    Design, Characterization, and 3D Printing of Cardiovascular Stents with Zero Poisson’s Ratio in Longitudinal Deformation

    2021-12-25 10:03:00ChengjinWangLeiZhangYongcongFangWeiSun
    Engineering 2021年7期

    Chengjin Wang, Lei Zhang,*, Yongcong Fang, Wei Sun,d

    a Biomanufacturing Center, Department of Mechanical Engineering, Tsinghua University, Beijing 100084, China

    b Biomanufacturing and Rapid Forming Technology Key Laboratory of Beijing, Beijing 100084, China

    c‘‘Biomanufacturing and Engineering Living Systems” Innovation International Talents Base (111 Base), Beijing 100084, China

    d Department of Mechanical Engineering, Drexel University, Philadelphia, PA 19104, USA

    ABSTRACT

    Keywords:Additive manufacturing 3D printing Screw extrusion Cardiovascular stent Zero Poisson’s ratio Inherent drawbacks associated with drug-eluting stents have prompted the development of bioresorbable cardiovascular stents. Additive manufacturing (3-dimentional (3D) printing) has been widely applied in medical devices. In this study, we develop a novel screw extrusion-based 3D printing system with a new designed mini-screw extruder to fabricate stents. A stent with a zero Poisson’s ratio (ZPR)structure is designed, and a preliminary monofilament test is conducted to investigate appropriate fabrication parameters.3D-printed stents with different geometric structures are fabricated and analyzed by observation of the surface morphology.An evaluation of the mechanical properties and a preliminary biological evaluation of 3D-printed stents with different parameters are carried out.In conclusion,the screw extrusion-based 3D printing system shows potential for customizable stent fabrication.

    1. Introduction

    With the application of a polymer coating that contains and controls the release of antiproliferative drugs, drug-eluting stents(DES)have become the gold standard for interventional treatment of coronary artery disease to date [1–3]. However, the backbone materials of DES are non-degradable metal materials that will cage the vessel permanently,which may lead to several inherent drawbacks, such as late restenosis, stent thrombosis, and chronic inflammation, and may impede the restoration of physiological function [4,5].

    Problems associated with DES have prompted the development of bioresorbable stents(BRS),which provide transient support and then disappear;thus BRS can reduce the risk of late stent thrombosis and avoid prolonged dual antiplatelet therapy [6,7]. The conventional manufacturing process for metal stents is laser micro cutting, which has also been applied in the processing of BRS.Stepak et al. [8] investigated the influence of laser parameters on the quality and geometry of polylactide-based polymer stents.Guerra et al. [9] explored the feasibility of a 1.08 μm wavelength fibre laser to cut polycaprolactone (PCL) sheets, as a feasibility study for stents manufacturing. Nevertheless, laser processing is a thermal process that may lead to thermal damage, such as heat-affected zones, micro-cracks, tensile residuals, and dross deposition [10,11]. More importantly, laser cutting, as a kind of subtractive manufacturing method,is not suitable for personalized customization.

    It has been observed that additive manufacturing (also known as 3-dimentional (3D) printing) has evolved into a versatile technology and economical solution for medical applications and customizable fabrication. Kaesemeyer et al. [12] used a rapid stent fabrication system with an extrusion system and a four-axis motion system to fabricate bioresorbable polystatin stents comprising lactide, glycolide, caprolactone, and lovastatin(60:15:10:15 parts by weight).Park et al.[13]evaluated the properties of a 3D-printed drug-coated PCL stent that was implanted into the porcine femoral artery.Guerra et al.[10]presented a novel 3D additive manufacturing machine to produce BRS with PCL filament material,and investigated the effects of the process parameters on the stent’s features.Wu et al.[14]manufactured polylactic acid (PLA) stents with a negative Poisson’s ratio (NPR) structure using a commercial fused deposition modeling (FDM) printer and PLA filaments, and studied the radial compressive property of the PLA stent. Although the results of these studies were promising,limitations remain regarding the stent’s shape. The BRS fabricated in the aforementioned studies have a cylindrical shape and a consistent diameter.In addition,stents generally undergo longitudinal foreshortening(a contraction in the length of the stent as the stent dilates).Clinicians must select stents with a longer length than the length of the plaque geometry to be stented. However, the degree of radial expansion and the final length of stents vary from person to person. Thus, the selection of a stent relies on the experienced judgment of the physician, which may go wrong and lead to malposition of the stent.Even more seriously,too much foreshortening would generate friction between the stent end and the artery,which would injure the delicate endothelium [15].

    In this study,we propose a method for fabricating cardiovascular stents;we also develop a novel screw extrusion-based 3D printing system that includes a newly designed mini-screw extruder.A novel design of a stent with zero Poisson’s ratio (ZPR) structure is introduced. Preliminary monofilament testing was conducted to investigate the fabrication parameters of the home-made extruder.Subsequently, stents with different materials, diameters, and geometric parameters were fabricated by changing the diameter,length, and shape of the mandrel. The surface morphology and the effects of geometric and fabrication parameters on the mechanical properties of the 3D-printed PCL stents were studied.

    2. Design of stents with a zero Poisson’s ratio structure

    At present,the majority of commercially available stents have a Z-shaped sequential-rings structure.This kind of design consists of two parts: Z-shaped rings (known as ‘‘struts”) and connectors(known as‘‘bridges”)[16].The Z-shaped rings are meant to expand during deployment and to support the vessel from collapse after the withdrawal of the balloon catheter. The connectors give the stent the ability to travel through the curvature of the arteries.

    To solve the longitudinal foreshortening problem, we designed a stent with a ZPR structure,as depicted in Fig.1.The structure unit(Fig. 1(a)) is similar to the semi-re-entrant honeycomb structure reported by Attard et al. [17], which exhibits a ZPR property. The Poisson’s ratio (υ) is defined as the ratio of the transverse strain(εi) and the axial strain (εj) in the direction of the loading force.The Poisson’s ratio can be expressed as follows:

    The structural unit of stent geometry can be defined by the following parameters:l represents the length of the inclined rib of the Z-shaped element;θ represents the angle of the Z-shaped element;h represents the length of the connector unit;and t represents the diameter (or width) of the strut. With the grid shown in Fig. 1(b)rolled up into a tube, as shown in Fig. 1(c), D represents the diameter of the stent and H represents the total length of the connectors. Here we introduce N1and N2as the number of structural units in the circumferential direction and axial direction,respectively.

    When stents are dilated or compressed, the force is loaded in the radial direction. In Figs. 2(a) and (b), the angle θ changes with an increase of r (the radius of the stent), which may be written as follows:

    For loading in the radial direction (Oxr), which results in an infinitesimal change dθ in the angle θ,the strain in the Oxidirection can be expressed as follows:

    where δxiis the deflection in Xidirection and i = r, l.

    As can be seen from Fig.2(c),the change in the diameter of the stent when it is dilated mainly depends on the deformation of the inclined rib of the Z-shaped ring; therefore, we focus here on the deformation of the inclined rib under loading. An inclined rib of length l can be likened to a cantilever beam that is loaded and guided at one end and fixed at the other(Fig.2(d))[18].The deflection of the rib in the x direction (δx) can be decomposed into δ⊥xgenerated by the bending moment and δ||xproduced by the component of the force F in the direction parallel to the rib.The deflection δ⊥xof the guided end and the deflection δ||xof the rib [19,20] are given by the following:

    where E0is the Young’s module of the material, I=πt4/64 is the second moment of the rib, t is the diameter of the stent strut, and A=πt2/4 is the area of the cross-section of the rib.

