.

    Motion artifact

    This phenomenon is seen clearly when the patient is unable to hold his/her breath or to change position quickly. Besides these extreme conditions,a motion artifact is usually seen when observing the left lobe of the liver,due to cardiac motion. Motion artifacts give rise to the characteristic image of a colorless area or a reddish area in the cursor,suggesting that SW values cannot be measured accurately in this area (Figure 5;see also,the section on tracking US beam-related problems below).

    Figure1 Schematic drawing of two-dimensional shear wave elastography in a healthy individual.

    MECHANISMS UNDERLYING ULTRASOUND ARTIFACTRELATED PROBLEMS IN TWO-DIMENSIONAL SHEAR WAVE ELASTOGRAPHY DETERMINED BY A COMPUTER SIMULATION MODEL

    US scanners reconstruct US images on the assumption that sound passes through all parts of the human body in a straight line and at a constant velocity,and apply this assumption to all scanning planes. Actually,however,these sound velocity in the human body varies with the composition of the tissue scanned. When a plane containing tissues with different velocities is scanned,sound refraction occurs at the interface of the tissues,according to Snell's law (see later in “Discussion”). Computer simulation analysis helps understand the global images of refraction in the plane,measuring the degree of refraction at each point of the interface,not only in grayscale US image but also in 2D-SWE image.

    2D-SWE in diffuse liver disease

    2D-SWE has been reported as useful for assessing liver fibrosis with high accuracy[8,9,34-36]. It is generally reported that SW speed increases as fibrosis advances[15,37-39];thus,heterogeneous distribution of different degrees of fibrosis will create a heterogeneous color mapping in 2D-SWE. Measurement of liver stiffness is based on the following assumptions: (1) All push-pulses emitted from the transducer advance directly without deviation,at an even distance from each other;and (2) All SWs created from these push-pulses propagate horizontally at the same distance from the transducer. However,in reality,like conventional US beams,push-pulses refract at the liver surface (according to Snell's law) because there is a large difference between the liver and the surrounding tissue and the hepatic surface is irregular.

    Figure6 shows the mode of propagation of push-pulses in the liver. As a result,SWs propagate in unassumed directions. This computer simulation model clearly teaches us an important lesson about measurement of SW speeds in liver cirrhosis.Furthermore,this speculation is in accordance with the findings from some studies with 2D-SWE measurements of healthy volunteers in whom the liver surface was smooth[40,41]. In those studies,measurements by novice operators were not significantly different from those of experienced operators[31,40,42]. Our computer simulation model shows that studies of normal liver do not contribute to resolving the problem of significant variability of SW values in liver cirrhosis. Although the measuring error is minimal,it is recommended to measure SW values through the least deformed hepatic surface possible.

    2D-SWE in liver tumors

    2D-SWE findings have been reported as useful for predicting tissue characterization of liver tumor,with markedly contradictory results[43-47]. The most widely accepted method for diagnosing liver tumors by 2D-SWE is measurement of maximal or median SW values within the lesion[46-49]. This method has long been performed but,as is shown in Figure 7,this idea is compromised by a combination of US refraction and reflection. The reported SW measurement of tumor stiffness is based on the following assumptions: (1) All SWs created from the push-pulses advance directly,without deviation,from the surrounding tissue toward the tumor and within the tumor;and(2) No reflection occurs at the interface between the surrounding tissue and liver tumor.

    Figure2 Reverberation artifact.

    However,in reality,like conventional US beams,SWs refract at the interface (again,according to Snell's law) because there is a large difference in propagation speed between the liver tumor and the surrounding tissue (Figure 7). Furthermore,the degree of refraction is thought to be more accentuated in 2D-SWE (1-10 m/s) than in conventional US (around 1540 m/s) (Figure 7B and C). The degree of reflection at the surrounding tissue-liver tumor interface is more accentuated in 2D-SWE,for the same reason. It is the deviation from the two above-mentioned assumptions that leads to the conclusion that accurate stiffness measurement of liver tumors is not possible by 2D-SWE,in most cases. This idea is in accordance with the American Institute of Ultrasound in Medicine recommendation that evaluation of liver tumor stiffness by 2D-SWE must be done with caution. Although,the SW value measurement inside the tumor is not highly accurate,this measurement is still valuable for judgement of relative stiffness of tumor and surrounding liver. However,conversely,this phenomenon helps in the detection of a liver tumor that is otherwise poorly visualized on B-mode (Figure 8) and in the differentiation of a pseudo-liver tumor(i.e.,focal fatty infiltration[43]or chronic hepatic porphyria[50]) from a true liver tumor.

