    Global mapping of live cell mechanical features using PeakForce QNM AFM

    2020-05-19 06:51:40JinrongHuShenbaoChenDandanHuangYanZhangShouqinLuMianLong
    Biophysics Reports 2020年1期

    Jinrong Hu, Shenbao Chen, Dandan Huang, Yan Zhang, Shouqin Lu¨?, Mian Long?

    1 Center of Biomechanics and Bioengineering, Key Laboratory of Microgravity (National Microgravity Laboratory),Beijing Key Laboratory of Engineered Construction and Mechanobiology,and CAS Center for Excellence in Complex System Mechanics, Institute of Mechanics, Chinese Academy of Sciences, Beijing 100190, China

    2 School of Engineering Science, University of Chinese Academy of Sciences, Beijing 100049, China

    Abstract Atomic force microscopy (AFM) has been widely used to acquire surface topography upon different scanning modes and to quantify mechanical properties of a cell using single-point ramp force mode.However,these traditional measurements need massive force curves originating from multiple points of a cell to exclude the potential errors resulted from limited and factitious selections of testing points,making the measurements time-consuming and highly localized. PeakForce Quantitative NanoMechanics (PF QNM) is a high-speed (faster by 3-4 order of magnitude) and global surface mechanical property mapping method with high spatial resolution, overcoming the drawbacks of traditional ramp mode especially used for a live cell with high heterogeneity.In this protocol,we elaborated how to run PF QNM measurements for live cells and relevant modification may be needed when extending this method to other cell-like soft materials.

    Keywords Global mechanical mapping, Live cell, Atomic force microscopy, PeakForce QNM

    INTRODUCTION

    Mechanical features of cells are significant to cellular behaviors and fate decision in many biological processes such as cell migration, growth, and differentiation responses to mechanical stimuli (Calzado-Mart?′n et al.2016; Geitmann and Ortega 2009; Qi et al. 2017; Wang et al.2014;Xu et al.2019).Thus,global and high-speed mechanical property mapping with high spatial resolution is important for understanding the dynamic behaviors of cells. Many approaches like magnetic tweezer, micropipette aspiration, cone plate viscometer and atomic force microscopy(AFM)are optional for cell mechanical measurements with respective advantages(Long et al.2011;Rodriguez et al.2013).AFM has been used widely with its mature operation and analysis system in the following two aspects of applications.Firstly, it has been used to acquire material surface topography, chemical and electrical properties upon different scanning modes (Barbee et al. 1994; Maivald et al. 1991; Zhong et al. 1993) and special probes(Rawlett et al. 2002; Sugimoto et al. 2007). Secondly,single-molecule force spectrum AFM upon ramp mode empowers us to investigate protein folding/unfolding and ligand-receptor interactions (Le et al. 2017; Marshall et al. 2003). In addition, localized mechanical properties of a cell or cells can also be extracted from corresponding force-distance curves of special sites based on contact mechanics theories (Sneddon 1965).Nowadays,AFM achieves its highest resolutions to subnanometer of scanning and picoNewton (pN) of force,makes it possible to resolve even atomic arrangement and interactions (Sugimoto et al. 2007).

    With the development of AFM technology, more and more detecting modes are exploited. Taking mechanical property acquisition as an example, Force volume (FV)mode AFM performs mechanical mapping of a cell or cells through combining topography scanning with single-point force-distance curve arrays in the selected ranges(Hansma et al.1997;Radmacher et al.1994).FV mode is a basic mode for all AFM instruments with its robustness in force-distance curve acquisition but is time-consuming with low sampling speed (usually <10 Hz). Fast force volume (FFV) and quantitative imaging (QI) modes are thus developed based on FV mode for much faster force-distance curve acquiring.While FFV or QI mode can map global mechanical property of a cell or cells with the advance of sampling speed and the spatial resolution of global mechanical property, accurate force control is still limited in those modes. PF QNM (PeakForce Quantitative NanoMechanics) is another mode released by Bruker for fast mechanical mapping based on PF tapping mode. With better force control than FFV and QI modes, PF QNM mode enables more precise global mechanical mapping with high spatial resolution and speed(see details in the following section ‘PRINCIPLES’).

    Nowadays, PF QNM mode has been attracting attentions in biological issues upon its unique advantages.For example,it is well applied in characterizing dynamic changes of the elastic modulus of plant leaves (Qi et al.2017) as well as in extensive applications in plant biomechanics. In recent years, PF QNM mode has also been extended to map global surface topography and mechanical features of mammalian live cells (Pittenger and Slade 2013; Schillers et al. 2016), including isolating tumorigenic cells from normal cells(Calzado-Mart?′n et al. 2016) and investigating mechanosensing of neutrophils onto HUVEC monolayer pre-cultured on stiffness-varied substrates (Xu et al. 2019). Here we present all the detailed protocols from experimental performance to data analysis in running PF QNM measurements for live cells. Actually, this protocol can also be used to perform mechanical measurements for other soft materials with suitable settings.

    MATERIAL, EQUIPMENT AND SOFTWARE

    Cell preparation

    Human umbilical vascular endothelial cell line (HUVEC)purchased from ATCC (American Type Culture Collection. Manassas, VA, USA) was used to represent typical mechanical mapping of live cells.HUVECs were cultured in Medium 199 supplemented with 20%FBS,100 U/mL penicillin, 20 mmol/L HEPES (HyClone, Logan, Utah,USA), 3 μg/mL thymidine (Sigma-Aldrich, Shanghai,China),1 mmol/L L-glutamine(Sigma-Aldrich,Shanghai,China), 14 U/mL heparin sodium (Sigma-Aldrich,Shanghai, China), 25 μg/mL amphotericin B (Amresco,Solon, OH, USA), and 5 ng/mL bFGF (R&D Systems,Minneapolis, MN, USA). They were plated on collagen I(Sigma-Aldrich, Shanghai, China) coated glass bottom dishes to form a monolayer before using.