    Fig.1. Design of stent with ZPR structure.(a)The structural unit;(b)2D expanded drawing of stent;(c)3D model of stent.l0:the half of the distance between the adjacent ribs.

    Fig.2. Schematic diagram of deformation of Z-shaped ring and an inclined rib.(a)A Z-shaped ring;(b)an inclined rib;(c)deformation of a structural unit cell under loading in 2D model; (d) deformation of an inclined rib considering moments and axial force.

    In this study, l, t, and θ are usually set as 1.5 mm, 0.2 mm, and 60°, respectively. Thus, we can find that

    This indicates that in this study, δ||xis negligible relative to δ⊥x,which means that the deformation stretching of the rib can be viewed as negligible. Thus, the deflection of the rib in the x direction (δx) can be written as follows:

    The force F can be written in terms of the applied stress as F =σth. Using the standard definition of the Young’s modulus,and substituting for I, the equivalent elastic modulus of the rib in the x direction is

    Here,we substitute l,t,h,and θ as 1.5 mm,0.2 mm,1.5 mm,and 60°, and find that

    From Eqs.(1)–(13),we can infer that during the stent expansion process, only deformation of the stent structure (the ribs) occurs,and the stent material itself will not deform. Thus, the Poisson’s ratio of structure in Fig. 1(c) for loading in the radial direction can be written as follows:

    Because dXl/dθ=0, it is obvious that vrl=0.

    Thus, in this application, the direction of the loading force is parallel to the circumferential direction (Oxl) in the twodimensional (2D) expanded drawing of the stent (Figs. 1(b) and 2(c)) when the stent is dilated. Introducing the ZPR structure(v =0) into stent geometry provides an ideal situation in that the length of the stent does not change in the expansion process.

    3. A 3D printing system based on micro-extrusion technology

    3.1. System configuration

    Many of the polymers that are widely used in medical devices,such as poly(L-lactide) acid (PLLA) and PCL, are thermoplastic materials. FDM as a kind of extrusion-based 3D printing technology that has been widely applied in processing polymer materials.Conventional FDM equipment is based on filament extrusion and the three-axis printing technology, so it requires raw materials to be formed into filaments before being printed. Furthermore, frequent failures of filament buckling during the filament extrusion process result in interruption of the extrusion and necessitate numerous operator interventions [21,22]. In contrast, screw extrusion equipment offers better flexibility in the choice of raw materials. Wang et al. [23] developed a precision extruding deposition (PED) system consisting of a screw-extruder to fabricate PCL tissue scaffolds; the presented PED system directly extrudes materials in its granulated or pellet form without the filament preparation, and deposits freeform. This configuration opens up the opportunity for the use of a wider range of materials.Nonetheless,when the hollow structure and complicated pattern of a stent are taken into consideration,this kind of layer-by-layer deposition method has some limitations, as it is inevitably necessary to print numerous support structures. In addition, it is worth mentioning that the molecular weight of the polymer material decreases with increasing residence time of the melt inside the extruder,so a short and miniature-sized extruder is desirable in the process of biopolymer extrusion [24].

    To alleviate these problems, we proposed a screw extrusionbased 3D printing system with the addition of a rotation axis(the fourth axis) and designed a mini-screw extruder. The complete process of stent fabrication is presented in Fig. 3(a).The screw extrusion-based 3D printing system consists of three main components: an x–y–z motion system; a rotation axis(rotating mandrel); and a mini-screw extruder (Fig. 3(b)). The x–y–z motion system adopts a gantry structure, and Panasonic AC servo motors (Japan) are applied for the x-, y-, and z-axes.The shape and length of the rotating mandrel are flexible and can be changed accordingly. To ensure deposition accuracy, the movement of the x–y–z motion system, the rotational velocity of the mandrel, and the rotational velocity of the miniextruder are synchronized based on a programmable multiple axis controller (PMAC) provided by OMRON’s Delta Tau Data Systems (USA).

    To ensure the processing quality, the mini-extruder should be taken into consideration as a crucial component in the fabrication system. The structure and main components of the mini-extruder designed in this study are shown in Fig. 3(c). The heating element is attached to the outer surface of the barrel to provide heat for the material. Thus, when the mini-extruder operates, materials in the form of pellets or powder can be added into the extruder through the material inlet and will then be transported downward by the motor-driven screw.The chamber between the screw and the barrel is filled with molten material.Finally,the pressure generated in the chamber helps the molten material to be extruded through the nozzle tip as filaments.

    3.2. Research on printing parameters

    To determine a suitable temperature to fabricate 3D-printed stents, extrusion of PCL (material safety data sheet (MSDS)440744, Sigma-Aldrich, USA) monofilament was first conducted using the home-made mini-extruder in order to observe the forming quality of the extruded filaments.Considering that the width of the stent strut is 100–400 μm,the diameter of the nozzle tip outlet(φnozzle) was set as 250 μm. Differential scanning calorimetry(DSC) of the raw PCL material was performed using a DSC Q2000 instrument (TA Instruments, Italy) in the temperature range of-80–200°C,at a heating rate of 10°C?min-1,and under a nitrogen flow. Uniaxial tensile tests of the PCL monofilament samples were carried out with a WDW 3020 universal testing machine (Changchun Kexin Test Instrument, China), with a load cell of 50 N and a crosshead displacement of 1 mm?min-1at room temperature.

    The DSC experiment with the raw PCL material shows that PCL has a low melting temperature (Tm) of about 60 °C and a wide range of processing temperatures (Fig. 4(a)). Generally, for the material to melt sufficiently,the heating temperature of the extruder (Textruder) should be set above the melting point (Tm) of the material.As shown in Fig.4(b),PCL monofilaments were fabricated at five different temperatures between 75 and 85 °C (75, 77.5, 80,82.5, and 85 °C), with the extruder’s rotation speed set at 11.2 r?min-1. The samples produced by these temperatures were termed T75, T77.5, T80, T82.5, and T85, respectively. As shown in Fig. 4(c), the filament diameters (φfilament) of samples T75, T77.5,T80, T82.5, and T85were determined to be (319.33 ± 5.65),(318.03±6.69),(311.14±2.21),(312.18±3.19),and(315.10±9.48)μm,respectively.The tensile strength of samples T75,T77.5,T80,T82.5,and T85was determined to be(16.97±0.80),(15.53±1.73),(16.22±0.56),(15.04±0.96),and(14.87±1.55)MPa,respectively.

    Fig.3. The custom-made screw extrusion-based 3D printing system.(a)Configuration of stent fabrication system;(b)schematic diagram of 3D printing system;(c)structure and main components of the new-designed mini-extruder.

    Fig. 4. Investigation of the influence of heating temperature on the extrusion of PCL monofilament. (a) DSC curve of PCL; (b) PCL monofilaments fabricated at five different temperatures between 75 and 85°C;(c) the effect of temperature on the tensile strength and diameter of monofilaments;(d)scanning electron microscope(SEM)image of PCL filament extruded at 80 °C (SEM, FEI Quanta 200, the Netherlands). Error bars in (c) represent means ± standard deviations. HV: high voltage; Mag: magnitude.