    Tracking US beam-related problems

    Finally,we would like to briefly mention refraction of tracking US beams. We mentioned earlier the influence of US artifacts on 2D-SWE,on the assumption that the tracking US pulses occur at regular intervals and advance straight from the transducer through the tissue. However,they deviate in some instances. Figure 9 shows such a case. US beams deviate twice when passing through a focal lesion,with one at the upper surrounding tissue-focal lesion interface and the other at the lower focal lesionsurrounding tissue interface. As a result,US beams propagating in tissue posterior to the focal lesion deviate;however,the US machine calculates SW values on the assumption that there is no deviation of US beams. Thus,this phenomenon suggests that SW measurements must be done in a plane that does not contain focal lesions.

    DISCUSSION

    In many reported studies,2D-SWE has been shown to have a sufficient level of sensitivity in evaluating the degree of hepatic fibrosis[7-9,36]but with significant heterogeneity of results. This heterogeneity has been considered due to variations among patient populations,study design,and devices used[49]. Some of these studies have pointed to technical flaws in earlier studies but recent studies also show some instable results[45,46,51]. These studies have brought into question whether they included the influence of US artifacts.

    Figure3 Attenuation artifact.

    This review has an important strength in that it is a theoretical analysis not found in similar trials reported in the literature. Computer simulation model yields a purely theoretical analysis. In the case of gray-scale US images,the sound refraction produces artifactual images,because the US scanner displays each point at the appropriate distance determined by the time taken for the echo to return to the transducer in the direction in which the transducer is pointing at the time,even when the US beam is refracted. In the case of 2D-SWE as well,the US scanner reconstructs 2D-SWE images on the assumption that SW passes horizontally in a straight line,without deviation,and the US scanner displays SW velocity mapping on the basis of data measured by tracking pulses (determined by the time/distance). Computer simulation model enables us to calculate accurately and automatically the degree of refraction at each point,and understand the global image of refraction in the plane.This method is especially useful for understanding the global mode of refraction at the curved interface (tumor-surrounding tissue interface or surrounding tissueirregular hepatic surface interface) (Figure 10)[52-54]. In short,it ensures that unfavorable factors,such as technical errors,differences in US machines used and different levels of 2D-SWE experience,or influence biased by additional clinical data,do not interfere in this analysis. It also maintains a high quality of interpretation. The SWE diagnosis is usually the so-called “final report” of the clinician and explains that it is necessarily correct. Our simulation model can be considered the basic thought of what the clinicians conclude. It is recommended,therefore,that as more practitioners are trained in 2D-SWE,our basic thought is used in clinical practice.

    CONCLUSION

    Two-dimensional shear wave elastography artifacts,although seen very frequently in a clinical setting,are poorly recognized. Our review emphasizes that interpretation of 2D-SWE images must incorporate knowledge of US artifacts.

    Figure4 Probe compression artifact.

    Figure5 Motion artifact.

    Figure6 Computer simulation model of push-pulses in a cirrhotic patient.

    Figure7 Two-dimensional shear wave elastography of liver tumor.

    Figure8 Representative case: intrahepatic cholangiocellular carcinoma.

    Figure9 Computer simulation model of deviated push-pulses.

    Figure10 Sound refraction at the interface.