    Equipment

    · AFM: Bioscope Catalyst (Bruker, Billerica, MA, USA)with PF QNM module installed.

    · AFM probes: Pre-calibrated PFQNM-LC probes (Bruker, Billerica, MA, USA) with spring constant of~0.07 N/m were used for mechanical mapping. Tip parameters are in a height of 17 μm and a tip radius of 65 nm.For comparison,a spherical probe(NOVASCAN, Boone, IA, USA) with a nominal spring constant of 0.07 N/m and a tip radius of 2.5 μm and a pyramid MLCT-E probe (Bruker, Billerica, MA, USA) with a nominal spring constant of 0.03 N/m, a tip height of 2-8 μm and a tip radius of 20 nm were used for live cells in a conventional ramp mode.

    Software

    · NanoScope 8.15 (Bruker, Billerica, MA, USA): online software for running the measurements.

    · NanoScope analysis v1.80 (Bruker, Billerica, MA,USA): offline software for performing data analysis.

    · MATLAB (MathWorks, Natick, MA, USA): software for user-defined data analysis codes (available at https://github.com/chenshenbao/AFM_PeakForce_QNM_mapping_data_analysis).

    · Prism 6(GraphPad,San Diego,CA,USA):software for statistical analysis and graph drawing.

    PRINCIPLES

    Principle of AFM test

    Schematic of working principle of an AFM device is shown in Fig. 1A. A piezoelectric translator (PZT) is used to drive a probe,composed by an elastic cantilever and a tip, approaching to, contacting with, and retracting from the sample. The interaction between the sample and the tip is characterized by the product of the deformation and spring constant of cantilever, and the deformation of cantilever is quantified by the deflection of the reflected laser from its back through a quad photodetector (QPD). For surface topography scanning,different scanning modes determine distinct criterions for controlling PZT movement upon feedback loop. For example, the criterion of a contact scanning mode is to apply a constant force between the tip and the sample.Thus, the deflection of the cantilever on each scanning site is first adjusted to be equal to the pre-set setpoint through the Z-axis movement of PZT and then move to the next scanning site,where the Z-axis displacement of each site is recorded as the representation of surface height. For force-distance curve acquisition, the tip is driven to approach to,contact with,and retract from the sample surface along Z-axis upon the pre-setting of contact force, and both the approach and retraction curves are recorded for further analyses. AFM device adopts a classical PID(Proportional-Integral-Derivative)feedback system to adjust the movement errors and track the sample surface accurately (?stro¨m and Ha¨gglund 1995).

    Principle of PF QNM mode

    PF QNM is based on Bruker patented technology of PF tapping released in 2009 and the most dominant feature of this mode is high-speed surface mechanical mapping.In this mode, Z-axis tip engagement is driven by the sinusoidal oscillation of the PZT with high frequency of 0.25-8 kHz (Fig. 1B), resulting in faster force-distance curve acquisition by 3-4 order of magnitude than that of classical ramp mode.The sinusoidal oscillation of the PZT in PF tapping mode also differs in its tip engaging trajectory along Z-axis from that of ramp mode. As shown in Fig. 1C, the PZT in ramp mode drives the tip moving up and down in fixed velocity (red line),resulting in high approaching velocity of the tip to the surface. Alternatively, the PF tapping mode enables the tip contact the sample surface in the valley of sinusoidal oscillation with approaching velocity of zero (purple line), protecting the sample with less damage. Furthermore, the oscillation frequency of the cantilever in PF tapping mode is far lower than the resonance frequency,keeping the stable tip engagement and the precise control of the indentation force (Fig. 1D). Above features of PF QNM enable this technology to be used for global mechanical mapping of a cell or cells with similar spatial resolution with that of surface topology scanning.

    Young’s modulus fitting

    Different theoretical models are developed for extracting mechanical property of the sample from AFM forcedistance curve.Sneddon model (Sneddon 1965)is used mostly to fit the curve for getting Young’s modulus E upon pyramid tip (Fig. 1E):

    wherein F denotes interaction force between tip and sample,ν is Poisson’s ratio,θ defines half angle of the tip and δ is indentation depth. Hertz model is used for fitting when a spherical probe is chosen:

    wherein R denotes the tip radius. Additionally, Hertzbased Derjaguin-Muller-Toropov (DMT) model is preferred when adhesion forces between the tip and the specimen are taken into consideration (Derjaguin et al.1975):

    wherein α denotes the contact radius and Fadhis the adhesion force (Fadh=-2πRΔγ and γ is the surface energy). As spherical probes are congenitally unsuitable to resolve the detailed surface topography and perform global mechanical mapping with poor spatial resolution, pyramid probes are used to run PF QNM measurements and Sneddon model is adopted to implement force-distance curve fitting in this protocol.Using PF QNM mode can collect several tens of thousands of force-distance curves for each run, as the density plot shown in Fig. 1F, and Young’s modulus for each point is extracted by fitting corresponding forcedistance curves. It should be noted that the exact number of force-distance curves collected depends on the scanning parameter settings of scan rate, sample/line and PeakForce frequency.

    Representative surface topography, 3D height image and height with E skin image of HUVECs were shown in Fig. 1G-I, respectively. Actin stress fibers at cell peripheries were easy to be discerned, and actin cap,namely stress fibers over the top of the nucleus, could also be found from the surface topography (Curry et al.2017; Efremov et al. 2018), even though fluorescence labeling is usually required to confirm the existence of actin bundles. The mechanical distributions of the whole scanning range were also visualized in E skin image with larger E values in the cell periphery and the cell-cell junctions than those of cell bodies. Thus, PF QNM mapping makes it possible to quantify the differences in mechanical properties of distinct cell regions with the same spatial resolution of scanning,which may be helpful for exploring the elaborate mechanical features at subcellular level especially for the highly heterogeneous cell.

    PROTOCOL

    The summary of the entire procedure is listed in Fig. 2.