    4. Results

    4.1. Stent fabrication

    As reported by Liu et al.[25],in the hot melt extrusion process,when the extruder motor is closed at the end of a line, the screw stops rotating immediately. However, material continues to be extruded from the nozzle tip because the pressure generated inside the extruder is not released.This phenomenon is called‘‘leakage of materials” and has an adverse effect on fabrication structure. Taking the design of the stent into consideration, the planning of the printing trajectory is crucial in the process of stent fabrication. If repeated starting and stopping of the extruder can be avoided—which also means that there is no need to suddenly move the extruder from one point to another during the printing process—then leakage of materials can be effectively eliminated.

    Accordingly, based on the design of the stent, auxiliary lines(gray lines in Fig. 5(a)) were introduced into the printing strategy.The 2D expanded drawings of the printing trajectory are shown in Fig. 5(a), where the direction of the arrows represents the moving direction of the extruder in the printing process. In this printing strategy, the molten filaments are deposited into a series of Zshaped lines (red lines) along the circumferential direction in the first layer. Adjacent Z-shaped lines (red lines) are connected with a ‘‘bridge” line (short black lines) at the end points. In the second layer,parallel straight lines(long black lines),which have a length of H, are printed along the axial direction, and adjacent straight lines (long black lines) are connected with auxiliary lines (gray lines). The materials fuse together at the overlap points of the Zshaped lines (red lines) and straight lines (long black lines).According to the forming method of intersection points,this printing method is called the ‘‘overlapping printing method” in this study.

    The designed screw extrusion-based 3D printing system with a mini-extruder was then employed to fabricate stents. The molten material filament was extruded from the nozzle and deposited onto the mandrel. With the synchronized movement of the x–y–z-axes and the rotation of the mandrel, the circumferential and axial filaments were collected successively to complete the fabrication of a stent. Stents were fabricated in a customized manner by changing the diameter and shape of the mandrel and the geometric structure of the stent model. A schematic of the stent fabrication process is shown in Fig. 5(b). The fabrication temperature was set at 80 °C. PLLA (RESOMER L210S, Evonik,Germany) stents and PCL (MSDS 440744, Sigma-Aldrich, USA)stents with different geometric structure were also fabricated(Figs. 5(c)–(e)).

    Table 1 The width of the stent strut based on different movement velocity and extruder rotation speed.

    Fig. 5. Fabrication of 3D-printed bioresorbable cardiovascular stents. (a) Printing trajectory of stents; (b) schematic of the stent fabrication process; (c) PLLA stents with different geometry structure; (d, e) tapered PCL stents; (f)–(j) images of stents listed in Table 2 (A1–E3).φ: the diameter of stents (unit: mm);ω: the angular velocity.

    To determine the influence of the stent diameter(φ),the length of the connector unit(h),and the extruder rotation speed(n)on the mechanical properties of the 3D-printed stents,five series(A–E)of 3D-printed PCL stents were fabricated (Figs. 5(f)–(j)). The experimental groups and parameters are shown in Table 2. For different groups of stents, l, θ, and H were constant. In this study, we set l=1.5 mm,l0=π/4-0.75,θ=60°,and H=15 mm.Thus,according to πφ=2N1(lcos(θ)+l0) and H =N2h (Fig. 1), when φ = 2, 3, and 4 mm, N1is 4, 6, and 8, respectively, and when h = 1.00, 1.25,and 1.50 mm, N2is 15, 12, and 10, respectively.

    4.2. Morphological analysis

    The surface morphology of the stents was observed by scanning electron microscopy (SEM, FEI Quanta 200, the Netherlands) at an accelerating voltage of 15 kV. Before SEM analysis, the samples were sputter-coated with gold for 60 s. As shown in Fig. 6(a), the stents developed with the‘‘overlapping printing method”generally showed good surface quality. The intersection points of the stents fabricated with the ‘‘overlapping printing method” are fused by two filaments. It can be seen from the SEM image (Fig. 6) that the PCL monofilaments melted and combined together at theintersection points,ensuring that the stent has good continuity and strength at these points.

    Table 2 Experimental groups and parameters of 3D-printed PCL stents.

    Fig. 6. SEM images of stent fabricated with ‘‘overlapping printing method.” (a) Mag 85×; (b, c) Mag 200×.

    In addition,a Keyence digital microscope(VHX-500,Japan)was used to observe the surface morphology and measure the width of the stent strut (t). The rotation speed of series D, A–C, and E was 8.4, 11.2, and 14 r?min-1. As depicted in Fig. 7, the average strut width of series D, A–C, and E was (220 ± 12), (362 ± 42), and(452 ± 35) μm. Specifically, the average width of with the stents in groups A1–A3 and groups B1–B3 was (349 ± 28) and(334 ± 23) μm, and the average width for the stents in groups C1–C3 was (404 ± 38)μm.

    4.3. Mechanical properties

    A three-point bending test and radial compression test were performed on the 3D-printed PCL stents listed in Table 2 (A1–E3)by installing the corresponding test apparatus (Figs. 8(a) and (b))on an INSTRON 3365 testing machine (USA) at room temperature according to American Society of Testing Materials (ASTM)F2606-08 and ASTM F3067-14, respectively [26,27]. In addition, a radial expansion test of the groups C1 and C2 stents was performed by means of a dilated balloon (Fig. 8(c)).

    4.3.1. Three-point bending test

    The three-point bending test is a common means of investigating stent’s bending flexibility, which affects the conformability of the stent (Fig. 9). A stent with good bending flexibility can easily be delivered through the curvature of the artery.Bending flexibility can be represented by bending stiffness: The higher the bending stiffness,the lower the flexibility.According to its definition,bending stiffness can be calculated by E ?I, where E and I are elastic modulus and area moment of inertia.E ?I is given by the following equation [26,28]:

    where F, L, and f are the bending force, bending span length, and deflection, respectively. In this test, L is set as 11 mm; thus, when the deflection is set at a certain value,stents with less bending load(F)have better bending flexibility.When the deflection is 2 mm,the loading force of the stents of groups A,B,and C is shown in Fig.9(d);the force of groups B, D, and E is shown in Fig. 9(h).

    In Figs. 9(a)–(d), the bending flexibility of different groups is B1 < A1 ≈C1, A2 < B2 < C2, and A3 < B3 < C3. As the length of the connector increases from 1 to 1.5 mm, for the φ2, φ3, and φ4stents, the ratio of the loading force between groups C1/A1, C2/A2, and C3/A3 is 1.01, 0.77, and 0.49, respectively. As shown in Figs. 9(e)–(h), the bending flexibility of different groups is E1 < B1 < D1, E2 < B2 < D2, and E3 < B3 < D3.

    Fig. 7. The width of stent strut of five series (A–E). (a)–(e) Photographs of stents with size marking taken by Keyence VHX-500 microscope; (f) the width of strut of different groups (A1–E3). Error bars in (f) represent means ± standard deviations.

    Fig. 8. Mechanical properties tests of stents. (a) Three-point bending test; (b) radial compression test; (c) radial expansion test.