    春色校园在线视频观看| 日韩欧美 国产精品| 亚洲av综合色区一区| 有码 亚洲区| 欧美日韩一区二区视频在线观看视频在线| 午夜免费男女啪啪视频观看| 赤兔流量卡办理| 九九爱精品视频在线观看| 麻豆乱淫一区二区| 青春草国产在线视频| 黄色视频在线播放观看不卡| 日韩欧美精品免费久久| 亚洲情色 制服丝袜| 国产高清国产精品国产三级| 色哟哟·www| 精品一区在线观看国产| 日本wwww免费看| 爱豆传媒免费全集在线观看| 国内揄拍国产精品人妻在线| 91在线精品国自产拍蜜月| 下体分泌物呈黄色| 十八禁高潮呻吟视频 | 精品一区二区免费观看| 亚洲国产欧美在线一区| 春色校园在线视频观看| 丁香六月天网| 亚洲电影在线观看av| 蜜臀久久99精品久久宅男| 免费久久久久久久精品成人欧美视频 | 水蜜桃什么品种好| 内地一区二区视频在线| 欧美区成人在线视频| 亚洲国产av新网站| 久久午夜综合久久蜜桃| 色视频www国产| 国产一级毛片在线| 在线精品无人区一区二区三| 国产亚洲5aaaaa淫片| 国产精品欧美亚洲77777| 男人和女人高潮做爰伦理| 三级国产精品片| a级一级毛片免费在线观看| 国产av精品麻豆| 又大又黄又爽视频免费| 国产免费一区二区三区四区乱码| 高清毛片免费看| 不卡视频在线观看欧美| 26uuu在线亚洲综合色| 九色成人免费人妻av| 卡戴珊不雅视频在线播放| 国产精品蜜桃在线观看| 国产精品一区二区三区四区免费观看| 国产 一区精品| a级毛片免费高清观看在线播放| 高清不卡的av网站| 99久久精品国产国产毛片| 男人和女人高潮做爰伦理| 男人狂女人下面高潮的视频| 国产免费又黄又爽又色| 国产av精品麻豆| 日韩不卡一区二区三区视频在线| 天堂8中文在线网| 国产精品一区www在线观看| 国产一区有黄有色的免费视频| av天堂中文字幕网| 视频区图区小说| 免费观看性生交大片5| 韩国av在线不卡| 在线观看免费视频网站a站| 在线天堂最新版资源| 国产成人免费观看mmmm| 久久午夜福利片| 国产91av在线免费观看| 国产黄片美女视频| 免费高清在线观看视频在线观看| 一级毛片久久久久久久久女| 成人二区视频| 精品酒店卫生间| 成人黄色视频免费在线看| av天堂久久9| 亚洲精品第二区| 秋霞伦理黄片| 亚洲精品aⅴ在线观看| 国产精品人妻久久久影院| 国产在视频线精品| 午夜免费观看性视频| 久久久久久久久久久久大奶| 久久影院123| 最近的中文字幕免费完整| 男女无遮挡免费网站观看| 国产熟女欧美一区二区| 国产精品免费大片| 99热全是精品| av视频免费观看在线观看| 日韩一本色道免费dvd| 国国产精品蜜臀av免费| 亚洲欧美中文字幕日韩二区| a级片在线免费高清观看视频| av在线观看视频网站免费| 成人影院久久| 青春草国产在线视频| 国产伦在线观看视频一区| 黄片无遮挡物在线观看| 校园人妻丝袜中文字幕| 十八禁高潮呻吟视频 | 午夜激情福利司机影院| 精品少妇黑人巨大在线播放| 少妇人妻一区二区三区视频| 精品少妇内射三级| 国产日韩欧美在线精品| 欧美日韩在线观看h| 亚洲欧美日韩卡通动漫| 91成人精品电影| 高清午夜精品一区二区三区| 成人特级av手机在线观看| 国产乱来视频区| 国产精品不卡视频一区二区| 日本与韩国留学比较| 国产日韩一区二区三区精品不卡 | 三上悠亚av全集在线观看 | 男人爽女人下面视频在线观看| 夜夜爽夜夜爽视频| 菩萨蛮人人尽说江南好唐韦庄| 性色avwww在线观看| 