    Cell preparation

    Firm adhesion of cells to substrate is necessary for performing PF QNM mapping. Generally, cells are seeded on the substrate and cultured at least overnight onto the dishes to ensure the firm adhesion. In addition to special culture medium for specific cell types, poly-L-lysine (PLL)-coated glass bottom dishes are also preferred for cell adhesion. The cells are washed 2-3 times using pre-warmed DPBS (or any other buffer suitable or even culture mediums) before AFM imaging for reducing tip contamination and laser interference.

    Equipment pre-settings

    (1) Turn on the computer and controller →(2) Mount the probe on the holder and mount the holder onto the detector →(3) Open the online software NanoScope 8.15 (or other versions) and choose PF QNM mode →(4) Initialize the stage →(5) Align the laser to the suitable site of back of the cantilever for a maximum laser SUM value →(6) Adjust the laser point. Technically, it is favorable to adjust PeakForce frequency to 0.5 kHz (or even lower to 0.25 kHz) for getting stable force-distance curves and set PeakForce amplitude to 300 nm for withdrawing the probe sufficiently from the sticky cell surface in each test circle.

    Probe calibration

    Different probes possess distinct inherent properties such as spring constant k and deflection sensitivity(DS).To ensure the consistency of measurements, the probe must be calibrated in situ before test even though the nominal spring constant is given.There are two kinds of relative and absolute methods for probe calibrating.The relative calibration depends on standard sample with known Young’s modulus, and the mechanical property of the sample to be tested should be comparable with that of standard sample. In general, the relative calibration method is suitable for measuring hard materials. Since the live cell is soft, the alternative absolute method is suitable for calibrating the probes with following steps:(1)Mount the cell dish onto the stage and then assemble the detector to the stage. (2) After aligning the laser point, calibrate DS first by indenting the probe to an extremely hard sample surface (like glass or sapphire) and then go to thermal tune panel(calibrate →thermal tune bottom for Bruker device)for k calibration. It should be noted that the laser point may not be stable with leaping around at beginning because of the thermal gradient between the probe and the culture medium,resulting that the calibration needs to be performed after the laser drifting through 5-10 min waiting.And the operator may also lower the probe closer to the sample surface before calibration using navigation in the control panel to save time in the followed engagement.To image the live cells stably,it is highly recommended to choose PFQNM-LC probe with pre-calibrated spring constant k.

    When using PFQNM-LC probes, DS can be backcalibrated using thermal tune based on energy equipartition theorem. As the probe can be treated as a harmonic oscillator, the mean kinetic energy of each degree of freedom(DOF)of the cantilever equals kBT/2,wherein kBis the Boltzmann constant and T is the thermodynamic temperature.Since the potential energy of the entire cantilever is proportional to the kinetic energy of each DOF, we can derive approximately,

    wherein x denotes the deformation of the cantilever.Supposing that one applies two runs of thermal tune using the same probe and the first run has false DS1then gives wrong k1but the second run has real DS2and gives real k2,the energy in each run is still the same and yields,

    in which x=DS·DF with DS in nm/V and DF (deflection) in V. Then we have:

    Substitute Eq. 6 into Eq. 5 and consider the same probe with same driving force, the deflection of the cantilever should be the same as DF1= DF2and then reads:

    Prior to the measurements, the operator can set real value of k as k2in thermal tune panel, and then assign an arbitrary/false DS as DS1. After running the thermal tune, one can get a fitted false k as k1and the real DS(namely DS2) of the cantilever can be obtained using Eq. 7. In addition, one can also use RS-15 M sample by Bruker(or any other roughness sample)to calibrate the tip radius before measurement. Constant-temperature incubator mounted on the stage may be used for those cells sensitive to temperature changes.

    Tip engagement

    Set the scan size to a small square region (mostly with the length of side ≤30 μm), scan rate to a small value(mostly <0.5 Hz), and sample/line to 128 × 128 or 256 × 256 pixels at first, collectively ensuring that the lateral tip velocity is set mostly ≤30 μm/s and the tip/sample is in less damage and less likely to crash.Then click engage on the control panel and the probe is forced to approach the cell surface. The operator can enhance the parameter of engage setpoint to help the probe find the surface easily. Here the default value of engage setpoint is 0.3 V and 0.1 V is always a good choice in practice. This is because the extremely higher engage setpoint value could damage the sample due to the larger contact force applied.When the tip is not able to find the surface, a false engage error is reported to indicate the quite low engage setpoint value.

    Parameter adjusting

    Once the tip finds the surface,scanning starts right after a short duration of device self-adjusting (about 20 s).Set the parameters of ScanAsyst auto control as individual and of ScanAsyst auto setpoint (SAS) as Off.Then adjust manually the tip-sample interaction force(in pN or nN),namely,SAS.Increase SAS value gradually to obtain well-defined force-distance curves (Fig. 1E),usually ranging from 200 pN to 1 nN for live cell mechanical imaging.Note that SAS value depends on the measurement site and depth, one may increase/decrease SAS value to capture the mechanical behaviors of deeper/shallower layer of the cell. Then adjust gradually scan size to the assigned size (mostly<100 μm on Catalyst).Note that the golden standard to judge if a SAS value is suitable is to monitor the quality of the force curves showing on the force monitor. Here this value is cell-type specific with a low value for flat fibroblasts but a high value for rugged HeLa cells and probe specific from one to another.In some cases,it is optional to manually tweak feedback gain for minimizing the peak force error and ensuring the tip track the surface correctly. Again, always keep the lateral tip velocity ≤30 μm/s to avoid tip crash error.Decrease scan rate to a lower value when a tip crash error reports.

    Scanning

    After completing the parameter settings, click frame down/up to initialize a target scanning. Make sure that the height, peak force error, Sneddon modulus and deformation channels are chosen. Notes: (1) Do not change any scanning parameters once the scanning starts to prevent the data recorded incorrectly; (2)When some errors like tip crash are encountered(without tip withdrawal), the operator can use capture now to capture the data manually.