    Fig. 9. Force-versus-deflection curves for 3D-printed PCL stents with different geometric unit parameter and fabrication parameter: for (a), (b), and (c), the length of connector(h)is 1,1.25,and 1.5 mm,respectively;(d)the loading force in(a),(b),and(c)of stents with different diameter and length of connector(h)when the deflection is 2 mm;for(e),(f),and(g),the extruder’s rotation speed(n)is 8.4,11.2 and 14 r?min-1,respectively;(h)the loading force in(e),(f),and(g)of stents with different diameters and fabrication parameters (n) when the deflection is 2 mm. Error bars in (d) and (h) represent means ± standard deviations.

    Table 3 Bending stiffness values of 3D-printed stents.

    The bending stiffness values (average value) of the stents at a deflection of 2 mm are given in Table 3. In contrast, the A1, C1,C3, and D1–D3 stents have relatively better bending flexibility. In a previous publication [29], the expanded Abbott Absorb GT1 stents (diameter: 3.0 mm, length: 18 mm) were shown to have a bending stiffness value of 4.20 N?mm2,which is close to the bending stiffness values of the A1, C1, C3 and D1–D3 stents. This indicates that within these 15 groups of stents, the A1, C1, C3, and D1–D3 stents have applicable bending flexibility.

    4.3.2. Radial compression test

    The Radial compression test was applied to evaluate the radial strength, which determines the radial support force of the stent.Stents with good radial strength can provide better support.Radial strength can be represented by radial stiffness: The higher the radial stiffness, the lower the radial strength. Schmidt et al. [30]suggested that radial stiffness is determined from the profile decreasing in a physiologically relevant range from 0 to 0.2 bar(1 bar=105Pa)outer pressure(Eq.(16)).ASTM F3067-14[27]suggests that when using an area normalized force,the instantaneous stent diameter should be used (instead of the starting diameter),multiplied by the initial length (Eqs. (17)–(18)). Accordingly, the radial stiffness is the compression ratio when the instantaneous pressure acting on the stents is 0.2 bar.

    A stent’s radial strength is mainly related to the width (t) and number of Z-shaped rings. In Figs. 10(a)–(d), it can be seen that the radial stiffness of the different groups is Bi > Ci > Ai (i = 1, 2,3); thus, the radial strength of the different groups is Bi < Ci < Ai(i = 1, 2, 3). From Figs. 10(e)–(h), it is obvious that the radial stiffness of the different groups is Di > Bi > Ei (i = 1, 2, 3); thus, the radial strength of the different groups is Di < Bi < Ei (i = 1, 2, 3).

    In a previous publication [30], the radial stiffness varied from 0.51% to 3.99%, which is close to the radial stiffness values of the A2, A3, C2, C3, and E3 stents. This indicates that within these 15 groups of stents, the A2, A3, C2, C3, and E3 stents have applicable bending flexibility.

    4.3.3. Radial expansion test

    Samples(the C1 and C2 stents)were expanded from the original mode until the expansion ratio became 10% and 20%, respectively.Group C1 stents with a diameter of 2 mm were expanded into about 2.2 and 2.4 mm, respectively. Group C2 stents with a diameter of 3 mm were expanded into about 3.3 and 3.6 mm,respectively. After that, the radial forces were removed and the stents recoiled in the radial direction.The diameter of each sample in the original mode (Oφ), expansion maximum mode (Mφ) and recoil mode(Rφ)is shown in Fig.11(a).The radial recoil ratio after removing the radial force can be calculated by the following:

    The average radial recoil ratio of the C1 stents expanded to 110% and 120% was 69% and 41%, respectively, while that of the C2 stents was 77% and 71%. The extent of the radial recoil of the 3D-printed PCL stents was unsatisfactory, which is related to the excellent elongation rate of PCL material. To resolve this problem,other biomaterials will be explored in future studies.

    The length of each sample in the original mode (OL), expansion maximum mode (ML), and recoil mode (RL) is shown in Fig. 11(b).The longitudinal foreshortening ratio when the stents were expanded into the maximum mode can be calculated by the following:

    The longitudinal foreshortening ratio of the C1 stents expanded to 110%and 120%was 0.33%and 0.13%,respectively, while that of the C2 stents was 0.40%and 0.09%.In contrast,current commercial stents have a shortening rate of 3%–5%[31].Thus,it can be verified that the stent geometry designed in this study can achieve a substantially steady axial length after being expanded.

    4.4. Biocompatibility

    Hemolysis analysis of the 3D-printed stents was conducted in accordance with ASTM F756-17[32].Fresh human blood obtained from voluntary donors was legally supplied by Beijing Huaxin Hospital, the First Affiliated Hospital of Tsinghua University. The hemolysis rate was determined by measuring hemoglobin release under static conditions,as previously reported[33].The hemolysis rate was calculated according to Eq. (21):

    where ODtrepresents the optical density of the test group,and ODpand ODnrepresent the absorbance of the positive and negative groups,respectively.According to ASTM F756-17 [32], the hemolysis of materials is classified as non-hemolytic(0–2%),slightly hemolytic (2%–5%), or hemolytic (> 5%). The hemolysis rate of the 3Dprinted PCL was 0.07% ± 0.02%, which is non-hemolytic.

    A cytocompatibility evaluation of the 3D-printed stents was performed according to ISO 10993–5:2009 [34]. Human umbilical vein endothelial cells (HUVECs; ATCC, USA) were cultured in clonetics endothelial cell growth medium-2 (EGM-2; cc3162, Lonza,Switzerland). Extracts of 3D-printed PCL were prepared by immersing samples in EGM-2 at 37 °C for 24 h. Cell viability was detected by adding 10 μL of cell counting kit-8 (CCK-8)(Dojindo Laboratories, Japan) in each well after the HUVECs were cultured in extracts for 96 h; next, then the optical density (OD)value for each well was read at a wavelength of 450 nm in order to determine the cell viability on a microplate reader (Multiskan,Thermo, USA). The cell viability was calculated according to Eq. (22):

    where ODtrepresents the OD value of the test group, and ODpand ODnrepresent the OD values of the positive and negative groups,respectively. According to the cell viability value, the cytotoxicity reactivity of the materials was classified as Grade 0(≥100%),Grade 1(80%–99%),Grade 2(50%–79%),Grade 3(30%–49%),or Grade 4(0–29%). When the grade is more than 2, a material is considered to have cytotoxic effects. The cell viability of the 3D-printed PCL was 90% ± 5%, which is Grade 1.

    Fig. 10. Force-versus-compression ratio curves for PCL stents with different geometric unit parameter and rotation speed (n). For stents in (a), (b), and (c), the length of connector(h)is 1,1.25,and 1.5 mm,respectively;(d)the radial stiffness of stents in(a),(b),and(c);for stents in(e),(f),and(g),rotation speed(n)is 8.4,11.2,and 14 r?min-1,respectively; (h) the radial stiffness of stents in (e), (f), and (g). Error bars in (d) and (h) represent means ± standard deviations.

    Fig. 11. Radial expansion results: (a) the change of stent diameter; and (b) the change of stent length. Error bars represent means ± standard deviations.

    The results of the hemolysis rate and cell viability confirmed the potential application of 3D-printed stents in vascular implantation.