大又大粗又爽又黄少妇毛片口| 夜夜看夜夜爽夜夜摸| 亚洲av成人精品一二三区| 夜夜骑夜夜射夜夜干| 国产成人精品一,二区| 免费观看在线日韩| 丰满乱子伦码专区| 国产69精品久久久久777片| 亚洲,欧美,日韩| 亚洲熟女精品中文字幕| 好男人视频免费观看在线| 高清毛片免费看| 麻豆成人av视频| 街头女战士在线观看网站| 在线观看国产h片| 亚洲丝袜综合中文字幕| 国产一区二区三区综合在线观看 | 99久久精品国产国产毛片| 51国产日韩欧美| 天堂俺去俺来也www色官网| 久久国产精品男人的天堂亚洲 | 啦啦啦中文免费视频观看日本| 一级黄片播放器| 一级毛片aaaaaa免费看小| 精品一区二区三卡| 黄色一级大片看看| 婷婷色综合大香蕉| 日韩av免费高清视频| 久久人人爽av亚洲精品天堂| 亚洲精品久久午夜乱码| 久久6这里有精品| 精品久久久精品久久久| 麻豆精品久久久久久蜜桃| av在线app专区| 尾随美女入室| 人妻夜夜爽99麻豆av| 日本午夜av视频| 久久人人爽人人爽人人片va| 一区二区三区四区激情视频| a 毛片基地| 中文字幕人妻熟人妻熟丝袜美| a级一级毛片免费在线观看| 国产av精品麻豆| 精品国产乱码久久久久久小说| 日韩强制内射视频| 中文字幕制服av| 99视频精品全部免费 在线| 日韩三级伦理在线观看| 青春草国产在线视频| 伦理电影大哥的女人| 亚洲精品乱久久久久久| 美女视频免费永久观看网站| 欧美xxⅹ黑人| 日韩强制内射视频| 丁香六月天网| 婷婷色综合大香蕉| av女优亚洲男人天堂| 国产精品人妻久久久久久| 亚洲天堂av无毛| 六月丁香七月| 两个人免费观看高清视频 | 中文字幕亚洲精品专区| 寂寞人妻少妇视频99o| 91久久精品国产一区二区三区| 久热久热在线精品观看| 交换朋友夫妻互换小说| 9色porny在线观看| 久久久久久久国产电影| 美女大奶头黄色视频| 亚洲,一卡二卡三卡| 91成人精品电影| 日韩av不卡免费在线播放| 天美传媒精品一区二区| 成人美女网站在线观看视频| 亚洲精品,欧美精品| 六月丁香七月| 日日摸夜夜添夜夜添av毛片| 我的女老师完整版在线观看| 久久精品久久久久久久性| 亚洲成色77777| 免费久久久久久久精品成人欧美视频 | 黄片无遮挡物在线观看| 欧美日韩国产mv在线观看视频| 午夜影院在线不卡| 欧美精品一区二区免费开放| 男人和女人高潮做爰伦理| 亚洲欧美成人精品一区二区| www.色视频.com| 成年人午夜在线观看视频| 国产毛片在线视频| 在线观看三级黄色| 久久精品国产亚洲网站| 插逼视频在线观看| 久久久久久久精品精品| 欧美精品高潮呻吟av久久| 亚洲精品久久久久久婷婷小说| 国产精品一区二区在线不卡| 欧美成人午夜免费资源| 午夜福利网站1000一区二区三区| 91在线精品国自产拍蜜月| 伊人久久国产一区二区| 国产黄频视频在线观看| 好男人视频免费观看在线| h视频一区二区三区| 国产精品成人在线| 一本久久精品| 99re6热这里在线精品视频| 我的女老师完整版在线观看| 国产精品久久久久久精品古装| 大陆偷拍与自拍| 男女国产视频网站| 久久久久久久久久久丰满| 国产永久视频网站| 一级毛片我不卡| h视频一区二区三区| 日韩精品有码人妻一区| 麻豆乱淫一区二区| 日本黄大片高清| 少妇裸体淫交视频免费看高清| 乱码一卡2卡4卡精品| 22中文网久久字幕| 国产免费一区二区三区四区乱码| 日日摸夜夜添夜夜添av毛片| 精品午夜福利在线看| 青春草国产在线视频| 99精国产麻豆久久婷婷| 