    Data processing

    The data obtained from the online software are then analyzed using either the official offline software of NanoScope analysis or in-house programmed MTALAB codes (Fig. 3).

    Processing using NanoScope analysis (Fig. 3A-C).After opening the data file to be processed in the software, one can click the Sneddon modulus (Young’s modulus E fitting using Sneddon model) channel as shown in Fig. 3A. To clarify E distribution along an arbitrary cross-section, one can use section tool on the toolbar to draw the arbitrary horizontal lines(white line in Fig. 3A)for producing various cross-sections,and the typical E values along the section are shown in Fig. 3B.This is a convenient way for users to exhibit quantitatively E distributions of a cell with different structures or regions. For example, the E values along cell-cell junctions (blue arrow) are much higher than those of stress fibers (cyan arrow) (Fig. 1I), consistent with the feature that cell-cell junctions are stiffer than the cell body or stress fibers across the nucleus (Ohashi et al.2002; Vargas-Pinto et al. 2013). It is noticed that the nominal stiffness of the nucleus measured in this protocol is the stiffness of the cytoplasm/cortex right above the nucleus with the indentation depth of about 400 nm, but not that of the nucleus itself. The actual stiffness of the nucleus should be measured upon deeper indentation even penetrating cell membrane and touching nucleus membrane directly (Liu et al. 2014).Also, one can use particle analysis tool to do the statistics of E in freely-drawn boxed regions (Fig. 3A,white dashed box). The software runs the statistics of each data points (one point one E value) inside the box and presents the results in frequency distributions in Fig. 3C,which can be used to calculate the mean E value for the region-of-interest (ROI) (Qi et al. 2017). In this case, one should be careful to select the parameter Xaxis in statistical histogram of E inside the box, where absolute but not relative must be chosen for correct data export (for Bruker offline processing software NanoScope analysis only).

    Processing using MATLAB (Fig. 3D-F). While the official offline software works well for conventional analyses and data extraction,only regular rectangular or square boxes can be drawn using the software and this limitation confines the elaborative analyses of any other ROIs with irregular boundary. Thus, we programmed MATLAB codes to pick out each E value in amorphous or closed regions (Fig. 3D) and to circle out the ROIs for extracting corresponding E values for further analyses(Xu et al. 2019). As an example, the two circled white contours in Fig. 3D defined cell body and junction regions and the corresponding statistics of E distributions of these two regions were shown in Fig. 3E and F,respectively. Apparently, the junction region is much stiffer with right shifted E distribution and higher mean value of 18.4 kPa than those of the cell body. Also, the mean E of 7.6 kPa in the cell body derived from MATLAB codes(Fig. 3E)was in consistence with that of 8.0 kPa from the commercial offline software (Fig. 3B),embodying the reliability of the in-house developed MATLAB programs.

    DISCUSSION

    In this protocol, we elaborate how to run mechanical mapping of live cells using AFM in PF QNM mode. This method is practical as it empowers us to acquire global mechanical properties of a whole cell in high-speed with the same spatial resolution as that of surface topology scanning (Qi et al. 2017; Xu et al. 2019). Since the PF QNM mode can capture surface topography and mechanical properties synchronously, it can be used to correlate surface features with their mechanical properties in situ as shown in Fig. 3. The entire profile of surface mechanical properties in high spatial resolution also furthers our understanding on biological issues like cell transmigration, cell repolarization and cell mechanosensing under mechanical stimuli (Livne et al.2014;Martinelli et al.2014;Roduit et al.2008;Schaefer and Hordijk 2015). This is especially meaningful by considering the heterogeneity of a cell with cell membrane, nucleus or nuclei, and numerous subcellular organelles. In this case, only one value of Young’s modulus at a specific site of a cell is not able to represent the global mechanical features of that cell.

    The impacts of probe geometry on mechanical measurements should not be overlooked. In general,spherical probes are more stable than pyramid probes(Fig. 4). For pyramid probes, it is difficult to estimate precisely the parameters of tip-sample contact area,non-linear elastic responses and probe geometry calibration (like tip radius), which are important for E fitting based on existing theories (Galluzzi et al. 2016).The paradox is that PF QNM mapping mode should use pyramid probes for high-resolution scanning. Thus, it needs to be emphasized that the acquired modulus data from PF QNM mode are contingent on probe geometries,fitting models, and scanning parameters. Several types of probes, like PFQNM-LC probes used in this protocol,are designed specifically for live cell mapping, and the main advantage is that this kind of probe suffers less hydrodynamic interference from high frequency excitation so as to favor more stable mechanical scanning(Schillers et al. 2016). Even using three distinct probes,the relative patterns of mechanical properties of HUVEC cultured on stiffness-varied substrates were similar,regardless of the differences in absolute E values(Fig. 4).

    In spite of the advantages in global and high-speed mechanical mapping of cell surface with high spatial resolution compared to the conventional ramp mode,there are several open issues to be clarified when applying PF QNM mode. The first issue is the high frequency excitation-induced cellular responses(HFEICR). Cellular viscoelasticity emerges obviously when high frequency excitation is exerted on a cell,presenting the stress-or strain-induced cell stiffening or softening at a high frequency excitation (Chaudhuri et al.2007;Gardel et al.2004;Schmoller et al.2010).It is noted that cellular stiffness upon PF QNM measurement is at most 100-fold higher than that upon the single-point ramp mode (Calzado-Mart?′n et al.2016).This huge difference can be minimized using the proper parameter settings for specific cell type.Actually,only one-fold increase of E was found in HUVECs in this protocol than those using similar pyramid probes under ramp mode (Fig. 4), attributed to the application of the lower PF frequency setting of 0.25 kHz and the usage of a PFQNM-LC probe.