    5. Discussion

    5.1. Effect of printing parameters on printing-ability and structural formability

    5.1.1. Heating temperature of the extruder

    Before stent fabrication, extrusion of PCL monofilaments was conducted to investigate the effect of the heating temperature of the extruder (Textruder) on the strength and dimensional stability of the extruded monofilament. By comparing the accuracy of the filament diameter of different groups, it was found that the PCL filaments had better dimensional stability when Textruderwas 80 and 82.5 °C (Fig. 4(c)), while samples extruded at 75 and 80 °C had the advantage in terms of tensile strength.Consequently,considering both strength and dimensional stability,the heating temperature of the extruder was set at 80 °C. This temperature was used in the subsequent stent fabrication process. As shown in Figs.4(b)and(d),the PCL filament extruded at 80°C showed good surface morphology.

    5.1.2. Movement velocity of the x-axis system, rotational velocity of the mandrel, and extruder rotation speed

    Table 1 shows that the width of the stent strut(t) can be changed by adjusting the movement velocity (VRx) and extruder rotation speed (n). As VRxincreases, the width of the stent strut becomes smaller and the printing stability decreases (the coefficient of variation increases). When extruder rotation speed (n)increases, the width of the stent strut increases and shows better precision (the coefficient of variation is lower). Therefore, for a comprehensive consideration of printing stability(e.g.,for a coefficient of variation less than 10%)and printing efficiency,the movement velocity(VRx)was set at 1 mm?s-1and the width of the strut was controlled by changing the extruder rotation speed (n).Accordingly, the movement velocity of the x-axis (Vx) is√---3/2 mm?s-1(0.866 mm?s-1) and the linear velocity of the rotating mandrel (VR) was 0.5 mm?s-1. Thus, the rotational velocity of the mandrel (nm, unit:r?min-1) was 30/(πφ) r?min-1(where φ represents the inner diameter of the stent). Extruder rotation speed (n)of 8.4, 11.2, and 14 r?min-1were then adopted in the subsequent experiments.

    5.1.3. Filament extrusion speed

    The filament extrusion speed depends on the rotation speed of the extruder. The higher the rotation speed is, the faster the filaments are extruded. The rotation speed of series D, A–C, and E was 8.4, 11.2, and 14 r?min-1, respectively. As depicted in Fig. 7,the stents of series A, B, and C show a middle value in the width of strut (t), while series D has a relatively lower t and series E has a greater t. It is clear that the width of the stent strut is positively related to the rotation speed of the extruder: The greater the rotation speed, the larger the width of the strut. The stents in the series A,B,and C were fabricated with the same rotation speed.However,compared with the stents in series A and B,the stents in series C had a relative larger dimension of the stent strut. Therefore, the stability of stent fabrication needs to be improved in the future studies.

    5.2. Effect of structural parameters on the mechanical properties of 3D-printed stents

    In the literature,various methods have been used to examine a stent’s mechanical properties.The bending flexibility property and radial strength are two important indicators of the mechanical properties of a stent, which are closely related to the geometric parameters of the stent[14,28].Based on the discussion in Sections

    4.3.1 and 4.3.2, it is noted that bending flexibility can be represented by bending stiffness,with a higher bending stiffness resulting in a lower flexibility. Radial strength can be represented by radial stiffness, with a higher radial stiffness resulting in a lower radial strength. In this study, the main parameters affecting the mechanical properties of stents were the length of the connector and the width of the stent strut.

    5.2.1. Length of connector (h)

    The difference between series A, B, and C is the length of the connector (h), which is 1, 1.25, and 1.5 mm, respectively. Because the length of the stent is constant, it is clear that the longer the connector h is, the fewer the number of Z-shaped rings will be,resulting in greater bending flexibility and lower radial strength.Thus, the theoretical result for bending flexibility and radial strength should be Ai < Bi < Ci and Ci < Bi < Ai (i = 1, 2, 3),respectively.

    In regard to bending flexibility,Figs. 9(a)–(d)and Table 3 show that the bending flexibility of the different groups is B1 Ci > Ai (i = 1, 2, 3);thus, the radial strength of different groups is Bi < Ci < Ai (i = 1,2,3),which is different from the theoretical result.This finding will be explained in Section 5.2.2.

    In conclusion,stents with a longer connector have an advantage in terms of bending flexibility.However,it is not conclusive in this part whether a longer connector is strictly unfavorable for radial strength.

    5.2.2. Width of stent strut (t)

    The difference between series D, B, and E is the extruder rotation speed (n), which is 8.4, 11.2, and 14 r?min-1, respectively.Because the width of the stent strut is positively related to the rotation speed of the extruder, the greater the rotation speed is,the larger the width of the strut will be,resulting in lower pending flexibility and greater radial strength. Thus, the theoretical result for bending flexibility and radial strength should be Ei < Bi < Di and Di < Bi < Ei (i = 1, 2, 3), respectively.

    As expected, Figs. 9(e)–(h), Table 2, and Figs. 10(e)–(h) show that the experimental results for bending flexibility and radial strength are consistent with the theoretical situation.These results confirm that the strut width(t)is a dominating factor for the radial strength of stents. Furthermore, as mentioned above, the radial strength of groups A–C is Bi < Ci < Ai (i = 1, 2, 3), which does not match with the theoretical result of Ci < Bi < Ai. This result can be explained in terms of the width of the stent strut (t). From Fig. 7(f), it can be determined that the strut width of the Ci stents is relatively larger than that of the Bi stents (i = 1, 2, 3).

    5.2.3. Brief summary

    Based on the discussion in Sections 5.2.1 and 5.2.2, it can be seen that bending flexibility and radial strength are a pair of contradictory properties. This is a result of a trade-off of stent structure parameters to realize integrated mechanical properties. For stents of the same diameter, the mechanical properties (bending flexibility and radial strength) of the stents are closely related to the length of the connector and the width of the stent strut. In regard to the bending flexibility of stents with the same diameter,it can be seen that stents with a longer connector and a smaller width of stent strut (t) have advantages; therefore, stents with a loose geometric structure tend to show better flexibility.In regard to the radial strength of stents with the same diameter, it can be inferred that the strut width(t)is a dominating factor.The answer can be obtained from Eq. (12), which indicates that the structural strength of the stent is proportional to t3and inversely proportional to h; this means that the properties of the stent are more susceptible to the strut width(t).It can be inferred that stents with a larger connector (h) and a moderate strut width are more conducive for a balance between bending flexibility and radial strength.

    6. Conclusions

    In this study, a home-made screw extrusion-based 3D printing system using a mini-screw extruder was developed to fabricate stents while adding a fourth rotation axis (the mandrel). A stent with ZPR structure was designed and was demonstrated to be steady in axial length after expansion.Preliminary testing was conducted to investigate appropriate fabrication parameters with monofilament experiments. Subsequently, stents (made of PLLA and PCL)with different shape and geometric structures were fabricated in a customized manner using the ‘‘overlapping printing method.”In addition,the mechanical properties of PCL stents with different parameters were determined.

    In conclusion, the screw extrusion-based 3D printing system reported here shows potential for customizable stent fabrication.PCL stents are well studied in terms of morphology and mechanical analysis. However, the insufficiency of stent flexibility indicates a need for more efforts in stent design in future research. Furthermore, in comparison with metal stents, PCL stents are not completely reliable in term of radial strength and radial recoil. PLLA or other biopolymers with greater strength and a lower elongation rate will be explored for stent fabrication.