校园人妻丝袜中文字幕| 国产淫语在线视频| 菩萨蛮人人尽说江南好唐韦庄| 狂野欧美激情性bbbbbb| 国产av精品麻豆| 伦理电影免费视频| 亚洲国产精品国产精品| 内地一区二区视频在线| 欧美亚洲 丝袜 人妻 在线| 能在线免费看毛片的网站| 国产亚洲av片在线观看秒播厂| 国内少妇人妻偷人精品xxx网站| 观看美女的网站| 妹子高潮喷水视频| 国产有黄有色有爽视频| 国产av国产精品国产| 免费少妇av软件| 一区在线观看完整版| 高清欧美精品videossex| 高清视频免费观看一区二区| 国产成人精品婷婷| 十八禁高潮呻吟视频 | 97在线人人人人妻| 观看av在线不卡| 18禁在线播放成人免费| 久久久国产欧美日韩av| 天天操日日干夜夜撸| 日本vs欧美在线观看视频 | 在线观看免费视频网站a站| 国产极品粉嫩免费观看在线 | 国精品久久久久久国模美| 亚洲国产成人一精品久久久| 国产男女内射视频| 久久久久网色| 啦啦啦啦在线视频资源| 国产欧美日韩精品一区二区| 成年人免费黄色播放视频 | 另类精品久久| 亚洲欧美日韩另类电影网站| 自线自在国产av| 两个人免费观看高清视频 | av免费在线看不卡| 亚洲欧洲日产国产| 丰满迷人的少妇在线观看| 最近2019中文字幕mv第一页| 午夜免费男女啪啪视频观看| 亚洲精品久久久久久婷婷小说| 国产一区二区三区av在线| www.色视频.com| 人妻人人澡人人爽人人| 国产日韩欧美亚洲二区| www.av在线官网国产| 日韩伦理黄色片| 一二三四中文在线观看免费高清| 国产精品一区www在线观看| 99久久精品一区二区三区| 在线看a的网站| 久久久久久久久大av| 亚洲,一卡二卡三卡| 免费看日本二区| 亚洲国产精品999| 亚洲精品日韩在线中文字幕| 青春草亚洲视频在线观看| 国产精品99久久99久久久不卡 | 欧美 亚洲 国产 日韩一| 成人午夜精彩视频在线观看| 91午夜精品亚洲一区二区三区| 男女边吃奶边做爰视频| 乱系列少妇在线播放| 成年女人在线观看亚洲视频| 制服丝袜香蕉在线| 亚洲自偷自拍三级| 99久国产av精品国产电影| 国产精品久久久久久久久免| 99国产精品免费福利视频| 久久99蜜桃精品久久| 一级二级三级毛片免费看| 中文字幕亚洲精品专区| 不卡视频在线观看欧美| 午夜av观看不卡| 久久国产亚洲av麻豆专区| 亚洲精品乱码久久久v下载方式| 夫妻性生交免费视频一级片| 亚洲av福利一区| 亚洲四区av| 人妻 亚洲 视频| 亚洲美女黄色视频免费看| 啦啦啦视频在线资源免费观看| 自线自在国产av| 欧美精品人与动牲交sv欧美| 亚洲精品日本国产第一区| 精品国产露脸久久av麻豆| 91精品伊人久久大香线蕉| 亚洲久久久国产精品| 免费av中文字幕在线| 日韩视频在线欧美| 国产成人精品婷婷| 亚洲三级黄色毛片| 国产国拍精品亚洲av在线观看| tube8黄色片| 少妇人妻 视频| 性色av一级| freevideosex欧美| 久久精品夜色国产| 在线观看av片永久免费下载| 国产亚洲精品久久久com| 99精国产麻豆久久婷婷| 亚洲综合精品二区| 你懂的网址亚洲精品在线观看| 一级毛片我不卡| 亚洲经典国产精华液单| 99久久综合免费| 国产成人精品福利久久| 亚洲欧美日韩卡通动漫| 看十八女毛片水多多多| 久久6这里有精品| 九草在线视频观看| 中文字幕av电影在线播放| 免费看光身美女| 色视频在线一区二区三区| 欧美人与善性xxx| 欧美性感艳星| 亚洲在久久综合| 搡老乐熟女国产| 久久婷婷青草| av一本久久久久| 蜜桃在线观看..