    Second issue arises from the bottom effect of cell culture substrate. No matter which type of probes/modes is adopted, most of culture substrates yields high stiffness, resulting in a rule of indentation depth/specimen thickness <0.1 when applying AFM indentation to soft material. To eliminate the potential artifacts stemming from the substrates, Bottom Effect Correction (BEC) was proposed theoretically (Gavara and Chadwick 2012), as a correction to Sneddon model when facing soft materials with pyramid or conical probes,

    wherein θ is half-opening angle of the tip, δ is indentation depth and h is specimen thickness. After obtaining cell height (Fig. 5A, pixel-to-pixel h) and indentation depth (Fig. 5B, pixel-to-pixel δ) in PF QNM mapping, pixel-to-pixel BEC was done (Fig. 5C) by substituting pixel-to-pixel δ/h into Eq. 8 and the actual modulus distribution was obtained from the correction of pixel-to-pixel bottom effect. These typical results showed that BEC reduces those overestimated moduli at cell periphery. Here BEC was based on the accurate calculation of regional thickness (height) when glass bottom was used as zero plane for calibration of height.Evidently, the glass bottom calibration also limits the application of this method especially for fully spread,thin cells.In fact,only small fraction of the glass bottom or substrate can be imaged clearly when the cells grow up to high confluence, which could lead to invalidity of height calibration. In this case, AFM device combined with in situ confocal microscopy to monitor pixelto-pixel sample thickness might be a better choice to help determine the cell height with sliced confocal images.Alternatively,the bottom effect is less significant or even neglected when cells are seeded on soft substrate like PA gels with similar cellular stiffness and sufficient thickness,which will not cause large deviation(Gavara and Chadwick 2012).

    Acknowledgements This work was supported by National Key Research and Development Program of China(2016YFA0501601),National Natural Science Foundation of China (31627804,91642203, 31661143044, and 31570942), Frontier Science Key Project of Chinese Science Academy (QYZDJ-SSW-JSC018), and Strategic Priority Research Program of Chinese Academy of Sciences (XDB22040101).

    Compliance with Ethical Standards

    Conflict of interest Jinrong Hu, Shenbao Chen, Dandan Huang,Yan Zhang, Shouqin Lu¨, and Mian Long declare that they have no conflict of interest.

    Human and animal rights and informed consent This article does not contain any studies with human or animal subjects performed by any of the authors.

    Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing,adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s)and the source,provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