    Acknowledgements

    The authors acknowledge funding support from the Beijing Municipal Natural Science Foundation,China(Z150001).This work was also supported by Beijing Anzhen Hospital.

    Compliance with ethics guidelines

    Chengjin Wang,Lei Zhang,Yongcong Fang,and Wei Sun declare that they have no conflict of interest or financial conflicts to disclose.

    久久精品综合一区二区三区| 女同久久另类99精品国产91| 巨乳人妻的诱惑在线观看| 久久精品影院6| 1024视频免费在线观看| 国产激情久久老熟女| 国产成人精品久久二区二区免费| 91字幕亚洲| 国产免费av片在线观看野外av| 欧美日韩亚洲综合一区二区三区_| 国产成人精品无人区| x7x7x7水蜜桃| 91在线观看av| 国产野战对白在线观看| 精品久久久久久久毛片微露脸| 国产熟女午夜一区二区三区| 天堂影院成人在线观看| 亚洲熟女毛片儿| 亚洲精品久久成人aⅴ小说| 一夜夜www| 97超级碰碰碰精品色视频在线观看| 成人一区二区视频在线观看| 在线观看www视频免费| 99久久精品热视频| 日本一二三区视频观看| 高清在线国产一区| av福利片在线观看| 色综合亚洲欧美另类图片| 亚洲一码二码三码区别大吗| 欧美中文综合在线视频| 黄色成人免费大全| 亚洲成人国产一区在线观看| 1024视频免费在线观看| 久久久精品大字幕| 久久久久久人人人人人| www.999成人在线观看| 男人的好看免费观看在线视频 | 99久久久亚洲精品蜜臀av| 国产精品免费一区二区三区在线| 日韩大码丰满熟妇| 国产精品永久免费网站| 精品不卡国产一区二区三区| 免费电影在线观看免费观看| 国产av又大| 久久国产精品人妻蜜桃| 女同久久另类99精品国产91| 亚洲精品美女久久久久99蜜臀| 国产成人影院久久av| 亚洲国产中文字幕在线视频| 小说图片视频综合网站| 88av欧美| 99在线视频只有这里精品首页| 久久久久久久久免费视频了| 成人av一区二区三区在线看| 精品久久久久久久毛片微露脸| 午夜激情福利司机影院| 成人永久免费在线观看视频| 亚洲国产精品999在线| 国产精品自产拍在线观看55亚洲| 性欧美人与动物交配| 搡老熟女国产l中国老女人| 97碰自拍视频| 欧美日韩乱码在线| 久久伊人香网站| 国语自产精品视频在线第100页| 国产一区二区在线观看日韩 | 国产成人系列免费观看| 久久精品91无色码中文字幕| 亚洲av成人精品一区久久| 亚洲人成网站在线播放欧美日韩| 国产精品一区二区三区四区久久| 欧美日韩一级在线毛片| 亚洲一区二区三区不卡视频| 国产三级在线视频| 亚洲黑人精品在线| 最新美女视频免费是黄的| 国产亚洲av高清不卡| 久久99热这里只有精品18| 免费人成视频x8x8入口观看| 丰满人妻熟妇乱又伦精品不卡| 婷婷精品国产亚洲av| 男女下面进入的视频免费午夜| av视频在线观看入口| 欧美极品一区二区三区四区| 婷婷亚洲欧美| 黄片小视频在线播放| 国产成人影院久久av| 久久国产精品人妻蜜桃| 亚洲av成人不卡在线观看播放网| videosex国产| 精品欧美一区二区三区在线| 国产亚洲av高清不卡| 黄色女人牲交| 欧美色视频一区免费| 日日爽夜夜爽网站| 国产三级在线视频| 久久精品91无色码中文字幕| 一级毛片精品| www日本黄色视频网| 最近视频中文字幕2019在线8| 中文字幕精品亚洲无线码一区| 在线国产一区二区在线| 国产一区在线观看成人免费| 亚洲国产精品成人综合色| 日本三级黄在线观看| 国产精品免费视频内射| 国产精品九九99| bbb黄色大片| 午夜日韩欧美国产| 亚洲国产看品久久| a在线观看视频网站| 最近最新中文字幕大全电影3| 国产亚洲精品久久久久久毛片| 久久久久久久久久黄片| 精品无人区乱码1区二区| 哪里可以看免费的av片| 热99re8久久精品国产| netflix在线观看网站| 午夜福利18| 亚洲精品中文字幕一二三四区| 色av中文字幕| 夜夜看夜夜爽夜夜摸| 中文字幕人妻丝袜一区二区| 最新在线观看一区二区三区| 在线a可以看的网站| www.999成人在线观看| 热99re8久久精品国产| 欧美高清成人免费视频www| 18禁观看日本| 亚洲av第一区精品v没综合| 亚洲avbb在线观看| 免费在线观看影片大全网站| 两个人视频免费观看高清| 精品久久久久久成人av| 亚洲成人免费电影在线观看| 欧美日韩中文字幕国产精品一区二区三区| 免费在线观看亚洲国产| 亚洲av五月六月丁香网| 波多野结衣高清作品| 国产成人av教育| 国产免费男女视频| 88av欧美| 中出人妻视频一区二区| 动漫黄色视频在线观看| 国产精品精品国产色婷婷| 免费看日本二区| 午夜激情av网站| 丰满人妻一区二区三区视频av | 19禁男女啪啪无遮挡网站| 两人在一起打扑克的视频| 日本熟妇午夜| 一二三四社区在线视频社区8| 狂野欧美白嫩少妇大欣赏| 国产午夜精品论理片| 国产精品av视频在线免费观看| 亚洲成av人片免费观看| 欧美乱色亚洲激情| 日本精品一区二区三区蜜桃| 欧美3d第一页| 国产精品免费视频内射| 国产精品自产拍在线观看55亚洲| 国产又黄又爽又无遮挡在线| 亚洲精品一区av在线观看| 日韩av在线大香蕉| 精品久久久久久久久久免费视频| 无遮挡黄片免费观看| 一级片免费观看大全| 99热这里只有是精品50| 天天躁狠狠躁夜夜躁狠狠躁| 俺也久久电影网| 亚洲av中文字字幕乱码综合| 亚洲美女黄片视频| 给我免费播放毛片高清在线观看| 亚洲国产精品合色在线| 19禁男女啪啪无遮挡网站| 又爽又黄无遮挡网站| 亚洲欧美精品综合一区二区三区| 亚洲狠狠婷婷综合久久图片| 久久人妻av系列| 久久精品人妻少妇| 中亚洲国语对白在线视频| 亚洲欧美日韩高清在线视频| 