| 啦啦啦在线观看免费高清www| 国产一区有黄有色的免费视频| av免费在线看不卡| 99热这里只有是精品50| 午夜久久久在线观看| 亚洲av成人精品一区久久| 日韩大片免费观看网站| av在线老鸭窝| 亚洲国产欧美在线一区| 精品一品国产午夜福利视频| 国产精品一区二区在线不卡| 国产成人一区二区在线| 天美传媒精品一区二区| 国产成人午夜福利电影在线观看| 视频区图区小说| 欧美日韩av久久| 国产成人freesex在线| 亚洲精品色激情综合| 少妇人妻精品综合一区二区| 女人久久www免费人成看片| 久久人人爽人人爽人人片va| 嘟嘟电影网在线观看| 国国产精品蜜臀av免费| 国产69精品久久久久777片| 国产av码专区亚洲av| 国产高清国产精品国产三级| 国产日韩欧美亚洲二区| 亚洲精品久久久久久婷婷小说| 午夜老司机福利剧场| 欧美+日韩+精品| 精品久久久久久电影网| 国产 精品1| 免费av不卡在线播放| 少妇的逼好多水| 成年av动漫网址| 2022亚洲国产成人精品| 日日摸夜夜添夜夜爱| 久久国产精品男人的天堂亚洲 | 精品99又大又爽又粗少妇毛片| 男人舔奶头视频| 黄色视频在线播放观看不卡| 汤姆久久久久久久影院中文字幕| 一级av片app| 久久国产乱子免费精品| 乱人伦中国视频| 我的女老师完整版在线观看| 大又大粗又爽又黄少妇毛片口| 免费久久久久久久精品成人欧美视频 | 国产亚洲91精品色在线| 欧美成人午夜免费资源| 一本—道久久a久久精品蜜桃钙片| 国产精品伦人一区二区| 日本wwww免费看| 中文在线观看免费www的网站| 啦啦啦在线观看免费高清www| h视频一区二区三区| 亚洲精品国产色婷婷电影| av专区在线播放| 久久ye,这里只有精品| 久久毛片免费看一区二区三区| 人人澡人人妻人| 免费看光身美女| 国产成人免费无遮挡视频| 免费av不卡在线播放| 美女cb高潮喷水在线观看| 国产男女超爽视频在线观看| 老司机亚洲免费影院| 久久久久久久久久久久大奶| 人妻系列 视频| 久久国产乱子免费精品| 亚洲人成网站在线播| 久久久久精品久久久久真实原创| 日本91视频免费播放| 国产欧美另类精品又又久久亚洲欧美| 久久毛片免费看一区二区三区| 亚洲欧洲精品一区二区精品久久久 | 极品少妇高潮喷水抽搐| 日韩大片免费观看网站| 亚洲国产成人一精品久久久| 国产亚洲5aaaaa淫片| av女优亚洲男人天堂| 男女无遮挡免费网站观看| 丰满人妻一区二区三区视频av| 久久人人爽人人片av| 丝瓜视频免费看黄片| 久久99蜜桃精品久久| 国产日韩欧美视频二区| 亚洲一区二区三区欧美精品| 99热6这里只有精品| 永久免费av网站大全| 亚洲国产最新在线播放| 亚洲av二区三区四区| 少妇猛男粗大的猛烈进出视频| 午夜视频国产福利| 久久久亚洲精品成人影院| 午夜福利视频精品| 乱码一卡2卡4卡精品| 街头女战士在线观看网站| 最近的中文字幕免费完整| 永久免费av网站大全| 午夜福利影视在线免费观看| 久久久久久久久久久免费av| 人人妻人人添人人爽欧美一区卜| 秋霞伦理黄片| 亚洲av福利一区| tube8黄色片| 99久久中文字幕三级久久日本| 亚洲av不卡在线观看| 蜜臀久久99精品久久宅男| 建设人人有责人人尽责人人享有的| 日韩av不卡免费在线播放| 在现免费观看毛片| 日产精品乱码卡一卡2卡三| 如何舔出高潮| 我的女老师完整版在线观看| a级一级毛片免费在线观看| 视频中文字幕在线观看| 