    国产一区二区在线av高清观看| 99在线人妻在线中文字幕| 永久网站在线| 亚洲欧美日韩东京热| 老司机深夜福利视频在线观看| 搡老熟女国产l中国老女人| 免费在线观看日本一区| 一本精品99久久精品77| 免费黄网站久久成人精品| 亚洲久久久久久中文字幕| 老司机午夜福利在线观看视频| 美女高潮的动态| 国产精华一区二区三区| 女人十人毛片免费观看3o分钟| 制服丝袜大香蕉在线| 18禁在线播放成人免费| 一个人免费在线观看电影| 九九热线精品视视频播放| 五月玫瑰六月丁香| 十八禁网站免费在线| 久久精品久久久久久噜噜老黄 | 亚洲av五月六月丁香网| 国产免费男女视频| 日韩精品中文字幕看吧| 婷婷精品国产亚洲av| 亚洲成人免费电影在线观看| 国产一级毛片七仙女欲春2| 亚洲成人免费电影在线观看| 国产在线精品亚洲第一网站| 国内久久婷婷六月综合欲色啪| 国产aⅴ精品一区二区三区波| 简卡轻食公司| 免费搜索国产男女视频| 日日撸夜夜添| 国产亚洲91精品色在线| 久久久成人免费电影| 美女黄网站色视频| 波多野结衣高清作品| 久久99热这里只有精品18| 免费电影在线观看免费观看| АⅤ资源中文在线天堂| 露出奶头的视频| 亚洲欧美日韩卡通动漫| 一本久久中文字幕| 国产欧美日韩一区二区精品| 国产成人av教育| 欧美丝袜亚洲另类 | 日韩欧美国产一区二区入口| 美女免费视频网站| 深夜a级毛片| 婷婷色综合大香蕉| 高清日韩中文字幕在线| 国产亚洲91精品色在线| 听说在线观看完整版免费高清| 我要看日韩黄色一级片| 听说在线观看完整版免费高清| 性插视频无遮挡在线免费观看| 日韩亚洲欧美综合| 午夜福利视频1000在线观看| 亚洲精品影视一区二区三区av| 禁无遮挡网站| 最近在线观看免费完整版| 国内精品宾馆在线| 精品一区二区三区视频在线| 午夜福利在线观看吧| 国内精品宾馆在线| 中文字幕精品亚洲无线码一区| 久久精品国产亚洲网站| 精品久久久久久成人av| 男女那种视频在线观看| 99精品在免费线老司机午夜| 免费一级毛片在线播放高清视频| 联通29元200g的流量卡| 日日夜夜操网爽| 免费一级毛片在线播放高清视频| 天天一区二区日本电影三级| 国语自产精品视频在线第100页| av.在线天堂| 亚洲成人久久性| 男人舔奶头视频| 精品久久久久久,| 国产伦精品一区二区三区视频9| 亚洲av免费在线观看| 欧美性猛交黑人性爽| 偷拍熟女少妇极品色| 乱人视频在线观看| 国产一区二区在线av高清观看| 舔av片在线| 在线观看午夜福利视频| 成人无遮挡网站| 午夜亚洲福利在线播放| 日日干狠狠操夜夜爽| 美女cb高潮喷水在线观看| 九色成人免费人妻av| 99热只有精品国产| 久久中文看片网| 97超视频在线观看视频| 国产私拍福利视频在线观看| 亚洲自偷自拍三级| 国产精品精品国产色婷婷| 校园人妻丝袜中文字幕| 一边摸一边抽搐一进一小说| 国产成人aa在线观看| 中出人妻视频一区二区| 国产大屁股一区二区在线视频| 国产激情偷乱视频一区二区| 亚洲最大成人中文| 精品福利观看| 国产高潮美女av| 99久久精品热视频| 日本免费a在线| 高清毛片免费观看视频网站| 久99久视频精品免费| 亚洲精品456在线播放app | 久久精品久久久久久噜噜老黄 | 亚洲精华国产精华精| 久久久国产成人免费| 人妻久久中文字幕网| 白带黄色成豆腐渣| 男女做爰动态图高潮gif福利片| 特级一级黄色大片| 成年版毛片免费区| 动漫黄色视频在线观看| 亚洲成av人片在线播放无| 看免费成人av毛片| 淫妇啪啪啪对白视频| 成人精品一区二区免费| 久久久久性生活片| 成人鲁丝片一二三区免费| 国产在线精品亚洲第一网站| 欧美潮喷喷水| 999久久久精品免费观看国产| 在线观看一区二区三区| 麻豆国产97在线/欧美| videossex国产| 校园春色视频在线观看| 国产亚洲精品久久久久久毛片| 深夜a级毛片| 亚洲熟妇中文字幕五十中出| 日本一二三区视频观看| 国产亚洲精品综合一区在线观看| 欧美日本亚洲视频在线播放| 色播亚洲综合网| 成熟少妇高潮喷水视频| 国产精品一区二区性色av| 久久精品影院6| 性色avwww在线观看| 亚洲av成人精品一区久久| 亚洲一区二区三区色噜噜| 亚洲熟妇中文字幕五十中出| 在线免费观看不下载黄p国产 | 亚洲av第一区精品v没综合| 美女高潮的动态| 日韩,欧美,国产一区二区三区 | 午夜福利高清视频| 色综合色国产| 国产高潮美女av| 亚洲精品亚洲一区二区| 自拍偷自拍亚洲精品老妇| 色吧在线观看| 深夜精品福利| 亚洲狠狠婷婷综合久久图片| 国产一区二区激情短视频| 国产黄色小视频在线观看| 少妇猛男粗大的猛烈进出视频 | 精品无人区乱码1区二区| 1024手机看黄色片| 久99久视频精品免费| 欧美xxxx性猛交bbbb| 成人av在线播放网站| 成人亚洲精品av一区二区| 日日夜夜操网爽| 欧美精品啪啪一区二区三区| 国产三级中文精品| 亚洲自偷自拍三级| 成年女人永久免费观看视频| 国产三级在线视频| 久久久久久久久久成人| 欧美+日韩+精品| 成人毛片a级毛片在线播放| 亚洲av.av天堂| 偷拍熟女少妇极品色| 欧美xxxx黑人xx丫x性爽| 免费看日本二区| 日韩一区二区视频免费看| 黄色日韩在线| 婷婷色综合大香蕉| 搡老妇女老女人老熟妇| 久99久视频精品免费| 真实男女啪啪啪动态图| 久久精品国产亚洲av涩爱 | 久久久久性生活片| 九色国产91popny在线| 欧美日韩瑟瑟在线播放| 欧美又色又爽又黄视频| 在线免费观看的www视频| 国产伦人伦偷精品视频| 日本一二三区视频观看| 亚洲av不卡在线观看| 国产精品电影一区二区三区| 天天一区二区日本电影三级| 日韩中文字幕欧美一区二区| 亚洲成人免费电影在线观看| 亚洲av中文av极速乱 | 国产成人av教育| 欧洲精品卡2卡3卡4卡5卡区| 日韩大尺度精品在线看网址| 欧美高清成人免费视频www| 色5月婷婷丁香| 女同久久另类99精品国产91| 国产女主播在线喷水免费视频网站 | 人妻丰满熟妇av一区二区三区| 久久久精品大字幕| 国产精品1区2区在线观看.