国产一区二区在线观看日韩 | 十八禁网站免费在线| 桃色一区二区三区在线观看| 亚洲精品在线美女| 亚洲人成77777在线视频| 欧美日韩亚洲国产一区二区在线观看| 国产爱豆传媒在线观看 | 久久久久久免费高清国产稀缺| 亚洲av电影在线进入| 亚洲精品国产一区二区精华液| 在线观看www视频免费| 叶爱在线成人免费视频播放| 91老司机精品| 国产男靠女视频免费网站| av超薄肉色丝袜交足视频| 免费在线观看日本一区| 欧美黑人精品巨大| 成人av在线播放网站| 亚洲va日本ⅴa欧美va伊人久久| 88av欧美| 成年女人毛片免费观看观看9| 全区人妻精品视频| 18禁美女被吸乳视频| 99热这里只有是精品50| av欧美777| 99精品久久久久人妻精品| 天天一区二区日本电影三级| 久久久水蜜桃国产精品网| 又爽又黄无遮挡网站| 亚洲国产高清在线一区二区三| 99国产综合亚洲精品| 国产精品永久免费网站| 丁香欧美五月| 亚洲成人久久爱视频| 少妇熟女aⅴ在线视频| 岛国视频午夜一区免费看| xxxwww97欧美| ponron亚洲| 亚洲国产欧美一区二区综合| 搡老熟女国产l中国老女人| 亚洲成人国产一区在线观看| 欧美另类亚洲清纯唯美| 国产野战对白在线观看| 久久久久九九精品影院| 夜夜躁狠狠躁天天躁| 妹子高潮喷水视频| 老鸭窝网址在线观看| 亚洲第一欧美日韩一区二区三区| 夜夜看夜夜爽夜夜摸| 欧美日韩乱码在线| 国产激情欧美一区二区| 美女黄网站色视频| 美女扒开内裤让男人捅视频| 国产精品影院久久| 99精品久久久久人妻精品| 亚洲色图av天堂| 琪琪午夜伦伦电影理论片6080| svipshipincom国产片| 国产三级中文精品| 久久九九热精品免费| 熟妇人妻久久中文字幕3abv| 麻豆国产97在线/欧美 | 亚洲国产看品久久| 夜夜夜夜夜久久久久| 一区二区三区激情视频| 亚洲国产精品成人综合色| 精品高清国产在线一区| 深夜精品福利| 久久久久精品国产欧美久久久| 欧美又色又爽又黄视频| 麻豆国产av国片精品| 欧美日韩瑟瑟在线播放| 嫩草影院精品99| xxx96com| 国产欧美日韩一区二区三| 欧美黄色淫秽网站| 亚洲av成人不卡在线观看播放网| 一级作爱视频免费观看| 全区人妻精品视频| 国产精品久久视频播放| 亚洲av五月六月丁香网| 国内毛片毛片毛片毛片毛片| 免费一级毛片在线播放高清视频| 色综合亚洲欧美另类图片| 性欧美人与动物交配| 一区二区三区高清视频在线| 两性夫妻黄色片| 国产精品久久久人人做人人爽| 欧美丝袜亚洲另类 | 长腿黑丝高跟| 正在播放国产对白刺激| 国内揄拍国产精品人妻在线| 少妇的丰满在线观看| 欧美一区二区精品小视频在线| 欧美成人一区二区免费高清观看 | 亚洲美女黄片视频| 免费在线观看成人毛片| 少妇粗大呻吟视频| 国产av又大| 国产精品98久久久久久宅男小说| av在线天堂中文字幕| 亚洲在线自拍视频| 亚洲激情在线av| 老司机福利观看| 日韩精品青青久久久久久| 亚洲成av人片免费观看| 日韩中文字幕欧美一区二区| 色精品久久人妻99蜜桃| 国产精品国产高清国产av| 午夜福利免费观看在线| 成人一区二区视频在线观看| 国产欧美日韩精品亚洲av| 国产成人啪精品午夜网站| 国产av在哪里看| 亚洲欧美日韩高清专用| 身体一侧抽搐| 禁无遮挡网站| 久久精品91蜜桃| 亚洲五月天丁香| 看片在线看免费视频| 国产精品一及| 国产精品永久免费网站| 两个人免费观看高清视频| 欧美黄色淫秽网站| 色在线成人网| 99久久精品国产亚洲精品| 老司机靠b影院| 十八禁人妻一区二区| 757午夜福利合集在线观看| 午夜福利免费观看在线| 午夜日韩欧美国产| 午夜福利免费观看在线| 精品高清国产在线一区| 久久久久久大精品| 丁香六月欧美| 亚洲精品国产精品久久久不卡| 欧美大码av| 婷婷六月久久综合丁香| 国产精品 欧美亚洲| 巨乳人妻的诱惑在线观看| 桃色一区二区三区在线观看| 午夜影院日韩av| 亚洲人与动物交配视频| 成人国产一区最新在线观看| 国产真实乱freesex| 一边摸一边做爽爽视频免费| 十八禁人妻一区二区| 亚洲欧美精品综合一区二区三区| 久久天躁狠狠躁夜夜2o2o| 日本熟妇午夜| av视频在线观看入口| 熟妇人妻久久中文字幕3abv| 51午夜福利影视在线观看| 女人爽到高潮嗷嗷叫在线视频| 视频区欧美日本亚洲| 亚洲 欧美一区二区三区| 日本精品一区二区三区蜜桃| 午夜精品一区二区三区免费看| 国产免费av片在线观看野外av| 亚洲av电影不卡..在线观看| 国内揄拍国产精品人妻在线| 日本三级黄在线观看| av在线播放免费不卡| 欧美3d第一页| 18禁黄网站禁片免费观看直播| 桃色一区二区三区在线观看| 99国产极品粉嫩在线观看| 成人av在线播放网站| 麻豆久久精品国产亚洲av| 男人的好看免费观看在线视频 | 午夜激情福利司机影院| 99国产极品粉嫩在线观看| 国产成人一区二区三区免费视频网站| ponron亚洲| 国产真实乱freesex| 免费在线观看影片大全网站| 国产真实乱freesex| 三级国产精品欧美在线观看 | 欧美久久黑人一区二区| 天堂动漫精品| 日日夜夜操网爽| 五月玫瑰六月丁香| 亚洲 欧美一区二区三区| 老汉色av国产亚洲站长工具| 亚洲 欧美一区二区三区| 一二三四社区在线视频社区8| 久久久精品大字幕| 日韩三级视频一区二区三区| 亚洲中文日韩欧美视频| 舔av片在线| 亚洲av片天天在线观看| 国产精品自产拍在线观看55亚洲| 无限看片的www在线观看| 日韩欧美免费精品| 日韩精品中文字幕看吧| 免费人成视频x8x8入口观看| 久热爱精品视频在线9| 1024视频免费在线观看| 久久精品91蜜桃| 91国产中文字幕| 欧美三级亚洲精品| 精品欧美国产一区二区三| 身体一侧抽搐| 午夜老司机福利片| 欧美日本亚洲视频在线播放| 亚洲欧美精品综合久久99| 搡老岳熟女国产| 成人三级做爰电影| 他把我摸到了高潮在线观看| 精品一区二区三区av网在线观看| 久久香蕉国产精品| 色综合婷婷激情| videosex国产| 欧美绝顶高潮抽搐喷水| 久久亚洲真实| 精品国产亚洲在线| 99在线视频只有这里精品首页| 少妇粗大呻吟视频| 亚洲人与动物交配视频| 