久久久久久久久久久免费av| 亚洲综合精品二区| 91午夜精品亚洲一区二区三区| 成人二区视频| 菩萨蛮人人尽说江南好唐韦庄| 欧美97在线视频| 伦理电影免费视频| 免费大片黄手机在线观看| 人妻人人澡人人爽人人| 亚洲激情五月婷婷啪啪| 国产深夜福利视频在线观看| 免费久久久久久久精品成人欧美视频 | 国产成人免费观看mmmm| 国产av一区二区精品久久| 91在线精品国自产拍蜜月| 九九在线视频观看精品| 最近的中文字幕免费完整| 美女主播在线视频| 国产日韩欧美亚洲二区| 91久久精品国产一区二区三区| 久久久久久久久久久免费av| 国产片特级美女逼逼视频| 日本与韩国留学比较| 伊人久久国产一区二区| 成人特级av手机在线观看| 久久久久久久久久久免费av| 下体分泌物呈黄色| 哪个播放器可以免费观看大片| 亚洲美女搞黄在线观看| 欧美精品国产亚洲| 国产探花极品一区二区| 春色校园在线视频观看| 在线观看免费日韩欧美大片 | 国产高清不卡午夜福利| 最近2019中文字幕mv第一页| 精品国产乱码久久久久久小说| 日本91视频免费播放| 搡老乐熟女国产| 人人妻人人澡人人爽人人夜夜| 亚洲欧美一区二区三区国产| 久久精品久久久久久噜噜老黄| 成人综合一区亚洲| 嫩草影院新地址| 肉色欧美久久久久久久蜜桃| 一级黄片播放器| 久久精品久久久久久噜噜老黄| 大又大粗又爽又黄少妇毛片口| 一区二区三区四区激情视频| 亚洲国产精品一区二区三区在线| 妹子高潮喷水视频| 中国三级夫妇交换| 精品少妇黑人巨大在线播放| 欧美 亚洲 国产 日韩一| 亚洲欧美精品专区久久| 成人美女网站在线观看视频| 一级二级三级毛片免费看| 国产av精品麻豆| 韩国av在线不卡| av免费在线看不卡| 国产伦理片在线播放av一区| 亚洲精品国产av成人精品| 免费人妻精品一区二区三区视频| 国产精品麻豆人妻色哟哟久久| av天堂久久9| 99精国产麻豆久久婷婷| 麻豆乱淫一区二区| 国产成人免费观看mmmm| 欧美+日韩+精品| 大香蕉97超碰在线| 一区二区三区精品91| av黄色大香蕉| 美女xxoo啪啪120秒动态图| 国产熟女欧美一区二区| 啦啦啦啦在线视频资源| 亚洲自偷自拍三级| 亚洲欧美日韩东京热| 日日爽夜夜爽网站| 国产精品一区二区在线观看99| 99久久精品国产国产毛片| 日本-黄色视频高清免费观看| 精品一区二区三区视频在线| 久久99一区二区三区| 亚洲国产精品成人久久小说| 亚洲电影在线观看av| 在线 av 中文字幕| 2022亚洲国产成人精品| 18禁动态无遮挡网站| 亚洲av.av天堂| 多毛熟女@视频| 在线播放无遮挡| 免费观看无遮挡的男女| 国产av精品麻豆| a 毛片基地| 欧美日韩在线观看h| 久久精品国产a三级三级三级| 中国三级夫妇交换| a级一级毛片免费在线观看| 视频中文字幕在线观看| 国产男女超爽视频在线观看| 中文乱码字字幕精品一区二区三区| 婷婷色麻豆天堂久久| 欧美成人精品欧美一级黄| 一级毛片 在线播放| 男女无遮挡免费网站观看| 国产综合精华液| 成人亚洲欧美一区二区av| 成年人午夜在线观看视频| 日本与韩国留学比较| 99久久人妻综合| 777米奇影视久久| av有码第一页| 99热这里只有是精品在线观看| 中文精品一卡2卡3卡4更新| 人人妻人人澡人人爽人人夜夜| 国产69精品久久久久777片| 日本猛色少妇xxxxx猛交久久| av天堂中文字幕网| 美女视频免费永久观看网站| 黑人猛操日本美女一级片| 久久人妻熟女aⅴ| 久久久久久久国产电影|