| 无遮挡黄片免费观看| 国国产精品蜜臀av免费| 成年人黄色毛片网站| 亚洲人成伊人成综合网2020| 日韩 亚洲 欧美在线| 韩国av在线不卡| 最后的刺客免费高清国语| 天天躁日日操中文字幕| 啦啦啦啦在线视频资源| 少妇的逼水好多| 动漫黄色视频在线观看| 久久这里只有精品中国| 亚洲七黄色美女视频| 久久草成人影院| 午夜日韩欧美国产| 日本-黄色视频高清免费观看| 一区福利在线观看| 欧美成人性av电影在线观看| 韩国av在线不卡| 日韩大尺度精品在线看网址| 全区人妻精品视频| 国产精品,欧美在线| 日韩欧美精品免费久久| 精品久久久久久成人av| 日韩国内少妇激情av| 国产高清激情床上av| 日韩av在线大香蕉| 免费看av在线观看网站| 欧美潮喷喷水| 欧美一区二区精品小视频在线| 免费不卡的大黄色大毛片视频在线观看 | 91精品国产九色| 在线国产一区二区在线| 欧美日本亚洲视频在线播放| 日本a在线网址| 免费看av在线观看网站| 婷婷亚洲欧美| 在线观看一区二区三区| 女人十人毛片免费观看3o分钟| 色av中文字幕| 亚洲av二区三区四区| 美女大奶头视频| 亚洲aⅴ乱码一区二区在线播放| 极品教师在线视频| 国语自产精品视频在线第100页| 婷婷六月久久综合丁香| 成人午夜高清在线视频| 国产精品永久免费网站| 99久久九九国产精品国产免费| www.色视频.com| 男插女下体视频免费在线播放| 国产精品一区二区免费欧美| 99久久久亚洲精品蜜臀av| 亚洲av一区综合| 亚洲人成网站在线播| 亚洲经典国产精华液单| 久久精品久久久久久噜噜老黄 | 成人av在线播放网站| 日日摸夜夜添夜夜添小说| 亚洲av中文av极速乱 | 日本熟妇午夜| 精品人妻视频免费看| 男女视频在线观看网站免费| 国产久久久一区二区三区| 国产白丝娇喘喷水9色精品| 少妇丰满av| 国产av不卡久久| 99九九线精品视频在线观看视频| 免费看日本二区| 欧美人与善性xxx| 国产精品福利在线免费观看| 哪里可以看免费的av片| 热99re8久久精品国产| netflix在线观看网站| av天堂在线播放| 免费大片18禁| 亚洲久久久久久中文字幕| 亚洲精品一卡2卡三卡4卡5卡| 日本 av在线| 99久久中文字幕三级久久日本| 最近在线观看免费完整版| 成人综合一区亚洲| 成人毛片a级毛片在线播放| 嫩草影院精品99| 久久99热6这里只有精品| 欧美激情国产日韩精品一区| 日韩欧美精品v在线| 最近视频中文字幕2019在线8| aaaaa片日本免费| 国产伦在线观看视频一区| 国产乱人视频| 国产黄a三级三级三级人| 免费在线观看成人毛片| 国产成人一区二区在线| 日本五十路高清| 欧美日韩乱码在线| 极品教师在线视频| 精品福利观看| 搡老岳熟女国产| 日韩欧美免费精品| 亚洲久久久久久中文字幕| 欧美日韩综合久久久久久 | 欧美成人a在线观看| xxxwww97欧美| 成人国产综合亚洲| 国产高清激情床上av| 日韩欧美精品免费久久| 观看免费一级毛片| 22中文网久久字幕| 国产一区二区三区在线臀色熟女| 麻豆精品久久久久久蜜桃| 日本熟妇午夜| 欧美日韩瑟瑟在线播放| 日本精品一区二区三区蜜桃| 久久午夜亚洲精品久久| 国内精品久久久久精免费| 日韩欧美 国产精品| 亚洲熟妇熟女久久| 成人性生交大片免费视频hd| 真人做人爱边吃奶动态| 午夜影院日韩av| 超碰av人人做人人爽久久| 蜜桃亚洲精品一区二区三区| 内射极品少妇av片p| 婷婷色综合大香蕉| 真人一进一出gif抽搐免费| 人人妻人人看人人澡| 久久久成人免费电影| 最近最新免费中文字幕在线| 国产又黄又爽又无遮挡在线| 日本-黄色视频高清免费观看| 舔av片在线| 亚洲一区高清亚洲精品| 国语自产精品视频在线第100页| 精品一区二区免费观看| 可以在线观看的亚洲视频| 欧美人与善性xxx| 久久婷婷人人爽人人干人人爱| 国产精华一区二区三区| 久久久久久久久中文| 3wmmmm亚洲av在线观看| 精品一区二区三区视频在线| 天堂av国产一区二区熟女人妻| 亚洲成人中文字幕在线播放| 18禁裸乳无遮挡免费网站照片| 波多野结衣巨乳人妻| 亚洲精品亚洲一区二区| 听说在线观看完整版免费高清| 97超级碰碰碰精品色视频在线观看| 欧美国产日韩亚洲一区| 夜夜看夜夜爽夜夜摸| 99在线人妻在线中文字幕| 欧美中文日本在线观看视频| 麻豆成人av在线观看| 十八禁网站免费在线| 啦啦啦啦在线视频资源| 国产乱人视频| 18禁裸乳无遮挡免费网站照片| 搡老熟女国产l中国老女人| 一进一出好大好爽视频| 久久精品国产99精品国产亚洲性色| 欧美日韩黄片免| 午夜亚洲福利在线播放| 国产色婷婷99| 18禁黄网站禁片免费观看直播| 国产精品美女特级片免费视频播放器| 亚洲专区国产一区二区| 欧美一区二区国产精品久久精品| 乱系列少妇在线播放| 国产在线精品亚洲第一网站| 一区二区三区激情视频| 99国产极品粉嫩在线观看| 99riav亚洲国产免费| 欧美潮喷喷水| 亚洲电影在线观看av| 丰满人妻一区二区三区视频av| 午夜免费激情av| 久久久久精品国产欧美久久久| 欧美一区二区精品小视频在线| 深爱激情五月婷婷| 18禁黄网站禁片午夜丰满| 中国美白少妇内射xxxbb| 免费不卡的大黄色大毛片视频在线观看 | 国产精品亚洲美女久久久| 国产私拍福利视频在线观看| 很黄的视频免费| 亚洲一区二区三区色噜噜| 69av精品久久久久久| 欧美最黄视频在线播放免费| 久久精品夜夜夜夜夜久久蜜豆| 在线免费十八禁| 在线a可以看的网站| 少妇裸体淫交视频免费看高清| 成人美女网站在线观看视频| 久久精品夜夜夜夜夜久久蜜豆| 日本撒尿小便嘘嘘汇集6| 国产老妇女一区| 男女之事视频高清在线观看| 