午夜福利18| 欧美一级毛片孕妇| 久久这里只有精品中国| 午夜久久久久精精品| 丰满人妻一区二区三区视频av | 两性夫妻黄色片| 两个人的视频大全免费| 无人区码免费观看不卡| 日韩欧美精品v在线| 性欧美人与动物交配| 国产一区在线观看成人免费| 高清在线国产一区| 亚洲精品美女久久久久99蜜臀| 国产精品电影一区二区三区| 淫秽高清视频在线观看| 欧美乱色亚洲激情| 一个人观看的视频www高清免费观看 | 精品高清国产在线一区| 欧美日韩乱码在线| 久热爱精品视频在线9| 在线观看免费日韩欧美大片| 国产精品综合久久久久久久免费| x7x7x7水蜜桃| 日韩 欧美 亚洲 中文字幕| 国产久久久一区二区三区| 国产蜜桃级精品一区二区三区| 精品久久久久久久久久久久久| 在线永久观看黄色视频| 亚洲精品中文字幕在线视频| 99精品久久久久人妻精品| 亚洲av成人不卡在线观看播放网| 欧美一级毛片孕妇| 国产精品一及| 日日摸夜夜添夜夜添小说| 成在线人永久免费视频| 老熟妇乱子伦视频在线观看| 日本一二三区视频观看| 伊人久久大香线蕉亚洲五| 国产69精品久久久久777片 | 国产精品,欧美在线| 99热这里只有是精品50| 90打野战视频偷拍视频| 欧美精品啪啪一区二区三区| 一本久久中文字幕| 好男人电影高清在线观看| 国产主播在线观看一区二区| 可以免费在线观看a视频的电影网站| 亚洲精品久久成人aⅴ小说| 精品第一国产精品| 他把我摸到了高潮在线观看| 窝窝影院91人妻| 亚洲狠狠婷婷综合久久图片| 亚洲一区中文字幕在线| 两性夫妻黄色片| 男人舔奶头视频| 国产v大片淫在线免费观看| 男女做爰动态图高潮gif福利片| 中文字幕最新亚洲高清| 欧美日韩中文字幕国产精品一区二区三区| 国产成+人综合+亚洲专区| 色播亚洲综合网| 久久久久久免费高清国产稀缺| 热99re8久久精品国产| 亚洲成人中文字幕在线播放| 久久人妻av系列| 久久久久久久午夜电影| 国产精品乱码一区二三区的特点| 午夜免费激情av| 久久精品人妻少妇| 欧美日韩亚洲国产一区二区在线观看| 99精品欧美一区二区三区四区| 麻豆一二三区av精品| 我要搜黄色片| 午夜两性在线视频| 国产成人aa在线观看| 丁香欧美五月| 日本一本二区三区精品| 一个人免费在线观看的高清视频| 成年版毛片免费区| 午夜精品久久久久久毛片777| 人人妻,人人澡人人爽秒播| 成年女人毛片免费观看观看9| 精品国产美女av久久久久小说| 可以在线观看毛片的网站| 中文字幕熟女人妻在线| 亚洲欧美精品综合一区二区三区| 日韩有码中文字幕| 亚洲国产看品久久| 国产探花在线观看一区二区| 亚洲狠狠婷婷综合久久图片| av福利片在线观看| 国产麻豆成人av免费视频| 中文字幕人成人乱码亚洲影| 欧美久久黑人一区二区| 一个人免费在线观看的高清视频| 国产精品98久久久久久宅男小说| av福利片在线| 国产精品香港三级国产av潘金莲| 久久久精品大字幕| 国产伦人伦偷精品视频| 精品国产超薄肉色丝袜足j| xxxwww97欧美| 91麻豆精品激情在线观看国产| 可以在线观看毛片的网站| 一夜夜www| 在线观看www视频免费| 97超级碰碰碰精品色视频在线观看| 国产精品av视频在线免费观看| 高潮久久久久久久久久久不卡| 久久精品国产亚洲av高清一级| 久久香蕉精品热| x7x7x7水蜜桃| 精品国产美女av久久久久小说| 黄色毛片三级朝国网站| 大型av网站在线播放| 黄色 视频免费看| 成年女人毛片免费观看观看9| 亚洲,欧美精品.| 亚洲国产欧美网| 亚洲黑人精品在线| 亚洲专区中文字幕在线| 蜜桃久久精品国产亚洲av| 黄色视频不卡| 亚洲国产精品合色在线| av福利片在线| 亚洲片人在线观看| 波多野结衣巨乳人妻| av视频在线观看入口| 我的老师免费观看完整版| 性欧美人与动物交配| 久久久久久久久久黄片| 亚洲自偷自拍图片 自拍| 国产成人影院久久av| 人人妻,人人澡人人爽秒播| 欧美中文综合在线视频| 哪里可以看免费的av片| 欧美中文日本在线观看视频| 他把我摸到了高潮在线观看| 精品国产乱子伦一区二区三区| 天堂√8在线中文| 欧美性长视频在线观看| 久9热在线精品视频| 午夜福利18| 亚洲av成人精品一区久久| 黄频高清免费视频| 麻豆久久精品国产亚洲av| 亚洲欧美精品综合久久99| 亚洲精品美女久久久久99蜜臀| 熟妇人妻久久中文字幕3abv| 亚洲色图 男人天堂 中文字幕| av国产免费在线观看| 国产99久久九九免费精品| 成人午夜高清在线视频| 18禁观看日本| 在线播放国产精品三级| 俄罗斯特黄特色一大片| 日韩中文字幕欧美一区二区| 免费看日本二区| 国产欧美日韩精品亚洲av| 又黄又爽又免费观看的视频| 麻豆成人午夜福利视频| 日韩 欧美 亚洲 中文字幕| 免费一级毛片在线播放高清视频| 中亚洲国语对白在线视频| 午夜免费成人在线视频| 国产在线精品亚洲第一网站| 欧美成人性av电影在线观看| 老司机靠b影院| 蜜桃久久精品国产亚洲av| 露出奶头的视频| 国产精品久久久久久久电影 | 午夜福利视频1000在线观看| 无遮挡黄片免费观看| 看免费av毛片| 母亲3免费完整高清在线观看| 波多野结衣高清无吗| 久久人妻av系列| 日本a在线网址| 国产爱豆传媒在线观看 | 视频区欧美日本亚洲| 国产视频一区二区在线看| 又爽又黄无遮挡网站| 99在线人妻在线中文字幕| 亚洲午夜精品一区,二区,三区| 免费在线观看黄色视频的| 亚洲片人在线观看| 黑人操中国人逼视频| 男插女下体视频免费在线播放| av国产免费在线观看| 成人精品一区二区免费| 免费看十八禁软件| 天天躁狠狠躁夜夜躁狠狠躁| 亚洲精品av麻豆狂野| 一边摸一边抽搐一进一小说| 色在线成人网| 一级毛片精品| 高清毛片免费观看视频网站| 亚洲欧美日韩无卡精品| 免费在线观看日本一区| 久久精品91无色码中文字幕| 日韩欧美精品v在线| 丁香六月欧美| 嫁个100分男人电影在线观看| 欧美在线黄色| 波多野结衣巨乳人妻| 午夜福利视频1000在线观看| av在线播放免费不卡| 中文资源天堂在线| 精品欧美国产一区二区三| 久久婷婷人人爽人人干人人爱| 成人永久免费在线观看视频| av福利片在线| 美女黄网站色视频| 曰老女人黄片|