免费一级毛片在线播放高清视频| 亚洲美女视频黄频| 女人被狂操c到高潮| 1000部很黄的大片| 色精品久久人妻99蜜桃| 尾随美女入室| 国产国拍精品亚洲av在线观看| 国产久久久一区二区三区| 精品人妻1区二区| 久久久久九九精品影院| 国产精品乱码一区二三区的特点| 99热只有精品国产| 内地一区二区视频在线| 亚洲精品亚洲一区二区| 亚洲av第一区精品v没综合| 中亚洲国语对白在线视频| av在线亚洲专区| 亚洲国产欧洲综合997久久,| 久久精品国产鲁丝片午夜精品 | 69人妻影院| 国产精品野战在线观看| 在线免费观看的www视频| 九九爱精品视频在线观看| 久久久午夜欧美精品| 精品一区二区三区视频在线| 亚洲成人免费电影在线观看| 免费人成视频x8x8入口观看| 91在线观看av| 亚洲熟妇熟女久久| 国产综合懂色| 久久精品国产亚洲av香蕉五月| 91精品国产九色| 欧美日本亚洲视频在线播放| 亚洲成人久久爱视频| 夜夜看夜夜爽夜夜摸| 精品久久久久久久末码| 国产免费男女视频| 欧美性猛交╳xxx乱大交人| 亚洲三级黄色毛片| 91麻豆精品激情在线观看国产| 欧美另类亚洲清纯唯美| 国产一区二区三区av在线 | 亚洲美女黄片视频| 欧美xxxx性猛交bbbb| 少妇人妻一区二区三区视频| 中文字幕免费在线视频6| 搡女人真爽免费视频火全软件 | 欧美高清成人免费视频www| 最近中文字幕高清免费大全6 | 一个人免费在线观看电影| 亚洲av五月六月丁香网| 成人无遮挡网站| 亚洲一区二区三区色噜噜| 免费在线观看成人毛片| 久久这里只有精品中国| 日日夜夜操网爽| 乱人视频在线观看| 免费av不卡在线播放| 别揉我奶头~嗯~啊~动态视频| 国产精品久久久久久亚洲av鲁大| 男人狂女人下面高潮的视频| 亚洲美女视频黄频| 国产av麻豆久久久久久久| 床上黄色一级片| 久久久久久久久久成人| 一个人观看的视频www高清免费观看| 亚洲第一区二区三区不卡| 国产精品美女特级片免费视频播放器| 一级a爱片免费观看的视频| 国产大屁股一区二区在线视频| 99在线人妻在线中文字幕| 国产伦在线观看视频一区| 久久天躁狠狠躁夜夜2o2o| 色哟哟哟哟哟哟| 最新在线观看一区二区三区| 国产精品一及| 赤兔流量卡办理| 亚洲无线在线观看| 日日摸夜夜添夜夜添小说| 亚洲av成人av| 别揉我奶头 嗯啊视频| 国产亚洲精品综合一区在线观看| 国产精品日韩av在线免费观看| 赤兔流量卡办理| 免费观看人在逋| 日韩精品中文字幕看吧| 日本爱情动作片www.在线观看 | 少妇的逼水好多| 成人特级av手机在线观看| 午夜日韩欧美国产| 超碰av人人做人人爽久久| 国产免费男女视频| 一个人免费在线观看电影| 午夜视频国产福利| 国产黄a三级三级三级人| 成人美女网站在线观看视频| 亚洲熟妇中文字幕五十中出| 亚洲欧美日韩高清在线视频| 国产精品一及| av在线老鸭窝| 国产蜜桃级精品一区二区三区| 亚洲精品国产成人久久av| 亚洲中文日韩欧美视频| 综合色av麻豆| 伊人久久精品亚洲午夜| 69av精品久久久久久| 变态另类成人亚洲欧美熟女| 国产黄色小视频在线观看| 熟妇人妻久久中文字幕3abv| av视频在线观看入口| 九九在线视频观看精品| 午夜老司机福利剧场| 国产极品精品免费视频能看的| 久久久久国内视频| 午夜福利在线观看免费完整高清在 | 狂野欧美白嫩少妇大欣赏| 久久香蕉精品热| 午夜精品久久久久久毛片777| 又粗又爽又猛毛片免费看| 在线免费观看不下载黄p国产 | 午夜爱爱视频在线播放| 国产一区二区三区视频了| 亚洲最大成人手机在线| 亚洲色图av天堂| 欧美zozozo另类| 国产伦一二天堂av在线观看| 国产精品爽爽va在线观看网站| 日韩一区二区视频免费看| 亚洲熟妇熟女久久| av专区在线播放| 99久久中文字幕三级久久日本| 老女人水多毛片| 国产91精品成人一区二区三区| 精品99又大又爽又粗少妇毛片 | 如何舔出高潮| 成人三级黄色视频| 日本黄色片子视频| 精品人妻1区二区| 狂野欧美白嫩少妇大欣赏| 神马国产精品三级电影在线观看| 韩国av在线不卡| eeuss影院久久| 亚洲三级黄色毛片| 男女视频在线观看网站免费| 大型黄色视频在线免费观看| 亚洲久久久久久中文字幕| 亚洲无线在线观看| 亚洲中文字幕日韩| 如何舔出高潮| 国国产精品蜜臀av免费| 国产精品98久久久久久宅男小说| 国产麻豆成人av免费视频| 国产精品久久久久久精品电影| 久久久久久久久中文| 99riav亚洲国产免费| 亚洲欧美日韩无卡精品| 亚洲在线观看片| 免费人成视频x8x8入口观看| 国产精品久久久久久亚洲av鲁大| 国产色爽女视频免费观看| 超碰av人人做人人爽久久| 麻豆久久精品国产亚洲av| 亚洲av不卡在线观看| 色综合站精品国产| 国产高清不卡午夜福利| 麻豆精品久久久久久蜜桃| 国产高清视频在线观看网站| 少妇熟女aⅴ在线视频| 日韩 亚洲 欧美在线| 久久久成人免费电影| 国产伦在线观看视频一区| 久久久国产成人免费| 国内揄拍国产精品人妻在线| 中文字幕高清在线视频| 欧美xxxx黑人xx丫x性爽| av在线观看视频网站免费| 又黄又爽又刺激的免费视频.| 岛国在线免费视频观看| 在线观看av片永久免费下载| 国内久久婷婷六月综合欲色啪| 超碰av人人做人人爽久久| 一区二区三区高清视频在线| 精品久久久久久成人av| 国产精品精品国产色婷婷| 我的老师免费观看完整版| 亚洲人成网站在线播| 国产精品精品国产色婷婷| 国产免费一级a男人的天堂| 中文字幕免费在线视频6| 乱人视频在线观看| 国产免费一级a男人的天堂| 亚洲第一电影网av| 嫩草影院新地址| 亚洲av中文av极速乱 | 亚洲av一区综合| 三级国产精品欧美在线观看| 亚洲人成网站在线播| 俄罗斯特黄特色一大片| 国产午夜福利久久久久久| 亚洲av电影不卡..在线观看| 舔av片在线| 日本-黄色视频高清免费观看| 九九爱精品视频在线观看|